Ceros® TCP Granules are indicated for use as a bone void filler in non-loadbearing areas in the oral and maxillofacial regions requiring cancellous rather than cortical bone material. Ceros® TCP Granules may, for example, be used for: · supplementing autogenous cancellous bone · supporting and stabilizing for guided bone regeneration, "GBR" · filling bone cavities after hemisection of teeth, bone removal, osteotomy, tumor resection, root apex resection, tooth extraction or cystectomy · filling bone defects resulting from traumatic or pathological origin · augmenting an atrophied mandibular ridge (only in conjunction with the "GBR" membrane technique) · filling defects after explantation of dental implants · filling small periodontal bone cavities as well as bi- and tri-furcations of teeth · the reconstruction of alveolar defects prior to providing a prosthesis · preparation of the implant bed (e.g. sinus lift) · filling bone defects around dental implant after immediate placement into extraction sockets Ceros® TCP Granules are intended to be gently packed or placed into the surgical site and may be combined with autogenous blood, bone marrow or saline.
Device Story
Synthetic, porous, osteoconductive, bioresorbable bone grafting material composed of beta-tricalcium phosphate (ß-TCP). Used in dental, oral, and maxillofacial surgery to fill, augment, or reconstruct bony defects. Clinician packs granules into surgical site; may be mixed with autogenous blood, bone marrow, or saline. Acts as scaffold for new bone growth; resorbs over time. Benefits patient by providing structural support for bone regeneration in non-loadbearing defects.
Clinical Evidence
Bench testing only. Testing included chemical composition (ICP, XRF), phase purity (quantitative XRD), trace element analysis (ICP/MS), particle size distribution (optical image analysis), surface morphology (SEM), and dissolution testing (ISO 10993-14).
Technological Characteristics
Synthetic beta-tricalcium phosphate (ß-TCP) granules. Conforms to ASTM F1088. Porous, osteoconductive, bioresorbable material. Provided in 0.5, 1.0, and 2.5 g volumes. Sterilized using standard methods.
Indications for Use
Indicated for patients requiring bone void filling in non-loadbearing oral and maxillofacial regions where cancellous bone is needed. Used for bone defects from trauma, pathology, tooth extraction, cystectomy, osteotomy, or tumor resection; alveolar reconstruction; sinus lifts; and dental implant site preparation.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K101426 Ceros® TCP Granules
### 510(k) Summary
## Mathys Ltd Ceros® TCP Granules
September 9, 2010
#### ADMINISTRATIVE INFORMATION
SEP 1 5 2010
| Manufacturer Name: | Mathys Ltd |
|----------------------------|-------------------------------------|
| | Güterstrasse 5 |
| | CH-2544 Bettlach |
| | Switzerland |
| Telephone: | +41 (0) 32 644 1 644 |
| Fax: | +41 (0) 32 644 1 161 |
| Official Contact: | Giancarlo Rizzoli, PhD |
| | Director Business Unit Biosurgery |
| Email: | giancarlo.rizzoli@mathysmedical.com |
| Representative/Consultant: | Linda K. Schulz, RDH, BSDH |
| | Kevin A. Thomas, PhD |
| | PaxMed International, LLC |
| | 11234 El Camino Real, Suite 200 |
| | San Diego, CA 92130 |
| Telephone: | +1 (858) 792-1235 |
| Fax: | +1 (858) 792-1236 |
| email: | lschulz@paxmed.com |
| | kthomas@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: |
|-----------------------------|
| Common Name: |
| Classification Regulations: |
| Product Code: |
| Classification Panel: |
| Reviewing Branch: |
Ceros® TCP Granules Bone Grafting Material, Synthetic 21 CFR 872.3930 LYC Dental Products Panel Dental Devices Branch
#### INTENDED USE
Ceros® TCP Granules are indicated for use as a bone void filler in non-loadbearing areas in the oral and maxillofacial regions requiring cancellous rather than cortical bone material. Ceros TCP Granules may, for example, be used for:
{1}------------------------------------------------
#### 510(k) Summary
K101426
Ceros® TCP Granules
- · supplementing autogenous cancellous bone
- · supporting and stabilizing for guided bone regeneration, "GBR"
- · filling bone cavities after hemisection of teeth, bone removal, osteotomy, tumor resection, root apex resection, tooth extraction or cystectomy
- · filling bone defects resulting from traumatic or pathological origin
· augmenting an atrophied mandibular ridge (only in conjunction with the "GBR" membrane technique)
- · filling defects after explantation of dental implants
- · filling small periodontal bone cavities as well as bi- and tri-furcations of teeth
- · the reconstruction of alveolar defects prior to providing a prosthesis
- · preparation of the implant bed (e.g. sinus lift)
- · filling bone defects around dental implant after immediate placement into extraction sockets
Ceros® TCP Granules are intended to be gently packed or placed into the surgical site and may be combined with autogenous blood, bone marrow or saline.
#### DEVICE DESCRIPTION
Ceros® TCP Granules are a synthetic, porous, osteoconductive, bioresorbable bone grafting material composed of beta-tricalcium phosphate, ß-TCP, [ß-C23]PO4]2] intended to fill, augment, or reconstruct bony defects in dental, oral or maxillofacial surgery. Ceros® TCP Granules are provided in volumes of 0.5, 1.0 and 2.5 g.
#### EQUIVALENCE TO MARKETED DEVICE
Mathys Ltd demonstrated that for the purposes of FDA's regulation of medical devices, Ceros TCP Granules are substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices:
Synthes (USA) chronOS™ - B-TCP cleared under K053022,
Curasan AG Cerasorb® DENTAL and Cerasorb® M DENTAL cleared under K051443 BioForm, Inc., Calcium Hydroxylapatite Implant cleared under K030682.
The subject device and the predicate devices have the same intended use and have the same technological characteristics. The subject device is made of of B-TCP, [B-Ca3(PO4)2] conforming to ASTM F1088 Standard Specification for Beta-Tricalcium Phosphate for Surgical Implantation. Predicate devices are made of B-TCP or a similar material, CaHA, with similar particle size, pore size and porosity to the subject device. Subject and predicate devices are packaged in similar materials and sterilized using similar methods.
Bench testing was performed to demonstrate conformance to ASTM F1088, including chemical composition by inductively coupled plasma and x-ray fluorescence, ß-TCP content by quantitative x-ray diffraction, and trace element analysis by inductively coupled plasma/mass spectroscopy. Testing was also performed to characterize physical characteristics including particle size and particle size distribution by optical image analysis and surface morphology by scanning electron microscopy. Dissolution testing was performed according to ISO 10993-14.
{2}------------------------------------------------
# 510(k) Summary
In summary, Ceros® TCP Granules have the following similarities to the predicate devices:
- . have the same intended use,
- . use the same operating principle,
- incorporate the same basic design, .
- incorporate the same or very similar materials, and .
- have similar packaging and is sterilized using the same materials and processes. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mathys Limited C/O Ms. Linda Schulz PaxMed International LLC 11234 El Camino Real, Suite 200 San Diego, California 92130-8587
SEP 1 5 2010
Re: K101426
Trade/Device Name: Ceros® TCP Granules Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: August 16, 2010 Received: August 17, 2010
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Ms. Schulz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other requireme
You must comply with all the Aot's resuire You must comply with all the Act's requirements, including, but not limited to: registration
and listing (21 CFR Part 807): lebeling (21 CER D . . (2002): and listing (21 CFR Part 807); labeling (21 CFR Part 801); mot limited to: regist
(reporting of medical device related education (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (21 CFR Part 820)
the Act): 21 CFR 1000-1050 the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801),
please go to please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH) DRHOthees/ucm] 15809.htm for for for for for for please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalTroblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet Assis.
http://www.fda.org/Mali.clD http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
hn for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
510(k) Summary
K101426
Ceros®TCP Granules
#### Indications for Use
510(k) Number: K101426
Device Name: Ceros® TCP Granules
SEP 1 5 2010
Indications for Use:
Ceros® TCP Granules are indicated for use as a bone void filler in non-loadbearing areas in the oral and maxillofacial regions requiring cancellous rather than cortical bone material. Ceros® TCP Granules may, for example, be used for:
- · supplementing autogenous cancellous bone
· supporting and stabilizing for guided bone regeneration, "GBR"
· filling bone cavities after hemisection of teeth, bone removal, osteotomy, tumor resection, root apex resection, tooth extraction or cystectomy
· filling bone defects resulting from traumatic or pathological origin
· augmenting an atrophied mandibular ridge (only in conjunction with the "GBR" membrane technique)
- · filling defects after explantation of dental implants
• filling small periodontal bone cavities as well as bi- and tri-furcations of teeth
- · the reconstruction of alveolar defects prior to providing a prosthesis
· preparation of the implant bed (e.g. sinus lift)
· filling bone defects around dental implant after immediate placement into extraction sockets
Ceros® TCP Granules are intended to be gently packed or placed into the surgical site and may be combined with autogenous blood, bone marrow or saline.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Turner
Page T of
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: حكومة الأمرات 2010)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.