K020046 · Nihon Kohden America, Inc. · CCK · Jul 25, 2002 · Anesthesiology
Device Facts
Record ID
K020046
Device Name
MULTIGAS UNIT, MODEL AG-920RA
Applicant
Nihon Kohden America, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Jul 25, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device currently marketed per K001693 is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring including arrhythmia detection and is available for use by medical personnel on patients within a medical facility including adults, children and infants. The optional AG-920RA module will measure carbon dioxide (CO2), nitrous oxide (N2O), oxygen (O2), and any of five anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane) of a patient undergoing anesthesia and display the results on a bedside monitor.
Device Story
BSM-4100A Bedside Monitor with AG-920RA option; monitors physiological data including cardiac signals and vital signs. AG-920RA module adds gas analysis capabilities; measures CO2, N2O, O2, and five anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane) for patients undergoing anesthesia. Device displays real-time data on bedside monitor; provides alarms for abnormal physiological states. Used by medical personnel in clinical settings. Facilitates clinical decision-making by providing continuous patient status updates during monitoring and anesthesia. Benefits include improved patient safety through continuous surveillance and early detection of physiological changes.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electromagnetic, environmental, safety, and performance procedures to verify operation. Software validation confirmed performance within specifications.
Technological Characteristics
Bedside patient monitor with integrated gas analysis module. Complies with IEC 60601-1, IEC 60601-1-2, and CISPR11 standards. Non-sterile. Not patient-contacting. Software-based monitoring and alarm system.
Indications for Use
Indicated for cardiac and vital signs monitoring, including arrhythmia detection and gas analysis (CO2, N2O, O2, and anesthetic agents), for adult, pediatric, and infant patients in medical facilities.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
NIHON KOHDEN AMERICA, INC. January 3, 2002
K 020046
510(k) NOTIFICATION BSM-4100A Series with AG-920RA Option
JUL 2 5 2002
#### SECTION 2 - 510(K) SUMMARY
Name and Address of Applicant Nihon Kohden America, Inc. Attn: Regulatory Affairs 90 Icon Street Foothill Ranch, California 92610
Phone: (949) 580-1555 Fax: (949) 580-1550
Common names for the device include Bedside Monitor, Patient Monitor, Cardiac Monitor and Vital Signs Monitor. The functions of the device added by the AG-920RA option are classified as Class II per the Anesthesiology Device Classification Panel per 21 CFR 868.1500, "Analyzer, Gas, Enflurane, Gaseous-Phase", CBQ; 21 CFR 868.1620, "Analyzer, Gas, Halothane, Gaseous-Phase", CBS; and 21 CFR 868.1700, "Analyzer, Gas, Nitrous-Oxide, Gaseous-Phase", CBR.
The AG-920PA module is equivalent to the 90518 Multi-gas Analyzer per K954962 and the SC9000/SC9015 Multi-gas Module per K965062.
The device currently marketed per K001693 is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring including arrhythmia detection and alarms within a medical facility. With the new AG-920PA option, the device will measure and display Carbon dioxide (CO2), nitrous oxide (N2O), oxygen (O2), and any of five anesthetic agents (Halothane, Isofilurane, Enflurane, Sevoflurane, and Desflurane).
The device complies with IEC 601-1 subclause 56.3(c) implemented by 21 CFR Part 898 Performance Standard for Electrode Lead Wires and Patient Cables. To date, no other special controls or performance standards are known or established for this device.
The device is not sterile.
The device does not directly contact patients, therefore, good laboratory practice studies were not required per 21 CFR part 58.
The device was subjected to electromagnetic, environmental, safety and performance testing procedures. These tests verified the operation of the device. Software validation tested the operation of the device software. The results confirmed that the device performed within specifications.
The device is designed to comply with the following voluntary industrial standards: IEC 60601-1 (1988-12), Amendment 1 (1991-11), Amendment 2 (1995-03), IEC 60601-1-2 (1993-05) and CISPR11 Group 1, Class A
Therefore, Nihon Kohden believes that the device is substantially equivalent to the predicate.
{1}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure, represented by three curved lines, suggesting a person in profile.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
**JUL 25 2002**
Ms. Serrah Namini Regulatory Affairs Manager Nihon Kohden America, Incorporated 90 Icon Street Foothill Ranch, California 92610
Re: K020046
Trade/Device Name: Multigas Unit, Model AG-920RA Regulation Number: 868.1400; 868.1720; 868.1700; 868.1500; 868.1620 Regulation Name: Carbon dioxide gas analyzer: Oxygen gas analyzer: Nitrous oxide gas analyzer; Enflurane gas analyzer; Halothane gas analyzer Regulatory Class: II Product Code: CCK; CCL; CBR; CBQ; NHO, NHP, NHQ; CBS Dated: May 2, 2002 Received: May 2, 2002
Dear Ms. Namini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
Page 2 - Ms. Namini
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice · requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
:Sincerely yours,
Timothy A. Ulatowski
Timothy A. I Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# NIHON KOHDEN AMERICA, INC. January 3, 2002
### 510(k) NOTIFICATION BSM-4100A Series with AG-920RA Option
G. Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
## Device Name: BSM-4100A Bedside Monitor with AG-920RA Option
The device currently marketed per K001693 is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring including arrhythmia detection and is available for use by medical personnel on patients within a medical facility including adults, children and infants.
The optional AG-920RA module will measure carbon dioxide (CO2), nitrous oxide (N₂O), oxygen (O₂), and any of five anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane) of a patient undergoing anesthesia and display the results on a bedside monitor.
Prescription Use
R.W. Wisham
Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.