K081168 · Brit Systems, Inc. · LLZ · Mar 23, 2009 · Radiology
Device Facts
Record ID
K081168
Device Name
BRIT PACS SYSTEMS
Applicant
Brit Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 23, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The BRIT PACS Systems is an image management system whose intended use is to provide (scaleable ) DICOM compatible PACS solutions for hospital and related institutions/sites, which will archive/ distribute/ retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM "For Presentations, and will include standard features and other Mammo tools. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA. Application areas include imaging centers, radiologist central reading rooms and any other locations where trained medical professionals would require access or desire patient images, demographic information or other patient medical information captured in the system.
Device Story
BRIT PACS Systems is a modular, DICOM-compliant image management solution for hospitals and clinics. It ingests medical images from modalities (CR, CT, DR, MR) and information systems. Components include: BRIT SCAN (film digitizer), BRIT Quality Workbench (QC), BRIT Roentgen RIS (scheduling/exam management), BRIT Roentgen Files (DICOM/Web server, archive, HL-7 interface), and various viewing workbenches (BRIT Vision, BRIT View, BRIT Lite). The system enables remote reading, image distribution, and diagnostic viewing. Radiologists and clinicians use the workstations to view, manipulate, and interpret images. The system supports mammography via 5MP monitors. It facilitates clinical decision-making by providing access to patient images and data across multiple locations. Benefits include improved workflow, scalability, and timely access to diagnostic information for healthcare providers.
Clinical Evidence
No clinical data provided. The submission relies on a comparison of technological characteristics and intended use to established predicate devices.
Technological Characteristics
DICOM-compliant PACS architecture; modular software components; PC-based workstations; supports 5MP monitors for mammography; HL-7 interface engine; web-based client for RIS; DICOM/Web server; film digitization hardware. Connectivity via network/DICOM standards.
Indications for Use
Indicated for use in hospitals, imaging centers, and reading rooms by trained medical professionals to archive, distribute, retrieve, and display medical images and data from modalities including CR, CT, DR, and MR. Includes display of structured reports and mammography images. Contraindications: Lossy compressed mammography images and digitized film screen images must not be used for primary interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# MAR 2 \$ 2009
Ka81168
Page 1 of 3
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21CFR 807.92)
### DATE PREPARED: 4/21/2008 I.
- II. SUBMITTER: BRIT Systems, Inc. 1909 Hi-Line Dr.
CONTACT PERSON: III.
Dallas, TX 75207
Robert Murry VP of Engineering bob@brit.com Phone: 214-630-0636 214-630-1638 Fax:
#### IV. DEVICE NAME:
Classification Name: Trade/Proprietary Name: Picture Archiving and Communications System (PACS) BRIT PACS SYSTEMS
- V. DEVICE CLASSIFICATION Class II CFR section: 892.2050 Product code: LLZ Panel: Radiology
#### VL PREDICATE DEVICES:
K062477 AMICAS Vision Series PACS Class II Decision Date: 10/27/2007
K041935 DR Systems PACS Class II Decision Date: 08/16/2004
{1}------------------------------------------------
KC81/68
Page 2 of 3
### SUBSTANTIAL EQUIVALENCE CONCLUSIONS: VII.
BRIT Systems concludes that the intended use for the BRIT PACS Systems is the same as that of the predicate devices, and that the technological characteristics demonstrate that they are equivalent to the predicate device. A comparison of the technological characteristics of the predicate and legally marketed devices available has been performed.
Thus, this premarket notification has demonstrated substantial equivalence.
# VIII. DEVICE DESCRIPTION AND INTENDED USE:
### System Description:
BRIT PACS Systems provides custom, turn-key solutions for PACS and teleradiology based systems on the latest DICOM standards. Using PACS, medical images can be made available at multiple locations - virtually anywhere at any time. Modular components make BRIT's solutions flexible, upgradeable, and scaleable.
BRIT SCAN is a high-quality film digitization system that allows standard radiographs to be converted into electronic images for transmission to PACS and teleradiolgy systems. BRIT Scan allows the users to integrate standard radiographic studies, which are often used for comparison studies, into the PACS strategy.
BRIT Quality Workbench is a DICOM based application that runs on a PC and is used to provide patient demographic and image data quality control functions within the PACS system.
BRIT Roentgen RIS provides a hospital or clinic with automated tools to electronically schedule and manage patient exam information. The Roentgen RIS utilizes a web based client that provides referring physicians the added flexibility to instantly view and schedule their own patients' radiographic services. The Roentgen RIS also accesses the same database as BRIT'S PACS solution.
BRIT Roentgen Files is the main product for exam and report storage in a PACS environment. It runs a DICOM Server and Web Server plus a selection of the following applications, all using the same database.
- DICOM Archive �
- Modality Worklist Server �
- HL-7 Interface Engine �
- Order Entry/Scheduling package ◆
BRIT Viewing Workbenches offer different viewing applications to address the differing needs of the users.
BRIT Vision is a diagnostic reading workbench which provides radiologists and high-power users a fully configured 12-bit DICOM image viewing workstation.
BRIT Vision Mammography Module is an optional diagnostic module utilizing 5MP monitors that runs in the BRIT Vision medical image viewing workstation.
BRIT View is a clinical viewing workstation or diagnostic reading workbench which provides high quality image viewing and manipulation capabilities in a fully configured 12-bit DICOM workstation, for use in radiology or non-radiological area of the facility such as ER, Surgery, ICU, and other areas.
1239
{2}------------------------------------------------
BRIT Lite is a viewing application that responds to the needs of healthcare organization that require access to images anywhere, anytime, but without the need for advanced visualization tools or large application on each workstation.
BRIT Roentgen Burner is a DICOM compliant automated solution for producing custom-labeled CD's and or DVD's that contain medical images plus BRIT's image viewing software. Radiological studies are sent to this device from any BRIT DICOM workstation on the network.
BRIT Roentgen Works is a workflow management service for radiologists to provide remote reading services for multiple health care facilities.
### Indications for Use:
The BRIT PACS Systems is an image management system whose intended use is to provide (scaleable) DICOM compatible PACS solutions for hospital and related institutions/sites, which will archive/ distribute/ retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR, and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM "For Presentation", and will include standard features and other Mammo tools. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Application areas include imaging centers, radiologist central reading rooms, and any other locations where trained medical professionals would require access or desire patient images, demographic information, or other patient medical information captured in the system.
#### IX. SAFETY INFORMATION:
The BRIT PACS Systems has no patient contact and is utilized only by trained professionals. Trained professionals allow sufficient review to afford identification and intervention in the event of a malfunction have evaluated the output of the device. Patient data is limited to authorized individuals.
#### X. CONCLUSION:
BRIT PACS Systems believes sufficient information is included to reach a determination of substantial equivalence. We conclude that the subject device is as safe and effective including the component and accessory devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird figure.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 3 2009
Mr. Robert Murry VP of QA/Engineering BRIT Systems 1909 Hi Line Drive DALLAS TX 75207
Re: K081168
Trade/Device Name: BRIT PACS Systems Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications systems Regulatory Class: II Product Code: LLZ Dated: February 18, 2009 Received: February 19, 2009
### Dear Mr. Murry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. - The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K081168
MAR 2 3 2009
510(k) Number (if known): N/A Kosilla
Device Name: BRIT PACS Systems
### Indications for Use:
The BRIT PACS Systems is an image management system whose intended use is to provide (scaleable ) DICOM compatible PACS solutions for hospital and related institutions/sites, which will archive/ distribute/ retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM "For Presentations, and will include standard features and other Mammo tools. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Application areas include imaging centers, radiologist central reading rooms and any other locations where trained medical professionals would require access or desire patient images, demographic information or other patient medical information captured in the system.
# (PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) Over the Counter Use
Tozu In Whan
(Division Sigi Division of Reproductive, Abdominal and Radiological Davices 510(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.