K981072 · Respironics, Inc. · CBK · Nov 12, 1998 · Anesthesiology
Device Facts
Record ID
K981072
Device Name
ESPRIT VENTILATOR
Applicant
Respironics, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Nov 12, 1998
Decision
ST
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ESPRIT ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications.
Device Story
Esprit is a microprocessor-controlled, electrically powered mechanical ventilator; provides continuous or intermittent ventilatory support. Features include integrated air source (eliminating need for central/stand-alone compressors), variable oxygen concentration (21-100%), and battery backup (30-min primary, 3-hour secondary). User interface utilizes touch screen and graphical display. Used in hospitals (ICU, surgical recovery), subacute facilities, and surgery centers; operated by qualified medical personnel. Device delivers prescribed flows and pressures to patients. Reliability enhanced via HALT and ESS manufacturing processes. Healthcare providers monitor patient status via ventilator output; supports clinical decision-making for respiratory management.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, environmental testing, and performance comparisons. Testing includes dielectric strength, electromagnetic emissions/susceptibility, vibration, shock, and ingress of liquids per IEC and MIL-STD standards. Waveform performance was compared to the 7200 ventilator predicate. Compliance with 'Draft Reviewer Guidance for Ventilators, July 1995' was demonstrated.
Technological Characteristics
Microprocessor-controlled, electrically powered mechanical ventilator. Features touch screen GUI. Integral air source (blower) or wall air supply. Oxygen mixing (21-100%). Power: AC, 30-min internal battery, 24 VDC secondary source. Tested to IEC 601-1, CISPR 11, IEC 801 series, and MIL-STD-461D. Manufacturing utilizes HALT and ESS. Software-controlled breathing system.
Indications for Use
Indicated for adult and pediatric patients requiring continuous or intermittent ventilatory support in invasive or non-invasive applications, as prescribed by a physician.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
NOV 1 2 1998
K981072
## 2. 510(k) Summary
This 510/k) is being submitted by: Respironics. 1261 Liberty Way Vista CA., 92083. Phone number : 760-727-3500 FAX number: 760-727-8349
The contact personnel for this submittal are:
Paul L. Woodring, primary contact 760-734-3202
Kathy E. Moore, secondary contact 760-734-3221
This summary was prepared on March 23, 1998.
This 510(k) is for the Esprit critical care ventilator, product code (21CFR 868.5895, 73 CBK) The legally marketed predicate devices are the 7200 ventilator offered for sale by Nellcor-Puritan-Bennett; the BIRD Tbird offered for sale by BIRD Medical Products; and the Siemens SV300 offered for sale by Siemens-Elema AB, Electromedical Systems Division. The 7200 ventilator is currently used in both invasive and noninvasive applications. A comparison is also made to the Quantum non-invasive ventilator offered for sale by Healthdyne Technologies (now part of Respironics). Detailed comparisons of the devices can be found in Section 8- Comparative Analysis to Predicate Devices.
The ESPRIT ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. The specific features of the product include:
- . Breathing system under microprocessor control.
- · User interface under microprocessor control. Uses "hyman interface" features associated with touch screen and graphical user interface technology. The specifics of the particular touch screen technology and its suitability to medical applications is covered in more detail in the principle of operation section.
- · Integral air source is built into the ventilator, eliminating the requirement for a central compressor and piped, medical grade wall air and/or an individual, stand alone compressor for each ventilator. The product will also be offered in a configuration without the blower in which case it will operate from a wall air supply.
- · Ability to provide variable oxygen concentrations (21% to 100% O2) from a 35 to 80 PSIG medical grade, oxygen gas source, including medical grade gas cylinders with suitable regulators.
- · Ability to operate on a re-chargeable primary battery for 30 minutes (nominal).
{1}------------------------------------------------
## K981072 2. 510(k) Summary
,
{2}------------------------------------------------
- · Ability to be powered by a secondary DC power source (24 VDC) for up to 3 hours.
- · Provide high product quality and reliability. ESPRIT has been designed using Highly Accelerated Life Testing (HALT), coupled with Environment Stress Screening (ESS) during manufacturing. Studies have shown that these techniques greatly eliminate reliability problems and are integral to the reliability of the product.
The Esprit Ventilator is intended for use by aualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications.
The ESPRIT ventilator is intended to be marketed worldwide as a new ventilator that shall address the needs of low and moderate acuity subacute facilities in the US, surgical recovery units in hospitals, in ICU applications where more expensive options such as trending and waveforms are not required, in stand alone surgery centers in the US and international markets, as well as high acuity intensive care units (ICUs) in undeveloped and emerging nation, rest-of-world (ROW) markets. No medical claims are made regarding the ventilator. It produces the flows and pressures detailed in Section 8 - Comparative Analysis to Predicate Devices. These performance characteristics have been validated with the testing that is summarized in Sections 7 -Table Of Ventilator Requirements, 8 - Comparative Analysis to Predicate Devices and 10 - Environmental Testing.
The determination of equivalence is also based on environmental and lab testina. The ventilator has been tested according to the following environmental standards:
{3}------------------------------------------------
| TEST | BASIS |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Dielectric Strength | IEC 601-1, Clause 20 |
| Leakage Current | IEC 601-1, Clause 19 |
| Radiated and Conducted<br>Electromagnetic Emissions | CISPR 11 |
| Magnetic Field Emissions | RE101 of MIL-STD 461D at 7 cm |
| Electrostatic Discharge | IEC 801-2 |
| Radiated Electromagnetic Fields<br>Susceptibility | IEC 801-3 |
| AC Voltage Fluctuation: Steady State,<br>Dropout, Slow Sags and Surges | IEC 801-11 |
| AC Voltage Fluctuation: Fast Transient<br>Bursts | IEC 801-4 |
| AC Voltage Fluctuation: Fast Surges | IEC 801-5 |
| Conducted Electromagnetic Energy<br>Susceptibility | MIL-STD-461D, CS114 |
| Magnetic Fields Susceptibility | RS101 (Army) of MIL-STD-461D |
| Quasi-static Electric Fields Susceptibility | Reviewer Guidance for Premarket<br>Notification Submissions, November,<br>1993 |
| Random Vibration | Reviewer Guidance for Premarket<br>Notification, Submissions,<br>November, 1993 |
| Shock | Test conditions based on IEC 68-2-<br>27 |
| Ingress of Liquids | IEC 601-1, Clause 44.6, and IEC 529 |
| Temperature/Humidity | Reviewer Guidance for Premarket<br>Notification Submissions, November,<br>1993. |
| Surface Temperature | Reviewer Guidance for Premarket<br>Notification Submissions, November,<br>1993 |
## Table 1 - Environmental Standards
Section 7 - Table Of Ventilator Requirements includes a table demonstrating how the Esprit ventilator meets the requirements called out in the "Draft Reviewer Guidance for Ventilators, July, 1995" Finally, Section 8 - Comparative Analysis to Predicate Devices includes a summary of performance testing comparing the Esprit ventilator to the 7200 predicate device.
A description of waveform performance is covered in 8.3 - Waveform Comparisons to 7200 Ventilator predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, overlaid with three horizontal bars. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 2 1998
Mr. Paul L. Woodring Respironics Inc. 1261 Liberty Way Vista, Ca. 92083
Re: K981072 Esprit Ventilator Requlatory Class: II (two) Product Code: 73 CBK Dated: August 14, 1998 Received: August 18, 1998
Dear Mr. Woodring:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Paul L. Woodring
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(K) Number (if known): K981072
Device Name: Esprit Ventilator
Indications for Use:
The ESPRIT ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kramer
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 1981072 510(k) Number _
prescriptions use
OTC
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.