ALBOGRAFT VASCULAR PROSTHESIS

K103080 · LeMaitre Vascular, Inc. · MAL · Jan 19, 2011 · Cardiovascular

Device Facts

Record IDK103080
Device NameALBOGRAFT VASCULAR PROSTHESIS
ApplicantLeMaitre Vascular, Inc.
Product CodeMAL · Cardiovascular
Decision DateJan 19, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3450
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K103080 · Jan 19, 2011ALBOGRAFT VASCULAR PROSTHESISLeMaitre Vascular, Inc.Retrospective clinical review of patient medical recordsThe sponsor provided a retrospective analysis of 66 consecutive patients treated with the AlboGraft vascular prosthesis to demonstrate clinical performance, including patency rates, limb salvage, and Fontaine stage improvement.Retrospective review; Clinical outcomes; Patency rates; Limb salvage

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective clinical review; Follow-up/Duration: Up to 24 months; Study Period: January 1, 2007 to March 31, 200966 consecutive patients with iliac artery aneurysmal or occlusive disease; Sample Size: 66Not applicable for this studyPrimary and secondary patency rates, limb salvage rate, Fontaine stage classification, ABI improvement

Indications for Use

The AlboGraft Double Velour Knitted and Woven Grafts are indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease.

Device Story

AlboGraft Vascular Prosthesis is a knitted or woven tubular polyester fabric impregnated with bovine collagen. Used by surgeons in clinical settings for arterial replacement or repair in patients with aneurysmal or occlusive disease. Device serves as a synthetic conduit to restore blood flow. Clinical benefit includes limb salvage and improved patency in peripheral arterial disease. No electronic or software components; purely mechanical implant.

Clinical Evidence

Retrospective clinical review of 66 patients (mean age 64.8, 66% male) with peripheral arterial occlusive disease (pAOD) undergoing iliac artery reconstruction. Primary endpoints included patency and limb salvage. Results at 24 months: primary patency 89.4%, secondary patency 95.1%, limb salvage 98.4%. ABI improved from 0.46 to 0.89 post-operatively. 30-day mortality 4.5%.

Technological Characteristics

Knitted or woven tubular polyester fabric impregnated with bovine collagen. Tested per ISO 7198 (burst, tensile strength, suture pull, permeability, dimensions) and ISO 10993-1 (biocompatibility).

Indications for Use

Indicated for replacement or repair of arteries affected by aneurysmal or occlusive disease in patients requiring vascular reconstruction.

Regulatory Classification

Identification

A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## SECTION 6: 510(k) SUMMARY KIO308 | Submitter: | LeMaitre Vascular, Inc.<br>63 Second Avenue<br>Burlington, MA 01803 | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Andrew Hodgkinson<br>Vice President, Clinical and Regulatory Affairs<br>Phone: 781-221-2266 x108<br>Fax: 781-425-5049<br>Email: ahodgkinson@lemaitre.com | | Date Prepared: | October 15, 2010 | | Trade Name: | Albograft™ Vascular Prosthesis | | Common Name: | Vascular Prosthesis | | Classification Name: | Graft, vascular, synthetic/biologic composite | | Predicate Device: | Hemashield Woven Double Velour Vascular Graft (K954848) | | Device Description: | Albograft Vascular Prosthesis is a knitted/woven tubular polyester<br>fabric, impregnated with bovine collagen. | | Intended Use: | The AlboGraft Double Velour Knitted and Woven Grafts are<br>indicated for use in the replacement or repair of arteries affected<br>with aneurysmal or occlusive disease. | | Summary of<br>Technological<br>Characteristics: | AlboGraft Polyester Vascular Prosthesis is a knitted or woven<br>tubular polyester fabric, impregnated with bovine collagen. | | Summary of Product<br>Testing: | The following tests have been performed per ISO 7198:<br>Burst Test<br>Longitudinal Tensile Strength<br>Radial Tensile Strength<br>Suture Pull-90°/45°<br>Wall Thickness<br>Integral Water Permeability<br>Visual Inspection<br>Usable Length<br>Relaxed Inner Diameter<br>Pressurized Inner Diameter | | Summary of Pre-<br>clinical Study: | The biocompatibility of the device was tested per ISO10993-1. | | Summary of<br>Retrospective<br>Clinical Review | Patient baseline data, preoperative risk factors, diagnosis and<br>secondary diseases:<br>Between Jan 1, 2007 and March 31, 2009, 66 consecutive patients have<br>received a surgical implant of a peripheral AlboGraft™ vascular<br>prosthesis in the iliac region (prox. anastomosis - common iliac artery /<br>distal anastomosis - proximal superficial femoral artery). The patients<br>consisted mainly of males (n= 44 / 66 %) with a mean age of 64.8 years<br>(Mean: H: 171 cm; W: 74 kg; BMI: 25.2 kg/cm²) and demonstrated typical | : . {1}------------------------------------------------ | risk factors for development of illiac artery peripheral arterial occlusive<br>disease (pAOD).<br>The majority of patients (88%) suffered from arterial hypertension and<br>histories of smoking in 92% of the cases (73% of all patients are still<br>active smokers). Almost half (49 %) of the patients had additional<br>coronary heart disease , while additional risk factors like diabetes mellitus<br>or hyperlipidemia was only present in 9% or 3% of the patients<br>respectively. Chronic obstructive pulmonary disease was present in 29%<br>of the patients and cerebro vascular disease in n= 10 (15%) of which n=8<br>had a previous stroke. Also almost half of the patients (49%) had<br>previous arterial reconstructions. | | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|--------------------------------| | Disease / Diagnosis:<br>4 patient suffered from aneurysmal disease (n=2 iliac artery aneurysms;<br>n=2 suture line aneurysms) and 62 patients reported pAOD of the iliac<br>artery with either severe claudication / Fontaine stage IIb (n=42/68%),<br>rest pain / Fontaine III (n=9/15%) or ulcerations or gangrene / Fontaine<br>stage IV (n=11/18%). (Table 1).<br>The severity of the iliac artery disease could be confirmed by a mean<br>pain free walking distance of only 89m and a mean Ankle Brachial Index<br>(ABI) on the affected side of 0.46.<br>Only patients with TASC II C (n=37 / 60%) and TASC II D (n=25 / 40%)<br>classification of disease morphology were operated on. | | | | Operative and postoperative results summary:<br>The 66 patients received either unilateral bypass reconstructions in n=40<br>(61%) or a cross-over bypass procedures in n= 26 (39%) of the cases. | | | | 30 day mortality was 4.5 % (n=3) and no graft infection has been<br>recorded during the complete follow up period. | | | | Postoperatively the ABI of the affected side improved significantly to 0.89<br>from 0.46 and the majority of patients were in Fontaine stage I and IIa at<br>their last follow up visit. (Table 1). The primary and secondary patency<br>rates at 12 and 24 month have been 88.9%/95.1 % and 88.9%/95.1%<br>respectively. The Limb Salvage Rate was 98.4% at 12 and 24 month<br>(Table 2). | | | | Table 1: Pre and Post Operative (last follow up visit) Fontaine stage<br>Classification for the n= 62 patients with Peripheral Arterial Disease<br>(pAOD). | | | | Fontaine stage | Pre-Op (n=62) | Last follow up<br>visit (n=57) | | I - asymptomatic | 0 (0.0%) | 39 (62.9%) | | Ila- Lifestyle limiting<br>Intermitted Claudication (IC) | 0 (0.0%) | 11 (17.7%) | | IIb - Severe Intermitted<br>Claudication (IC) | 42 (67.7%) | 7 (11.3%) | | III - Restpain | 9 (14.5%) | 0 (0.0%) | | IV - Ulcerations or Gangrene | 11 (17.7%) . | 0 (0.0%) | 11 - 11 . {2}------------------------------------------------ | Table 2: Patency and Limb Salvage Rates in % analyzed by Kapla<br>Meier Survival Analysis: | | | |--------------------------------------------------------------------------------------------|--------------------------------------|-----------------------------------| | n = 66 patients (SE = Standard Error) | | | | | 12 month | 24 month | | Primary Patency<br>Rate | 89.4%<br>(SE 4%) n = 51 at<br>risk | 89.4%<br>(SE 4%) n = 19 at risk | | Secondary<br>Patency Rate | 95.1%<br>(SE 2.8%) n = 55 at<br>risk | 95.1%<br>(SE 2.8%) n = 20 at risk | | Limb Salvage<br>Rate | 98.4%<br>(SE 2.8%) n = 57 at<br>risk | 98.4%<br>(SE 2.8%) n = 22 at risk | Vascular Prosthesis is substantially equivalent to the predicate Conclusion: device based on its indications for use and fundamental scientific technology. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the bird's body and wings. The bird is positioned to the right of the text. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 LeMaitre Vascular, Inc. c/o Mr. Andrew Hodgkinson 63 Second Avenue Burlington, MA 01803 Re: K103080 AlboGraft Double Velour Knitted or Woven Vascular Prosthesis (Peripheral indication) Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular Graft Prosthesis Regulatory Class: Class II Product Code: MAL Dated: October 18, 2010 Received: October 19, 2010 Dear Mr. Hodgkinson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. JAN 1 9 2011 {4}------------------------------------------------ Page 2 - Mr. Andrew Hodgkinson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. . You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours, Duma R. Kirchner Bram D. Zuckerman, M.D. V Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K103080 Device Name: _LeMaitre AlboGraft Double Velour Knitted or Woven Vascular Prosthesis (peripheral indication) Indications For Use: The AlboGraft Double Velour Knitted or Woven Vascular Prosthesis (peripheral indication) are indicated for use in the replacement or repair of arteries affected with aneurismal or occlusive disease. Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) uma R. V. hmes (Division Sign-Off) (Division of Cardiovascular Devices 510(k) Number_K 103080 Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...