K092733 · Arm Medical Devices, Inc. · BSK · Feb 26, 2010 · Anesthesiology
Device Facts
Record ID
K092733
Device Name
PYTON
Applicant
Arm Medical Devices, Inc.
Product Code
BSK · Anesthesiology
Decision Date
Feb 26, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To measure, and regulate intra-cuff pressures of endotracheal, supraglottic airways, and tracheostomy tubes. The PYTON is intended for use on patients who are intubated. The PYTON is to be used under medical supervision in hospital (EMS), extended care facilities and outpatient clinics where a patient may be intubated.
Device Story
PYTON is an ETT cuff pressure regulator; measures and automatically maintains cuff pressure at user-selected levels. Input: air pressure from ETT cuff via pilot balloon connector. Operation: uses pressure sensor for monitoring; small air pump increases pressure; dump valve decreases pressure. Output: LED back-lit LCD displays pressure; user interface includes four pushbuttons for target pressure selection and battery check. Used in hospitals, EMS, extended care, and clinics by respiratory practitioners. Includes mechanical pressure relief valve (1.5 p.s.i. limit) and 10-hour battery backup for patient transport. Benefits: maintains cuff pressure within target range (typically 20-30 cmH2O) to prevent complications associated with over- or under-inflation.
Clinical Evidence
Bench testing only. Electrical safety (EN 60601-1), EMC (EN 60601-1-2), and mechanical/environmental testing performed per FDA guidance. Measurement accuracy validated at +/- 1 cmH2O. No clinical data presented.
Technological Characteristics
Electrically powered (battery/wall cube) ETT cuff pressure regulator. Features: pressure sensor, air pump, dump valve, LED back-lit LCD, mechanical pressure relief valve (1.5 p.s.i. +/- 20%). Connectivity: standalone. Sterilization: not specified. Materials: plastic tubing set with male Luer adapter.
Indications for Use
Indicated for intubated patients requiring measurement and regulation of intra-cuff pressures for endotracheal, supraglottic, or tracheostomy tubes. Used under medical supervision in hospitals, EMS, extended care, and outpatient clinics.
Regulatory Classification
Identification
An inflatable tracheal tube cuff is a device used to provide an airtight seal between a tracheal tube and a patient's trachea.
Predicate Devices
Cuff Mate 2 Endotracheal Cuff Inflator and Monitor (K902114)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a series of handwritten digits and a letter. The characters are K092733. The numbers are written in black ink on a white background. The handwriting is somewhat stylized.
- 1. Official Contact Information:
Anthony Calderoni President ARM Medical Devices, Inc. 190 Dino Rd FEB 2 6 2010 Bristol, CT 06010 Telephone Cell (860) 593-6004 E-mail: calderoni207@aol.com
- Alternate Daytime e-mail: tonyc@mgminstruments.com
- 2. Proprietary or Trade Name: PYTON
- 3. Common / Usual Name: ETT Cuff Pressure Regulator
- Classification Name: Cuff, Tracheal Tube, Inflatable 4.
- Regulation Number: 21 CFR 868.5750 5.
- Classification Product Code: BSK 6.
- 7. Predicate Devices:
- 7.1.Cuff Mate 2 Endotracheal Cuff Inflator and Monitor,
- 7.1.1.Manufacturer: Diemolding Corporation
- 7.1.2.FDA 510k number: K902114
- 7.1.3.Classification Product Code: BSK
- 7.2. Posey Cufflator
- 7.2.1.Manufacturer: Posey Co.
- 7.2.2.FDA 510k number: K912723
- 7.2.3.Classification Product Code: BSK
- 7.3. Rusch Endotest Cuff Pressure Monitor
- 7.3.1.Manufacturer: Rusch Intl.
- 7.3.2.FDA 510k number: K951046
- 7.3.3.Classification Product Code: BSK
- 7.4. Pressure Easy Cuff Pressure Controller
- 7.4.1.Manufacturer: Smiths Healthcare Manufacturing
- 7.4.2.FDA 510k number: K83327 (There have been several companies that have marketed this device over the years since it first was approved.)
..
- 7.4.3.Classification Product Code: BSK
- 7.5. Model GV-10 Lanz Pressure Valve
- 7.5.1.Manufacturer: Extracorporeal Medical Specialties, Inc.
- 7.5.2.FDA 510k number: K791045
- 7.5.3.Classification Product Code: BTR
- 7.6. CuffAlert
- 7.6.1.Manufacturer: SunMed
- 7.6.2.FDA 510k number K081805
- 7.6.3.Classification Product Code: BSK
{1}------------------------------------------------
- Functional Summary of the PYTON measures the ETT cuff pressure, and automatically 8. adjusts the air pressure inside the ETT Cuff to a level selected by the user who is intended to be a professional respiratory practitioner. Room air is the "fluid" that is utilized to inflate the ETT Cuff, so there is not any use of (or contact to) any of the patient's biological fluids by the PYTON.
- 8.1. Monitoring Function: The monitoring function is achieved by use of a pressure sensor.
- 8.2. ETT Cuff Pressure Regulation: The automatic ETT cuff pressure regulation is achieved by activation of either a small air pump (of the same type used in many automatic blood pressure cuff measurement systems) in order to raise the ETT cuff pressure, or by activation of a "dump" valve to lower the ETT cuff pressure.
- 8.3. User Interface: The PYTON provides an LED back-lit LCD digital display to report the ETT cuff pressures to the user, and four pushbuttons that provide the user with the ability to select the target ETT cuff pressure, and check battery charge level.
- 8.4. Safety Valve: The PYTON includes a mechanical pressure relief valve as a safety precaution that ensures a maximum ETT cuff pressure (in the case of a PYTON failure) to a level of 1.5 p.s.i. (105 cmH2O) +/-20%.
- 8.5. Battery Back-Up: The PYTON includes a battery back-up power supply allowing the PYTON to be used for up to 10 hours without mains power. This feature allows (as most of the predicates do) for the PYTON to be used while the patient is moved within the hospital to X-Ray for example, or during transport of the patient.
## 9. Intended Use:
- 9.1. To measure, and regulate intra-cuff pressures of endotracheal, supraglottic airways, and tracheostomy tubes.
- 9.2. The PYTON is intended for use on patients who are intubated.
- 9.3. The PYTON is to be used under medical supervision in hospital (EMS), extended care facilities and outpatient clinics where a patient may be intubated.
- 10. Prescription Use: Caution: US Federal law restricts this device to sale by or on the order of a physician.
- 11. Steps in the Application of the PYTON: The intended use instructions for respiratory practitioners are briefly summarized below:
- 11.1.1. Connection of the PYTON to the ETT Cuff Pilot Balloon: The PYTON connects through a plastic tubing set to the ETT Cuff pilot balloon connector with a male Luer adapter exactly like predicate devices.
- 11.1.2. Set Cuff Pressure: The respiratory practitioner next powers on the PYTON device and selects the target ETT Cuff pressure with a few key strokes on the PYTON's front panel. Generally accepted ETT cuff pressure levels for high volume/low pressure cuffs are in the range of 20 to 30 cmH2O. The PYTON includes a default "SET" cuff pressure value of 25 cmH2O.
- 11.1.3. Monitor/Regulation: The PYTON automatically adjusts the ETT Cuff pressure to the target "SET" point pressure, and continues to monitor, and adjust the cuff pressure to the "SET" point pressure.
.
{2}------------------------------------------------
- 12. Summary of Technological Characteristics of the PYTON ETT Cuff Pressure Regulator as Compared to Predicate Devices: The PYTON ETT Cuff Pressure Regulator performs two functions. It monitors the ETT Cuff Pressure, and secondly, it regulates the ETT cuff pressure to a "SET" point pressure within +/- 2 cmH2O. All the predicate devices perform the monitoring function, and some also provide pressure regulation, as well. A brief discussion of these points is given below.
- The automatic pressure regulation found in the predicate devices listed above like the 12.1. PressureEasy Cuff Pressure Controller, and the Model GV-10 Lanz Pressure valve are equivalent to the PYTON's pressure regulation.
- 12.2. The PYTON's monitoring function is equivalent to the monitoring functions of all the predicate devices listed above. The PYTON is electrically powered as is the CuffAlert and the Cuff Mate 2 Endotracheal Cuff Inflator, and Monitor. Both the PYTON and the Cuff Mate 2 incorporate LCD digital displays of the ETT cuff pressure in units of "xx" cmH2O. The Cuff Mate 2, however, does not employ automatic pressure regulation as found in the PYTON.
- 12.3. The CuffAlert device is also electrically powered, and it is similar in its monitoring function to the PYTON in that it incorporates and monitoring device that alerts the user by illuminating a red LED if the tracheal cuff pressure is too high.
- Determination of Substantial Equivalence: (See Table 1) 12.4. :
- 12.4.1. The monitoring mode of the PYTON is similar to each of the predicate devices.
- 12.4.2. The pressure regulation feature of the PYTON is similar to the following predicate devices:
- 12.4.2.1. The PressureEasy Cuff Pressure Controller
- The Model GV-10 Lanz Pressure Valve 12.4.2.2.
- 12.4.3. The PYTON has the same intended use as each of the predicate devices. (Allowing for manual regulation of the Cuff Pressures by some of the predicate devices.)
- 12.4.4. The PYTON has the same intended patient population and use environments that each of the predicate devices have.
- 12.4.5. The PYTON does not raise new questions as to the safety and effectiveness of cuff pressure regulator devices.
- 12.4.6. The PYTON has similar instructional information, including warning and caution statements as compared to the predicate devices.
.
{3}------------------------------------------------
# Table 1: Non-Confidential Summary of Safety and Effectiveness
| Attributes/Description | PYTON (Proposed Device) | Cuff Mate 2 | CuffAlert | Posey Cufflator and Rusch Endotest | PressureEasy and Model GV-10 Lanz Pressure Valve |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| FDA 510k number | To be assigned | K902114 | K081805 | K912723-<br>Posey<br>K951046-<br>Rusch | K833327 - PressureEasy<br>K791045-GV-10 Lanz<br>Pressure Valve |
| Indications for use | To measure and<br>regulate intra-cuff<br>pressures of<br>endotracheal,<br>supraglottic airways,<br>and tracheostomy<br>tubes | Same,<br>except that<br>regulation is<br>performed<br>manually. | Same,<br>except that<br>regulation is<br>performed<br>manually. | Same,<br>except that<br>regulation is<br>performed<br>manually. | To measure and regulate<br>intra-cuff pressures of<br>oral/nasal endotracheal<br>tubes. |
| Environments of Use | To be used under medical<br>supervision in hospitals,<br>pre-hospitals (EMS),<br>extended care facilities<br>and outpatient clinics<br>where a patient may be<br>intubated. | (Same) | (Same) | (Same) | (Same) |
| Patient Population | Intubated patients | (Same) | (Same) | (Same) | (Same) |
| Single Patient or Reusable? | Reusable | (Same) | (Same) | (Same) | Single-patient |
| Range of<br>measured Cuff<br>Pressure | 0 to 85 cmH2O | 0 to 99<br>cmH20 | 10 to 40<br>cmH20 | 0 to 120<br>cmH20-<br>Posey<br>0 to 120<br>cmH2O -<br>Rusch | 20-30 cmH2O -<br>PressureEasy<br>30 to 34 cmH20 - GV-10<br>Lanz Pressure Valve. |
| Accuracy of Cuff<br>Pressure<br>Measurement | +/-1 cmH2O (see<br>Bench Tests for<br>validation.) | +/- 2.4<br>cmH2O ( 2<br>Std Dev<br>from study<br>done in<br>Feb 2004) | +/- 3<br>cmH2O @<br>Cuff<br>Pressures<br>20 to 30<br>cmH2O | +/- 3.8<br>cmH2O-<br>Posey<br><br>+/- 2.6<br>cmH2O-<br>Rusch<br><br>( 2 Std Dev<br>from study<br>done in<br>Feb 2004) | Not Specified |
| Power | Battery or Wall Cube<br>supply | Battery | Battery | Manual | Manual |
・
··
{4}------------------------------------------------
#### 12.5. Summary of Performance Testing:
- 12.5.1. Electrical safety testing was conducted per EN 60601-1.
- 12.5.2. EMC testing was performed per EN60601-1-2.
- 12.5.3. Software validation testing was performed per FDA's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices.
- 12.5.4. Electrical, mechanical and environmental testing of the PYTON was conducted in accordance with the FDA Draft Reviewer Guidance for Premarket Notification Submissions (1993).
- 12.5.5. Conclusion: The results of all performance tests demonstrate that the PYTON meets its design and system requirements.
- 13. Conclusions: The technological characteristics of the PYTON ETT Cuff Pressure Regulator and the results of bench tests do not raise new questions of safety or effectiveness when compared to the legally marketed predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human services.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Anthony Calderoni President : ARM Medical Devices, Incorporated 190 Dino Road Bristol, Connecticut 06010
FEB 2 6 2010
Re: K092733
Trade/Device Name: PYTON Regulation Number: 21 CFR 868.5750 Regulation Name: Inflatable Tracheal Tube Cuff Regulatory Class: II Product Code: BSK Dated: February 22, 2010 Received: February 24, 2010
Dear Mr. Calderoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Calderoni
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{7}------------------------------------------------
# Indications for Use Statement
510(k) Number:
WALFS 3 (To be assigned)
PYTON Device Name:
# Indications for Use:
To measure, and regulate intra-cuff pressures of endotracheal supraglottic airways or tracheostomy tubes.
The PYTON is intended for use on patients who are intubated.
The PYTON is to be used under medical supervision in hospitals, pre-hospital (EMS), extended care facilities and outpatient clinics where a patient may be intubated.
Prescription Use: YES Over-the-Counter Use: NO or (Part 21 CFR 801 Subpart D) 21 CFR 807 Subpart C)
(Please do not write below this line- Continue on another page if needed)
Concurrence of CDRH, Office of device Evaluation (ODE)
Z. Rhetten
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
092733 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.