The Spacelabs Smart Disclosure System, Model 92810 is indicated for use in clinical situations where there is a need for review of physiological waveform information and alarm events up to 72 hours after the fact. Smart Disclosure is also indicated in those situations where a retrospective analysis of monitoring patients' ECG waveform data, that can be annotated and edited, is desired. The intended use of the Spacelabs Smart Disclosure is to interface with the Spacelabs monitoring network, providing the user with a means of recalling waveform information and performing retrospective analysis. The most recent 72 hours of monitored patient ECG waveform data can be analyzed, with each analysis limited to 24 hours or less.
Device Story
Software-only medical device residing on server or client workstation; interfaces with Spacelabs monitoring network. Inputs: stored ECG and physiological waveforms/alarm events from network database. Function: enables retrospective review of up to 72 hours of data; provides shape-based retrospective analysis of ECG waveforms. Outputs: visual display of waveforms/events on workstation; printed reports via network-connected laser printer. Used in clinical settings by healthcare providers to evaluate patient history, review infrequent events, and identify onset/termination of events. Facilitates clinical decision-making through retrospective analysis and annotation of patient data.
Clinical Evidence
No clinical data. Performance testing was limited to bench testing and software verification/validation activities. Results confirmed compliance with predetermined specifications and applicable standards.
Technological Characteristics
Software-only device; operates on client-server architecture within Spacelabs monitoring network. Features shape-based retrospective ECG analysis algorithm. Connectivity via network-connected laser printer for reporting. Software developed per FDA guidance for medical device software, off-the-shelf software, and general principles of software validation.
Indications for Use
Indicated for clinical review of physiological waveforms and alarm events up to 72 hours post-event; retrospective analysis, annotation, and editing of patient ECG waveform data for monitored patients.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Spacelabs Medical Inc. / Full Disclosure System, Model 91810 (K063490)
Submission Summary (Full Text)
{0}------------------------------------------------
# K110779
### 510(k) Premarket Notification Spacelabs Healthcare Spacelabs Smart Disclosure System, Model 92810 510(k) Summary
APR 1 9 2011
| Submission Date: | 14 March 2011 | | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|--------------------------------|
| Submitter: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029 | | |
| Submitter Contact: | Mr. David J. Geraghty<br>Manager, Regulatory & Quality<br>+1 (425) 657-7200<br>david.geraghty@spacelabs.com | | |
| Official Contact: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 | | |
| Manufacturing Site: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029 | | |
| Trade Name: | Spacelabs Smart Disclosure System, Model 92810 | | |
| Common Name: | Computer, Diagnostic, Programmable | | |
| Classification Name: | Programmable Diagnostic Computer | | |
| Classification<br>Regulation: | 21 CFR §870.1425 | | |
| Product Code: | DQK | | |
| Substantially<br>Equivalent Devices: | New Spacelabs Model | Predicate 510(k) Number | Predicate Manufacturer / Model |
| Spacelabs Smart Disclosure<br>System, Model 92810 | K063490 | Spacelabs Medical Inc. /<br>Full Disclosure System,<br>Model 91810 | |
{1}------------------------------------------------
## K川0779
### 510(k) Premarket Notification Spacelabs Healthcare Spacelabs Smart Disclosure System, Model 92810 510(k) Summary
| Device Description: | The Spacelabs Healthcare (Spacelabs) Smart Disclosure System (Smart<br>Disclosure), Model 92810, is a software only medical device that<br>resides on a server or a client workstation. It is an update to the<br>predicate device, the Spacelabs Full Disclosure, Model 91810, cleared<br>by FDA in 510(k) submission K063490. Smart Disclosure is presented<br>as an icon that, when selected, opens Smart Disclosure on a client<br>workstation.<br><br>Smart Disclosure allows the recall and retrospective review of up to 72<br>hours of electrocardiogram (ECG) and other physiological waveforms<br>and alarm events that are stored in the Spacelabs Network Database.<br>The clinician can review vital signs information which enhances the<br>ability to evaluate a patient's history. Smart Disclosure is able to<br>reproduce the waveform data and events on the client workstation<br>graphic display as well as in printed reports. The clinician can review<br>infrequent events and ensures that onsets and terminations of the events<br>are captured, and make standard-sized tracings using a network-<br>connected laser printer. Additionally, the clinician can perform a shape-<br>based retrospective analysis of the ECG waveform data using the Smart<br>Disclosure shape-based retrospective algorithm. |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The Spacelabs Smart Disclosure System, Model 92810 is indicated for<br>use in clinical situations where there is a need for review of<br>physiological waveform information and alarm events up to 72 hours<br>after the fact. Smart Disclosure is also indicated in those situations<br>where a retrospective analysis of monitoring patients' ECG waveform<br>data, that can be annotated and edited, is desired.<br><br>The intended use of the Spacelabs Smart Disclosure is to interface with<br>the Spacelabs monitoring network, providing the user with a means of<br>recalling waveform information and performing retrospective analysis.<br>The most recent 72 hours of monitored patient ECG waveform data can<br>be analyzed, with each analysis limited to 24 hours or less. |
| Technology<br>Comparison: | The Smart Disclosure employs the same technological characteristics as the<br>predicate device. |
{2}------------------------------------------------
## KII0779
### 510(k) Premarket Notification Spacelabs Healthcare Spacelabs Smart Disclosure System, Model 92810 510(k) Summary
#### Summary of Performance Testing:
| Performance Testing | The Smart Disclosure was tested for performance in accordance with internal<br>requirements. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Test results indicated that the Smart Disclosure complies with its<br>predetermined specification and with the applicable Standards. |
| Software Testing | Software device modifications made to the Smart Disclosure were designed<br>and developed according to a robust software development process, and were<br>rigorously verified and validated. |
| | Software information is provided in accordance with: |
| | • FDA guidance: The content of premarket submissions for software<br>contained in medical devices, 11 May 05; |
| | • FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99;<br>and |
| | • FDA guidance: General principles of software validation; Final guidance<br>for industry and FDA staff, 11 Jan 02. |
| | Test results indicate that the Smart Disclosure complies with its predetermined<br>specification. |
| Conclusion | Verification and validation activities were conducted to establish the<br>performance and safety characteristics of the software device modifications<br>made to the Smart Disclosure. The results of these activities demonstrate that<br>the Smart Disclosure is safe and effective when used in accordance with its<br>intended use and labeling. |
| | Therefore, the Smart Disclosure is considered substantially equivalent to the<br>predicate device. |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Spacelabs Helathcare c/o Mr. Thomas Kroenke Speed to Market, Inc. PO Box 3018 Nederland, CO 80466
APR 1 9 2011
Re: K110779
Trade Name: Spacelabs Smart Disclosure System, Model 92810 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: March 14, 2011 Received: March 21, 2011
Dear Mr. Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misburanding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Thomas Kroenke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
| 510(k) Number (if known): | K 110779 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Spacelabs Smart Disclosure System, Model 92810 |
| Indications for Use: | The Spacelabs Smart Disclosure System, Model 92810 is<br>indicated for use in clinical situations where there is a need for<br>review of physiological waveform information and alarm events<br>up to 72 hours after the fact. Smart Disclosure is also indicated<br>in those situations where a retrospective analysis of monitoring<br>patients' ECG waveform data, that can be annotated and edited,<br>is desired.<br><br>The intended use of the Spacelabs Smart Disclosure is to<br>interface with the Spacelabs monitoring network, providing the<br>user with a means of recalling waveform information and<br>performing retrospective analysis. The most recent 72 hours of<br>monitored patient ECG waveform data can be analyzed, with<br>each analysis limited to 24 hours or less. |
X ____________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------ak
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K110779
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.