SENTEC DIGITAL MONITOR, V-SIGN SENSOR AND V-SIGN SENSOR 2 WITH DIGITAL SENSOR ADAPER CABLE
K101690 · Sentec AG · LKD · Dec 3, 2010 · Anesthesiology
Device Facts
Record ID
K101690
Device Name
SENTEC DIGITAL MONITOR, V-SIGN SENSOR AND V-SIGN SENSOR 2 WITH DIGITAL SENSOR ADAPER CABLE
Applicant
Sentec AG
Product Code
LKD · Anesthesiology
Decision Date
Dec 3, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2480
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SenTec Digital Monitoring System - consisting of the SenTec Digital Monitor (SDM). Sensors and Accessories - is indicated for continuous, non-invasive patient monitoring. The SenTec Digital Monitoring System is indicated for use in hospital-type facilities, intra-hospital transport environments, and - if under clinical supervision - home environments. The SenTec Digital Monitoring System is for prescription use only. V-Sign™ Sensor, model VS-A/P, and V-Sign™ Sensor 2, model VS-A/P/N, are indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are reguired for adult and pediatric patients. In neonatal patients the use of V-Sign™ Sensor and of V-Sign™ Sensor 2 is indicated for carbon dioxide tension monitoring only. SenTec's Ear Clip, model EC-MI, is intended for use with V-Sign™ Sensor or V-Sign™ Sensor 2 when continuous, non-invasive carbon, oxygen saturation and pulse rate monitoring are required. The Ear Clip is for single-patient use and is indicated to attach V-Sign™ Sensor or V-Sign™ Sensor 2 to the ear lobe of the patient. The use of the Ear Clip is contraindicated for patients whose earlobes are too small to ensure adequate sensor application. SenTec's Multi-Site Attachment Rings, model MAR-SF and model MAR-MI, are intended to attach V-Sign™ Sensor or V-Sign™ Sensor 2 to conventional measurement sites for carbon dioxide tension monitoring when continuous, non-invasive carbon dioxide tension monitoring is required for adult, pediatric, and neonatal patients. They are intended to attach the V-Sign™ Sensor 2 to the forehead or the cheek when continuous, non-invasive carbon dioxide tension, oxygen saturation, and pulse rate monitoring is required for adult and pediatric patients. The Multi-Site Attachment Rings, model MAR-MI, are for single use. SenTec's multi-compatible and reusable SpO2 Soft Sensors, models RSS-L, RSS-M and RSS-S, are indicated for use with the monitoring devices indicated in the respective sensor directions for use when continuous non invasive monitoring of oxygen saturation, and pulse rate are required for patients weighing more than 20 kg. V-STATS™ is an optional PC-based software which is indicated for use with the SenTec Digital Monitor (SDM) when remote monitoring and/or trend reporting and statistical analysis of the data measured by the SDM is required. V-STATS™ is not intended to provide diagnosis, it is intended to supplement and not to replace any part of the SDM monitoring procedures. Caution: Federal law restricts this device to sale by or on the order of a physician.
Device Story
System monitors PCO2, SpO2, and pulse rate via ear lobe, forehead, cheek, or digit sensors. Uses heated Severinghaus-type PCO2 electrode and 2-wavelength transmissive/reflective SpO2 sensors. SDM monitor processes signals, manages sensor temperature/site time, and provides alarms. V-STATS PC software enables remote monitoring, trend reporting, and statistical analysis. Used in hospitals, transport, and supervised home settings by clinicians. Output aids clinical decision-making through continuous physiological data and trend analysis. Benefits include non-invasive, continuous monitoring of respiratory and circulatory status.
Clinical Evidence
Clinical studies conducted on healthy adult volunteers under progressive induced hypoxia. Reference: arterial blood samples via CO-Oximeter. Results confirmed SDM saturation values meet accuracy requirements for all supported sensors and sites.
Technological Characteristics
Heated PCO2/SpO2 sensors (Severinghaus-type, 2-wavelength reflective/transmissive). Materials compliant with ISO 10993-1:2003. Connectivity: SDM monitor with V-STATS PC software for data management. Firmware-controlled sensor temperature and site time management. Software: Moderate level of concern.
Indications for Use
Indicated for continuous, non-invasive monitoring of PCO2, SpO2, and pulse rate in adult and pediatric patients. Neonatal patients indicated for PCO2 monitoring only. SpO2 Soft Sensors indicated for patients >20 kg. Contraindicated for patients with earlobes too small for Ear Clip application.
Regulatory Classification
Identification
A cutaneous carbon dioxide (PcCO2) monitor is a noninvasive heated sensor and a pH-sensitive glass electrode placed on a patient's skin, which is intended to monitor relative changes in a hemodynamically stable patient's cutaneous carbon dioxide tension as an adjunct to arterial carbon dioxide tension measurement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Cutaneous Carbon Dioxide (PcCO2) and Oxygen (PcO2) Monitors; Guidance for Industry and FDA.” See § 868.1(e) for the availability of this guidance document.
Bernoulli™ Management System, Oxinet III and Lifegard Vue (K091461)
Submission Summary (Full Text)
{0}------------------------------------------------
י . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Section 5
# Section 5 – Summary of Safety & Effectiveness อย - 3 200
# 510(k) Summary
Submitter [807.92(a)(1)]:
Address
Phone FAX
SenTec AG Ringstrasse 39 4106 Therwil - Switzerland +41 61 726 97 60
Contact Person [807.92(a)(1)]:
Ms. Sabine Nieba, Ph.D. Quality & Regulatory Affairs Manager nieba@sentec.ch
Date prepared [807.92(a)(1)]:
April 13, 2010
+41 61 726 97 61
# Trade, Common and Classification Name [807.92(a)(2)]:
| Trade / Proprietary Name | Common /<br>Usual Name | Classification<br>Name | Product<br>Code | Class | Regulation<br>Number |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|----------------------------------------|-----------------|-------------------------|-------------------------|
| SenTec Digital Monitoring<br>System (SDMS)<br>SenTec Digital Monitor (SDM),<br>sensors, accessories for sensor<br>application/maintenance; V-<br>STATS TM PC-based trend reporting,<br>data analysis and remote monitoring<br>Software for the SDM | Cutaneous Carbon<br>Dioxide Monitor | Monitor, Carbon-<br>Dioxide, Cutaneous | 73 LKD | II | 21 CFR Part<br>868.2480 |
| Pulse Oximeter | Oximeter | 74 DQA | II | 21 CFR Part<br>870.2700 | |
| Pulse Oximeter | Oximeter, Ear | 74 DPZ | II | 21 CFR Part<br>870.2710 | |
# Substantially Equivalent to [807.92(a)(3)]:
| | Device | Manufacturer | Comment |
|---------|----------------------------------------------------|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K071672 | SenTec Digital Monitoring<br>System | SenTec AG | This device is predicate for PCO2, SpO2, and PR<br>monitoring using the ear lobe as monitoring site in<br>adult and pediatric patients, for PCO2 only<br>monitoring in adult, pediatric and neonatal patients<br>using conventional PCO2 measurement sites as well<br>as for the new V-Sign™ Sensor 2. |
| K043003 | TCM4/40 Monitoring<br>System | Radiometer Medical<br>ApS | The TOSCA 500 is predicate for PCO2, SpO2, and PR<br>monitoring using the forehead and cheek as a<br>monitoring site in adult and pediatric patients as well<br>as for the new parameter 'Pulsation Index'. |
| K063434 | TOSCA 500 tcPCO2, SpO2<br>and PR Monitoring system | Radiometer Basel AG | These two devices are predicates for the expanded<br>sensor temperature and site time management as<br>well as for the new parameter 'Heating Power' being<br>introduced by the modified SDM firmware. |
| K060675 | Reusable EnviteC SoftTipY<br>SpO2 Sensors | EnviteC-Wismar GmbH | These two devices are predicate for conventional<br>SpO2 and PR monitoring on a patient's digit using<br>SenTec's new SpO2 Soft Sensor being connected to<br>the SDM with SenTec's new SpO2 Adapter Cable. |
| K991823 | N-395 Pulse Oximeter | Nellcor Puritan<br>Bennett, Inc. | |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the word "sentec" in a bold, sans-serif font. The letters are closely spaced together, creating a compact and unified appearance. The overall impression is clean and modern.
# SenTec Digital Monitoring System 510(k) Notification
| | Device | Manufacturer | Comment |
|---------|-----------------------------------------------------------------|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K961450 | Score Software<br>(PC Software) | Nellcor Puritan<br>Bennett, Inc. | These two devices are predicate for V-STATS™<br>(SenTec's optional PC software). Score Software is<br>predicate for V-STATS™'s claim to download data<br>from the internal memory of the SDM for subsequent<br>display, analysis and reporting. Oxinet III software is |
| K091461 | Bernoulli™ Management<br>System, Oxinet III and<br>Lifegard Vue | Cardiopulmonary<br>Corporation | predicate for V-STATS™'s claim for remote<br>monitoring. [Note: Nellcor's OxiNet III is a private<br>label of the Bernoulli Software for pulse oximeters] |
# Reason For Submission: Additional or Expanded Indications
# Description of the device [807.92(a)(4)}:
The SenTec Digital Monitoring System (SDMS) as listed under K071672 is a device consisting of a standalone monitor [SenTec Digital Monitor (SDM)], a digital sensor [V-Sign™ Sensor (model VS-A/P)], connecting cables, and accessories for sensor application and maintenance. The SDMS is designed for the continuous and non-invasive monitoring of carbon dioxide partial pressure (PCO2), functional oxygen saturation (SpO2), and pulse rate (PR) using the ear lobe as monitoring site in adult and pediatric patients. It is furthermore designed for PCO2-only monitoring in adult, pediatric and neonatal patients using conventional PCO2 measurement sites.
This new 510(k) introduces the following additional features/functions to the SDMS:
- V-Sign™ Sensor 2 (model VS-A/P/N) [successor of V-Sign™ Sensor (model VS-A/P) with improved . embedded hardware/software] for PCO2, SpO2, PR monitoring using the ear lobe, the forehead or the check as monitoring site in adult and pediatric patients as well as for PCO2-only monitoring in adult, pediatric and neonatal patients using conventional PCO2 measurement sites
- SpO2 Soft Sensor (models RSS-L, RSS-M, RSS-S) with the SpO2 Adapter Cable (model SC-XX; where . XX=150, 250, or 750) for continuous and non-invasive SpO2/PR monitoring in patients weighing more than 20 kg. These conventional oximetry sensors, which are applied to a patient's digit, must be connected to the SDM by using the new SpO2 Adapter Cable.
- V-STATS™, an optional PC software application for remote monitoring and/or trend reporting and . statistical analysis of the data measured with the SDMS.
This new 510(k) furthermore introduces the following modifications/enhancements:
- Modified intended use for the Multi-Site Attachment Rings: In K071672 the Multi-Site Attachment . Rings (model MAR-MI, formerly designated MAR-A/P; and model MAR-SF, formerly designated MAR-A/P/N) were listed for continuous and non-invasive PCO2-only monitoring. This new 510(k) submission expands the intended use for both models to additionally include continuous and noninvasive SpO2 and PR monitoring if V-Sign™ Sensor 2 (model VS-A/P/N) is applied to the forehead or cheek of adult or pediatric patients. In K071672 the Multi-Site Attachment Ring (model MAR-MI) was listed for use in adult and pediatric patients. This new 510(k) submission expands the intended use for the Multi-Site Attachment Ring (model MAR-MI) to additionally include the neonatal population. Proposed labeling states that model MAR-MI can be used for patients with mature and intact skin, whereas for patients with sensitive/fragile skin model MAR-SF must be used.
- Modified firmware for the SenTec Digital Monitor: the modified firmware of the SDM introduces an . expanded sensor temperature / site time management two new monitoring parameters (Pulsation expanded benser comports the new sensors being introduced with introduced with this new 510(k).
{2}------------------------------------------------
sentec
# Intended Use [807.92(a)(5)]:
The SenTec Digital Monitoring System - consisting of the SenTec Digital Monitor (SDM). Sensors and Accessories - is indicated for continuous, non-invasive patient monitoring. The SenTec Digital Monitoring System is indicated for use in hospital-type facilities, intra-hospital transport environments, and - if under clinical supervision - home environments. The SenTec Digital Monitoring System is for prescription use only.
V-Sign™ Sensor, model VS-A/P, and V-Sign™ Sensor 2, model VS-A/P/N, are indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are reguired for adult and pediatric patients. In neonatal patients the use of V-Sign™ Sensor and of V-Sign™ Sensor 2 is indicated for carbon dioxide tension monitoring only.
SenTec's Ear Clip, model EC-MI, is intended for use with V-Sign™ Sensor or V-Sign™ Sensor 2 when continuous, non-invasive carbon, oxygen saturation and pulse rate monitoring are required. The Ear Clip is for single-patient use and is indicated to attach V-Sign™ Sensor or V-Sign™ Sensor 2 to the ear lobe of the patient. The use of the Ear Clip is contraindicated for patients whose earlobes are too small to ensure adequate sensor application.
SenTec's Multi-Site Attachment Rings, model MAR-SF and model MAR-MI, are intended to attach V-Sign™ Sensor or V-Sign™ Sensor 2 to conventional measurement sites for carbon dioxide tension monitoring when continuous, non-invasive carbon dioxide tension monitoring is required for adult, pediatric, and neonatal patients. They are intended to attach the V-Sign™ Sensor 2 to the forehead or the cheek when continuous, non-invasive carbon dioxide tension, oxygen saturation, and pulse rate monitoring is required for adult and pediatric patients. The Multi-Site Attachment Rings, model MAR-MI, are for single use.
SenTec's multi-compatible and reusable SpO2 Soft Sensors, models RSS-L, RSS-M and RSS-S, are indicated for use with the monitoring devices indicated in the respective sensor directions for use when continuous non-invasive monitoring of oxygen saturation, and pulse rate are required for patients weighing more than 20 kg.
V-STATS™ is an optional PC-based software which is indicated for use with the SenTec Digital Monitor (SDM) when remote monitoring and/or trend reporting and statistical analysis of the data measured by the SDM is required. V-STATS™ is not intended to provide diagnosis, it is intended to supplement and not to replace any part of the SDM monitoring procedures.
Caution: Federal law restricts this device to sale by or on the order of a physician.
# Technological Characteristics [807.92(a)(6)]:
The SenTec Digital Monitoring System has the same indications for use as well as the same indicated patient population as the specified combination of the predicate devices.
Compared to the SDMS configuration cleared under K071672 the V-Sign™ Sensor (model VS-A/P), existing accessories (Digital Sensor Adapter Cable, Ear Clip, Multi-Site Attachment Rings, V-Sign™ Membrane Changer, Service Gas, Contact Gel) as well as the hardware and the majority of software features of the SenTec Digital Monitor (SDM) remain unchanged and therefore are identical to the SDMS configuration listed under K071672.
{3}------------------------------------------------
、
SenTec's new V-Sign™ Sensor 2 (model VS-A/P/N) is the successor of V-Sign™ Sensor (model VS-A/P) cleared under K071672. It uses the exact same basic principles of operation and the same technology as the predicate device. The differences are that it makes use of improved embedded hardware/software. These differences do not raise additional issues reqarding safety and effectiveness.
SenTec's new RSS series 'SpO2 Soft Sensor' with the SC series SpO2 Adapter Cable is a reusable, conventional oximetry sensor which is applied to a patient's digit. It utilizes an existing EnviteC sensor design cleared under K060675.
All relevant patient monitoring features of the SenTec Digital Monitoring System are equivalent to the SDMS configuration listed under K071672, the TOSCA 500, the TCM 4 (as used with Electrode E5260 [PCO2-only]), and the Reusable EnviteC SoftTipY SpO2 Sensors (being used with Nellcor's N-395 pulse oximeter). The key features are:
# A. Monitoring with heated PCO2/SpO2 or PCO2 Sensor
- same principles of operation (reflective 2 wavelength SpO2 measurement, and Severinghaus-type ● PCO2 electrode technology)
- sensors heated to an operator selectable SET temperature which is selectable by the operator .
- equivalent or similar means to control sensor temperature and application time at the site Timer) .
- equivalent or similar sensor application means (adhesive Ear Clip and/or adhesive attachment rings) .
- equivalent or similar sensor calibration unit integrated in monitor ●
- equivalent or similar accessories for sensor maintenance: membrane changer, contact gel, service gas .
# B. Monitoring with unheated, conventional SpO2 Sensor
- same principles of operation (transmissive 2 wavelength SpO2 measurement) ●
- equivalent recommendations for application time (Site Timer controlled in case of the SDMS) .
- . equivalent sensor application (sheath-like rubber tube being pushed onto finger tip)
# C. General Monitoring features
- equivalent or similar methods to calculate/display the monitored parameters .
- equivalent or similar alarms, messages and menu structure as predicate devices .
- equivalent or similar interfacing and data management possibilities .
- equivalent or similar configuration/management of parameters settings .
V-STATS™ PC Software application combines and includes relevant features of the predicate Score Software and of the Oxinet III Software. In particular these are:
# A. Downloading, analyzing, reporting trend data
- equivalent or similar method to download trend data stored in monitor's internal memory .
- equivalent or similar trend data display, event markers, scrolling and zooming functions .
- equivalent or similar operator-adjustable data analysis criteria .
- equivalent or similar printable report .
# B. Remote Monitoring and Alarm Surveillance
- equivalent or similar central station (a PC running the respective software) .
- equivalent or similar method to establish/control data communication between central station and . multiple monitors
- equivalent or similar method to administer patients (admit/transfer/discharge a patient) .
- equivalent or similar information provided in monitoring cell/window of an individual patient .
- equivalent or similar alarm system .
{4}------------------------------------------------
sentec
、
# Non-Clinical Performance data [807.92(b)(1)]:
..
# Standards Testing (Electrical, Mechanical and Environmental)
The SenTec Digital Monitoring System was tested to applicable standards for medical device Electrical Safety, Electromagnetic Compatibility, Shock and Vibration, and Environmental Temperature and Humidity. Additionally the device was tested in accordance with applicable alarm standards as well as FDA Guidance Documents and particular standards applicable for pulse oximeters and cutaneous PCO2 monitors.
# Biocompatibility Testing
Biocompatibility testing has been conducted for all patient contact materials in compliance with ISO 10993-1:2003. All materials met Biocompatibility requirements.
## Risk Analysis ،
Detailed risk, hazard, and failure analyses were performed on the SenTec Digital Monitoring System in consideration of the additions and modifications being introduced with this new 510(k). All hazards were mitigated to ALARP levels (as low as reasonably possible) and residual risks were determined to be acceptable.
## Software Development and Testing
The SenTec Digital Monitoring System software was developed in accordance with FDA quidelines for MODERATE level of concern devices. The software was verified to requirements and validated to meet the specified intended use(s).
## Bench Performance testing
Bench performance testing verified that the SenTec Digital Monitoring System measures pulse rate values within specified accuracy claims for all supported sensor types.
# Clinical Performance data [807.92(b)(2)]:
Clinical studies were performed using the SenTec Digital Monitoring System with healthy adult volunteer subjects who were subjected to progressive induced hypoxia arterial hemoglobin oxygen saturation determined from arterial blood samples with a CO-Oximeter as reference. The results from the clinical studies show for all supported sensors and claimed monitoring sites that the reported saturation values from the SDMS meet specified accuracy requirements.
# Conclusion [807.92(b)(3)]:
The results of all laboratory tests demonstrate that the SenTec Digital Monitoring System meets specified requirements.
The clinical and non-clinical testing performed demonstrates that the SenTec Digital Monitoring System is safe, effective and performs as well as the predicate devices, and therefore, it is substantial equivalent to the predicate devices.
# Other information [807.92(d)}:
Not applicable.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the department's name around the perimeter. Inside the circle is a stylized representation of a human figure embracing or protecting another person, symbolizing the department's mission of promoting health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SENTEC AG C/O Mr. Stephen H. Gorski IMAGENIX, Incorporated S65 W35739 Piper Road Eagle, Wisconsin 53119
DEC - 3 2010
Re: K101690
Trade/Device Name: SenTec Digital Monitoring System (SDMS) Regulation Number: 21 CFR 868.2480 Regulation Name: Cutaneous Carbon Dioxide (PcCO) monitor Regulatory Class: II Product Code: LKD, DQA, DPZ Dated: November 23, 2010 Received: November 24, 2010
Dear Mr. Gorski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Mr. Gorski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
# 510(k) Number (if known):
DEC: - 3 2010
#### SenTec Digital Monitoring System (SDMS) Device Name:
# Indications for use:
The SenTec Digital Monitoring System - consisting of the SenTec Digital Monitor (SDM), Sensors and Accessories - is indicated for continuous, non-invasive patient monitoring. The SenTec Digital Monitoring System is indicated for use in hospital-type facilities, intra-hospital transport environments, and - if under clinical supervision - home environments. The SenTec Digital Monitoring System is for prescription use only.
V-Sign™ Sensor, model VS-A/P, and V-Sign™ Sensor 2, model VS-A/P/N, are indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, and pulse rate are required for adult and pediatric patients. In neonatal patients the use of V-Sign™ Sersor and of V-Sign™ Sensor 2 is indicated for carbon dioxide tension monitoring only.
SenTec's Ear Clip, model EC-MI, is intended for use with V-Sign™ Sensor or V-Sign™ Sensor 2 when continuous, non-invasive carbon dioxide tension, oxygen saturation and pulse rate monitoring are required. The Ear Clip is for single-patient use and is indicated to attach V-Sign™ Sensor or V-Sign™ Sensor 2 to the ear lobe of the patient. The use of the Ear Clip is contraindicated for patients whose ear-lobes are too small to ensure adequate sensor application.
SenTec's Multi-Site Attachment Rings, model MAR-MI, are intended to attach V-Sign™ Sensor or V-Sign™ Sensor 2 to conventional measurement sites for carbon dioxide tension monitoring when continuous, non-invasive carbon dioxide tension monitoring is required for adult, pediatric, and neonatal patients. They are intended to attach the V-Sign™ Sensor 2 to the forehead or the cheek when continuous, non-invasive carbon dioxide tension, oxygen saturation, and pulse rate monitoring is required for adult and pediatric patients. The Multi-Site Attachment Rings, model MAR-SF and model MAR-MI, are for single use.
SenTec's multi-compatible and reusable SpO2 Soft Sensors, models RSS-L, RSS-M and RSS-S, are indicated for use with the monitoring devices indicated in the respective sensor directions for use when continuous noninvasive monitoring of oxygen saturation, and pulse rate are required for patients weighing more than 20 kg.
V-STATS™'Is an optional PC-based software which is indicated for use with the SenTec Digital Monitor (SDM) when remote monitoring and/or trend reporting and statistical analysis of the data measured by the SDM is required. V-STATS™ is not intended to provide diagnosis, it is intended to supplement and not to replace any part of the SDM monitoring procedures.
Caution: Federal law restricts this device to sale by or on the order of a physician.
| Prescription Use | X | AND / OR |
|-----------------------------|---|----------|
| (Part 21 CFR 801 Subpart D) | | |
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Per Infection Control. Dental Device
510(k) Number: K101690
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.