6F TRIPLE LUMEN POWERLINE CENTRAL VENOUS CATHETER, 5F DUAL LUMEN POWERLINE CENTRAL VENOUS CATHETER
K093927 · C.R. Bard, Inc. · LJS · Feb 4, 2010 · General Hospital
Device Facts
Record ID
K093927
Device Name
6F TRIPLE LUMEN POWERLINE CENTRAL VENOUS CATHETER, 5F DUAL LUMEN POWERLINE CENTRAL VENOUS CATHETER
Applicant
C.R. Bard, Inc.
Product Code
LJS · General Hospital
Decision Date
Feb 4, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PowerLine™ catheters are indicated for short or long term access to the central venous system for intravenous therapy and blood sampling.
Device Story
PowerLine™ central venous catheters are open-ended, reverse-taper polyurethane catheters available in 5F dual-lumen and 6F triple-lumen configurations. Designed for short or long-term central venous access, they facilitate IV therapy, blood sampling, parenteral nutrition, and power injection of contrast media. The device features a tissue ingrowth cuff for stabilization, depth markings for insertion reference, and purple colorants to identify power-injectability. Used in clinical settings by healthcare professionals, the catheter allows for central venous pressure monitoring. By providing reliable venous access, the device supports critical care and diagnostic imaging procedures, enabling efficient delivery of fluids and medications while minimizing the need for repeated venipuncture.
Clinical Evidence
Bench testing only. Verification and validation activities were performed on sterilized, finished devices in accordance with ISO 10555-1, ISO 10555-3, and ISO 10993 standards. Design validation confirmed the device met all predetermined performance specifications. No clinical data was required or presented.
Technological Characteristics
Polyurethane catheter; reverse taper design; 5F dual-lumen or 6F triple-lumen configurations; 50 cm usable length; tissue ingrowth cuff; purple colorants for power-injection identification; sterile tray packaging. Standards: ISO 10555-1, ISO 10555-3, ISO 10993-1, ISO 10993-7, ISO 14971.
Indications for Use
Indicated for short or long-term central venous access in patients requiring IV fluid, blood product, drug, or parenteral nutrition administration, blood withdrawal, power injection of contrast media, or central venous pressure monitoring. Recommended lumen size for CVP monitoring is 20 gauge or larger; max infusion rate 5ml/sec.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Bard Access Systems
PowerLine Central Venous Catheters
Special 510(k) Premarket Notification
K093927
Section 5 – 510(k) Summary
Page 2 of 5
## Section 5
## 510(k) Summary 21 CFR 807.92(a)
FEB - 4 2010
PowerLine™ Central Venous Catheters
| General<br>Provisions | Submitter of 510(k)<br>Premarket Notification: | Bard Access Systems, Inc. (BAS)<br>[Subsidiary of C.R. Bard, Inc.]<br>Salt Lake City, Utah 84116<br>Phone: (801) 522-5696<br>Fax: (801) 522-5425 | | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--|
| | Contact Person: | Sarabjyot Mankoo<br>Regulatory Affairs Specialist | | |
| | Device Trade Name:<br>Device Generic Name: | 5 F DL & 6 F TL PowerLine™<br>Central Venous Catheter (CVC) | | |
| Predicate<br>Device | Primary Predicate: | | | |
| | Trade Name: | PowerLine™ | | |
| | Common/Usual Name: | Central Venous Catheter | | |
| | Classification Name: | 80LJS - Long term Intravascular Catheter | | |
| | CFR Reference: | 21 CFR §880.5970- Class II | | |
| | Classification Panel: | General Hospital | | |
| | Premarket Notification: | See table below | | |
| | Secondary Predicate: | | | |
| Trade Name: | PowerPICC™ | | | |
| Common/Usual Name: | Peripherally Inserted Central Catheter (PICC) | | | |
| Classification Name: | 80LJS – Long term Intravascular Catheter | | | |
| CFR Reference: | 21 CFR §880.5970– Class II | | | |
| Classification Panel: | General Hospital | | | |
| Premarket Notification: | See table below | | | |
| | Predicate Device Name | 510(k) | Concurrence Date | |
| | 5 F Single Lumen (SL) PowerLine™ | K050185 | May 26, 2005 | |
| | 6 F Dual Lumen (DL) PowerLine™ | K051417 | June 30, 2005 | |
| | CVP Monitoring - PICC & CVC | K051991 | October 20, 2005 | |
| | 5F Dual Lumen (DL) PowerPICC™ | K051672 | November 23, 2005 | |
| | 6F Triple Lumen (TL) PowerPICC™ | K053501 | January 13, 2006 | |
| Classification | 21 CFR §880.5970- Class II<br>80LJS – Long term Intravascular Catheter | | | |
| Performance<br>Standards | Performance standards have not been established by FDA under section 514<br>of the Federal Food, Drug and Cosmetic Act | | | |
なる。
ﺮ ﺩ
{1}------------------------------------------------
| Intended Use | PowerLine™ catheters are indicated for short or long term access to the<br>central venous system for intravenous therapy and blood sampling. |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The PowerLine™ catheter is indicated for short or long term access to the<br>central venous system. PowerLine™ catheters are designed for the<br>administration of I.V. fluids, blood products, drugs, parenteral nutrition<br>solutions, as well as blood withdrawal. In addition, PowerLine™ catheters<br>allow for power injection of contrast media and central venous pressure<br>monitoring. The maximum recommended infusion rate is 5ml/sec. For<br>central venous pressure monitoring, it is recommended that a catheter lumen<br>of 20 gauge or larger be used. |
| Device<br>Description | The PowerLine™ catheters are open-ended polyurethane catheters with a<br>reverse taper design. Catheters are available in 5 F Dual-Lumen and the 6 F<br>Triple-Lumen configurations. The usable length of the catheter is 50 cm.<br>Catheter shaft tubing is marked with depth indicators, with "0" indicated to<br>serve as a reference for the catheter insertion point. The catheters have a<br>tissue ingrowth cuff bonded to the catheter shaft. Purple colorants added to<br>the catheter extension legs and shaft aid in distinguishing the catheter as<br>power-injectable. The molded hub is labeled to identify the catheter as<br>PowerLine™. The catheter extension leg and clamp are labeled with<br>information to facilitate proper use of the device. The PowerLine™ catheters<br>are provided in sterile tray configurations. |
| Technological<br>Characteristics | Technological similarities between the subject PowerLine™ catheters and<br>the predicate device remain identical. There are no new questions raised<br>regarding safety or efficacy of the subject PowerLine™ catheters. |
| Verification &<br>Validation<br>Activities | Verification and validation activities were designed and performed to<br>demonstrate that the subject PowerLine™ catheters met predetermined<br>performance specifications. Tests were performed on sterilized, finished<br>devices. The following guidance documents and standards in conjunction with<br>in-house protocols were used to determine appropriate methods for<br>evaluating the performance of the device:<br>Guidance on Premarket Notification [510(k)] Submission for Short-<br>Term and Long-Term Intravascular Catheters, March 16, 1995 ISO 10555-1:1997, Sterile, single-use intravascular catheters, Part 1.<br>General requirements, Amendment 2 BS/EN/ISO 10555-3:1997, Sterile, single-use intravascular catheters,<br>Part 3. Central venous catheters AAMI/ANSI/ISO 10993-1:2003, Biological Evaluation of Medical<br>Devices Part 1: Evaluation and Testing, and the FDA Modified ISO<br>10993 Test Profile AAMI/ANSI/ISO 10993-7:1995, Biological Evaluation of Medical<br>Devices Part 7: Ethylene Oxide Sterilization Results |
| | The subject PowerLine™ catheters met all predetermined acceptance criteria<br>derived from the above mentioned references. Design validation was<br>conducted on the subject PowerLine™ configuration and yielded acceptable<br>results. |
| | Risk management, including a failure modes and effects analysis (FMEA), of<br>the subject device was conducted in accordance with ISO 14971:2007,<br><i>Medical Devices - Risk Management for Medical Devices</i> . No new types of<br>safety or efficacy questions were identified for the subject PowerLine™<br>catheters. |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and safety and<br>performance testing, the subject PowerLine™ catheters met the minimum<br>requirements for its intended use/indications for use, and is substantially<br>equivalent in design, materials, sterilization, principles of operation and<br>indications for use to current commercially available catheters/cited<br>predicates. |
.
{2}------------------------------------------------
·
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Sarabjyott Mankoo Regulatory Affairs Specialist C. R. Bard, Incorporated Bard Access Systems 605 North 5600 West Salt Lake City, Utah 84116
FEB - 4 2010
Re: K093927
Trade/Device Name: PowerLine™ Central Venous Catheters Regulation Number: 21 CFR 880.5970 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II
Product Code: LJS Dated: December 17, 2010 Received: January 5, 2010
Dear Mr. Mankoo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Mr. Mankoo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours,
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Bard Access Systems PowerLine Central Venous Catheters Special 510(k) Premarket Notification Section 4 - Indications for Use Statement
Page 14 of 92
510(k) Number (if known):
Device Name:
PowerLine™ Central Venous Catheters
Indications for Use:
The PowerLine™ catheter is indicated for short or long term access to the central venous system. PowerLine™ catheters are designed for the administration of I.V. fluids, blood products, drugs, parenteral nutrition solutions, as well as blood withdrawal. In addition, PowerLine "" catheters allow for power iniection of contrast media and central venous pressure monitoring. The maximum recommended infusion rate is 5ml/sec. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.
Prescription Use V (Part 21 CFR §801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR §801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093227
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.