CALCIUM HYDROXYLAPATITE IMPLANT, ICC AND 0.5CC, MODELS 8038-3, 8037-3
K030682 · Bioform, Inc. · LYC · Jun 27, 2003 · Dental
Device Facts
Record ID
K030682
Device Name
CALCIUM HYDROXYLAPATITE IMPLANT, ICC AND 0.5CC, MODELS 8038-3, 8037-3
Applicant
Bioform, Inc.
Product Code
LYC · Dental
Decision Date
Jun 27, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Calcium Hydroxylapatite Implant is intended to fill and/or augment dental intraosseous and oral/maxillofacial defects including: - Periodontal Defects - Ridge Augmentation - Extraction Sites - Craniofacial Augmentation - Cystic Defects
Device Story
Calcium Hydroxylapatite Implant is a sterile, non-pyrogenic, cohesive paste consisting of calcium hydroxylapatite particles suspended in a gel of glycerin, water, and sodium carboxymethylcellulose. The device is supplied in a syringe for direct, controlled deposition into dental or oral/maxillofacial defects. The gel binder facilitates placement and prevents particle migration; it is resorbed post-operatively, leaving the non-resorbable calcium hydroxylapatite particles to serve as an osteoconductive scaffold for bone infiltration. The device is used by clinicians in dental or oral/maxillofacial settings. By providing a ready-to-use, pre-mixed paste, the device eliminates the need for intraoperative mixing with blood or saline, reducing contamination risk and improving procedural convenience for the physician.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on the history of use of the components in medical devices, preclinical testing, and the established biological mechanism of action (osteoconductive scaffold for tissue ingrowth) shared with predicate devices.
Technological Characteristics
Paste containing calcium hydroxylapatite particles (ASTM F1185) and a binder (glycerin, water, sodium carboxymethylcellulose). Resorbable binder provides controlled porosity for bone ingrowth; non-resorbable particles provide osteoconductive scaffold. Delivered via syringe. Sterile, non-pyrogenic.
Indications for Use
Indicated for filling or augmenting dental intraosseous and oral/maxillofacial defects, including periodontal defects, ridge augmentation, extraction sites, craniofacial augmentation, and cystic defects.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
## JUN 2 7 2003
Calcium Hydroxylapatite Implant 510(k) Premarket Notification Submission
KOZOLOSA
June 25, 2003
### SECTION 4
# 510(k) Summary of Safety and Effectiveness
#### Manufacturer and Submitter Information: 4.1
BioForm Inc. 4133 Courtney Road, #10 Franksville, WI 53126 Telephone: (262) 835-9800 Contact: William G. Hubbard Summary Preparation Date: June 25, 2003
| 4.2 | Device Trade Name: | Calcium Hydroxylapatite Implant |
|-----|------------------------|---------------------------------------------------------------------------------------------|
| | Common Name: | Synthetic Bone Graft Material |
| | Classification Name: | Bone Filling and Augmentation Material |
| | Device Class: | Unclassified |
| | Product Code: | LYC |
| | Medical Specialty: | Dental (DE) |
| | Performance Standards: | No applicable mandatory performance standards or<br>special controls exist for this device. |
#### Substantial Equivalent Predicate Device: 4.3
The following are the predicate devices that are substantially equivalent to the Calcium Hydroxylapatite Implant:
| K882682 | Calcitite<br>Calcitek, Inc., 4125-B Sorrento Valley Boulevard, San<br>Diego, CA 92121 |
|---------|------------------------------------------------------------------------------------------------------------------------------|
| K852742 | Osteograf AR Alveolar Ridge Hydroxylapatite 18-40<br>Coors Biomedical Company, 12860 West Cedar Drive,<br>Lakewood, CO 80228 |
| K852765 | HA-2000<br>Orthomatrix, Inc., 6968 Sierra Court, Dublin CA 94568 |
| K992416 | Perioglas-BioGlass Bone Graft Particulate<br>U.S. Biomaterials Corp., One Progress Boulevard, Alachua,<br>FL 32615 |
| K000149 | Novabone-Bioglass Bone Graft Particulate<br>U.S. Biomaterials Corp., One Progress Boulevard, Alachua,<br>FL 32615 |
| K952922 | Biogran Bioactive Glass Synthetic Bone Graft Material |
{1}------------------------------------------------
### June 25, 2003
| | Orthovita Co., 212 Carnegie Center Drive, Suite 206, |
|---------|----------------------------------------------------------|
| | Princeton, NJ 08540 |
| K921468 | PermaMesh Hydroxylapatite Matrix |
| | CeraMed Corp., 12860 West Cedar Drive, Lakewood, |
| | CO 80228 |
| K862061 | Osteograf/AR+ Permaridge Hydroxylapatite, 18-40 |
| | Coors Biomedical Co., 12860 West Cedar Drive, Suite 210, |
| | Lakewood, CO 80228 |
| K910432 | HAPSET Hydroxylapatite Bone Graft Plaster |
| | Lifecore Biomedical, Inc., 1050 Connecticut Avenue, N.W. |
| | Washington Square, Suite 1100, Washington, D.C. 20036 |
Calcium Hydroxylapatite Craniofacial Implants are identical to Coaptite® Laryngeal Augmentation System (K013243) and Coaptite® Tissue Marker (K012955) manufactured by BioForm Inc. The Calcium Hydroxylapatite Implant is substantially equivalent to the predicate devices sited above.
#### Device Description: 4.4
Calcium Hydroxylapatite Implant is a sterile, non-pyrogenic, flexible, semi-solid, cohesive paste containing calcium hydroxylapatite particles. Calcium Hydroxylapatite Implant is intended to fill and/or augment dental intraosseous and oral/maxillofacial defects. Calcium Hydroxylapatite Implant contains calcium hydroxylapatite particles in a pasty gel of glycerin, water and sodium carboxymethylcellulose that acts as a binder for the particles. The calcium hydroxylapatite meets ASTM F1185. The gel ingredients are pharmacoutical grade excipients and are listed as GRAS materials.
#### 4.5 Intended Use:
Calcium Hydroxylapatite Implant is intended to fill and/or augment dental intraosseous and oral/maxillofacial defects including:
- Periodontal Defects 0
- Ridge Augmentation
- Extraction Sites ●
- Craniofacial Augmentation .
- Cystic Defects .
#### 4.6 Technological Characteristics:
Calcium Hydroxylapatite Implant (CaHA Implant) is a paste of calcium hydroxylapatite (calcium phosphate) particles and a binder. The binder is resorbable and acts to hold the particles together. The resorption of the binder provides the controlled porosity for bono ingrowth. These 'porcs' are the interconnected pathways around and between the calcium hydroxylapatite particles that are non-absorbable. The calcium hydroxylapatite particles provide
{2}------------------------------------------------
June 25, 2003
an osteoconductive scaffold for bone infiltration. The CaHA Implant functions as an osteoconductive scaffold for bone infiltration, the same mechanism of action as the predicate devices. The premixed and ready to use characteristic facilitates the delivery of CaHA Implant by allowing direct and controlled deposition using the syringe. This may be enhanced for implantation in difficult to access locations by the placing of a needle on the syringe. In addition, the binder serves to reduce or prevent migration during the application and in the post-operative healing period prior to incorporation of the particles by tissue ingrowth.
The Calcium Hydroxylapatite Implant (CaHA Implant) is substantially equivalent in composition and intended use to the predicate devices. The CaHA Implant and the predicate devices are all intended to fill and/or augment dental intraosseous and oral/maxillofacial defects. All of the predicate devices and the CaHA Implant are composites of resorbable (binder) and non-resorbable (calcium phosphate particles) components. The predicate devices use sterile water, saline, the patient's own blood or absorbable sutures to bind the particles together to facilitate delivery of the particles and minimize migration of the particles from the implant site. These other binding agents are resorbed within weeks or months depending upon the used and how it is prepared. All of these are also substantially equivalent in their mechanism of action as they provide a calcium phosphate particle that provides an osteoconductive scaffold for bone infiltration. The bone infiltration provides the final form of the implant.
The principal component of the Calcium Hydroxylapatite Implant, calcium hydroxylapatite is identical to the calcium hydroxylapatite used in most of the predicates and is very similar to the calcium phosphate glasses used in the rest of the predicate devices. All of these meet the same biocompatibility requirements. It is substantially equivalent in terms of biocompatibility and biological mechanism of action in that all of these products utilize tissue ingrowth to support and sustain augmentation.
In terms of risk versus benefit, BioForm believes that the CaHA Implant has obvious advantages compared to the predicate devices. The CAHA implant is provided as a sterile and ready to use paste that is more convenient for the physician, minimizes the potential for contamination during the mixing processes required by other devices, eliminates the risk and pain associated with drawing blood and is more convenient to place. The CaHA Implant also is convenient to use because the amount of filling or augmentation can be easily controlled through direct application.
{3}------------------------------------------------
June 25, 2003
#### 4.7 Conclusion:
In summary, Calcium Hydroxylapatite Implant is substantially equivalent to the cited predicate devices and, in some instances, identical to certain predicate devices. All have the same intended use designed to fill and/or augment dental intraosseous and oral/maxillofacial defects. The components used in the Calcium Hydroxylapatite Implant and the predicate devices are biocompatible, based on the history use in many medical devices as well as from preclinical testing and clinical experience. All function as composites that utilize a resorbable binder and a non-resorbable calcium phosphate particulate. The binder facilitates placement and the particles function as an osteoconductive matrix. All reach their final form after tissue infiltration. The Calcium Hydroxylapatite Implant is substantially equivalent in intended use, technical characteristics and are as safe as the predicate devices cited.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 7 2003
Mr. William G. Hubbard BioForm Incorporated 4133 Courtney Road, #10 Franksville, Wisconsin 53126
Re: K030682
Trade/Device Name: Calcium Hydroxylapatite Implant Regulation Number: None Regulation Name: Bone Filling and Augmentation Material Regulatory Class: Unclassified Product Code: LYC Dated: June 9, 2003 Received: June 10, 2003
Dear Mr. Hubbard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 -- Mr. Hubbard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Susan Quinn
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
March 4, 2003 Revised June 9, 2003
### SECTION 2
# Indications for Use/Intended Use Statement
510(k) Number (if known): K030682
Device Name: Calcium Hydroxylapatite Implant
Indications for Use:
BioForm's Calcium Hydroxylapatite Implant is intended to fill and/or augment dental intraosseous and oral/maxillofacial defects including:
- · Periodontal Defects
- · Ridge Augmentation
- · Extraction Sites
- · Craniofacial Augmentation
- · Cystic Defects
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) - - - - - - - - - - - - -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use Over-the-Counter Use or (Per 21 CFR 801.109)
Optional Format 1-2-96
Ka Muluy for RSR
General Hospital,
Number K030682
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.