The D-SPECT™ Cardiac Scanner System is an emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. The D-SPECT™ Cardiac Scanner System is primarily intended for cardiac applications. The D-SPECT™ Cardiac Scanner System supports radionuclides within the energy range of 40 -170 keV. The D-SPECT™ Cardiac Scanner System supports acquisition and imaging modes such as, but not limited to, SPECT (Single Photon Emission Computed Tomography), Multi-Gated SPECT and is featuring low count rate (low dose) acquisition, with no lose in image quality. The D-SPECT™ Cardiac Scanner System is intended for Processing, Reporting, Archiving, Display, Communication and analysis of emission computerized tomography data using tools for imaging and fully automated review and quantification of Cardiac SPECT data including Quantitative Perfusion of CCT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS).
Device Story
D-SPECT™ Cardiac Scanner System performs myocardial perfusion imaging; utilizes detector head, gantry, and patient chair; detector modules rotate within closed head; no external moving parts surrounding patient. System uses RFID tags on patient wrist for positive identification of patient and radiopharmaceutical. Acquisition station console controls operation. Processing workstation performs data reconstruction, archiving, display, communication (DICOM), and analysis. Includes automated review and quantification software (QPS, QGS, QBS). Used in clinical settings by healthcare professionals. Benefits include shorter scan times and low-dose acquisition without loss of image quality; automated tools assist clinicians in cardiac SPECT data interpretation and decision-making.
Clinical Evidence
No clinical data provided; substantial equivalence supported by technological comparison and bench testing of system modifications.
Technological Characteristics
Emission computed tomography system; detector head with rotating internal modules; supports 40-170 keV radionuclides. Connectivity via NEMA DICOM protocol for PACS integration. Includes RFID-based patient/agent identification. Software includes automated quantification algorithms (QPS, QGS, QBS).
Indications for Use
Indicated for patients requiring emission computed tomography imaging, primarily cardiac applications, using gamma ray radionuclides (40-170 keV). Supports SPECT and Multi-Gated SPECT acquisition modes for diagnostic imaging, review, and quantification.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for Spectrum Dynamics. The logo consists of a triangular shape on the left and the words "SPECTRUM DYNAMICS" on the right. Below the words "SPECTRUM DYNAMICS" is the phrase "HIGH SPEED FUNCTIONAL IMAGING".
JUL 1 3 2011
110567
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92
## Submitter Information
Spectrum Dynamics (Israel) Ltd. 22 Bareket Street North Industrial Park POB 3033 Caesarea 30889, Israel Tel: (+972) 73 7374500, Fax: (+972) 73 737450 Submission contact person: Ilan Sharon P.O.Box 4262, Zichron Yaacov 30900, Israel TEL: (+972) 52 8704904, Fax: (+972) 73 737450
#### Device Classification
| | Proprietary Device Name: | D-SPECT™ Cardiac Scanner System |
|--|----------------------------|-----------------------------------------------------------------------------------------------|
| | Common name: | Emission Computed Tomography System |
| | Product Code: | KPS (subsequent code LLZ) |
| | Classification Name: | Emission computed tomography system (Subsequent – Picture archiving and communications system |
| | Classification Regulation: | 21 CFR §892.1200 (subsequent 21 CFR §892.2050) |
| | Regulatory Class: | II |
## Identification of Legally Marketed Predicate Devices
| Spectrum Dynamics: D-SPECT™ Cardiac Scanner System - | K072468 |
|------------------------------------------------------|---------|
| ELGEMS: eNTEGRA workstation - | K003264 |
| ADAC Laboratories: AutoQUANT - | K980715 |
| UltraSPECT: WBR Half Dose - | K080784 |
#### Device Description
Spectrum Dynamics' D-SPECT™ Cardiac Scanner System is a SPECT device, which is designed to perform myocardial perfusion imaging. The device is comprised of a detector head, gantry, and patient chair. Device operation is controlled from an acquisition station console. The system is supported by use of data-transfer accessories (RFID tags), which are attached to the patient's wrist, for patient and radiopharmaceutical agent positive identification. The cardiac gamma camera is designed such that there are no external moving parts that surround the patient. Detector modules rotate within the closed detector head. The special scanning geometry and detector technology enable shorter scan times. Additionally, a new revision of software was developed, adding Archiving, Display and Communication capabilities (NEMA standard DICOM protocol) with external processing stations PACS systems. The new software package also features correction of insignificant bugs discovered since last release. The modified D-SPECT™ Cardiac Scanner System include a Processing workstation. The Processing Workstation, which is part of the imaging system, is equipped with software package for Processing, Archiving, Display, Communication and analysis of emission computerized tomography data using tools for imaging and fully automated review and quantification of Cardiac SPECT data including the Cedars Sinai Quantitative Perfusion SPECT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS) software.
{1}------------------------------------------------
### Intended Use of Device
The D-SPECT™ Cardiac Scanner System is an emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data.
The D-SPECT™ Cardiac Scanner System is primarily intended for cardiac applications.
The D-SPECT™ Cardiac Scanner System supports radionuclides within the energy range of 40 -170 keV. The D-SPECT™ Cardiac Scanner System supports acquisition and imaging modes such as, but not limited to, SPECT (Single Photon Emission Computed Tomography), Multi-Gated SPECT and is featuring low count rate (low dose) acquisition, with no lose in image quality.
The D-SPECT™ Cardiac Scanner System is intended for Processing, Reporting, Archiving, Display, Communication and analysis of emission computerized tomography data using tools for imaging and fully automated review and quantification of Cardiac SPECT data including Quantitative Perfusion of CCT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS).
## Safety & Effectiveness
The intended use and indications of the submitted modified D-SPECT™ Cardiac Scanner is identical to the legally marked devices: Spectrum Dynamics: D-SPECT™ Cardiac Scanner System (K072468), ELGEMS: eNTEGRA workstation (K003264), ADAC Laboratories: AutoQUANT (K980715) and UltraSPECT: WBR Half Dose (K080784).
Similarity of intended use, indications and technological features demonstrated, therefore, the risks and benefits are comparable to the predicate devices.
We therefore believe that there are no new questions of safety or effectiveness raised by the introduction of modified D-SPECTTM Cardiac Scanner System.
#### Rational for Substantial Equivalency
Similarities:
The intended use and indications of the proposed modified D-SPECT™ Cardiac Scanner System are similar to the legally marketed predicate devices.
The technology, performance and most of the specifications of the proposed modified D-SPECT™ Cardiac Scanner System are similar to the legally marketed predicate devices Differences:
The proposed modified system includes wider indications in reference to legally marked D-SPECT™ Cardiac Scanner System (K072468) indication for use that was limited to Cardina SPECT aquicition and reconstruction.
The proposed System intended use and indications for use are similar to the above mentioned predicate devices (Adding PACS and fully automated review and quantification of Cardiac SPECT data).
## Substantial Equivalence Statement
Based on the above, it is Spectrum Dynamics's opinion that the proposed modified D-SPECT™ Cardiac Scanner System is substantially equivalent in terms of design, functional features and safety & effectiveness to the unmodified D-SPECT™ Cardiac Scanner System (K072468) legally marketed device and to the legally marked predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Spectrum Dynamics (Israel) Ltd. % Mr. Ilan Sharon QA & RA Consultant Ilan Sharon - Regulatory Affairs Consultant 1 Ha'rakefet St. Zichron Yaacov, 30900 ISRAEL
JUL 1 3 2011
Re: K110507
Trade/Device Name: D-SPECT™ Cardiac Scanner System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: February 20, 2011 Received: February 22, 2011
## Dear Mr. Sharon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: D-SPECTTM Cardiac Scanner System
Indications for Use:
The D-SPECT™ Cardiac Scanner and System is an emission computed tomography system intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. The D-SPECTTM Cardiac Scanner System includes display equipment, patient and equipment supports, component parts, and accessories.
D-SPECT™ Cardiac Scanner System is primarily intended, but not limited to, cardiac applications.
D-SPECT™ Cardiac Scanner System supports radionuclides within the energy range of 40-170 keV.
The D-SPECT™ Cardiac Scanner System supports acquisition and imaging modes such as, but not limited to, SPECT (Single Photon Emission Computed Tomography), Multi-Gated SPECT and features low count rate (low dose) acquisition, with no lose in image quality.
D-SPECT™ Cardiac Scanner System is intended for Processing, Reporting, Archiving, Display, Communication and analysis of SPECT data, using tools for imaging and fully automated review and quantification of Cardiac SPECT data including Quantitative Perfusion SPECT (QPS), Quantitative Gated SPECT (QGS) and Quantitative Blood Pool Gated SPECT (QBS).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S. Patel
Division of Radiologica Vitro Diagnostic D
510K K110507
Page 1 of 1
: " | |
时 ﮯ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.