K103803 · Hamilton Medical AG · CBK · Nov 23, 2011 · Anesthesiology
Device Facts
Record ID
K103803
Device Name
HAMILTON-G5
Applicant
Hamilton Medical AG
Product Code
CBK · Anesthesiology
Decision Date
Nov 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The HAMILTON-G5 ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients, and optionally to infant and neonatal patients. The device is intended for use in the hospital and institutional environment where healthcare professionals provide patient care, including use as a patient bedside for intra-facility transport, provided compressed gas is supplied. The device is not intended for transportation outside the hospital or for use in the home environment.
Device Story
Electronically controlled pneumatic intensive care ventilator; provides ventilatory support using oxygen, air, or heliox. Inputs: gas supply; user-defined settings via LCD touch screen, keys, and press-and-turn knob. Outputs: controlled gas delivery for ventilation, nebulization of physician-prescribed medications, and continuous cuff pressure monitoring/adjustment. Used in hospital/institutional settings by trained personnel under physician supervision; supports intra-facility transport. Features include nCPAP-PS mode for neonates and NIV-ST mode for adult/pediatric patients. Benefits: provides respiratory support, manages airway cuff pressure, and enables medication delivery via integrated nebulization.
Clinical Evidence
No clinical testing was conducted or required in support of this premarket clearance notification.
Technological Characteristics
Electronically controlled pneumatic system; AC power with battery backup. Interface: LCD touch screen, keys, press-and-turn knob. Features: integrated AERONEB nebulization, cuff pressure controller, nCPAP-PS and NIV-ST ventilation modes. Standards: IEC60601-1, IEC60601-1-2, IEC 60601-2-12, ASTM F1100-90. Gas sources: oxygen, air, or heliox.
Indications for Use
Indicated for positive pressure ventilatory support in adult, pediatric, infant, and neonatal patients within hospital or institutional settings. Contraindicated for use in the presence of flammable anesthetic agents, ignition sources, or MRI environments.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten string of characters. The characters appear to be "K1038803". The characters are written in a cursive style, with some of the numbers being connected. The image is in black and white.
# 510(k) Summary
| Submitter: | HAMILTON MEDICAL AG |
|-----------------------------|----------------------------------------------|
| | Via Crusch 8 |
| | 7402 Bonaduz, Switzerland |
| Contact Person: | Name: Frederike Brühschwein |
| | Phone: +41 81 660 6568 |
| | Fax: +41 81 660 6020 |
| Preparation Date: | 2011-11-02 |
| Trade Name: | HAMILTON-G5 |
| Common Name: | Continuous Ventilator |
| Classification Name: | Ventilator, Continuous, Facility Use (21 CFR |
| | 868.5895, Product Code: CBK) |
| Legally marketed devices to | HAMILTON-C2 (K092148) |
| | HAMILTON-GALILEO (K080181) |
| which equivalence is being | Dräger EvitaXL with NeoFlow and SpO2 |
| claimed: | options (K072412, K983219) |
| | ARM PYTON ETT Cuff Pressure Regulator |
| | (K092733) |
| | GE Datex-Ohmeda Engström Carestation |
| | (K081842) |
#### Device Description
The HAMILTON-G5 is an electronically controlled pneumatic intensive care ventilator ventilation system.
It uses oxygen and air or heliox to ventilate adults, pediatrics, infants, and neonates. It is powered by ac with battery backup to protect against power failure or unstable power and to facilitate intrahospital transport.
The HAMILTON-G5's pneumatics deliver gas, and its electrical systems control pneumatics, monitor alarms, and distribute power.
The user interface consists of a LCD-display with touch screen, keys, and a press-and-turn knob. The new nebulization function with AERONEB nebulizers is for use with mechanically ventilated patients to aerosolize physician prescribed medications for inhalation. The new cuff pressure controller implemented in the HAMILTON-G5 is for continuous monitoring and adjustment of the cuff pressure of a tracheal or tracheostomy tube. nCPAP-PS is a new ventilation mode which applies nCPAP with additional pressure support by nasal interfaces with reduced dead space on neonates. NIV-ST is a new mode available for adult and pediatric patients. It delivers
{1}------------------------------------------------
pressure-controlled, time-cycled mandatory breaths and pressure-supported, flow-cycled spontaneous breaths by a mask or other noninvasive patient interface.
#### Intended use
The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where healthcare professionals provide patient care. The HAMILTON-G5 ventilator is intended for use by properly trained personnel under the direct supervision of a licensed physician. The HAMILTON-G5 ventilator may be used for transport within a hospital or hospital type facility provided compressed gas is supplied. The device is not to be used in the presence of flammable anesthetic agents or other ignition sources. The ventilator is not to be used in an environment with magnetic resonance imaging (MRI) equipment.
The device is not intended for transportation outside the hospital or for use in the home environment.
In the USA, federal law restricts this device to sale by or on the order of a physician.
# SUMMARY OF THE TECHNOLOGY AND PERFORMANCE SPECIFICATIONS COMPARISON WITH THE PREDICATED DEVICES
The indication statements for the HAMILTON-G5 ventilator are comparable to those for the predicate devices.
Technological characteristics and performance specifications of the HAMILTON-G5 ventilator are substantially equivalent to those of the predicate devices.
The new nebulization function is comparable to the nebulization function in the predicate device Engström Carestation.
The new cuff pressure controller is considered to be substantial equivalent to the PYTON ETT Cuff Pressure Regulator.
The new NIV-ST mode is equivalent to the NIV-ST mode in the HAMILTON-C2 and the new nCPAP-PS mode can be compared with the PCV+/Assist, mask ventilation mode of the Evita XL.
HAMILTON MEDICAL has demonstrated the modified HAMILTON-G5 ventilator to be safe and effective.
The ventilator is considered to be substantial equivalent to currently marketed predicate devices which have been previously cleared by FDA.
#### SUMMARY OF NON-CLINICAL PERFORMANCE TESTS
Safety testing of the HAMILTON-G5 with the new options was conducted according to IEC60601-1, IEC60601-1-2, IEC 60601-2-12 and other applicable standards. The test results show that the device is safe and effective for its intended use.
The ventilator was further subject to wave-form performance testing as described in the standard ASTM F1100-90. The data provided from these tests, were shown to be substantially equivalent to a legally marketed device.
{2}------------------------------------------------
The results of the software verification and validation testing demonstrate that all specified requirements have been implemented correctly and completely.
### SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
No clinical testing was conducted or required in support of this premarket clearance notification.
### SUMMARY OF OTHER INFORMATION
This submission included comparison of intended use statements, proposed product labeling and summary information and labeling on predicate devices.
### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
The results of verification, validation, and testing activities demonstrate that the modified HAMLTON-G5 ventilator is as safe, as effective, and performs as well as or better than the legally marketed devices identified above.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the graphic.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
HAMILTON MEDICAL AG C/O Mr. Brain Edwards Senior Consultants, Regulatory Services MDGI 49 Plain Street North Attleboro, Massachusetts 02764
Re: K103803
Trade/Device Name: HAMILTON-G5 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: November 21, 2011 Received: November 22, 2011
Dear Mr. Edwards:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subiect to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
NOV 2 3 2011
{4}------------------------------------------------
Page 2 - Mr. Edwards
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. nut
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
| 510(k) Number: | |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | HAMILTON-G5 |
| Indication for Use: | The HAMILTON-G5 ventilator is intended to provide positive<br>pressure ventilatory support to adult and pediatric patients, and<br>optionally to infant and neonatal patients.<br>The device is intended for use in the hospital and institutional<br>environment where healthcare professionals provide patient care,<br>including use as a patient bedside for intra-facility transport,<br>provided compressed gas is supplied.<br>The device is not intended for transportation outside the hospital or<br>for use in the home environment. |
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Penn
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
103803 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.