K103134 · Fukuda Denshi USA, Inc. · MHX · Mar 11, 2011 · Cardiovascular
Device Facts
Record ID
K103134
Device Name
FUKUDA DENSHI
Applicant
Fukuda Denshi USA, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Mar 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The DS-7000 Series Telemetry Monitoring System is intended to be used as central station monitoring system for the evaluation of the cardiovascular system. It is intended to be used by or on the order of a physician or similarly qualified healthcare professional. The DS-7000 Series Telemetry Monitoring System is intended to be used in hospital environments: ER, ICU, a clinic or similar settings. The DS-7000 Series Telemetry Monitoring System is intended to be used in those situations where the patient is being monitored by a Fukuda Denshi DS-5000-7000 Series bedside monitor, or patient worn telemetry transmitter where remote, central station monitoring is desired. This system is not intended for home use.
Device Story
Patient-worn telemetry transmitter; collects ECG, respirogram, SpO2, plethysmograph, and pulse rate data. Uses digital FSK technology to transmit physiological signals to a Fukuda Denshi Central Monitor via 608-614 MHz WMTS frequencies. Front LCD displays real-time waveforms and numeric values. Operated by clinicians in hospital environments (ER, ICU, clinic). Integrates OEM SpO2 modules (Konica Minolta or Nellcor). Provides continuous monitoring to assist healthcare providers in cardiovascular assessment; enables remote observation of patient status.
Clinical Evidence
Bench testing only. Performance and functional testing conducted to ensure specifications were met. Device host-tested at OEM engineering facilities for SpO2 module integration. No clinical data presented.
Technological Characteristics
WMTS-compliant transmitter (608-614 MHz); digital FSK modulation. Dimensions: 72x98x24.8 mm; weight: 190g. Power: 2 AA alkaline batteries. Interfaces with Fukuda Denshi Central Monitors. SpO2 sensing via integrated Konica Minolta or Nellcor modules. ECG lead selection (3 or 5 electrode).
Indications for Use
Indicated for patients in hospital settings (ER, ICU, clinic) requiring cardiovascular system evaluation via remote central station monitoring. Not for home use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Fukuda Denshi central telemetry receiver (K980728)
Submission Summary (Full Text)
{0}------------------------------------------------
### REVISED 510(k) Summary 1-30-2011
MAR 1 1 2011
## Fukuda Denshi Model LX-7230KM/7230N Transmitter
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR part 870.92.
The assigned 510(k) number is: K103134.
- Submitter: Fukuda Denshi USA, Inc. 17725 NE 65th St. Building C Redmond WA, 98052 Tel: 425-881-7737 Fax: 425-869-2018
- Contact Person: Doug Blakely . Director- Regulatory Affairs -Fukuda Denshi USA, Inc. 17725 NE 65th St. Building C Redmond WA, 98052 Tel: 210-240-4521 Fax: 408-547-4521 Email: DBlakely(@fukuda.com
- Date Prepared: October 8, 2010 .
#### Device Name:
- Proprietary Name : Fukuda Denshi Model LX-7230KM/7230N Transmitter .
- . Common Name: Telemetry Transmitter
- Product Code: MHX ●
- Regulation Number: 21 CFR Part 870.1025 .
II
- Class: .
- . Review Panel: Cardiovascular
{1}------------------------------------------------
### REVISED 510(k) Summary 1-30-2011
# Fukuda Denshi Model LX-7230KM/7230N Transmitter
### Legally Marketed Device:
Fukuda Denshi model LX-5630 Transmitter, #K033711
In addition, the SpO2 measurement function of the LX-7230KM/7230N Transmitter utilizes technology incorporated into previously cleared devices and several OEM manufactured modules that have received separate clearance from the FDA as follows:
The SpO2 measurement technology used in the LX-7230KM is the same as that used in the Konica Minolta model "Pulsox-300/300i" cleared under 510(k) # K053419. We have made no modifications to Konica Minolta SpO2 module or sensors under K053419.
The SpO2 measurement technology used in the LX-7230N is the same as that used in the Nellcor Puritan Bennett model "OxiMax N-600x Pulse Oximeter" cleared under 510(k) # K060576. We have made no modifications to the Nellcor SpO2 module or sensors under K060576.
### Description:
The Fukuda Denshi model LX-7230KM/7230N is a patient worn Transmitter that transmits physiological data such as ECG, respirogram, arterial oxygen saturation (SpO2), plethysmograph, and pulse rate from a patient to a Fukuda Denshi Central Monitor. The front LCD display information such as ECG, heart rate, respirogram, respiration rate, SpO2, plethysmograph, pulse rate, pulse amplitude level, battery level, and the conditions of the ECG electrodes and SpO2 sensor. For the SpO2 measurement, the LX-7230N utilizes Nellcor SpO2 module technology (K060576) and the LX-7230KM utilizes Konica-Minolta SpO2 module technology (K053419). Both transmitters can only be used as an interface device of the previously cleared Fukuda Denshi Central Monitor (K970585, K000746, K020084) utilizing the central telemetry receiver (K980728). Both transmitters utilize digital FSK (frequency shift keying) technology and operate in the WMTS 608 to 614 MHz transmission frequencies. One or two channel ECG waveforms are selectable with lead selection available using the two buttons (Enter and ▽) on the front panel. (In case of using a 3-electrode lead cable or a 5electrode chest lead cable). Both transmitters are battery powered using 2 AA alkaline batteries with available continuous operation for 6 days (LX-7230KM) or 3 days (LX-7230N). The LX-7230KM/7230N is small and lightweight at 190 grams including batteries. The physical dimensions of the device are 72 mm (W) x 98 mm (H) x 24.8(D) mm.
Additional details regarding the device description can be found in the device's Operation Manual (Exhibit # 3) and Section 10, "Device Information (Modified Device) for Fukuda Denshi model LX-7230KM/7230N Transmitter*. Also included in Section 10 are the subject device's Engineering Drawings, such as external appearance.
{2}------------------------------------------------
3/4
## REVISED 510(k) Summary 1-30-2011
## Fukuda Denshi Model LX-7230KM/7230N Transmitter
The modifications to the Fukuda Denshi model LX-5630 Transmitter include the replacement of a later version of Konica Minolta SpO2 module, adding an optional SpO2 module from Nellcor, changes to the overall dimensions of the device, and changes to the ECG lead wire set.
The new modified device, LX-7230KM, utilizes the same fundamental technology of ECG and Respiration measurement as the current patient worn transmitter LX-5630 (K033711) and, for SpO2 measurement, utilizes the latest version SpO2 module built and manufactured by Konica-Minolta and cleared as the PULSOX-300/300i Oximeter (K053419). We have made no modifications to Konica Minolta SpO2 module or sensors under K053419.
The new modified device, LX-7230N, utilizes the same fundamental technology of ECG and Respiration measurement as the current transmitter LX-5630 (K033711) and, for SpO2 measurement, utilizes the SpO2 module built and manufactured by Nellcor and cleared as the N-600x Pulse Oximeter (K060576). We have made no modifications to Nellcor SpO2 module or sensors under K060576
The LX-7230KM/7230N Transmitter weighs 190 grams including batteries and has dimensions of 72(W) x 98(H) x 24.8(D) mm. The modified device is also powered by two (2) AA size batteries as before.
The LX-7230KM/7230N Transmitter is a WMTS compliant device and operates only preprogrammed channels in the 608 to 614 MHz bandwidth as before.
#### Statement of Intended Use:
The DS-7000 Series Telemetry Monitoring System is intended to be used as central station monitoring system for the evaluation of the cardiovascular system. It is intended to be used by or on the order of a physician or similarly qualified healthcare professional. The DS-7000 Series Telemetry Monitoring System is intended to be used in hospital environments: ER, ICU, a clinic or similar settings. The DS-7000 Series Telemetry Monitoring System is intended to be used in those situations where the patient is being monitored by a Fukuda Denshi DS-5000-7000 Series bedside monitor, or patient worn telemetry transmitter where remote, central station monitoring is desired. This system is not intended for home use.
The intended use of the modified device as described in the labeling has not changed as a result of the modifications.
{3}------------------------------------------------
### REVISED 510(k) Summary 1-30-2011
# Fukuda Denshi Model LX-7230KM/7230N Transmitter
### Technological Characteristics:
The Fukuda Denshi model LX-7230KM/7230N Transmitter incorporates the identical technology as the predicate devices. The device provides a means with interfacing with a patient, transmitting specific physiological data to a Fukuda Denshi Central Monitor (K970585, K000746, K020084) utilizing the central telemetry receiver (K980728). and processing the data for display of numeric values and waveforms on the front LCD.
The technology characteristics of the LX-7230KM/7230N Transmitter do not affect the safety or efficacy of the device. Any safety issues raised by a software control medical device are either the same issues already addressed by the predicate devices or are addressed the system hazard analysis. or in the system validation.
#### Testing:
The Fukuda Denshi model LX-7230KM/7230N Transmitter has been subjected to extensive safety, environmental and performance testing. Final testing for the device included various performance tests for the device designed to insure that all functional and performance specifications were met. We have made no modifications to Konica Minolta SpO2 module or sensors under K053419 or to Nellcor SpO2 module or sensors under K060576. Additionally the device was host tested at the previously noted OEM engineering test facility to insure that performance and functional specifications for their supplied module were met.
### Conclusion:
In conclusion, drawing from laboratory testing, validation and risk Analysis, the Fukuda Denshi model LX-7230KM/7230N Transmitter demonstrates that this device is as safe and effective as and performs as well as the legally marketed predicate device, the Fukuda Denshi model LX-5630 Transmitter 510(k) #K033711.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the eagle. The eagle is facing left, and the stripes are angled upwards. The text is in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Fukuda Denshi USA, Inc. c/o Doug Blakely Regulatory Director 17725 NE 65th Street, Bldg. C Redmond, WA 98052
MAR 1 1 2011
Re: K103134
Trade/Device Name: Fukuda Denshi Model LX-7230KM/7230N Transmitter Regulation Number: 21 CFR 870.1025 Regulation Name: Telemetry Transmitters Regulatory Class: Class II (two) Product Code: MHX Dated: February 25, 2011 Received: March 3, 2011
Dear Mr. Blakely,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate – commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I odoral battle and step including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 – Mr. Doug Blakely
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
R. Kirchner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K103134 Device Name: _ Fukuda Denshi Model LX-7230KM/7230N Transmitter Indications For Use:
> The DS-7000 Series Telemetry Monitoring System is intended to be used as central station monitoring system for the evaluation of the cardiovascular system. It is intended to be used by or on the order of a physician or similarly qualified healthcare professional. The DS-7000 Series Telemetry Monitoring System is intended to be used in hospital environments: ER, ICU, a clinic or similar settings. The DS-7000 Series Telemetry Monitoring System is intended to be used in those situations where the patient is being monitored by a Fukuda Denshi DS-5000-7000 Series bedside monitor, or patient worn telemetry transmitter where remote. central station monitoring is desired. This system is not intended for home use.
Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
sumar R.V.limes
Page 1 of
(Division Sign-Off) (Division Sign-Off)
Division of Cardiovascular Devices
Division of Cardiovascular
510(k) Number للكتروما كل
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.