K043003 · Radiometer Medical Aps · LKD · Mar 1, 2005 · Anesthesiology
Device Facts
Record ID
K043003
Device Name
TCM40 MONITORING SYSTEM
Applicant
Radiometer Medical Aps
Product Code
LKD · Anesthesiology
Decision Date
Mar 1, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.2480
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The TCM4 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia. The TCM40 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures as well as of oxygen saturation of arterial hemoglobin and pulse rate. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia.
Device Story
TCM4/40 Monitoring Systems consist of a monitor unit, combined oxygen/carbon dioxide interface module, electrode assembly, and (for TCM40) an oxygen saturation/pulse module. Device inputs include transcutaneous gas signals and pulse oximetry data. The system processes these inputs to provide continuous real-time monitoring of pO2, pCO2, SpO2, and pulse rate. Used in clinical settings by healthcare providers to monitor patient physiological status. Output is displayed on a touch screen, allowing clinicians to assess patient respiratory and circulatory health. Benefits include non-invasive, continuous monitoring of gas exchange and oxygenation status, aiding in clinical decision-making for patients not under gas anesthesia.
Clinical Evidence
No clinical data. Safety and performance were established through bench testing and compliance with consensus standards including IEC-60601-1, IEC-60601-1-2, IEC-60601-2-23, IEC-60601-3-1, EN-865, and IEC-60601-1-8.
Technological Characteristics
Monitor unit includes cabinet, touch screen, electronic controls, power supply, and battery. Uses combined tcpO2/tcpCO2 electrodes and pulse/SpO2 sensors. Connectivity includes integrated monitoring unit. Safety standards: IEC-60601-1, IEC-60601-1-2, IEC-60601-2-23, IEC-60601-3-1, EN-865, IEC-60601-1-8.
Indications for Use
Indicated for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures (TCM4/40) and oxygen saturation/pulse rate (TCM40 only) in neonates, pediatrics, and adults not under gas anesthesia.
Regulatory Classification
Identification
A cutaneous carbon dioxide (PcCO2) monitor is a noninvasive heated sensor and a pH-sensitive glass electrode placed on a patient's skin, which is intended to monitor relative changes in a hemodynamically stable patient's cutaneous carbon dioxide tension as an adjunct to arterial carbon dioxide tension measurement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Cutaneous Carbon Dioxide (PcCO2) and Oxygen (PcO2) Monitors; Guidance for Industry and FDA.” See § 868.1(e) for the availability of this guidance document.
Predicate Devices
TCM400 Monitoring System
TCM3 Monitoring System
OxiMAX N-550 Pulse Oximetry System
TOSCA PCO2, SpO2 and Pulse Rate Monitoring System
Perimed Transcutaneous PO2 and PCO2 Monitor (PF5040)
MicroGas 7650 Transcutaneous Monitor
Siemens Transcutaneous TpO2/CO2 Gas Module
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Premarket Notification; TCM4/40; K043003
RADIOMETER COPENHAGEN
February 16, 2005
MAR ] - 2005
# 510(k) Summary - TCM4/40 Monitoring Systems
| Submitter: | Radiometer Medical ApS |
|-----------------|-----------------------------------------|
| Address: | Åkandevej 21, DK-2700 Brønshøj, Denmark |
| Phone: | +45 3827.3827 |
| Fax: | +45 3827.2736 |
| Contact Person: | Pierre Pelletier |
| Trade Name: | TCM4/40 Monitoring Systems |
|-----------------------|------------------------------------------------------|
| Common name: | Transcutaneous pO2/pCO2/SpO2/pulse monitoring system |
| Classification Names: | 21CFR§868.2500: Monitor, Oxygen, Cutaneous |
| | 21CFR§868.2480: Monitor, Carbon Dioxide, Cutaneous |
| | 21CFR§870.2700: Oximeter |
| Predicate Devices: | TCM400 Monitoring System |
| | TCM3 Monitoring System |
| | OxiMAX N-550 Pulse Oximetry System |
| | TOSCA PCO2, SpO2 and Pulse Rate Monitoring System |
| | Perimed Transcutaneous PO2 and PCO2 Monitor (PF5040) |
| | MicroGas 7650 Transcutaneous Monitor |
| | Siemens Transcutaneous TpO2/CO2 Gas Module |
#### Device Description
The TCM4/40 Monitoring Systems essentially consist in a monitor unit, a combined oxygen/carbon dioxide interface module along with the combined electrodes assembly and (for the TCM40 only) an oxygen saturation/pulse module with a selection of three sensors.
#### Intended use
The TCM4 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia.
The TCM40 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures as well as of oxygen saturation of arterial hemoglobin and pulse rate. It is indicated for use on neonates, pediatrics, and adults not under gas anesthesia.
{1}------------------------------------------------
RADIOMETER COPENHAGEN
# February 16, 2005
## Technological characteristics
The monitoring unit (including software, cabinet, touch screen, electronic controls, power supply and battery) is an adaptation of Radiometer's TCM400 Monitoring System, with only one combined tcp02/tcpCO2 electrode and (for the TCM40 only) one pulse/SoO2 sensor (the TCM400 has up to 6 separate tcg0> electrodes). The combined tcp02/tcpCO2 electrode is identical to the electrode used in Radiometer's previous generation TCM3 Monitoring System. The pulse/SpO2 sensors are identical to the sensors used with Puritan Nellcor Bennett's OxiMAX N-550 Pulse Oximetry System.
## Safety and performance evaluation studies
In order to ensure TCM4/40 Monitoring Systems is safe and effective for its intended use, it has been designed and tested to the requirements of the following standards:
- . IEC-60601-1:1988 Medical electrical equipment - Part 1: General requirements for safety
- IEC-60601-1-2:2001 . Medical Electrical Equipment - Part 1-2: General Requirements for Safety Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- IEC-60601-2-23:1999 . Medical electrical equipment - Part 2-23: Particular reguirements for the safety, including essential performance, of transcutaneous partial pressure monitoring equipment
- IEC-60601-3-1:1996 . Medical Electrical Equipment Part 3-1: Essential performance requirements for transcutaneous oxygen and carbon dioxide partial pressure monitoring equipment
- EN-865:1997 . Pulse oximeters - Particular requirements
. IEC-60601-1-8:2003
Medical electrical equipment - Part 1-8: General requirements for safety Collateral standard: General requirements, tests and quidance for alarm systems in medical electrical equipment and medical electrical systems
Based upon the extensive use of consensus standards covering all aspects of safety and performance, no clinical evaluation was deemed necessary to assess the product performance levels.
## Conclusions from safety and performance evaluation studies
Compliance with the requirements in the above-named safety and performance standards demonstrates that the TCM4/40 Monitoring Systems are state-of-the-art transcutaneous oxygen/carbon dioxide monitors and pulse oximeters that have the necessary safety mechanisms and performs up to the requirements of the relevant international consensus standards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract design of an eagle's head and neck, rendered in black lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle design. The text is in a sans-serif font and is also black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 - 2005
Mr. Pierre Pelletier Regulatory Affairs Manager Radiometer Medical ApS Åkandevej 21, Brønshøj Denmark DK-2700
Re: K043003
Trade/Device Name: TCM4/40 Monitoring Systems Regulation Number: 21 CFR 868.2480 Regulation Name: Cutaneous Carbon Dioxide (PcCO2) Monitor Regulatory Class: II Product Code: LKD, LPP, KLK, DQA Dated: January 20, 2005 Received: January 24, 20045
Dear Mr. Pelletier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Ecderal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Pelletier
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Runne
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K043003
TCM4/40 Monitoring Systems Device Name:
#### Indications for Use:
The TCM4 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and The TCM4 Monitoring System is mended for continued for use on neonates, pediatrics, and adults not under gas anesthesia.
The TCM40 Monitoring System is intended for continuous transcutaneous monitoring of oxygen and carbon dioxide partial pressures as well as of oxygen saturation of arterial hemoglobin and pulse Carbon uloxide partial pressures as well as of thyger of the mot under gas anesthesia.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
and the comments of the comments of the comments of the comments of the comments of the comments of
Cicia Fowler
sion Sign-O
son of Anesthesiology, General Hospital.
action Control. Dental Devices
in Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.