K102159 · C.R. Bard, Inc. · LJS · Nov 17, 2010 · General Hospital
Device Facts
Record ID
K102159
Device Name
POWERPICC SV CATHETERS
Applicant
C.R. Bard, Inc.
Product Code
LJS · General Hospital
Decision Date
Nov 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerPICC® SV catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.
Device Story
PowerPICC® SV catheter is a peripherally inserted central catheter (PICC) designed for power injection of contrast media, intravenous therapy, blood sampling, and central venous pressure monitoring. Available in 3F single lumen and 4F dual lumen configurations. Device is inserted peripherally to access the central venous system. Catheter extension leg, luer hub, and junction feature printed markings to facilitate proper use. Provided in sterile kit configurations. Used by clinicians in clinical settings. Enables power injection of contrast media at specified flow rates (1 mL/sec for 3F; 2.5 mL/sec for 4F). Benefits include reliable central venous access for various therapies and diagnostic procedures.
Clinical Evidence
Bench testing only. Verification and validation activities performed on sterilized, finished devices per ISO 10555-1, ISO 10555-3, and ISO 10993 standards. FMEA conducted per ISO 14971. Results met all predetermined acceptance criteria.
Technological Characteristics
Percutaneous, long-term intravascular catheter. Materials evaluated per ISO 10993. Sterilization via ethylene oxide per ISO 11135. Configurations: 3F single lumen and 4F dual lumen. Features printed markings on extension leg, luer hub, and junction. Power injection capable.
Indications for Use
Indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and central venous pressure monitoring (recommended 20 gauge or larger lumen).
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for BARD ACCESS SYSTEMS. The word "BARD" is in large, bold, sans-serif font, with a unique design where the letters are interconnected. Below "BARD" is the text "ACCESS SYSTEMS" in a smaller, sans-serif font.
.
:
…
K102159 NOV 1 7 2010
## 510(k) Summary 21 CFR 807.92(a)
## PowerPICC® SV Catheter
| General<br>Provisions | Submitter Name:<br>Address: | Bard Access Systems, Inc.<br>605 North 5600 West<br>Salt Lake City, UT 84116 |
|-----------------------|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Jessica Agnello |
| | Telephone Number:<br>Fax Number:<br>Date of Preparation: | Regulatory Affairs Specialist<br>(801) 522-5651<br>(801) 522-5425<br>November 4, 2010 |
| Subject<br>Device | Trade Name:<br>Common Name:<br>Classification Name: | PowerPICC® SV Catheter<br>Peripherally Inserted Central Catheter (PICC)<br>Percutaneous, implanted, long-term intravascular<br>catheter<br>LJS, 21 CFR § 880.5970 Class II |
| Predicate<br>Devices | Trade Name:<br>Classification Name: | Poly Per-Q-Cath® PICC Catheter<br>Percutaneous, implanted, long-term intravascular<br>catheter |
| | Premarket Notification:<br>Manufacturer: | LJS, 21 CFR §880.5970 Class II<br>K034019, K051991<br>Bard Access Systems, Inc. |
| | Trade Name:<br>Classification Name: | 6F Triple Lumen Poly Per-Q-Cath® PICC Catheter<br>Percutaneous, implanted, long-term intravascular<br>catheter |
| | Premarket Notification:<br>Manufacturer: | LJS, 21 CFR §880.5970 Class II<br>K043502, K051991<br>Bard Access Systems, Inc. |
| | Trade Name:<br>Classification Name: | 4F Single Lumen PowerPICC® Catheter<br>Percutaneous, implanted, long-term intravascular<br>catheter, |
| | Premarket Notification:<br>Manufacturer: | LJS, 21 CFR §880.5970 Class II<br>K070996<br>Bard Access Systems, Inc |
| | Trade Name:<br>Classification Name: | 5F Single Lumen PowerPICC® Catheter<br>Percutaneous, implanted, long-term intravascular<br>catheter, |
| | Premarket Notification:<br>Manufacturer: | LJS, 21 CFR §880.5970 Class II<br>K033389, K051991<br>Bard Access Systems, Inc |
:5
·
{1}------------------------------------------------
| Trade Name: | 6F Triple Lumen PowerPICC® Catheter | | | | | | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| Classification Name: | Percutaneous, implanted, long-term intravascular catheter,<br>LJS, 21 CFR §880.5970 Class II | | | | | | |
| Premarket Notification: | K053501, K051991 | | | | | | |
| Manufacturer: | Bard Access Systems, Inc | | | | | | |
| Trade Name: | PowerLine® Catheters | | | | | | |
| Classification Name: | Percutaneous, implanted, long-term intravascular catheter,<br>LJS, 21 CFR §880.5970 Class II | | | | | | |
| Premarket Notification: | K093927 | | | | | | |
| Manufacturer: | Bard Access Systems, Inc | | | | | | |
| Trade Name: | Morpheus® CT PICC Catheter | | | | | | |
| Classification Name: | Percutaneous, implanted, long-term intravascular catheter,<br>LJS, 21 CFR §880.5970 Class II | | | | | | |
| Premarket Notification: | K072196 | | | | | | |
| Manufacturer: | AngioDynamics, Inc | | | | | | |
| Device Description | The subject PowerPICC® SV catheters are members of Bard Access<br>Systems' PowerPICC® series of power injectable catheters. The<br>PowerPICC® SV catheters are available in 3F single lumen and 4F dual<br>lumen configurations. The catheter extension leg, luer hub and junction are<br>printed with description markings to facilitate proper use of the device.<br>The PowerPICC® SV catheters are provided in sterile kit configurations. | | | | | | |
| Intended Use | The PowerPICC® SV catheters are intended for short or long term peripheral<br>access to the central venous system for intravenous therapy and blood<br>sampling. | | | | | | |
| Indications For Use | The PowerPICC® SV catheter is indicated for short or long-term peripheral<br>access to the central venous system for intravenous therapy, blood sampling,<br>power injection of contrast media, and allows for central venous pressure<br>monitoring. For central venous pressure monitoring, it is recommended that a<br>catheter lumen of 20 gauge or larger be used. | | | | | | |
| | Catheter Size Maximum Flow Rate 3 F Single Lumen 1 mL/sec 4 F Dual Lumen 2.5 mL/sec | | | | | | |
| Technological Characteristics | Technological characteristics of the subject PowerPICC® SV catheters are<br>equivalent with respect to the basic catheter design and function to those of<br>the predicate Poly Per-Q-Cath® Catheters. Technical characteristics based<br>on the subject device's power injection capabilities are comparable to the<br>predicate 4F SL PowerPICC® catheter and the Morpheus® CT PICC<br>catheter. Differences do not raise any new questions regarding safety and | | | | | | |
:
.
the submit the state of the state of the states and
{2}------------------------------------------------
| Safety &<br>Performance<br>Tests | Verification and validation activities were designed and performed to demonstrate that the subject PowerPICC® SV catheters met predetermined performance specifications. Tests were performed on sterilized, finished devices. The following guidance documents and standards in conjunction with in-house protocols were used to determine appropriate methods for evaluating the performance of the device:<br>• Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995<br>• ISO 10555-1:2009, Sterile, single-use intravascular catheters, Part 1. General requirements<br>• BS/EN/ISO 10555-3:1996/Cor 1:2002, Sterile, single-use intravascular catheters, Part 3. Central venous catheters<br>• AAMI/ANSI/ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, and the FDA Modified ISO 10993 Test Profile<br>• AAMI/ANSI/ISO 10993-7:2008, Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Results<br>• AAMI/ANSI/ISO 11135:2007, Sterilization of Healthcare Products Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The subject PowerPICC® SV catheters met all predetermined acceptance criteria derived from the above mentioned references. Design validation was conducted on the subject PowerPICC® SV configuration and yielded acceptable results. |
| | Risk management, including a failure modes and effects analysis (FMEA), of the subject device was conducted in accordance with ISO 14971:2007, Medical Devices - Risk Management for Medical Devices. No new types of safety or efficacy questions were identified for the subject PowerPICC® SV catheters. |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and safety and performance testing, the subject PowerPICC® SV catheters met the minimum requirements for its intended use/indications for use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available catheters/cited predicates. |
.
.
.
:
7
.
.
.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol composed of three curved lines that resemble a human figure or abstract design. The logo is black and white and appears to be a scanned or printed image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Jessica Agnello Regulatory Affairs Specialist C.R. Bard, Incorporated 605 North 5600 West Salt Lake City, Utah 84116
NOV . 1 7 2010
Re: K102159
Trade/Device Name: PowerPICC® SV Catheter Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: November 4, 2010 Received: November 5, 2010
Dear Ms. Agnello:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Ms. Agnello
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
I. Hata for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, general Hospital, Infection Control and dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
NOV 1 7 2010
510(k) Number (if known):
K102159
Device Name:
PowerPICC® SV Catheter
Indications for Use:
The PowerPICC® SV catheter is indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.
| Catheter Size | Maximum Flow Rate |
|------------------|-------------------|
| 3 F Single Lumen | 1 mL/sec |
| 4 F Dual Lumen | 2.5 mL/sec |
V Prescription Use (Part 21 CFR §801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR §801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rhd C. Chape 11/17/10
Page 1 of 1 (Division Sinn-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Numbar.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.