FUKUDA DENSHI DYNASCOPE MODEL DS-5000 CENTRAL TELEMETRY SYSTEM
K033711 · Fukuda Denshi USA, Inc. · MHX · Jun 3, 2004 · Cardiovascular
Device Facts
Record ID
K033711
Device Name
FUKUDA DENSHI DYNASCOPE MODEL DS-5000 CENTRAL TELEMETRY SYSTEM
Applicant
Fukuda Denshi USA, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jun 3, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The DS-5000 Series Telemetry Monitoring System is intended to be uses as a central station monitoring system for the evaluation of the cardiovascular system. It is intended to be used by or on the order of a physician or similarly qualified health care professional. The DS-5000 Series Telemetry Monitoring System is intended to be used in hospital environments; ER, ICU, a clinic or similar settings. The DS-5000 Series Telemetry Monitoring System is intended to be used in those situations where the patient is being monitored by a Fukuda Denshi DS-5000 -7000 Series bedside monitor, or patient worn telemetry transmitter where remote, central station monitoring is desired. This system is not intended for home use.
Device Story
Telemetry monitoring system for cardiovascular evaluation; consists of central receiver, bedside monitors, and patient-worn transmitters (LX-5630). Input signals: ECG, respiration, and SpO2 (via integrated Konica Minolta OEM module). System transmits physiological data to central monitor for display, trending, arrhythmia/ST monitoring, and alarm management. Used in hospital settings (ER, ICU, clinic) by physicians or qualified healthcare professionals. Central monitor provides real-time visualization of waveforms and numeric data; alerts clinicians to patient status changes to facilitate clinical decision-making. Benefits include remote, continuous patient monitoring.
Clinical Evidence
Bench testing only. Verification and validation included electrical safety, EMC, and radio telemetry testing. Pulse oximetry laboratory accuracy testing conducted to verify compliance with OEM predicate device specifications.
Technological Characteristics
Processor-based software-controlled telemetry system. Components: central receiver, bedside monitors, and patient-worn transmitters (LX-5630). LX-5630 integrates Konica Minolta OEM SpO2 module. Connectivity via LAN or direct connection to central monitor. Data transmission via radio telemetry. Software-controlled arrhythmia and ST monitoring.
Indications for Use
Indicated for patients requiring central hemodynamic monitoring in medically supervised healthcare environments (ER, ICU, clinic). Not for home use. Prescription use only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Fukuda Denshi DS-5000 Telemetry Monitoring System (K980728)
Fukuda Denshi DS-5000 series Central Patient Monitor (K970585)
Fukuda Denshi DS-5000 series Central Patient Monitor (K000746)
Fukuda Denshi DS-5000 series Central Patient Monitor (K020084)
Submission Summary (Full Text)
{0}------------------------------------------------
# JUN - 3 2004
## Fukuda Denshi Model DS-5000 Telemetry Monitoring System Special 510(k) Device Modification
# Exhibit B
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR part 870.92
The assigned 510(k) number is (k) 033711
- Submitter: Fukuda Denshi U.S.A. Inc. 17725 NE 65th St. Building C Redmond, WA 98052 Tel: 425-881-7737 Fax: 425-869-2018
| Contact Person: | Larry Walker<br>Regulatory Affairs Manager<br>Redmond, WA 98052<br>Tel: 425-881-7737<br>Fax: 425-869-2018 |
|-----------------|-----------------------------------------------------------------------------------------------------------|
|-----------------|-----------------------------------------------------------------------------------------------------------|
- Date Prepared: May 3, 2004
# Device Name:
| | Proprietary Name: Fukuda Denshi DynaScope<br>Model DS-5000 Central Telemetry System |
|--------------------------|----------------------------------------------------------------------------------------------------------|
| Common Name: | Central Telemetry System |
| Classification: | Arrhythmia Detector and Alarm (§870.1025)<br>Class III |
| Legally Marketed Device: | Fukuda Denshi DS-5000 Telemetry Monitoring<br>System (K 980728)<br>Minolta PULSOX-3Li Oximeter (K010413) |
{1}------------------------------------------------
Exhibit B 510(k) Summary
### Device Description:
The Fukuda Denshi model DS-5000 series Telemetry Monitoring System consists of a series of interface devices which include a central transmitter receiver, an HLX-501 Multi-parameter transmitter, LX-5120 patient worn ECG/Respiration transmitters, and a Fukuda Denshi DS-5000 series Central Patient Monitor (K970585, K000746, K020084). The unit's are processor based software control devices. The receiver module can receive data for 4 or 8 patients and can be connected either directly or by local area network (LAN) to the DS-5000 series Central Patient Monitor. Each patient data when received at the central transmitter receiver is considered as a separate network node. Input signals are provided from the patient worn LX series transmitters or from the HLX-501 Multi-parameter transmitter when connected to a Fukuda Denshi DS-5000 series patient monitor.
Patient physiological data displays, controls, recordings and alarms are controlled from the Central Patient Monitor. Recordings can also be initiated from the bedside monitor or from the patient worn transmitters. System functions such as trending, arrhythmia and ST monitoring and data access are available to the user from the central monitor.
The HLX-501 multi-parameter transmitter may provide up to six waveforms and numeric data from the bedside monitor. Parameter monitored may include ECG, SpO2, Resp, BP, NIBP and Temp.
The patient worn LX-5120 transmitter provides monitoring of ECG and Respiration parameters.
The modified transmitter subject to this submission, the patient worn LX-5630 provides ECG and Respiration monitoring identically as the LX-5120 and adds SpO2 transmission through the integration of an Pulse Oximetry OEM module designed and manufactured by Konica Minolta Sensing Inc and cleared as the Minolta Pulseox-3 LI (K010413).
{2}------------------------------------------------
## Exhibit B 510(k) Summary
#### Statement of Intended Use:
The DS-5000 Series Telemetry Monitoring System is intended to be uses as a central station monitoring system for the evaluation of the cardiovascular system. It is intended to be used by or on the order of a physician or similarly qualified health care professional. The DS-5000 Series Telemetry Monitoring System is intended to be used in hospital environments; ER, ICU, a clinic or similar settings. The DS-5000 Series Telemetry Monitoring System is intended to be used in those situations where the patient is being monitored by a Fukuda Denshi DS-5000 -7000 Series bedside monitor, or patient worn telemetry transmitter where remote, central station monitoring is desired. This system is not intended for home use.
#### Technological Characteristics:
The addition of the model LX-5630 patient worn transmitter, to the DS-5000 series Telemetry Monitoring System does not change the fundamental technology of the unmodified system. Data is transmitted from a telemetry transmitter to a Fukuda Denshi Central Patient Monitor utilizing the same central transmitter receiver. The modification allows the addition of SpO2 monitoring from a patient worn transmitter. The monitoring of SpO2 has been available as part of the DS-5000 series Telemetry Monitoring System, through the use of the multi-parameter transmitter (model HLX-501) when connected to a DS-5000 series bedside monitor. The LX-5630 adds SpO2 monitoring, through the integration of the OEM supplied SpO2 module designed and manufactured by Konica Minolta Sensing Inc as cleared as the Minolta PULSOX - 3Li Pulse Oximetery module (K010413), to the ECG and Respiration monitoring capabilities found in the model LX-5120 patient worn transmitter which was cleared in the original submission.
The technology characteristics of the modified transmitter, Model LX-5360, do not affect the safety or efficacy of the DS-5000 series Telemetry Monitoring System. Any safety issues that may be raised by a software control medical device are either the same issues already addressed during the submission of the unmodified devices or are addressed in the hazard analysis and verification/validation process.
{3}------------------------------------------------
## Exhibit B 510(k) Summary
## Testing:
Verification and validation testing were conducted to establish the performance and reliability characteristics of the modified device. Testing involved safety testing from the risk analysis including electrical safety testing, EMC testing and radio telemetry testing. Additionally pulse oximetry laboratory accuracy testing was conducted to verify compliance to clinical testing results submitted with the OEM produced predicate oximetry device. Acceptance criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence
#### Conclusion:
The conclusion drawn from the verification and validation testing . of the modified DS-5000 Series Telemetry Monitoring System demonstrates that the device is safe and effective and performs as well or better than the legally marketed predicate device(s).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 3 2004
Fukuda Denshi USA, Inc. c/o Mr. Larry D. Walker Regulatory Affairs Manager 17725 N.E. 65th St., Building C Redmond, WA 98052
Re: K033711
KU33711
Trade Name: Fukuda Denshi DS-5000 Telemetry Monitoring System and LX-5630 Transmitter Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: II (two) Product Code: MHX Dated: May 3, 2004 Received: May 5, 2004
Dear Mr. Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the indication we nave reviewed your Security (the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for the stated in the encrosule) to regally management date of the Medical Device American Canada Frank Frank Frank commerce prior to May 20, 1976, the exaccidance with the provisions of the Federal Food, Drug, devices that have been recadsined in accessful of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket a Act . The You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, therefore, market the as received, of the Act include requirements for annual registration, listing of general controls provisions of alle ties labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it II your device is classified (see aberty) me existing major regulations affecting your device can inay be subject to suen additional controllations, Title 21, Parts 800 to 898. In addition, FDA may be found in the extrements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Larry D. Walker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase be actised that I termination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I oderal the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quand of our of troll provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product in allow you to begin marketing your device as described in your Section 510(k) I the letter withing of substantial equivalence of your device of your device to a legally marketed prodicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Neil R.P. Oden
Bram D. Zuckerman, M.D. (
Bram D. Zuckerman, Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Exhibit E Indication for Use
| 510(k) Number if<br>Known | K033711 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Fukuda Denshi DS-5000 Telemetry Monitoring<br>System |
| Indication for Use | The Fukuda Denshi DS-5000 series Telemetry<br>Monitoring System is indicated in those situations<br>where centrally located hemodynamic monitoring<br>(central station monitoring) of one or more patients<br>cardiovascular condition is desired and those<br>patients are currently monitored by a Fukuda<br>Denshi bedside monitor, or is wearing a Fukuda<br>Denshi telemetry transmitter. Use of this device is<br>indicated only in a medically supervised healthcare<br>environment (e.g. ER, ICU or clinic). It is not<br>intended for home use. This device is intended for<br>Prescription Use Only |
× Prescription Use (Part 21 CFR 801 Subpart D)
And/Or
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDHR, Office of Device Evaluation (ODE)

for BD2
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K053711 |
|---------------|---------|
|---------------|---------|
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.