ALMA LASERS THERMOXEL CO2 LASER SYSTEM AND DELIVERY DEVICE ACCESSORIES

K080463 · Alma Lasers , Ltd. · GEX · Jul 23, 2008 · General, Plastic Surgery

Device Facts

Record IDK080463
Device NameALMA LASERS THERMOXEL CO2 LASER SYSTEM AND DELIVERY DEVICE ACCESSORIES
ApplicantAlma Lasers , Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateJul 23, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialties as defined in Section 4 of the present submission.

Device Story

ThermoXEL™ CO2 laser system; console includes optical bench, microcontroller, power supply, cooling, and air-purge system. Input: electrical power; Output: CO2 laser energy delivered via 7-joint articulated arm and various handpieces (single spot, scanner, pixel). Operated by physicians in clinical settings (OR, clinic). System transforms electrical energy into laser beam for soft tissue ablation, vaporization, excision, incision, and coagulation. Healthcare providers use handpieces to target specific tissue; output provides precise surgical control for procedures like skin resurfacing, tumor resection, and debridement. Benefits include controlled tissue removal with coagulation, reducing bleeding and surgical time.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices. Bench testing only.

Technological Characteristics

CO2 laser system; 7-joint articulated arm; delivery handpieces (single spot, scanner, pixel). Includes LCD control panel, keypad, footswitch, microcontroller, high voltage power supply, cooling system, and air-purge system. Energy source: CO2 laser. Software-controlled operation. Standalone device.

Indications for Use

Indicated for surgical ablation, vaporization, excision, incision, and coagulation of soft tissue across multiple specialties including dermatology, plastic surgery, podiatry, otolaryngology, gynecology, neurosurgery, orthopedics, general/thoracic surgery, dental/oral surgery, and genitourinary procedures. Patient population includes adults and children requiring soft tissue treatment for conditions ranging from benign/malignant lesions, tumors, and cysts to aesthetic skin resurfacing and surgical debridement. Prescription use only.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Section 5-510(k) Summary or 510(k) Statement #### I. General Information Submitter: Alma Lasers, Ltd. 14 Halamish Street (PO Box 3021) Industrial Park, Caesarea, 38900 ISRAEL Contact Person: Summary Preparation Date: June 2, 2008 Regulatory Affairs Manager Tatiana Enstein #### II. Names Alma Lasers ThermoXEL™ CO2 Laser System and Device Names: Delivery Device Accessories Primary Classification Names: Laser Instrument, Surgical Powered #### III. Predicate Devices - . Lasering SLIM Evolution Family of CO2 Lasers and Delivery Device Accessories (K063001) - PhotoMedex LaserPro CO2 Carbon Dioxide Laser System (K040234) . - Modified Lumenis Family of UltraPulse SurgiTouch CO2 Laser Systems (K030147) . - Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and Delivery Device . Accessories (K022060) #### IV. Product Description The Alma Lasers ThermoXEL™ CO2 Laser System and delivery device accessories consist of the following major components: - 1. Laser system console (containing the optical bench assembly and laser, the microcontroller control electronics and system software, the high voltage power supply, the laser cooling system, the compressed air-purge system, and the service panel) - a. LCD control panel - b. Keypad - 2. 7-joint articulated arm - 3. Footswitch - 4. Variety of Delivery Device Handpieces (single spot, scanner, pixel) #### V. Indications for Use The Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialties as defined in Section 4 of the present submission. {1}------------------------------------------------ #### VI. Rationale for Substantial Equivalence The Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories share the same or similar indications for use, operation, overall technical and functional capabilities, and therefore are substantially equivalent to the predicate devices. #### VII. Safety and Effectiveness Information The review of the indications for use and technical characteristics provided demonstrates that the Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories are substantially equivalent to the predicate devices. #### VIII. Conclusion The Alma Lasers ThermoXELTM CO2 Laser System and Delivery Device Accessories were found to be substantially equivalent to the predicate devices. The Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories shares the same or similar indications for use, similar design features, and functional features with, and thus is substantially equivalent to, the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Alma Lasers, Ltd. % A. Worden Consulting Ms. Anne Worden Regulatory Consultant 3637 Bernal Avenue Pleasanton, California 94566 JUL 2 3 2008 Re: K080463 Trade/Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical insgrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: June 2, 2008 Received: June 5, 2008 Dear Ms. Worden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Ms. Anne Worden CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Milliman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): K08 1 Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use: The Alma Lasers ThermoXFLM CO2 Laser System and Delivery Device Accessories are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knec). The Alma Lasers ThermoXEL™ CO2 Laser System is cleared for use for the particular indications as follows: ## Dermatology & Plastic Surgery The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of: - . laser skin resurfacing - . laser derm-abrasion - . laser burn debridement. Laser skin resurfacing (ablation and/or vaporization) for the treatment of: - wrinkles, rhytids, and furrows (including fines lines and texture irregularities) . Clinical literature demonstrates that skin resurfacing of wrinkles, rhytids, and furrows with CO2 laser increases the amount of sub-epidermal collagen. V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) 023 Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | 12080463 | |---------------|----------| |---------------|----------| Page 1 of 10 {5}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: # Dermatology & Plastic Surgery, continued Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: - keratoses, including actinic and seborrheic keratosis, seborrhoecae vulgares, sebortheic . wart, and verruca seborrheica; - vermillionectomy of the lip; . - . cutaneous horns; - . solar/actinic elastosis; - chelitis, including actinic chelitis; - lentigines, including lentigo maligna or Hutchinson's malignant freckle; . - uneven pigmentation/ dyschromia: - . acne scars; - surgical scars; ● - . keloids including acne keloidalis nuchae; - hemangiomas (including Buccal, port wine and pyogenic granulomas/granuloma . pyogenicum/granuloma telagiectaticum); - . tattoos; - . telangiectasia; - . removal of small skin tumors, including periungual (Koenen) and subungual fibromas; - superficial pigmented lesions; . - . adenosebaceous hypertrophy or sebaceous hyperplasia; - rhinophyma reduction; - . cutaneous papilloma (skin tags); - milia: . - debridement of eczematous or infected skin; . - basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions; | Prescription Use | <div> <span style="text-decoration: overline;"></span> <span>✓</span> </div> | |-----------------------------|------------------------------------------------------------------------------| | (Part 21 CFR 801 Subpart D) | AND | D/OR Over-The-Counter Use (21 CFR 801 Subpart C) 024 Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number 1<080463 Page 2 of 10 {6}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXEJ.TM CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: ## Dermatology & Plastic Surgery, continued - . nevi, including spider, epidermal and protruding; - neurofibromas; . - laser de-epithelialization; . - . tricoepitheliomas; - . xanthelasma palpebrarum; - . syringoma Laser ablation, vaporization and /or excision for complete or partial nail matrixectorny. Vaporization/coagulation of: - . benign/malignant vascular/avascular skin lesions; - � Moh's Surgery; - . Lipectomy; - Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungaual . warts. Laser incision and /or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty. Laser incision and /or excision of soft tissue for the creation of recipient sites for hair transplantation. | Prescription Use<br>(Part 21 CFR 801 Subpart D) | ✓ | | |-------------------------------------------------|--------|------------------------------------------------| | | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | 11040463 | |---------------|----------| |---------------|----------| Page 3 of 10 025 {7}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: ### Podiatry Laser ablation, vaporization and /or excision of soft tissue for the reduction, removal, and/or treatment of: - verrucae vulgares/plantar (warts), including paronychial, periungal, and subungual warts; . - fungal nail treatment; . - porokeratoma ablation; - ingrown nail treatment; ◆ - neuromas/fibromas, including Morton's neuroma; . - debridement of ulcers; ● - other soft tissue lesions. ● Laser ablation, vaporization and /or excision in podiatry for complete or partial matrixectomy. # Otolaryngology (ENT) Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology the treatment of: - . choanal atresia, - leukoplakia, including oral, larynx, uvula, palatal, upper lateral pharyngeal tissue; . - . nasal obstruction; - adult and juvenile papillomatosis polyps; . - polypectomy of nose and nasal passages; . - . lymphangioma removal; - removal of vocal cord/fold nodules, polyps and cysts; . - removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx and . trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords; Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) 026 | (Division Sign-Off) | | |------------------------------------------------------------|--------------| | Division of General, Restorative, and Neurological Devices | | | 510(k) Number | L080463 | | | Page 4 of 10 | {8}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: ## Otolaryngology (ENT), continued - . Iaser/tumor surgery in the larynx, pharynx, nasal, ear and oral structures and tissue; - Zenker's Diverticulum/pharyngoesophageal diverticulum Jendoscopic laser-assisted . esophagodivertuculostomy (ELAED)}; - stenosis, including subglottic stenosis; � - tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil ● ablation/tonsillotomy; - . pulmonary bronchial and tracheal lesion removal; - . benign and malignant nodules, tumors and fibromas (larynx, pharynx, trachea, trachcobronchial/endobronchial); - . benign and malignant lesions and fibromas (nose and nasal passages); - . benign and malignant tumors and fibromas; (oral); - . stapedotomy/stapedectomy; - acoustic neuroma in the ear; . - superficial lesions of the ear, including chondrodermatitis nondularis chronica . helices/Winkler's disease; - . telangiectasia/hemangioma of larynx, pharynx and trachea (includes uvula, palatal or upper lateral pharyngeal tissue): - cordectomy, cordotomy (for the treatment of vocal fold paralysis/vocal fold motion . impairment), and cordal lesions of larynx, pharynx and trachea; - ◆ myringotomy/tympanostomy (tympanic membrane fenestration); - . uvulopalatoplasty (LAUP. laser UPP): - turbinectomy and turbinate reduction/ablation); ◆ - septal spur ablation/reduction and septoplasty; ● - . partial glossectomy; - tumor resection on oral, subfacial and neck tissues; � - rhinophyma: . - verrucae vulgares (warts); - gingivoplasty/gingivoctomy. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | 1104046 | |---------------|---------| |---------------|---------| Page 5 of 10 {9}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: ## Gynecology (GYN) Laser incision, excision, ablation and/or vaporization and of soft tissue in gynecology (GYN) for the treatment of; - conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and � vaginal intraepithelial neoplasia (VIN, VAIN); - . condyloma acuminate, including cervical, genital, vulvar, perincal, and Bowen's disease, (Erythroplasia of Queyrat) and Bowenoid papulosa (BP) lesions; - . leukoplakia (vulvar dystrophies); - . incision and drainage (I&D) of Bartholin's and nubuthian cysts; - herpes vaporization; ● - urethral caruncle vaporization; � - � cervical dysplasia; - benign and malignant tumors; - . hemangiomas. # GYN Laparoscopy Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic and laparoscopic surgery, including GYN laparoscopy, for treatment of: - . endometrial lesions, including ablation of endometriosis; - . excision/lysis of adhesions; - . salpingostomy; - . oophorectomy/ovariectomy: - . fimbroplasty: - metroplasty: . - microsurgery (tubal); . - uterine myomas and fibroids; . Prescription Use > (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) 028 Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | |-----------------------------------| | Division of General, Restorative, | | and Neurological Devices | | 510(k) Number | L08046 | |---------------|--------| |---------------|--------| | Page 6 of 10 | |--------------| |--------------| {10}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: # GYN Laparoscopy, continued - ◆ ovarian fibromas and follicle cysts; - . uterosacral ligament ablation; - . hysterectomy. # Neurosurgery Laser incision, excision, ablation and/or vaporization and of soft tissue in neurosurgery for the treatment of: # Cranial - . posterior fossa tumors; - . peripheral neurectomy; - benign and malignant tumors and cysts (e.g. gliomas, menigiomas (including basal . tumors), acoustic neuromas, lipomas and large tumors); - arteriovenous malformation; . - . pituitary gland tumors (transphenoidal approach). # Spinal Cord - incision/excision and vaporization of benign and malignant tumors and cysts; . - . intra- and extradural lesions; - . laminectomy/ laminotomy/ microdiscectomy. V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | IC080463 | |---------------|----------| |---------------|----------| Page 7 of 10 029 {11}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXELTM CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: ## Orthopedics Incision/excision and vaporization and of soft tissue in orthopedic surgery. Applications include: ## Arthroscopy - . menisectomy; - . chodromalacia; - . chondroplasty; - . ligament release (lateral and other); - . excision of plica; - partial synovectomy. ● ### General - . deridement of traumatic wounds; - debridement of decubitus and diabetic ulcers; . - . microsurgery; - artificial joint revision; . - PMMA removal. � ## General/Thoracic Surgery Incision, excision and vaporization and of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include: - . debridement of decubitus ulcers, stasis, diabetic and other ulcers; - mastectomy; - . debridement of burns; - rectal and anal hemorrhoidectomy; - . breast biopsy; Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CORH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative. and Neurological Devices 510(k) Number Page 8 of 10 030 {12}------------------------------------------------ 510(k) Number (if known): K08 Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: # General/Thoracic Surgery, continued - ◆ reduction mammoplasty; - ◆ cytoreduction for metastatic disease; - . laparotomy and laparoscopic applications; - . mediastinal and thoracic lesions and abnormalities; - . skin tag vaporization; - . atheroma; - . cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips; - pilonidal cyst removal and repair; . - . abscesses; - . other soft tissue applications. ## Dental/Oral Surgery Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include: - . gingivectomy- removal of hyperplasias; - gingivoplasty; - . incisional and excisional biopsy; - . treatment of ulcerous lesions, including aphthous ulcers; - . incision of infection when used with antibiotic therapy; - . frenectomy (frenum release); - . excision and ablation of benign and malignant lesions; - . homeostasis: - operculectomy; . - . crown lengthening; Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) 031 Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | K080463 | |---------------|---------| | Page | 9 of 10 | {13}------------------------------------------------ K08 510(k) Number (if known): Device Name: Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories Indications for Use - Continued from the previous page: # Dental/Oral Surgery, continued - . removal of soft tissue, cysts and tumors; - oral cavity tumors and hemangiomas; . - abscesses; - . extraction site hemostasis; - salivary gland pathologies; - preprosthetic gum preparation; . - leukoplakia; - partial glossectomy; . - . periodontal gum resection. # Genitourinary Incision/excision and vaporization and of soft tissue in genitourinary procedures. Applications include: - . benign and malignant lesions of external genitalia; - . condyloma; - . phimosis - . erythroplasia. The Alma Lasers ThermoXEL™ CO2 Laser System Pixel Handpieces are indicated for use in soft tissue for: ### Dermatology & Plastic Surgery - . Skin resurfacing Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** LCO10463 Page 10 of 10 032
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...