GF-210R MULTI-GAS MODULE FOR NOHON KOHDEN BEDSIDE MONITORS
K110594 · Nihon Kohden Corp. · CCK · Oct 28, 2011 · Anesthesiology
Device Facts
Record ID
K110594
Device Name
GF-210R MULTI-GAS MODULE FOR NOHON KOHDEN BEDSIDE MONITORS
Applicant
Nihon Kohden Corp.
Product Code
CCK · Anesthesiology
Decision Date
Oct 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nihon Kohden GR-210R Multi-Gas Module is intended to measure carbon dioxide (CO2), nitrous oxide (N2O), oxygen (O2), and following anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane) of a patient undergoing anesthesia and display the results on a Nihon Kohden bedside monitor. The GF-210R can measure two anesthetic agents simultaneously. The system is intended for use by qualified medical personnel within a hospital or clinical environment and is available for use on any patients as determined by the qualified medical personnel.
Device Story
GF-210R Multi-Gas Module is an integrated anesthesia gas detection unit; samples gas from anesthetic/respiration circuits via exhaust tube. Uses infrared ray absorption for CO2, N2O, and anesthetic agents; paramagnetic detection for O2. Measures gas concentrations on real-time, breath-by-breath basis; detects inspired/expired phases from CO2 concentration; calculates respiration rate. Contains OEM Draeger gas analyzer module. Connects to Nihon Kohden bedside monitors for display of numerics, waveforms, and alarms. Used by qualified medical personnel in hospitals/clinics. Provides continuous monitoring of patient gas levels during anesthesia; assists clinicians in assessing patient respiratory status and anesthetic depth.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness demonstrated through software and hardware verification and validation, and compliance with safety standards.
Technological Characteristics
Enclosure containing OEM Draeger gas analyzer. Measurement methods: Non-Dispersive Infrared (NDIR) for CO2, N2O, and anesthetic agents; Paramagnetic for O2. Connectivity: Multi-link connector to Nihon Kohden bedside monitors. Power: 100-240 VAC. Dimensions: 280 x 100 x 200 mm. Weight: 4.2 kg. Operating temperature: 10-40°C.
Indications for Use
Indicated for patients of any age undergoing anesthesia in hospital or clinical settings. Used to monitor CO2, N2O, O2, and anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane) and respiration rate.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
K110594
Nihon Kohden America, Inc. September 14, 2011
## 510(k) Summary
| Company Name: | Nihon Kohden Corporation<br>90 Icon Street<br>Foothill Ranch, CA 92610 |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | GF-210R Multi-Gas Module for Nihon Kohden Bedside Monitors |
| 510(k) Sponsor,<br>Contact: | Nihon Kohden America, Inc.<br>90 Icon Street<br>Foothill Ranch, CA 92610<br><br>Steve Geerdes<br>Director Quality Assurance and Regulatory Affairs<br>Phone: (949) 580-1555 Ext. 3325<br>Fax: (949) 580-1550 |
| Summary Date: | 9/14/2011 |
| Common Name: | Gas Analyzer, carbon dioxide, oxygen, nitrous oxide, enflurane, and<br>halothane Gas analyzer |
## Classification Name:
| 868.1400 | Analyzer, Gas, Carbon dioxide, | was a final from the first of the county of the county of<br>CCK |
|-------------------------------------------------------|--------------------------------|------------------------------------------------------------------|
| · Production Million March Justice March March Market | Gaseous-Phase | |
## Predicate Device(s):
| Nihon Kohden AG-920RA | K#020046 |
|--------------------------|----------|
| Draeger SC 6802XL w/Scio | K#040188 |
### 1.0 Description of Device
The GF-210R Multi-Gas Module is a fully integrated anesthesia gas detection unit that measures Carbon dioxide (CO2), nitrous oxide (N20), oxygen (O2), and five anesthetic agents and displays the results on a Nihon Kohden bedside monitor.
The GF-210R is an enclosure that contains an OEM Gas Analyzer (Draeger Module # 6871620). There is a sample gas exhaust port that allows the GF-210R to sample gas from a anesthetic or respiration circuit through exhaust gas tube. A multi-link connector allows the GF-210R to comunicate the measured gas value and display the value on a Nihon kohden bedside monitor
### Clinical Application 1.1
{1}------------------------------------------------
The GF-210R Multi-Gas Module is used to measure Carbon dioxide (CO2), nitrous oxide (N2O), The OI -2101 Maili Oas Mesthetic agents and display the results on a bedside monitor.
This anesthesia gas detection unit samples and measures the concentration of gases administered to and respired by the patient during anesthesia. Infrared ray absorption detection method is used for carbon dioxide (CO2), nitrous oxide (N2O), and anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane). Paramagnetic detection method is used for oxygen (O2). Gases are monitored on a real time, breath-by-breath basis and the measured inspired and (32). Case an entrations are displayed on the Nohon Kohden bedside monitor. The inspired and expired phases are detected from the instantaneous CO2 concentration and respiration rate is counted from these phases.
The measured data is transferred to and displayed on a Nihon Kohden bedside monitor. User interface for controlling and displaying alarms, setting sampling rates and calibration are provided through the bedside monitor. All measured data (numerics and waveforms) are displayed and recorded through the bedside monitor.
### Intended Use of Device 2.0
The Nihon Kohden GR-210R Multi-Gas Module is intended to measure carbon dioxide (CO2), nitrous oxide (N2O), oxygen (O2), and following anesthetic agents (Halothane, Isoflurane, mirous only (120), only get (22ffurane) of a patient undergoing anesthesia and display the results on a Nihon Kohden bedside monitor. The GF-210R can measure two anesthetic agents simultaneously.
The system is intended for use by qualified medical personnel within a hospital or clinical environment and is available for use on any patients as determined by the qualified medical personnel.
### Technical Characteristics 3.0
The technical characteristics of the GF-210R are equivalent to those of the predicate devices. The following tables summarizes equivalence.
| Comparison feature<br>or specification | Predicate device<br>(K#020046)<br>Nihon Kohden AG-<br>920RA | Predicate device<br>(K040188) Draeger<br>SC 6802XL w/Scio | New device<br>Nihon Kohden<br>GF-210R | Comments |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <general></general> | | | | |
| Host | Nihon Kohden<br>"LifeScope" patient | Infinity Nodular<br>Monitors | Nihon Kohden<br>"LifeScope" | |
| | monitors | ( SC7000, SC8000, SC9000XL) | patient monitors | |
| Power input | 100 to 240 VAC | 100-240 V, 50/60<br>Hz | Same | |
| Power<br>Consumption | 40VA | from specified<br>power supply | 70VA | |
| Operating<br>Temperature | 10 to 35 deg C | 10 to 40 deg C | 10 to 40 deg C | |
| Dimension | 180 x 140 x 220 mm | 115 x 190 x 270<br>mm | 280 x 100 x 200<br>mm | |
| Weight | 3.4 kg | 3.457Kg | 4.2 kg | |
| measurement> | | | | |
| OEM manufacturer<br>of gas module | Artema Medical | | Draeger Medical | |
| Measurement<br>Parameters | Fi/ET CO2, N2O, O2,<br>Anesthetic agents<br>(Halothane, Isoflurane,<br>Enflurane, Sevoflurane,<br>Desflurane),<br>Respiration Rate | Fi/ET CO2, N2O,<br>O2, Anesthetic<br>agents<br>(Halothane,<br>Isoflurane,<br>Enflurane,<br>Sevoflurane,<br>Desflurane),<br>Respiration Rate | Same | |
| Sampling Rate | Selectable : 70 to 200<br>mL/min<br>Auto : 100 or 200<br>mL/min | 200 mL/min | 200 mL/min | |
| Warm-up time | 45 Sec (Startup)<br>10 min (Full accuracy) | 1 min (CO2 Startup)<br>6 min (Full<br>accuracy) | 1 min (CO2<br>Startup)<br>6 min (Full<br>accuracy) | |
| CO2 Measurement | | | | |
| Measurement<br>method | Non-Dispersive<br>Infrared Ray absorption | Non-Dispersive<br>Infrared Ray<br>absorption | Same | |
| Measuring range | 0 to 76 mmHg | 0 to 10 % | 0 to 10 Vol.%<br>(Ref. 0 to 76<br>mmHg @ambient<br>pressure<br>760mmHg) | The value using Vol.<br>% is the spec by<br>Draeger.<br>See the value, which<br>is converted to<br>mmHg for reference. |
| Measuring<br>accuracy | ± 2 mmHg (0 to 40<br>mmHg)<br>± 3 mmHg (40 to 55<br>mmHg)<br>± 4 mmHg (55 to 76<br>mmHg) | ± (0.5 Vol% or<br>±12% rel) | ± (0.43 Vol.% +<br>8 % rel.)<br>(Ref.±(3.3<br>mmHg + 8 % rel.)<br>@ambient<br>pressure 760<br>mmHg) | The value using Vol.<br>% is the spec by<br>Draeger.<br>See the value, which<br>is converted to<br>mmHg for reference.<br>Essential<br>performance in<br>EMC standard |
| Rise Time (10 to<br>90%) | 250 ms @ 200mL/min | < 350 ms @<br>150mL/min | 350 ms | |
| N2O Measurement | | | | |
| Measurement<br>method | Non-Dispersive<br>Infrared Ray absorption | Non-Dispersive<br>Infrared Ray<br>absorption | Same | |
| Measuring range | 0 to 100 % | 0 to 100 % | Same | |
| Measuring<br>accuracy | ± 3 vol.% | ± (2 Vol. % + 8 %<br>rel.) | ± (2 Vol.% + 8<br>% rel.) | |
| Rise Time (10 to<br>90%) | 250 ms @ 200mL/min | < 500 ms | 500 ms | |
| 02 Measurement | | | | |
| Measurement<br>method | Paramagnetic | Paramagnetic sensor | Same | |
| Measuring range | 0 to 100 % | 0 to 100 % | 5 to 100 % | |
| Measuring<br>accuracy | +/- 2 vol.% (0 to 55 %)<br>+/- 3 vol.% (55 to 100<br>%) | +/- 3 vol.% | +/- (2.5 Vol.% +<br>2.5 % rel.) | Essential<br>performance in<br>EMC standard |
| Rise Time (10 to<br>90%) | 500 ms | < 600 ms<br>@150mL/min | Same | |
| Anestetic Agent<br>Measurement | | | | |
| Measurement<br>method | Non-Dispersive<br>Infrared Ray absorption | Non-Dispersive<br>Infrared Ray<br>absorption | Same | |
| Agent<br>Identification | Auto | Auto | Same | |
| Mixed Agent<br>Measure. | Primary and Secondary | Primary and<br>Secondary | Same | |
| Measuring range | | | | |
| Halothane | 0 to 5 % | 0 to 8.5 % | 0 to 8.5 % | |
| Isoflurane | 0 to 5 % | 0 to 8.5 % | 0 to 8.5 % | |
| Enflurane | 0 to 5 % | 0 to 10 % | 0 to 10.0 % | |
| Sevoflurane | 0 to 8 % | 0 to 10 % | 0 to 10.0 % | |
| Desflurane | 0 to 18 % | 0 to 20 % | 0 to 20.0 % | |
| Measuring<br>accuracy | +/- 0.2 vol.% (0 to 5 %)<br>+/- 0.4 vol.% (5 to 10<br>%)<br>+/- 0.6 vol.% (10 to 15<br>%)<br>+/- 1.0 vol.% (15 to 18<br>%) | +/- (0.15 Vol% +15<br>% rel.) | +/- (0.2 vol%<br>+15.0 % rel.) | Essential<br>performance in<br>EMC standard |
| Rise Time (10 to<br>90%) | 300 ms (Halothane,<br>Isoflurane, Sevoflurane,<br>Desflurane) @<br>200mL/min<br>500 ms (Enflurane) @<br>200mL/min | < 500 ms @<br>150mL/min | 500 ms | |
{2}------------------------------------------------
{3}------------------------------------------------
{4}------------------------------------------------
Nihon Kohden America, Inc. September 14, 2011
## Traditional 510(k) Notification GF-210R Multi Gas Module
.
{5}------------------------------------------------
| Respiration Rate<br>Measurement | | | | |
|---------------------------------|---------------------------------------------|---------------------------------------------------------------------------|-----------|--|
| Measuring range | 4 to 60 counts/min | 0 to 90 Counts/min | Same | |
| Measuring<br>accuracy | 1 counts/min | 1 Counts/min (0 to<br>60 counts/min)<br>not specified (>60<br>counts/min) | Same | |
| MAC Calculation | available (only when<br>used with BSM-6000) | available | available | |
### 4.0 Data Summary
Testing of the Nihon Kohden System with the MS-120BK was performed in compliance with the Nihon Kohden Corporation design control process. Testing included:
- 1. Software and hardware verification and validation, and
- 2. Safety standard compliance
A summary of data supporting the safety of the Nihon Kohden GF-201R was provided.
### 5.0 Conclusions
The safety and effectiveness of the Nihon Kohden GF-210R was demonstrated by testing in compliance with the Design Control process. The intended use and technology of the Nihon Kohden GF-201R is equivalent to the predicate devices. No new questions of safety or effectiveness are raised.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle-like symbol with three curved lines representing the bird's body and wings. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
OCT 2 8 2011
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Steve Geerdes Director Quality Assurance and Regulatory Affairs Nihon Kohden Corporation 90 Icon Street Foothill Ranch, California 92610
- Re: K110594
Trade/Device Name: Nihon Kohden GF-210R Multi-Gas Module Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: September 14, 2011 Received: September 20, 2011
Dear Mr. Geerdes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2 - Mr. Geerdes
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K110594
Device Name: _______ Nihon Kohden GF-210R Multi-Gas Module_
Indications for Use:
The Nihon Kohden GR-210R Multi-Gas Module is intended to measure carbon dioxide (CO2), nitrous oxide (N2O), oxygen (O2), and following anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane) of a patient undergoing anesthesia and display the results on a Nihon Kohden bedside monitor. The GF-210R can measure two anesthetic agents simultaneously.
The system is intended for use by qualified medical personnel within a hospital or clinical environment and is available for use on any patients as determined by the qualified medical personnel..
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultheis
Division Sign-Off Office of In Device Evaluation 510(k)
signature
Page 1 of
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k/10594
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.