K110625 · Safetymate, Inc. · OVR · Jun 17, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110625
Device Name
SAFETYMATE TALKING FIRST AID
Applicant
Safetymate, Inc.
Product Code
OVR · General, Plastic Surgery
Decision Date
Jun 17, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4014
Device Class
Class 1
Attributes
Pediatric
Indications for Use
SafetyMate® Talking First Aid, SM200 Series is an electronic device intended for use to provide verbal information from established sources to refresh how to handle common medical emergencies through utilization of button options, voice queries and prompts which remind the user of important points in the management of identified emergency situations for infant, child, and adult emergencies.
Device Story
SafetyMate SM200 is a handheld, battery-powered electronic audio guide for first aid. User selects one of eight emergency categories (e.g., CPR, choking, bleeding, burns) via physical buttons. Device uses voice queries and prompts to guide users through age-appropriate first aid steps and provides audio cues for CPR cadence. Used in home or office settings by laypeople. Output is verbal instructions and audio beeps; intended to refresh user knowledge of established first aid procedures. Does not deliver therapy or contain drugs. Benefits include providing structured, step-by-step guidance during emergencies to potentially improve care appropriateness and reduce confusion.
Clinical Evidence
Clinical usability and human factors studies conducted. Evaluated whether audio prompting caused treatment delays or led to inappropriate care in simulated emergencies. Results showed 100% of users were able to use the device correctly without prompting; no significant delays in treatment noted. Protocol refinements were implemented based on initial study feedback.
Technological Characteristics
Handheld electronic device; plastic body with eight user-selection buttons. Battery-powered. Software-controlled audio output. Meets EN 55014-1/2, EN 55022, EN 61000-4-2, and FCC Part 15 standards for electromagnetic compatibility and radiated emissions.
Indications for Use
Indicated for infant, child, and adult populations as an adjunct to first aid training and information sources to provide verbal guidance and reminders for managing common medical emergencies.
Regulatory Classification
Identification
A nonresorbable gauze/sponge for external use is a sterile or nonsterile device intended for medical purposes, such as to be placed directly on a patient's wound to absorb exudate. It consists of a strip, piece, or pad made from open woven or nonwoven mesh cotton cellulose or a simple chemical derivative of cellulose. This classification does not include a nonresorbable gauze/sponge for external use that contains added drugs such as antimicrobial agents, added biologics such as growth factors, or is composed of materials derived from animal sources.
{0}------------------------------------------------
510(k) Pre-Market Notification SafetyMate, Inc.
# EXHIBIT C
# 510(k) Summary
JUN 1 7 2011
SafetyMate, Inc. 1633 Monrovia Ave. Costa Mesa, California 92627
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K110625.
### Submitter's Identification:
| Submitter's name: | SafetyMate, Inc. |
|-------------------------|------------------------------|
| Address: | 1633 Monrovia Avenue |
| | Costa Mesa, California 92627 |
| Phone: | 949.722.1121 |
| Fax: | 949.764.1774 |
| Name of contact person: | Sharon Cohen, Ph.D. |
| | Vice President |
| | sharon@safetymate.com |
| Phone: | 949.722.1121 |
| Date Summary Prepared: | June 15, 2011 |
#### 2. Name of the Device:
Name of device: Classification name: New Product Code: Class: Regulation Number: Definition:
Kit, first aid, talking OVR 1 878.4014 The device provides verbal instructions pertaining to various first aid and common medical emergencies. The device does not contain any drug or supporting device that could be applied to the patient.
#### 3. Common or Usual Name:
SafetyMate Talking First Aid
#### 4. Predicate Device Information:
The legally marketed device to which we are claiming equivalence [807.92(a)(3)]:
SafetyMate, SM200 Series
{1}------------------------------------------------
| Reference # | Device Name | Applicant |
|-------------|------------------------------|-----------|
| Exempt | 510(k) Exempt First Aid Kits | |
| | Without Drugs | |
### 5. Device Description:
SafetyMate® Talking First Aid, SM200 Series is a hand-held, interactive talking first aid guide for adult, child or infant emergencies. It includes eight themed first aid categories including CPR/Unconscious, Choking/Breathing, Allergic Reaction, Bleeding, Burns, Falls/Bone Injury, Poison/Bites/Stings, and Seizures. Besides providing guidance for what to do in case of an emergency, SafetyMate also highlights what not to do to prevent further injury. SafetyMate is designed to reinforce and remind users of the correct steps and procedures to provide established first aid care.
#### 6. Intended Use:
SafetyMate® Talking First Aid, SM200 Series is an electronic device intended for use to provide verbal information from established sources to refresh how to handle common medical emergencies through utilization of button options, voice queries and prompts which remind the user of important points in the management of identified emergency situations for infant, child, and adult emergencies.
#### Comparison to Predicate Device: 7.
### Table #1
| Subject Area | SafetyMate, SM200 Series | First Aid Kit Without Drug<br>Predicates (510(k) Exempt) | Similar or<br>Different |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Product Code | OVR (*the Agency created a<br>new product code to better<br>accommodate / define the subject<br>device) | OHO | Different |
| Product<br>Classification | Class I | Class I | Similar |
| Classification Name | Kit, talking, first aid | Non-Resorbable gauze<br>/sponge for external Use | Different |
| Proprietary Name | SafetyMate, SM200 Series | First Aid Kit Without Drug | |
| Intended Use | Talking First Aid<br>assistance | Supplies to be used | Similar |
| Indications for Use | Intended for use as an<br>adjunct to first aid /<br>responder training and<br>information sources<br>through utilization of | Intended for use as an<br>adjunct to first aid offering<br>the user an external means of<br>absorbing body fluids in the<br>management of emergency | Different |
| | voice queries and prompts<br>which remind the user of<br>important points in the<br>management of identified<br>emergency situations. | situation | |
| Intended<br>Population | Infant, child and adult | Infant, child and adult | Similar |
| Technological<br>features - Mode of<br>Operation | Electronic talking first aid<br>device. User selects one of<br>eight first aid categories.<br>Voice queries prompt<br>user to select age-<br>appropriate instructions. | No technological features | Different |
| Design | Electronic audio aid in<br>performing CPR and<br>other first aid activities.<br>Audio beeps proper CPR<br>cadence with voice<br>prompt for breaths.<br>Software controlled.<br><br>Has eight user selection<br>buttons which guide the<br>user through the proper<br>first aid steps with audio<br>instructions. Voice queries<br>prompt user to select age<br>category. | Porous material used to soak<br>up liquid/bodily fluids.<br>Not software controlled. | Different |
| Materials: | Plastic body with eight<br>button selections. Battery<br>powered. | Porous material. Not battery<br>powered. | Different |
| Performance | First aid assistant device<br>with eight major first aid<br>arcas. | First aid assistant materials. | Different |
| Specifications | Hand held device | Hand held product(s) | Different |
| Mechanical Safety | None | None | Similar |
| Chemical Safety | N/A | N/A | Similar |
| Anatomical Sites | N/A | Wherever bodily fluid is<br>gushing | Different |
| Human Factors | Designed for ease of use | Designed for ease of use | Similar |
| Energy<br>Used/Delivered | Electrical none delivered | Electrical none delivered | Similar |
| Compatibility with<br>the environment<br>and other devices | Standard luer connections<br>to administration devices | N/A | Different |
| Where used | Home or office | Home or office | Similar |
| Standards met | See Section I -- CDRH<br>Premarket Review<br>Submission Cover Sheet | Unknown | |
| Electrical Safety | See Exhibit J | Unknown | |
| Thermal safety | N/A | N/A | |
| Radiation safety | N/A | N/A | |
{2}------------------------------------------------
.
・
ﺴﻴ
{3}------------------------------------------------
# 510(k) Pre-Market Notification
SafetyMate, Inc.
# 8. Discussion of Non-Clinical Tests Performed:
| SM202 | Standard | Title |
|-------|-------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| SM202 | EN 55014-1<br>EN 55022 | Electromagnetic compatibility –<br>Requirements for household<br>appliances and similar apparatus –<br>Part 1 Emissions |
| SM202 | EN 55014-2<br>EN 61000-4-2 | Electromagnetic compatibility –<br>Requirements for household<br>appliances and similar apparatus –<br>Part 2 Immunity |
| SM202 | FCC Part 15<br>Section 15.109 | Radiated Emission |
### Table #2
### 9. Discussion of Clinical Tests Performed:
Clinical usability and human factors studies were conducted to assess if SafetyMate Talking First Aid Device impairs appropriateness of care and/or speed to treat. Conclusions indicated that overall, no significant delays in treatment were noted. While there were a few instances in which changes to the protocols needed to be made to address points of confusion, either over word choice or clarity of instructions, those changes have been implemented and the issues have been answered.
#### 10. Conclusions:
{4}------------------------------------------------
Based on the design, technology, performance, functional testing, and intended use, the SafetyMate, SM200 series is substantially equivalent to the devices currently marketed under the Federal Food, Drug and Cosmetic Act. Usability studies indicate that 100% of the users were able to use the device correctly without any prompting from the observer. The human factors study addressed two questions: first, whether the use of an audio prompting device (in this case, a SafetyMate Talking First Aid device), in a simulated first aid emergency situation, caused a delay in treatment when compared to the same situation without a prompting device; and second, whether the use of a prompting device lead the subjects to render inappropriate care, in a simulated first aid emergency situation. While there were a few instances in which changes to the protocols needed to be made, those changes have been implemented and the issues have been answered.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle or other bird with outstretched wings. The bird is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Sharon Cohen Ph.D Vice President SafetyMate, Incorporated 1633 Monrovia Avenue Costa Mesa, California 92627
## JUN 1 7 2011
Re: K110625
Trade/Device Name: SafetyMate® Talking First Aid, SM200 Series Regulation Number: 21 CFR 878.4014 Regulation Name: Nonresorbable Gauze/Sponge for External use Regulatory Class: I Product Code: OVR Dated: May 10, 2011 Received: May 12, 2011
Dear Dr. Cohen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2- Dr. Cohen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutlFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Antony D. Austin
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Exhibit E
### Indications for Use
Page __ 1__ of__ 1_
510(k) Number (if known): K110625
SafetyMate® Talking First Aid, SM200 Series Device Name:
Indications For Use:
SafetyMate® Talking First Aid, SM200 Series is an electronic device intended for use to provide verbal information from established sources to refresh how to handle common medical emergencies through utilization of button options, voice queries and prompts which remind the user of important points in the management of identified emergency situations for infant, child, and adult emergencies.
Prescription Use (Pet 21 CFR 801 Subpart D)
OR
Over-The Counter Use X (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
NAT 6115/2011
"Weifing Gm for RZC. 6/1/2011
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K110625
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.