LIGASURE 5MM BLUNT TIP LAPAROSCOPIC SEALER-DIVIDER
K092879 · Valleylab · GEI · Oct 16, 2009 · General, Plastic Surgery
Device Facts
Record ID
K092879
Device Name
LIGASURE 5MM BLUNT TIP LAPAROSCOPIC SEALER-DIVIDER
Applicant
Valleylab
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 16, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider is a bipolar electrosurgical accessory. It is intended for use with the ForceTriad™ Electrosurgical Generator in general and gynecological, laparoscopic, surgical procedures where the ligations of vasculature (vessels and lymph) is desired. The device creates a scal by the application of RF electrosurgical energy to vascular structures interposed between the jaws of the instrument. The sealed vessels and other tissue structures may be divided by the deployment of a mechanical knife that resides within the shaft of the instrument and extends forward in a slot within the jaws. Indications for use include general laparoscopic procedures including urologic, vascular, thoracic and thoracoscopic, and gynecologic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectorny, Nissen fundoplication, colectomy, adhesiolysis, oophorecotmy, etc. The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures. The LigaSure 5mm Blunt tip Laparoscopic Sealer-Divider can be used on vessels and lymphatics up to and including 7mm and tissue bundles as large as will fit in the jaws of the instrument. The LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider can be used to seal Pulmonary Vasculature when used with the ForceTriad™ Electrosurgical generator.
Device Story
Bipolar electrosurgical instrument for laparoscopic surgery; used with ForceTriad generator. Input: vascular structures/tissue bundles clamped between jaws. Operation: applies RF energy to seal vessels/lymphatics; surgeon-actuated mechanical blade divides tissue. Features: 5mm shaft; 37cm length; 179-degree jaw rotation; blunt oval-shaped jaws with textured surface. Operated by surgeon in OR. Output: sealed and divided tissue. Benefits: enables vessel ligation and tissue division in minimally invasive procedures. Connects to generator via 'Smart' connector for instrument identification.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence established via non-clinical testing, including ex-vivo tissue studies, in-vivo porcine and canine models, and performance verification against design specifications.
Technological Characteristics
Bipolar electrosurgical instrument; 5mm diameter, 37cm length. Materials: stainless steel/polymer components. Energy source: RF electrosurgical energy via ForceTriad generator. Connectivity: 'Smart' connector for generator identification. Sterilization: supplied sterile for single use. Mechanism: bipolar RF sealing with integrated mechanical blade.
Indications for Use
Indicated for general, gynecologic, urologic, vascular, and thoracic laparoscopic procedures requiring ligation and division of vessels and lymphatics up to 7mm. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
LigaSure 5mm Laparoscopic Sealer and Divider (K031011)
Submission Summary (Full Text)
{0}------------------------------------------------
K092879
# 510(k) Summary or 510/k) Statement
510(k) Summary
OCT 1 6 2009
- Submitter Information: 1.
Valleylab, a Division of Tyco Healthcare Group LP 5920 Longbow Drive Boulder, CO Contact: Philip E. Ake Senior Regulatory Associate Phone: 303-581-6934 Fax: 303-516-8307 Email: Philip.Ake@Covidien.com
2. Name of Device
Trade name: LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider Common: Bipolar Laparoscopic Electrosurgical Instrument Classification name: Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400)
3. Predicate Devices
The LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider is substantially equivalent to the LigaSure 5mm Laparoscopic Sealer and Divider submitter under K031011. The substantial equivalence letter was dated 5/29/03. This Special 510(k) releases a blunt tip version of 5mm Laparoscopic sealer divider. The primary difference between the LigaSure 5mm Blunt tip Laparoscopic Sealer-Divider and the predicate LigaSure 5mm is the jaw. The jaw of the Blunt Tip version is shaped in a more oval fashion, has a textured surface to improve tissue grasping, and has a longer cut length than the original.
4. Device Description
The LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider is a multifunctional electrosurgical instrument for use with the ForceTriad™ generator when performing laparoscopic surgery. The instrument is capable of sealing vessels and lymphatics, dividing tissue (including vessels) clamped between the jaws, grasping tissue, and blunt dissection. The outer diameter of the instrument shaft is 5mm, with a working length of 37 cm. The following controls are located on the instrument handle
{1}------------------------------------------------
- A lever for opening and closing the instrument jaws. The . mechanism incorporates a latch to hold the jaws in the closed position during vessel sealing and cutting
- . An activation button for generator power to initiate vessel sealing
- . A trigger for actuating the cutter. The cutter can only be actuated when the jaws are closed and latched.
- . A knob to rotate the instrument jaws. The jaws can rotate 179 degrees to facilitate surgeon access and visibility.
All controls can be operated with either the right or left hand. Vessel scaling can be initiated using the activation button, or utilizing a footswitch connected to the generator.
The instrument attaches to the ForceTriad generator via a ten foot cord with a "Smart" connector that identifies the instrument type to the generator. The instrument is supplied sterile for single use.
5. . Intended Use
The LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider is a bipolar electrosurgical accessory. It is intended for use with the ForceTriad™ Electrosurgical Generator in general and gynecological, laparoscopic, surgical procedures where the ligations of vasculature (vessels and lymph) is desired.
The device creates a scal by the application of RF electrosurgical energy to vascular structures interposed between the jaws of the instrument. The sealed vessels and other tissue structures may be divided by the deployment of a mechanical knife that resides within the shaft of the instrument and extends forward in a slot within the jaws.
Indications for use include general laparoscopic procedures including urologic, vascular, thoracic and thoracoscopic, and gynecologic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectorny, Nissen fundoplication, colectomy, adhesiolysis, oophorecotmy, etc. The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
The LigaSure 5mm Blunt tip Laparoscopic Sealer-Divider can be used on vessels and lymphatics up to and including 7mm and tissue bundles as large as will fit in the jaws of the instrument.
{2}------------------------------------------------
The LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider can be used to seal Pulmonary Vasculature when used with the ForceTriad™ Electrosurgical generator.
#### 6. Summary of Technology Characteristics
This is a Special 510(k). The changes associated with this 510(k) are primarily related to the device jaw and increased force delivery. This modified 5mm Blunt Tip Laparoscopic Sealer-Divider has the same basic technological characteristics as the predicate LigaSure device. The device will seal vessels and lymph using bipolar RF energy and can mechanically divide the sealed areas or tissue with a mechanical cutting device incorporated into the shaft and jaws. The device can only operate with the ForceTriad Generator.
### Summary Non-clinical testing 7.
The Non-clinical testing consisted of ex-vivo tissue studies, in-vivo tissue studies (porcine and canine model), and performance verification. The purpose of the testing was to ensure that the device functions as intended, and meets design specifications. This also included testing to the relevant safety standards. The data obtained showed that the device is substantially equivalent to the predicate and meets the safety and effectiveness criteria.
8. Summary of Clinical testing
No clinical testing was conducted. Clinical testing was not necessary to establish substantial equivalence.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an eagle-like emblem with three stylized lines representing the bird's body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Valleylab, a Division of Tyco Healthcare Group LP % Mr. Philip E. Ake, Senior Regulatory Associate 5920 Longbow Drive Boulder, CO 80301
OCT 1 6 2009
Re: K092879
Trade/Device Name: LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 17, 2009 Received: September 18, 2009
Dear Mr. Ake:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
### Page 2 - Mr. Philip E. Ake
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Mellon
Mark N. Melkerson Director Division of Surgical, Orthopedic. and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): _
Device Name: LigaSure 5mm Blunt Tip Laparoscopic Sealer-Divider
Indications for Use:
The LigaSure 5mm. Blunt tip. Laparoscopic Sealer- Divider is a bipolar electrosurgical instrument intended for use the ForceTriad Generator in general and gynecologic surgical procedures where ligation and division of vessels and lymph is desired. The instrument creates a seal by application of RF electrosurgical energy to vascular structures (vessels and lymph) interposed between the jaws of the instrument. A blade within the instrument is surgeon actuated to divide tissue.
Indications for use include general laparoscopic procedures including urologic, vascular, thoracic and thoracoscopic, and gynecologic procedures where ligation and division of vessels is performed. These procedures include: laparoscopically assisted vaginal hysterectomy, Nissen fundoplication, colectomy, adhesiolysis, oophorecotmy etc. The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
The LigaSure 5mm Blunt tip Laparoscopic Sealer-Divider can be used on vessels and lymphatics up to and including 7mm and tissue bundles as large as will fit in the jaws of the instrument.
Over-The-Counter Use Prescription Use __ X (Part 21 CFR 801 Subpart AND/OR 21 CFR 801 Subpart D) C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
FOR M-MELKERSON
(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K092879
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.