K110332 · Clearcanvas, Inc. · LLZ · Mar 24, 2011 · Radiology
Device Facts
Record ID
K110332
Device Name
CLEARCANVAS RIS / PACS
Applicant
Clearcanvas, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 24, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The ClearCanvas RIS/PACS is an image management system whose intended use is to provide scaleable DICOM compatible PACS solutions for hospitals and related institutions and sites, which will archive, distribute, retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR, and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM "For Presentation", and will include standard features and other tools for analyzing mammography images. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega-pixel resolution and meets other technical specifications approved by the FDA.
Device Story
ClearCanvas RIS/PACS is a software-based Picture Archiving and Communication System (PACS) for medical image management. It ingests DICOM-compliant images from various modalities and HL7-compliant patient data from information systems. The system comprises interrelated components: Workstation (client for display/analysis), Webstation (browser-based viewer), ImageServer (storage/distribution), RIS (workflow management), IntegrationServer (HL7 interoperability), and EnterpriseServer (authentication/auditing). Used in radiology reading rooms, clinics, and referring physician offices by trained professionals. It provides tools for image enhancement, manipulation (W/L, zoom, pan, rotation, fusion, MPR), and annotation. Output is displayed on clinical monitors to assist healthcare providers in diagnostic decision-making. The system operates on generic PC hardware, facilitating scalable, enterprise-wide access to medical imaging information.
Clinical Evidence
Bench testing only. The device underwent non-clinical software verification and validation testing, including human factors testing under simulated use conditions. Testing confirmed compliance with DICOM and JPEG standards and verified that design outputs met design inputs. No clinical data was required or provided.
Technological Characteristics
Software-only PACS; operates on off-the-shelf PC hardware. Architecture: client-server (Workstation, Webstation, ImageServer, RIS, IntegrationServer, EnterpriseServer). Connectivity: DICOM (PS 3.3-3.12) and HL7 compliant. Image processing: compression, magnification, value-of-interest manipulation, orientation, fusion, multiplanar reformat. Standards: ISO 13485:2003, ISO 62304. No patient contact; no life-sustaining function. OS: Windows.
Indications for Use
Indicated for hospitals and clinical institutions requiring management, archiving, retrieval, and display of DICOM-compliant medical images (CR, CT, DR, MR, etc.) and associated data. Includes mammography image display and analysis tools. Contraindicated for primary interpretation of lossy compressed mammography or digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K110332
# MAR 2 4 2011
.
# 510(k) Summary of Safety and Effectiveness
### Date Prepared
November 22, 2010
# Submitter
.
:
ClearCanvas Inc. 439 University Ave., Suite 1920 Toronto, ON M5G 1Y8 " | ways . : が、「私 .. . " -、" Canada
・・
# Contact Person '
Clinton Chau Email: clinton@clearcanvas.ca Phone: (647) 547-2987 Fax: (866) 584-5688
# Device Name and Classification
| Trade Name | ClearCanvas RIS/PACS |
|----------------------------|----------------------------------------------------|
| Common Name | Picture Archiving and Communications System (PACS) |
| Device Classification Name | System, Image Processing, Radiological |
| Device Class | Class II |
| Regulation Number | 892.2050 |
| Product Code | LLZ |
| Classification Panel | Radiology |
### Predicate Devices
| Trade Name | BRIT PACS SYSTEMS | IntelePACS(tm) |
|-----------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Model Number | -- | -- |
| Common Name | Picture Archiving and<br>Communication System (PACS) | Picture Archiving and<br>Communication System (PACS) |
| 510(k) Submitter/<br>Holder | BRIT Systems Inc.<br>1909 Hi-Line Dr.<br>Dallas, TX 75207 | Intelerad Medical Systems Inc.<br>Anibal Jodorcovsky<br>460 Ste-Catherine West, Suite 210<br>Montreal, QC Canada H3B 1A7 |
| 510(k) Number | K081168 | K070080 |
| Regulation Number | 892.2050 | 892.2050 |
| Panel | Radiology | Radiology |
| Product Code | LLZ | LLZ |
{1}------------------------------------------------
#### Device Description and Intended Use
ClearCanvas RIS/PACS is a Picture Archiving and Communication (PACS) software system for the management and review of medical image data, and other digital images. Such data can be received from all DICOMcompliant imaging modalities or imported directly into the system. This data can then be stored, archived, distributed, processed, enhanced, and displayed for review. Enhancement, processing and analysis of the image data includes but is not limited to compression, magnification, value-of-interest manipulation, orientation changes, image fusion and multiplanar reformat. Additional health data, such as patient demographics, from other information systems can be received from HL7-compliant sources.
The ClearCanvas RIS/PACS is a system of interrelated software components:
- . ClearCanvas Workstation – a desktop client that is primarily used to retrieve, display, enhance and analyze medical images. Access to patient demographic data is also possible through the Workstation when working with the RIS server.
- . ClearCanvas Webstation – a browser-based web application that allows the user to retrieve images from the ImageServer for review. Primarily intended for referring physicians and as an image viewer for EMR systems, the images are JPEG-compressed.
- . ClearCanvas ImageServer – a server application that receives images from imaging modalities, stores and archives them, and distributes them to client applications.
- . ClearCanvas RIS – a server application that acts as an online transaction processor for the management of patient and workflow information.
- . ClearCanvas IntegrationServer – a server application composed of optional modules that implement integration and interoperability, such as an inbound and outbound HL7 processor.
- . ClearCanvas EnterpriseServer -- a server application that provides enterprise-wide facilities such as user authentication, authorization and auditing.
Working in concert or stand-alone, each component plays a valuable role in the management information in the radiology reading room, the clinic, the referring physician's office, or any other patient care setting where access to medical imaging information is important.
ClearCanvas RIS/PACS software is primarily written in C#, employing the most current best practices in objectoriented and component-oriented software architecture resulting in a highly scalable and extensible design. The system is also designed to work on generic PC hardware that meets the minimum system requirements.
#### Indications for Use
The ClearCanvas RIS/PACS is an image management system whose intended use is to provide scaleable DICOM compatible PACS solutions for hospitals and related institutions and sites, which will archive, retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR, and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM "For Presentation", and will include standard features and other tools for analyzing mammography images.
Page 2 of 8
{2}------------------------------------------------
Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at ieast 5 mega-pixel resolution and meets other technical specifications approved by the FDA.
## Technological Characteristics
The ClearCanvas RIS/PACS has no patient contact and does not control any life sustaining devices. It is a software PACS device used in the management of medical imaging information. This device is used by trained and qualified professionals affording ample opportunity for competent human intervention in interpreting the images and information presented. These technological characteristics are identical to those in the predicate devices. Furthermore, the following comparison table summarizes the specific characteristics of the ClearCanvas RIS/PACS and its predicate devices that demonstrate substantial equivalence:
| | Trade Name | Subject Device<br>ClearCanvas RIS/PACS | Predicate Device<br>BRIT PACS SYSTEM | Predicate Device<br>IntelePACS | |
|----|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--|
| | Intended Use | | | | |
| 1 | To be used as a Picture Archive and<br>Communication System (PACS) | Y | Y | Y | |
| 2 | To be used for storing and<br>distributing images | Y | Y | Y | |
| 3 | To be used for displaying,<br>analyzing, manipulating and<br>enhancing of images | Y | Y | Y | |
| | Indications For Use Statement | | | | |
| 4 | Statement uses the term, “picture<br>archive and communication<br>system” or “PACS” | Y | Y | Y | |
| 5 | Statement references the display<br>of medical images | Y | Y | Y | |
| 6 | Statement references receiving<br>images from image sources | Y | Y | Y | |
| | | Subject Device | Predicate Device | Predicate Device | |
| 7 | Statement references multi-planar reformat | N | N | Y | |
| 8 | Statement references mammographic images | Y | Y | Y | |
| | | Standards Met | | | |
| 9 | Conforms to the essential requirements of the DICOM standard for data exchange (PS 3.3-PS 3.12) | Y | Y | Y | |
| 10 | Conforms to the JPEG standard for image compression of digital medical images | Y | Y | Y | |
| | | Design | | | |
| 11 | Software device that operates on off-the-shelf hardware | Y | Y | Y | |
| 12 | Device is a system composed of multiple components that work stand-alone or in concert | Y | Y | Y | |
| 13 | Device components are organized in a client-server architecture | Y | Y | Y | |
| 14 | Medical image viewer, the client application, is a key component of the device system | Y | Y | Y | |
| 15 | Image storage and archive server, the server application, is a key component of the device system | Y | Y | Y | |
| | Subject Device | Predicate Device | Predicate Device | | |
| | Functions and Capabilities | | | | |
| 16 | Query, import, send, receive<br>DICOM images | Y | Y | Y | |
| 17 | Image analysis tools W/L, zoom,<br>pan, stack, rotation, flip | Y | Y | Y | |
| 18 | Measurement and annotation | Y | Y | Y | |
| 19 | Multi-monitor awareness | Y | Y | Y | |
| 20 | Image layout | Y | Y | Y | |
| 21 | Image Thumbnails | Y | Y | Y | |
| 22 | Reference lines and spatial locator<br>for tomographic images | Y | Y | Y | |
| 23 | User annotations of images | Y | Y | Y | |
| 24 | Synchronized stacking | Y | Y | Y | |
| 25 | Probe tool | Y | N | Y | |
| 26 | Shutters | Y | N | Y | |
| 27 | Key Image marking | Y | Y | Y | |
| 28 | Multiplanar Reformat (MPR) | Y | Y | Y | |
| 29 | Lossy and lossless image<br>compression | Y | Y | Y | |
| 30 | Automatic routing of images | Y | Y | Y | |
| 31 | Image archiving | Y | Y | Y | |
| | | Subject Device | Predicate Device | Predicate Device | |
| 32 | Patient information and workflow management components | Y | Partial | Y | |
| 33 | Image streaming from server | Y | Y | N | |
| 34 | Operating System/Platform | Windows | Linux and Windows | Linux and Windows | |
| 35 | DICOM image printing | Y | Y | Y | |
| 36 | Export DICOM images to optical media | Y | Y | Y | |
| 37 | PET/CT fusion | Y | Y | N | |
| | | Other Areas of Comparison | | | |
| 38 | Target population | No restrictions | No restrictions | No restrictions | |
| 39 | Anatomical sites | No restrictions | No restrictions | No restrictions | |
| 40 | Environment for use | No restrictions; to be used by trained professionals only | No restrictions; to be used by trained professionals only | No restrictions; to be used by trained professionals only | |
| 41 | Energy used and/or delivered | Electric power supply to computer hardware only | Electric power supply to computer hardware only | Electric power supply to computer hardware only | |
| 42 | Human factors | Standard windows-based or web-based graphical user interfaces, with standard input devices | Standard windows-based or web-based graphical user interfaces, with standard input devices | Standard windows-based or web-based graphical user interfaces, with standard input devices | |
| | | Subject Device | Predicate Device | Predicate Device | |
| 43 | Performance | Real-time performance requirements not applicable | Real-time performance requirements not applicable | Real-time performance requirements not applicable | |
| 44 | Materials | Not applicable | Not applicable | Not applicable | |
| 45 | Biocompatibility | Not applicable | Not applicable | Not applicable | |
| 46 | Compatibility with environment and other devices | Interoperation with other devices based on consensus standards on data exchange (DICOM) | Interoperation with other devices based on consensus standards on data exchange (DICOM) | Interoperation with other devices based on consensus standards on data exchange (DICOM) | |
| 47 | Sterility | Not applicable | Not applicable | Not applicable | |
| 48 | Electrical safety | Not applicable | Not applicable | Not applicable | |
| 49 | Mechanical safety | Not applicable | Not applicable | Not applicable | |
| 50 | Chemical safety | Not applicable | Not applicable | Not applicable | |
| 51 | Thermal safety | Not applicable | Not applicable | Not applicable | |
| 52 | Radiation safety | Not applicable | Not applicable | Not applicable | |
| | Product Labelling | | | | |
| 53 | Instructions for searching and retrieval of images | Y | Y | Y | |
| 54 | Instructions for displaying of images | Y | Y | Y | |
| | Subject Device | Predicate Device | Predicate Device | | |
| 55 | Instructions for manipulating<br>orientation of images | Y | Y | Y | |
| 56 | Instructions for enhancing images | Y | Y | Y | |
| 57 | Instructions for annotating images | Y | Y | Y | |
Page 3 of 8
.
{3}------------------------------------------------
. ·
Page 4 of 8
{4}------------------------------------------------
.
.
.
.
Page 5 of 8
.
.
{5}------------------------------------------------
·
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
.
. .
{6}------------------------------------------------
.
.
.
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
.
Page 7 of 8
{7}------------------------------------------------
#### Testing
The ClearCanvas RIS/PACS complies with the voluntary DICOM and JPEG standards for device performance and is designed and manufactured according to the process and management system standards set in ISO 13485:2003 and ISO 62304, as discussed in the 510(k) submission.
Performance testing was conducted to verify the design output met the design input requirements and to validate the device conforms to the defined user needs and intended uses. Nonclinical software testing, as well as human factors testing, was conducted under simulated use conditions. Predefined acceptance criteria were met and demonstrated that the device is as safe and as effective as the predicate devices.
#### Substantial Equivalence Conclusions
ClearCanvas concludes that the intended use for the ClearCanvas RIS/PACS is the same as that of the predicate devices, and that the technological characteristics demonstrate that they are equivalent to the predicate devices. A comparison of the technological characteristics of the predicate and legally marketed devices available has been performed.
#### Conclusion
The 510(k) Pre-Market Notification for ClearCanvas RIS/PACS contains adequate information and data to determine that ClearCanvas RIS/PACS is as safe and effective as the legally marketed predicate devices
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
ClearCanvas. Inc. % Ms. Paula Wilkerson Sr. Staff Engineer-Medical Devices Intertek Testing Services 2307 East Aurora Rd., Unit B7 TWINSBURG OH 44087
MAR 2 4 2011
Re: K110332
Trade/Device Name: ClearCanyas RIS/PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 15, 2011 Received: March 16, 2011
Dear Ms. Wilkerson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{9}------------------------------------------------
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k), remarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, perceits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (01) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of advertal Reents under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): _K110332
Device Name: ClearCanvas RIS/PACS
Indications for Use:
The ClearCanvas RIS/PACS is an image management system whose intended use is to provide scaleable DICOM compatible PACS solutions for hospitals and related institutions and sites, which will archive, distribute, retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR, and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM."For Presentation", and will include standard features and other tools for analyzing mammography images.
Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega-pixel resolution and meets other technical specifications approved by the FDA.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Patel
Division Sign-Öff Office of In Vitro Diagnostic Device Evaluation and Safety 10332
. ...
... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.