V200 VENTILATOR WITH APRV MODE, ESPRIT VENTILATOR WITH APRV MODE
K110083 · Respironics California, Inc. · CBK · Jan 13, 2012 · Anesthesiology
Device Facts
Record ID
K110083
Device Name
V200 VENTILATOR WITH APRV MODE, ESPRIT VENTILATOR WITH APRV MODE
Applicant
Respironics California, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jan 13, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Esprit and V200 Ventilators are microprocessor controlled, electrically powered, mechanical ventilators. They are intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult, pediatric, and neonatal patients as prescribed by a physician. The Esprit and V200 Ventilators are intended for use in either invasive or non-invasive applications. The Esprit and V200 Ventilators with APRV Mode are intended for use for invasively ventilated adult and pediatric patients as prescribed by a physician
Device Story
Esprit and V200 are microprocessor-controlled, electrically powered mechanical ventilators. Modification adds Airway Pressure Release Ventilation (APRV) mode via software upgrade. APRV delivers gas via endotracheal or tracheostomy tube at two pressure levels (High/Low), allowing spontaneous or supported breathing at both levels. Operated by qualified medical personnel in clinical settings. Software upgrade installed via I-button (factory or field); adds GUI button for mode activation. Output provides ventilatory support; assists clinicians in managing patient respiration. Benefits include expanded ventilation options for invasive support in adult/pediatric populations.
Clinical Evidence
Bench testing and clinical simulation conducted. No clinical trial data on human subjects presented. Testing confirmed APRV mode performance and safety.
Technological Characteristics
Microprocessor-controlled, electrically powered mechanical ventilator. APRV mode implemented as software upgrade via I-button. GUI-based control. Firmware-based operation. No changes to hardware, ergonomics, or other ventilation modes.
Indications for Use
Indicated for invasive ventilatory support in adult and pediatric patients as prescribed by a physician. Excludes neonatal patients and non-invasive applications for the APRV mode.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
K110083
# 510(k) Summary
JAN 1 3 2012
[Refer to 21 CFR 807.92]
| Owner: | Respironics California, Inc.<br>2271 Cosmos Court<br>Carlsbad, CA 92011 | | |
|-------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------|
| Contact Person: | Tamatha Ley<br>Regulatory Affairs Specialist<br>Phone: (760) 918-1026<br>Fax: (760) 918-0169 | | |
| Date Prepared: | January 10, 2011 | | |
| Proprietary Name: | Esprit Ventilator with APRV Option<br>V200 Ventilator with APRV Option | | |
| Common Name: | Ventilator | | |
| Classification<br>Name: | Continuous Ventilator (21 CR 868.5895, Product Code 73 CBK) | | |
| Predicate Devices: | Manufacturer<br>Respironics California, Inc.<br>Respironics California, Inc<br>CareFusion | Device Name<br>Esprit Ventilator<br>V200 Ventilator<br>Avea Ventilator | 510(k) Number<br>K981072<br>K102054<br>K013642 |
### Intended Use of the Device:
The Esprit and V200 Ventilators are microprocessor controlled, electrically powered, mechanical ventilators. They are intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult, pediatric, and neonatal patients as prescribed by a physician. The Esprit and V200 Ventilators are intended for use in either invasive or non-invasive applications.
The Esprit and V200 Ventilators with APRV Mode are intended for use for invasively ventilated adult and pediatric patients as prescribed by a physician
The intended use is the same as that of the predicate devices, except that the APRV Mode is for use on a subset of the patient population the original devices are cleared for (e.g. neonatal patients and non-invasive applications are excluded).
## Device Description:
The Esprit and V200 Ventilators are microprocessor controlled, electrically powered, mechanical ventilators. This modification to the currently marketed Esprit Ventilator and V200 Ventilators is the addition of the APRV Mode.
{1}------------------------------------------------
The APRV Mode is an optional software upgrade. It is both a breath type and ventilation mode intended for invasively ventilated adult and pediatric patient populations. APRV enables the ventilator to deliver gas via an endotracheal tube or tracheostomy tube at two levels of pressure (Press High and Press Low), and allows for spontaneous or supported breathing at both levels.
The APRV Mode is activated via a software download through an I-button and is integrated into the Esprit and V200 Ventilators in the same way as other currently released software options. It can either be installed in the factory or in the field as an upgrade to existing Esprit and V200 ventilators. Downloading this option will add a "button" to the Graphical User Interface (GUI), which is used to turn APRV on and off.
## Substantial Equivalence:
The Esprit and V200 Ventilators with APRV Mode have the same intended use as the unmodified ventilators and similar design and technological characteristics as the other standard, currently marketed ventilators.
The APRV Mode incorporates a new mode of ventilation and a new breath type similar to the existing CareFusion Avea and unmodified Esprit and V200 Ventilators. All software activities, including verification and validation have been successfully completed in accordance with Respironics California, Inc. policies and procedures and the FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices dated May 11, 2005. In addition to the software verification and validation activities, performance testing and a clinical simulation were conducted and support the assertion that the APRV Mode does not raise any new questions regarding safety and effectiveness.
The technological characteristics of the currently marketed Esprit and V200 Ventilators with respect to the control mechanism, operating principle, energy type, ergonomics of the patient interface, firmware, environmental specifications, and performance specifications remain unchanged. Changes to the operational software were to include the APRV Mode only. No changes are being made to any other mode of ventilation. Therefore, all other modes of ventilation are identical to those in the currently marketed Esprit and V200 Ventilators.
This submission contains comparative information, including documentation related to the aforementioned activities to conclude that the Esprit and V200 with APRV Mode is substantially equivalent to currently marketed devices cleared by the FDA. These changes do not raise any new questions regarding safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, with three wavy lines representing the staff and a stylized head at the top.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 1 3 2012
Ms. Mary Funk Senior Manager, Regulatory Affairs Respironics California, Incorporated 2271 Cosmo Court Carlsbad, California 92130
Re: K110083
Trade/Device Name: Esprit Ventilator with Airway Pressure Release Ventilation (APRV) Mode, V200 Ventilator with Airway Pressure Release Ventilation (APRV) Mode
Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: January 9, 2012 Received: January 10, 2012
Dear Ms. Funk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Funk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Rh foc
t
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Esprit Ventilator with Airway Pressure Release Ventilation (APRV) Mode Device Name:
Indications for Use:
The Esprit Ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult, pediatric, and neonatal patients as prescribed by a physician. The Esprit Ventilator is intended for use in either invasive or non-invasive applications and contains the following modes of ventilation:
- Assist/Control (A/C) .
- Synchronized Intermittent Mandatory Ventilation (SIMV) ●
- Continuous Positive Airway Pressure (CPAP) ●
×
- Spontaneous (Spont) .
- Spontaneous/Timed (Spont/T) ●
- Apnea (Backup Mode) .
The Esprit Ventilator with Airway Pressure Release Ventilation (APRV) Mode is intended for use only on invasively ventilated adult and pediatric patients as prescribed by a physician.
Prescription Use
AND/OR
Over-the-Counter Use
(Part 21 CFR 801, Subpart D)
(Part 21 CFR 807, Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign | |
|------------------------------------|-----------------------|
| Division of Anesu. | Iowa General Hospital |
| Infection Control, Dental Services | |
510(k) N
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): K110083
V200 Ventilator with Airway Pressure Release Ventilation (APRV) Mode Device Name:
### Indications for Use:
The V200 Ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult, pediatric, and neonatal patients as prescribed by a physician. The V200 Ventilator is intended for use in either invasive or non-invasive applications and contains the following modes of ventilation:
- Assist/Control (A/C) .
- Synchronized Intermittent Mandatory Ventilation (SIMV) ●
- Continuous Positive Airway Pressure (CPAP) .
- Spontaneous (Spont) .
- Spontaneous/Timed (Spont/T) .
- Apnea (Backup Mode) .
The V200 Ventilator with Airway Pressure Release Ventilation (APRV) Mode is intended for use only on invasively ventilated adult and pediatric patients as prescribed by a physician.
× Prescription Use
AND/OR
Over-the-Counter Use
(Part 21 CFR 801, Subpart D)
(Part 21 CFR 807, Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
n Sigi. Til (Div. · Anest · sicingy, General Hospital DIVIs . . trol, Denices
510(k) Numb
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.