K051038 · Ciriticare Systems, Inc. · DRT · Aug 12, 2005 · Cardiovascular
Device Facts
Record ID
K051038
Device Name
506 PATIENT MONITOR
Applicant
Ciriticare Systems, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Aug 12, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarms and patient data analysis will be available to the care provider from the monitor.
Device Story
506 Patient Monitor measures/displays real-time physiological data: NIBP, SpO2, and temperature. Inputs: NIBP cuffs, SpO2 sensors (Nellcor module), and temperature probes (Kendall Filac FasTemp module). Device processes signals to provide numerical data, pulse bar plethysmogram, limit alarms, and trend storage. Used in healthcare facilities by professional healthcare providers; stationary or portable during patient translocation. Output displayed on alphanumeric/LED screen; provides strip chart recordings. Enables clinicians to monitor patient status, assess vital signs, and make clinical decisions based on real-time data and trends. Benefits include continuous or predictive temperature monitoring and motion-resistant SpO2 tracking.
Clinical Evidence
Bench testing only. Compliance with national and international standards including UL 60601-1 (electrical safety), EN 60601-1-2 (EMC), IEC 60601-2-49 (multiparameter monitor safety), ISO 10993-5.10-11 (biocompatibility), IEC 60601-2-30/EN 1060-1/EN 1060-3/AAMI SP-10 (NIBP performance), and EN 865 (oximetry performance).
Technological Characteristics
Molded plastic housing; alphanumeric and LED numeric displays; membrane key panel interface. Parameters: NIBP, SpO2 (Nellcor module), Temperature (Kendall Filac FasTemp module). Standards: UL 60601-1, EN 60601-1-2, IEC 60601-2-49, ISO 10993, IEC 60601-2-30, EN 1060-1/3, AAMI SP-10, EN 865/ASTM F 1415. Connectivity: Strip chart printer output. Power: Medical electrical.
Indications for Use
Indicated for monitoring NIBP, SpO2, and temperature in adult or pediatric patients in any healthcare environment. Temperature measurement via oral, axillar, or rectal placement. No specific contraindications listed.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
K051038
Image /page/0/Picture/1 description: The image shows a symbol that appears to be a combination of a circle, a line, and two short horizontal dashes. A diagonal line intersects the circle, extending beyond it at both ends. The circle is positioned around the middle of the line. Two short, horizontal dashes are placed on either side of the circle and line, roughly aligned with the horizontal center of the symbol.
15 March 2005
# AUG 1 2 2005
510K Summary
## Model 506 Patient Monitor
Alex Kaplan Contact: Director of QA & RA Criticare Systems, Inc. 20925 Crossroads Circle Waukesha, WI 53186 USA 262-798-8282 Voice 262-798-8290 FAX
Trade Name: 506 Patient Monitor
Common Name: Vital Signs Monitor
Classification Name: Monitor, Physiological, Patient (74 MWI)
Substantial Equivalence is claimed to : 507EL Vital Signs Monitoring System (K022435).
## Device Description:
The 506 monitor measures and displays real time physiological data of the patient, including a pulse bar plethysmogram and numerical data. The 506 can be used to monitor one or more of the following parameters: Noninvasive BP (NIBP), SpO2, and Temperature. For all these vital parameters, the 506 will be capable of limit alarms and alerts, printing of strip chart recordings and storing trends for retrospective review.
#### Intended Use:
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarms and patient data analysis will be available to the care provider from the monitor.
#### Comparison with predicate device:
Criticare Systems Inc. has developed and distributed physiological monitoring devices worldwide since its inception in 1984. The 506 monitor utilizes existing core technologies from the predicate 507EL monitor for patient monitoring of NIBP, SpO2,
> @2005 Criticare Systems, Inc All Rights Reserved
Yz
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a handwritten symbol. The symbol consists of a circle with a line going through it. There are two short horizontal lines on either side of the symbol.
and Temperature. Temperature measurement is provided by Kendall's Filac FasTemp module which has been integrated into the 506 monitor. This temperature measurement technology allows predictive or continuous temperature measurement on adult or pediatric patients in oral, axillar, or rectal placements. SpO2 can optionally be provided by Nellcor's SpO2 module. This SpO2 technology has been listed with the capability to be used in an environment where resistance to clinical motion is required. The patient data collected by the 506 monitor is displayed for the user on an alphanumeric display and LEDs equivalent to the predicate device. The 506 monitor utilizes alphanumeric display technology in combination with LED numeric displays. Membrane key panels provide a user interface equivalent to the predicate device. The packaging design of the 506 monitor is molded plastic and allows for it to be either a stationary monitor or to be used during patient translocation within the healthcare facility, as did the predicate 506EL.
## Determination of Substantial Equivalence:
The 506 monitor performance for each monitoring modality has been confirmed to be equivalent to the predicate device. Additionally, the 506 complies with applicable safety and performance standards (detailed below) for each monitoring modality and verification of compliance has been completed. The patient monitoring technologies present in the 506 monitor have been in clinical use for at least six years in the 507EL monitor and it's predicates. CSI's field experience with these modalities in the predicate devices has been satisfactory. This combination of equivalence testing, applicable objective standards compliance and field experience substantiates a high level of confidence in the safety and efficacy of the 506 monitor.
Therefore, the 506 monitor is substantially equivalent to the predicate devices.
#### Compliance to standards and regulations:
The 506 Vital Signs Monitor complies with the following national and international standards:
#### Safety
UL 60601-1 Medical Electrical Safety EN 60601-1-2 EMC Compliance IEC 60601-2-49 Multiparameter Monitor Safety ISO 10993-5.10-11 Biocompatibility
#### Performance
IEC 60601-2-30 NIBP Safety EN1060-1 NIBP Performance EN 1060-3 NIBP Performance {including EN 475 Alarm Performance} AAMI SP-10 NIBP Performance EN 865 Oximetry Performance (Equivalent to ASTM F 1415)
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.
AUG 1 2 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Criticare Systems, Inc. c/o Mr. Alex Kaplan Director of QA and RA 20925 Crossroads Circle Waukesha, WI 53186
K051038 Re: Trade Name: 506 Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: DRT Dated: August 5, 2005 Received: August 9, 2005
Dear Mr. Kaplan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2- Mr. Alex Kaplan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Freast oc advised that I DTT 3155aanse or our device complies with other requirements of the Act that I DA has made a acterinaulations administered by other Federal agencies. You must of any I cochar statutes and regaranents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFN in the quality systems (DS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product tadiation only to begin marketing your device as described in your Section 510(k) I his iciter will anow you to cogm mains of substantial equivalence of your device to a legally premarket notification " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spocific at noliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micrimational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
B.Brimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications For Use
510(k) Number (if known): K051038
Device Name: 506 Patient Monitor
---
Indications for Use:
The 506 Patient monitor interprets and displays real time physiological data of the patient I he 300 I attent inomitor interpred and safigured to monitor Noninvasive BP (NIBP), S.O., and Temperature. For each patient vital parameter, the 506 will be capable of providing limit Temperature. Tor each patient that pat recordings and storing data trends for retrospective alarins and arets, primatig of carpexisting core technologies from the predicate 507EL monitor for patient monitor unfized embining of Niemperature. Temperature measurement is provided 101 patient momtoring of Nibly Op on thich has been integrated into the 506 monitor. This of Rendan's I has Fear enterest and one predictive or continuous temperature measurement temperature measurements in oral, axillar, or rectal placements. SpO2 can optionally be provided by Nellcor's SpO2 module.
provided by Nentor of DPOD model.
This system is intended to monitor physiological parameters of patients within any healthcare I mis system is intended to interpret the monitored data made available, will be a environinent. The user, respendere w sistem alarms and patient data analysis will be available to the care provider from the monitor.
Prescription Use (Part 21 CFR 801.109)
AND/OR
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
B.B. Summer
Division Slan-Om Division of Cardiovascular Devices 510(k) Number Korloßo
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.