K060758 · Siemens Medical Solutions USA, Inc. · MOS · May 5, 2006 · Radiology
Device Facts
Record ID
K060758
Device Name
OR HEAD COIL 1.5 T
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MOS · Radiology
Decision Date
May 5, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of the Noras OR head coil 1.5 T is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the human brain just before, during and at the end of the brain surgery in the operating room. The coil can also be used as a standard diagnostic head coil for diagnostic examinations and fMRI (Functional Magnetic Resonance Imaging). Used in the 1.5T MAGNETOM Systems, it is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Device Story
OR Head Coil 1.5T is an 8-channel phased array coil designed for 1.5T MRI systems; consists of top and bottom arrays (4 channels each) tuned to 63.6 MHz (1H spin). Used in operating rooms for brain surgery imaging and standard diagnostic/fMRI head exams; requires Siemens OR Head Holder. Device captures MR signals from the head; transmits data via cables to MAGNETOM system for image reconstruction. Physician interprets resulting transverse, sagittal, coronal, and oblique images to assist in clinical diagnosis. Benefits include enabling intra-operative visualization of brain structures during neurosurgical procedures.
Clinical Evidence
Bench testing only. Performance and safety parameters verified against NEMA standards and IEC 60601-2-33:2002. Testing conducted on syngo MR2004A software platform, with risk analysis confirming applicability to syngo MR2006A and MAGNETOM 1.5T systems.
Technological Characteristics
8-channel phased array coil; 1.5T magnetic field strength; 63.6 MHz resonant frequency. Comprised of top and bottom arrays. Connectivity via dedicated cables to MAGNETOM MRI systems. Conforms to NEMA standards and IEC 60601-2-33:2002.
Indications for Use
Indicated for patients requiring MR examination of the human brain, including pre-operative, intra-operative, and post-operative brain surgery assessment, as well as standard diagnostic head examinations and fMRI.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MRI Devices Corporation (Now Known as Invivo Corporation) High Resolution Head Coil -Model HRH-63-8 (K013159)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "NORAS" in large, bold, black letters. The letters are slightly slanted to the right. There is a black line underneath the word, and the bottom of the image is cut off.
Image /page/0/Picture/1 description: The image shows a logo with the words "RÖNTGEN- UND MEDIZINTECHNIK" in bold, black letters. Below the text is a thick, black line. Underneath the line, there is some illegible text that appears to be handwritten, with the number 0758 being the only part that is legible.
### 510(k) Summary
## OR Head Coil 1.5T
MAY - 5 2006
Date of Summary Preparation: March 15, 2006
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
- General Information 1.
- Importer/Distributor Name and Address Siemens Medical Solutions, Inc. 51 Valley Stream Parkway Malvern, PA 19355
Establishment Registration Number 2240869
Manufacturing Site Name and Address Noras Röntgen- und Medizintechnik GmbH Leibnizstrasse 4 97204 Höchberg Germany
Establishment Registration Number Establishment Registration 3004929307
Owner/Operator 9071737
#### Contact Person 2.
Judy Campbell Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, E-50 Malvern, PA 19355 Telephone: (610) 448-4918 (610) 448-1787 Fax: Email Address: judith.campbell@siemens.com
CONFIDENTIAL
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "NORAS" in large, bold, black letters. Below the word, there is some text that is partially obscured. The text appears to be related to the word "NORAS", possibly providing a definition or context. The overall image has a high-contrast look, with the black text standing out against a lighter background.
Image /page/1/Picture/1 description: The image shows the words "RÖNTGEN- UND MEDIZINTECHNIK" in bold, black font. The words are stacked on top of each other, with "RÖNTGEN- UND" on the top line and "MEDIZINTECHNIK" on the bottom line. The background is white.
Trade Name: Common Name: Classification Name: Classification Panel: CFR Number: Device Class: Product Code:
OR Head Coil 1. OR Head Coil 1.5 Magnetic Resonance Diagnostic Device Radiology 21 CFR § 892.1000 ============================================================================================================================================================================== 90MOS
#### 4. Device Description
The OR Head Coil 1.5T is an 8-channel phased array coil. The coil is divided into a bottom and a top array of 4 channels each. Each channel is tuned to the Larmor frequency of the 1H spin in a magnetic field of 1.5T, 63.6 MHz. Each coil is connected to the MAGNETOM system by a separate cable.
Image /page/1/Picture/6 description: The image shows two curved, gray objects with multiple holes. Each object has a cable attached to it, and the cables are connected to rectangular connectors. The connectors appear to be of different types. The background is black, which makes the objects stand out.
Use of the OR Head Coil 1.5T requires the Siemens OR Head Holder. The pictures below show the complete set.
Image /page/1/Picture/8 description: The image contains two photos of a football helmet. The photo on the left shows the helmet from the side, and the photo on the right shows the helmet from the front. The helmet is dark in color and has a face mask. There are some white dots on the top of the helmet.
Traditional 510(k) Noras Rontgen- Und Medizintechnik GmbH OR Head Coil 1.5T March 16, 2006
CONFIDENTIAL
000 05
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the text "RÖNTGEN- UND MEDIZINTECHNIK" in bold, black letters. The text is stacked in two lines, with "RÖNTGEN- UND" on the top line and "MEDIZINTECHNIK" on the bottom line. To the left of the text is a stylized graphic that resembles a triangle with a horizontal line extending from its base. The logo appears to be for a company or organization related to X-ray and medical technology.
The intended use of the Noras OR Head Coil ST is the MR examination of the human brain just before, during and at the end of the brain surgery in the operating room. It can also be used as a standard diagnostic head coil for diagnostic examinations and fMRI.
#### 6. Substantial Equivalence
Noras and Siemens believe that, within the meaning of the Safe Medical Devices Act of 1990, the 8-channel phased array OR head coil 1.5T for the Magnetom Systems is substantially equivalent to the following coil:
| Coil Name | Pemarket Notification | Clearance Date |
|--------------------------------------------------------------------------------------------------------------------|-----------------------|------------------|
| MRI Devices<br>Corporation (Now<br>Known as Invivo<br>Corporation)<br>High Resolution Head<br>Coil -Model HRH-63-8 | K013159 | October 16, 2001 |
### Summary of Technological Characteristics of the Principal Device as 7. Compared with the predicate Device
Although the MRI Devices High Resolution Head Coil is limited for use outside the operating room, we believe that both coils are substantially equivalent.
#### General Safety and Effectiveness Concerns 8.
The OR Head Coil 1.5T will conform to the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the IEC standards for safety issues with the Magnetic Resonance Imaging Devices, IEC 60601-2-33: 2002. This will assure that the performance of this device can be considered safe and effective when used with the currently available MAGNETOM 1.5T systems.
The NEMA Tests can be found in Appendix C. The NEMA Tests were done on Software platform syngo MR2004A. We have included the Common Risk Analysis for syngo MR2006A, which is the same functionality as syngo platform MR 2004A. We believe the NEMA tests on the Symphony are applicable to all MAGNETOM 1.5T systems.
Traditional 510(k) Noras Rontgen- Und Medizintechnik GmbH OR Head Coil 1.5T March 16, 2006
CONFIDENTIAL
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NORAS Röntgen- und Medizintechnik. The word "NORAS" is in large, bold, black letters. To the right of "NORAS" is a black line that forms a V shape, and to the right of that is the text "RÖNTGEN- UND MEDIZINTECHNIK" in smaller, bold, black letters. A black line underlines the text.
#### Conclusion as to Substantial Equivalence 9.
Noras and Siemens believe that, within the definition of the Safe Medical Devices Act of 1990, the Noras 8-channel phased array OR head coil 1.5T is substantially equivalent to the predicate device listed above.
Auebert Wares
Hubert Noras President
March 16, 2006
Traditional 510(k) Noras Rontgen- Und Medizintechnik GmbH OR Head Coil 1.5T March 16, 2006
CONFIDENTIAL
000 07
7
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
MAY - 5 2006
Ms. Judith Campbell Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway, E-50 MALVERN PA 19355
Re: K060758
Trade/Device Name: Noras OR Head Coil 1.5 T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 16, 2006 Received: March 21, 2006
Dear Ms. Campbell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that, FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation. control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation prumber at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known):
Device Name:
Noras OR Head Coil 1.5 T
Indications for Use:
The intended use of the Noras OR head coil 1.5 T is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the human brain just before, during and at the end of the brain surgery in the operating room. The coil can also be used as a standard diagnostic head coil for diagnostic examinations and fMRI (Functional Magnetic Resonance Imaging).
Used in the 1.5T MAGNETOM Systems, it is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head.
When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONT NUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--------|
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | ✓ |
| | AND/OR |
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
*(Division Sign-Off)*
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K060258 |
|---------------|---------|
| Page | of |
Traditional 510(k)
Noras Rontgen- Und Medizintechnik GmbH
CONFIDENTIAL
Image /page/6/Picture/9 description: The image shows the numbers 000 02 in a bold, sans-serif font. The numbers are printed in black ink on a white background. The numbers are slightly blurred, and there is some noise in the image. The right side of the image is obscured by a dark, blurry shape.
March 16, 200€
**CONFIDENTIAL**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.