K101788 · Ohmeda Medical · FMZ · Jul 21, 2010 · General Hospital
Device Facts
Record ID
K101788
Device Name
GIRAFFE OMNIBED
Applicant
Ohmeda Medical
Product Code
FMZ · General Hospital
Decision Date
Jul 21, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Device Story
Giraffe OmniBed is a combination neonatal incubator and radiant warmer. The subject modification adds the 'Giraffe Shuttle,' a mobile power source accessory. The Shuttle attaches to the bed base to provide electrical power during intra-hospital transport (e.g., between NICU, Radiology, OR). It features a mechanical locking mechanism (pedal-operated) and an internal power source consisting of two 12V 42Ah lead-acid batteries. The Shuttle includes sensors for interference detection and attachment status, with an LED display for battery health and runtime. The device is operated by hospital staff. By providing continuous power, the Shuttle enables uninterrupted thermal regulation and monitoring during patient movement, reducing the need for transfer to separate transport incubators and minimizing handling-related clinical risks. It is not intended for use outside the hospital building.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, requirements/design reviews, module and system-level verification, performance/safety testing, and usability validation.
Technological Characteristics
Combination incubator/radiant warmer with mobile power accessory (Giraffe Shuttle). Shuttle utilizes two 12V 42Ah sealed lead-acid batteries in series (24V nominal). Features mechanical locking pedal, LED display for battery status, and sensors for interference/attachment detection. Connectivity is limited to power delivery to the bed and auxiliary equipment. Biocompatibility and safety testing performed per voluntary standards.
Indications for Use
Indicated for neonates unable to thermo-regulate based on their own physiology. Provides controlled heat via incubator or radiant warmer modes and optional servo-controlled oxygen delivery (21-65%).
Regulatory Classification
Identification
A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.
{0}------------------------------------------------
K10/788
1 of 4
JUL 2-1 2010
Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The circle has decorative swirls around the letters, and small dots are placed around the perimeter of the circle.
GE Healthcare Special 510(k) Premarket Notification Submission 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | June 16 2010 |
|-----------------------------------|-----------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare, Ohmeda Medical<br>8880 Gorman Rd.<br>Laurel, MD 20723 |
| Primary Contact Person: | Agata Smieja<br>8880 Gorman Rd.<br>Laurel, MD 20723<br>Tel: 410-456-0329<br>Fax: 410 888 0544 |
| Secondary Contact Person: | Andrew Ahn<br>8880 Gorman Rd.<br>Laurel, MD 20723<br>Tel: 410-888-5332<br>Fax: 410 888 0544 |
| Device:<br>Trade Name: | Giraffe OmniBed |
| Common/Usual Name: | OmniBed |
| Classification Names: | Incubator, Neonatal: FMZ (880.5400) |
| Product Code: | Warmer, Infant Radiant: FMT (880.5130) |
| Predicate Device(s): | Giraffe OmniBed |
| Device and Change<br>Description: | Legally Marketed Device<br>Giraffe OmniBed is a combination device |
hat can function as an incubator (in the closed mode) or as an infant radiant warmer (in the open mode), based on the user's selection. The device cannot function in both modes at the same time.
Giraffe OmniBed can be used with the Giraffe Uninterruptible Power Supply (UPS) accessory. The UPS is an accessory that mounts to the Giraffe OmniBed as well as the Giraffe Incubator and the Giraffe and Panda Warmers. Giraffe UPS provides uninterruptible power to the bed for quick trips. It also allows thermally supported, cordless mobility during periods of power uncertainty. Whether moving between bed places, or between care areas, the UPS for Giraffe and Panda offers uninterrupted power to the bed and display for continuous monitoring of the baby's situation, and bed controls.
When used with the UPS, the Giraffe OmniBed are is not intended for use as transport incubator or to be taken outside of the hospital
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. There are decorative swirls or flourishes around the outer edge of the circle, adding a stylized touch to the design.
K101788
2 of 4
GE Healthcare
Special 510(k) Premarket Notification Submission building.
Description of Device Modification
The proposed modification of the Giraffe OmniBed is the addition of the Giraffe Shuttle accessory.
The Giraffe OmniBed can be used with the Giraffe Shuttle, a mobile power source that allows for transport of the patient between care areas within the hospital building and provides power to the OmniBed. These areas include, but are not limited to Labor and Delivery, NICU, Radiology, and Operating Room.
This eliminates the need to transfer the infant to and from a transport incubator for transport within the hospital building, reducing the potential for clinical problems associated with patient touch, handling and movement. This reduces the potential for clinical problems associated with intra-hospital transport that result from interrupted patient thermal regulation. .
Similarly to the UPS accessory, the Shuttle facilitates the mobility of the Giraffe OmniBed within the hospital building. Giraffe OmniBed when used with the UPS or with the Shuttle is not a transport incubator and is not intended to be used outside of the hospital building.
The Shuttle has two primary active functions: Locking to a bed and providing transportable power to the bed and accessories.
Attaching the Shuttle to a bed is accomplished by guiding the Shuttle between the legs of the bed and stepping on the Lock pedal. This effectively locks the Shuttle and bed together as one unit.
To detach the Shuttle from the bed, the locking arms are rotated back to their unlocked position by stepping on the Unlock pedal. Once unlocked, the Shuttle can be moved away from the bed.
The internal power source of the Shuttle consists of two 12 volt 42 amp hour lead acid batteries. The batteries are high capacity, sealed, no-maintenance batteries and are connected in series to provide a nominal 24 volts supply to the power generation module.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters and the inner part of the circle are black, while the outer part of the circle is white.
K101788
3 of 4
## GE Healthcare
## Legally Marketed Device Intended Use:
Special 510(k) Premarket Notification Submission The intended use of the legally marketed device and the proposed modified device is identical. The intended use is as follows:
The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.
Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment.
This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Device Modification Technology:
The Giraffe Shuttle is a transportable power source that is an accessory to the Giraffe OmniBed. The Shuttle connects to the bed and provides electrical power to the bed and other auxiliary equipment, required for patient care during transport.
The Shuttle is configured to attach to the Giraffe OmniBed and provide electrical power to the bed and selected accessories. The Shuttle makes it possible to deliver continuous baby care during transport from one hospital area to another (e.g. from L&D to NICU).
The Shuttle is designed to accept all Shuttle-specific accessories from GE Healthcare, including: the cord wrap bracket and gas cylinder holder.
The Shuttle has two primary active functions: Locking to a bed and providing transportable power to the bed and accessories.
The Shuttle contains two sensor systems:
- One system detects an interference condition.
- The other system determines the attaching status of the device.
The Shuttle features an LED Display Board, which contains the battery runtime indicator and the battery health indicator.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. There are four decorative elements resembling small swirls or flourishes positioned symmetrically around the circle, enhancing the logo's visual appeal. The logo is presented in black against a white background, providing a clear contrast.
K101788
4 of 4
GE Healthcare Special 510(k) Premarket Notification Submission
Summary of Non-Clinical Tests: Determination of Substantial Equivalence:
> The Giraffe OmniBed and its applications with the Giraffe Shuttle, complies with voluntary standards as detailed in Section 9, 11 and 17 The following quality assurance of this premarket submission. measures were applied to the development of the system:
- . Risk Analysis
- Requirements Reviews .
- . Design Reviews
- Testing on unit level (Module verification) ●
- � Integration testing (System verification)
- . Performance testing (Verification)
- . Safety testing (Verification)
- Usability testing (Validation) .
- Biocompatibility testing
## Clinical Tests:
The subject of this premarket submission, Giraffe OmniBed used with Shuttle, did not require clinical studies to support substantial equivalence.
Conclusion: GE Healthcare considers the modified Giraffe OmniBed, used with the Giraffe Shuttle accessory to be as safe, as effective, and performance is substantially equivalent to the legally marketed Giraffe OmniBed.
25
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three wavy lines extending from its wing, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## JUL 2-1 2010
Mr. Agata Smieja Regulatory Affairs Director OHMEDA Medical 8880 Gorman Road Laurel, Maryland 20723
Re: K101788
Trade/Device Name: Giraffe OmniBed Regulation Number: 21 CFR 880.5400 Regulation Name: Neonatal Incubator Regulatory Class: II Product Code: FMZ Dated: June 16, 2010 Received: June 25, 2010
Dear Mr. Smieja:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Smieja
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. watson, B.S., M.S.m M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and connected, creating a recognizable emblem for the company.
GE Healthcare Special 510(k) Premarket Notification Submission
510(k) Number (if known):
INSERT 510(k). NUMBER IF KNOWN
Device Name: Giraffe OmniBed
Indications for Use:
The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.
Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment.
This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Prescription Use__x___________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
e of Device Evaluation (ODE)
7/20/10
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Numb
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.