K110411 · Monteris Medical, Inc. · MOS · Apr 8, 2011 · Radiology
Device Facts
Record ID
K110411
Device Name
ATAMA SYSTEM
Applicant
Monteris Medical, Inc.
Product Code
MOS · Radiology
Decision Date
Apr 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of the AtamA™ System (comprised of two major subsystems: a MR head coil apparatus and a patient stabilization apparatus), in conjunction with a magnetic resonance (MR) imaging system, is the collection of MR data and images of the human brain before, during, and at the end of brain surgery, in a standard operating room, diagnostic MR rooms, or in a MR intra-operative room, while stabilizing the patient's head during neurosurgical procedures and imaging. The patient head-stabilizing apparatus can also be used for other neurosurgical procedures requiring the patient's head to be stabilized or fixed. The head coil can also be used as a standard diagnostic head coil for diagnostic MR imaging. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Device Story
AtamA System combines MR head coil and patient stabilization apparatus; used in neurosurgery and diagnostic imaging. Head coil subsystem: split-array 8-channel receive coil; two-part design allows rotation around patient head for surgical access without repositioning patient or moving coils. Stabilization subsystem: Head Fixation Ring (HFR), cradle for fixation relative to MRI bore, and patient board for transport. System maintains spatial registration for stereotactic procedures. Operated by clinical staff in OR or MR suites. Provides high-quality MR images for physician interpretation to assist in diagnosis and surgical guidance. Benefits include expedient data collection, minimized patient movement, and stable head fixation.
Clinical Evidence
No clinical investigations or clinical performance testing were conducted. Substantial equivalence is supported by bench testing and biocompatibility analysis per ISO 10993.
Technological Characteristics
Split-array 8-channel MR receive coil; two-part design. Materials are functionally similar to predicates. System includes Head Fixation Ring, cradle, and patient board. Operates with MR imaging systems; utilizes specific coil configuration data files. Complies with NEMA MRI standards and ISO 10993 biocompatibility requirements.
Indications for Use
Indicated for patients undergoing neurosurgical procedures or diagnostic MR imaging requiring head stabilization and/or MR data collection of the brain. Used in standard ORs, diagnostic MR rooms, or intra-operative MR suites.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Intra-operative Coil and Head Fixation Device (major component of Neuro II-SE Intra-operative Imaging System) (K071099)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Monteris Medical. The logo consists of a stylized graphic element to the left of the text "MONTERIS" in all caps. Below "MONTERIS" is the word "MEDICAL" in a smaller font size. The logo is in black and white.
#### Section 5: 510(k) Summary
APR - 8 2011
KI10411
### Device Information:
| Category | Comments |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Monteris Medical, Inc.<br>100 - 78 Innovation Drive<br>Winnipeg, Manitoba<br>CANADA R3T 6C2<br>Tel: 204-272-2220<br>Fax: 204-272-2219<br>www.monteris.com |
| Establishment Registration Number | 3005840757 |
| Correspondent Contact Information: | K. Jeff Wilson, Ph.D.<br>Monteris Medical, Inc.<br>100 - 78 Innovation Drive<br>Winnipeg, MB R3T 6C2 CANADA<br>Tel: 204-272-2220<br>Fax: 204-272-2219 |
| Device Common Name: | Head Coil and Stabilization System |
| Device Classification & Product Code: | Class II, MOS |
| Device Classification Name & Citation | Magnetic resonance diagnostic device<br>21CFR892.1000 |
| Device Proprietary Name: | Monteris Medical AtamA™ System |
### Predicate Devices Information:
| Predicate Device: | Noras OR Head Coil 1.5 T |
|----------------------------------------|-------------------------------------------------------|
| Manufacturer: | Siemens Medical Solutions USA, Inc. (Noras) |
| K# | K060758 |
| Common Name: | Noras OR Head Coil 1.5 T |
| Device Classification Name & Citation: | Magnetic resonance diagnostic device<br>21CFR892.1000 |
| Device Classification & Product Code: | Class II, MOS |
| Predicate Device: | Noras OR Head Holder |
|----------------------------------------|----------------------------------------------------------|
| Manufacturer: | Siemens Medical Solutions USA, Inc. (Noras) |
| K# | K071179 |
| Common Name: | OR Head Holder |
| Device Classification Name & Citation: | Neurosurgical head holder (skull clamp)<br>21CFR882.4460 |
| Device Classification & Product Code: | Class II, HBL |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Monteris Medical. The logo consists of a stylized graphic to the left of the company name, "MONTERIS", with the word "MEDICAL" underneath. The graphic is a black, abstract shape that resembles a stylized "M" or a mountain range with a curved element on top.
AtamA™ Premarket Notification: Traditional 510(k)
| Predicate Device: | Intra-operative Coil and Head Fixation<br>Device (major component of Neuro II-SE<br>Intra-operative Imaging System) |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | IMRIS |
| KH | Major Component of K071099 |
| Common Name: | Head Coil Kit |
| Device Classification Name & Citation: | Magnetic resonance diagnostic device<br>21CFR892.1000 |
| Device Classification & Product Code: | Class II, LNH |
# b. Date Summary Prepared - Mage Summ 10.Feb.2011
# c. Description of Device of Device
The Monteris medical AtamA™ System is comprised of two major subsystems: (i) a MR head coil functionality, and (ii) a patient stabilization functionality. These two subsystems have been designed to work independently and in unison to meet the varied needs of neurosurgical procedures utilizing MRI.
The Head Coil Subsystem is comprised of a novel design of a standard split-array multichannel (8 channel) receive coil in the form of a common Lower Coil and two interchangeable Upper Coils (only one used at any time) and the appropriate coil configuration data, known as "coil files", for the selected MR system to be used.
The design of the two-part Head Coil system is such that, in conjunction with the rest of the AtamA System, openings in the two-part coil can be rotated around the patient's head to allow surgical access to essentially any supratentorial aspect of the patient's head, and still allow for the collection of MR data without repositioning the patient or moving the coils within the MR system. This will allow for more expedient collection of high quality MR data, minimize moving the patient, and maintenance of spatial registration (required for or supporting stereotactic procedures).
The Stabilization subsystem is comprised of three major components and their accessories. The major components are:
- A Head Fixation Ring (HFR) apparatus, .
- . A Cradle that fixates the HFR and the two-part Coil relative to each other and the MRI system bore, and
- . A patient Board that allows for the effective transport of the patient to and from various procedures and procedure locations while maintaining the stability and spatial registration of the patient.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Monteris Medical. The logo consists of a stylized graphic to the left of the company name. The word "MEDICAL" is printed in a smaller font size below the company name.
## Intended Use . . . . . . . .
The intended use of the AtamA™ System (comprised of two major subsystems: a MR head coil apparatus and a patient stabilization apparatus), in conjunction with a magnetic resonance (MR) imaging system, is the collection of MR data and images of the human brain before, during, and at the end of brain surgery, in a standard operating room, diagnostic MR rooms, or in a MR intra-operative room, while stabilizing the patient's head during neurosurgical procedures and imaging.
The patient head-stabilizing apparatus can also be used for other neurosurgical procedures requiring the patient's head to be stabilized or fixed.
The head coil can also be used as a standard diagnostic head coil for diagnostic MR imaging.
When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
The intended use of the AtamA System is essentially the same as the intended use of the predicate devices, as are the indications for use.
# Technological Characteristics
The technical modes of action and technical principles are materially the same as the predicate devices, including: split array 8-channel MR head coil design and construction, materials being functionally similar in all cases (identical in several aspects), and relying upon the same principles of head stabilization/immobilization. The design differences are in how the AtamA System utilizes and integrates these technical characteristics in an expedient and convenient system.
Non-Clinical Performance Data
Performance testing was performed to recognized standards for the Coil functionality and to actual values of predicate devices measured in parallel of predicate devices (when available), or to values reported by the manufacturers of the predicates. NEMA Magnetic Resonance Imaging standards, where appropriate were followed. Biocompatibility, per ISO 10993, was demonstrated.
## Clinical Performance Data - - - - - - - - - - - - - - - - - - - - - - - - - - - -
This application does not rely on any clinical investigations or clinical performance testing in support of substantial equivalence to the stated predicate devices.
### Summary of Performance Data
Biocompatibility analysis demonstrates that the AtamA system is in compliance with ISO 10993.
Bench testing has demonstrated that the device is in compliance with the product specification, the expectations of the medical community and the product labeling.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
K. Jeff Wilson, Ph.D. Vice President. Clinical and Regulatory Affairs Monteris Medical 100-78 Innovation Drive Winnipeg, MB R3T 6C2 CANADA
APR - 8 2011
Re: K110411
Trade/Device Name: Monteris Medical AtamATM System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: February 10, 2011 Received: February 14, 2011
Dear Dr. Wilson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Monteris Medical. The logo consists of a stylized graphic to the left of the word "MONTERIS" in all caps. Below "MONTERIS" is the word "MEDICAL" in a smaller font size, also in all caps. The logo is black and white.
#### SECTION 4: INDICATIONS FOR USE STATEMENT
| | Indications for Use |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known): | K110411 |
| Device Name: | Monteris Medical AtamA™ System |
| Indications for Use: | The intended use of the AtamA™ System (comprised of two major subsystems: a MR head coil apparatus and a patient stabilization apparatus), in conjunction with a magnetic resonance (MR) imaging system, is the collection of MR data and images of the human brain before, during, and at the end of brain surgery, in a standard operating room, diagnostic MR rooms, or in a MR intra-operative room, while stabilizing the patient's head during neurosurgical procedures and imaging. The patient head-stabilizing apparatus can also be used for other neurosurgical procedures requiring the patient's head to be stabilized or fixed. The head coil can also be used as a standard diagnostic head coil for diagnostic MR imaging. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. |
| Prescription Use | X AND/OR Over-The-Counter Use |
| (Part 21 CFR 801 Subpart D) | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ___
Mary S Pastel
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.