V-TRUST TD-8002 MULTI-PARAMETER SPOT-CHECK MONITOR MODEL TD-8002
Applicant
Taidoc Technology Corporation
Product Code
MWI · Cardiovascular
Decision Date
Jun 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The V-TRUSTTD-8002 Multi-Parameter Spot-Check Monitor measures systolic and diastolic pressure, pulse rate, ear temperature, blood glucose and oxygen saturation of arterial hemoglobin (SpO2). The device is intended to be used by clinicians and medically qualified personnel. For glucose measurements, it quantitatively determines glucose levels using capillary and venous whole blood. For blood pressure measurements, it is intended to be used to measure noninvasively the systolic and diastolic blood pressure and pulse rate of adults. For pulse oximeter measurements, is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2 ) and pulse rate for adults. For ear thermometer, it is intended for the intermittent measurement and monitoring human body temperature.
Device Story
V-TRUST TD-8002 is a multi-parameter spot-check monitor for clinical use. Main unit integrates oximeter and NIBP monitor; external infra-red ear thermometer and blood glucose meter connect via USB. Inputs: NIBP, SpO2, pulse rate, ear temperature, blood glucose. Device processes physiological signals to display real-time data; stores monitoring data for retrospective review; provides limit alarms for SpO2. Operated by clinicians/medically qualified personnel in clinical settings. Output allows providers to assess patient vital signs and glucose levels, facilitating clinical decision-making and patient monitoring.
Clinical Evidence
Laboratory and clinical studies were conducted to evaluate performance. Results demonstrate that the system meets its intended use and is equivalent to the predicate devices.
Technological Characteristics
Multi-parameter monitor with integrated oximeter and NIBP modules. External peripherals (ear thermometer, glucose meter) connect via USB. Uses standard physiological sensing principles consistent with predicate devices. Capable of data storage and limit alarms.
Indications for Use
Indicated for adults for noninvasive measurement of systolic/diastolic blood pressure, pulse rate, and functional SpO2. Indicated for intermittent human body temperature monitoring. Indicated for quantitative glucose determination using capillary and venous whole blood. Intended for use by clinicians and medically qualified personnel.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
KI01259 P//3
### 3. 510(k) Summary
# JUN 1 7 2010 -
#### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The Assigned 510(k) number is:
- 1. Submitter's Identification:
TaiDoc Technology Corporation 3F, 5F, No.127, Wugong 2nd Rd., Wugu Township, Taipei County, 248, Taiwan
Correspondence: Debra Liang Regulatory Affairs Specialist Tel: +886-2-6625-8188 Fax: +886-2-6625-0288
Prepared date: April 29, 2010
2. Device name:
Proprietary name: V-TRUSTTD-8002 MULTI-PARAMETER SPOT CHECK MONITOR
Regulatory information:
- Regulation section: 21 CFR 870.2300 A.
- Classification: Class II മ്പ്
- Product Code: C. MWI, monitor, physiological, patient (without arrhythmia detection or alarms)
- D. Panel: Cardiovascular
#### 3. Intended Use:
The V-TRUSTTD-8002 Multi-Parameter Spot-Check Monitor measures systolic and diastolic pressure, pulse rate, ear temperature, blood glucose and oxygen saturation of arterial hemoglobin (SpO2).
{1}------------------------------------------------
- . The device is intended to be used by clinicians and medically qualified personnel.
- . For glucose measurements, it quantitatively determines glucose levels using capillary and venous whole blood.
- . For blood pressure measurements, it is intended to be used to measure noninvasively the systolic and diastolic blood pressure and pulse rate of adults.
- . For pulse oximeter measurements, is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2 ) and pulse rate for adults.
- t For ear thermometer, it is intended for the intermittent measurement and monitoring human body temperature.
#### 4. Device Description:
The V-TRUST TD-8002 Multi-Parameter Spot-Check Monitor, Model TD-8002 (TD-8002) is a patient monitor that measures and displays real time physiological data of the patient, including blood glucose measurement. The TD-8002 can be used to monitor one or more of the following parameters: Noninvasive blood pressure (NIBP), blood oxygen saturation (SpO2), pulse rate, body temperature, and blood glucose. For all these vital parameters, the TD-8002 will be capable of limit alarms for SpO2, and storing monitoring data for retrospective review.
The TD-8002 has two monitoring devices built in the main unit: oximeter and NIBP monitor. The other two monitoring devices, infra-red ear thermometer and blood glucose meter are connected to the main unit via USB cable. All of the devices for monitoring vital signs are cleared by the FDA.
The oximeter sensor connected to the main unit of TD-8002 is the same as the Solaris Compatible Reusable Adult SpO2 Finger Sensor (Model S100A-090103 or Model S100A-300103) cleared under k061931.
- 5. Substantial Equivalence Information:
- Predicate device name and k number: A.
- 506 Patient Monitor (K051038) ●
- . Oximeter: 503 PULSE OXIMETER (K911124)
- . Blood Glucose Meter: Clever Check TD-1231 Blood Glucose Monitoring System (K063212)
- Blood Glucose Test Strip: TaiDoc Pro I Glucose Test Strip (K082169) .
Page 2
3- 2 of 3
్లో,
{2}------------------------------------------------
- . 1R Thermometer: Fora ComfortScan Ear Thermometer (KU81445)
- �: NIBP: U-RIGHT TD-3132 Blood Pressure Monitor (K092106)
- Comparison with predicate: ದ್ದ
The V-TRUST TD-8002 Multi-Parameter Spot-Check Monitor has equivalent technological characteristics and the similar intended use as the predicate devices.
## 6. Test Principle:
The V-TRUST TD-8002 Multi-Parameter Spot-Check Monitor uses the same test principle as the predicate devices.
- 7. Performance Characteristics:
The laboratory and clinical studies for the performance of V-TRUST TD-8002 Multi-Parameter Spot-Check Monitor demonstrate that the performance of this system meets its intended use and is equivalent to the predicate devices.
- 8. Conclusion:
Based on the information provided in this submission, the V-TRUST TD-8002 Multi-Parameter Spot-Check Monitor is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
## JUN 1 7 2010
Ms. Debra Liang Regulatory Affairs, Senior Specialist TaiDoc Technology Corporation 6F, No.127, Wugong 2nd Rd., Wugu Township Taipei County, Taiwan 24888
Re: K101259
Device Name: V-TRUST Model TD-8002 Multi-Parameter Spot Check Monitor Regulation Number: 21 CFR 870.2300
Regulation Name: Patient Physiological Monitor (Without Arrhythmia Detection or Alarms)
Regulatory Class: Class II (Two) Product Code: MWI Dated: April 29, 2010 Received: May 4, 2010
Dear Ms. Liang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Ms. Debra Liang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.tga.gov.au/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number: K101259
Device Name: V-TRUST TD-8002 MULTI-PARAMETER SPOT CHECK MONITOR, MODEL TD-8002
Indications for IJse:
The TD-8002 Multi-Parameter Spot-Check Monitor measures systolic and diastolic pressure, pulse rate, ear temperature, blood glucose and oxygen saturation of arterial hemoglobin (SpO2).
- The device is intended to be used by clinicians and medically qualified રુ personnel..
- . For glucose measurements, it quantitatively determines glucose levels using capillary and venous whole blood.
- . For blood pressure measurements, it is intended to be used to measure noninvasively the systolic and diastolic blood pressure and pulse rate of adults.
- . For pulse oximeter measurements, is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2 ) and pulse rate for adults.
- . For ear thermometer, it is intended for the intermittent measurement and monitoring human body temperature.
Preseription Use ਪ੍ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
Division of Cardiovascular Devices
Page 1 of 1
ござ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.