A clinical literature review was conducted to evaluate design areas and determine the optimum configuration for the Spiral Laminar Flow™ Vascular Arteriovenous Graft.
Literature review; Design optimization
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical Literature Review; Literature review
—
Not applicable for this study
Optimum configuration for graft design
Indications for Use
The Spiral Laminar Flow™ Vascular Arteriovenous Graft is a vascular prosthesis, which is intended for use as a subcutaneous arteriovenous conduit for vascular access during hemodialysis. ONLY trained and qualified physicians and/or surgeons, under the controlled conditions of a hospital operating theatre environment are indicated for use of this device for implantation.
Device Story
Device is a vascular prosthesis for hemodialysis access; consists of straight ePTFE tubular graft with external spiral flow inducer and inducer indicator made of ChronoFlex C-80A polyurethane. Inducer segment propagates spiral laminar flow into distal circulation to improve hemodynamics; indicator ring marks proximal end of inducer to prevent cannulation in that segment. Used in hospital operating theatres by surgeons; implanted subcutaneously. Benefits include improved flow characteristics compared to standard straight grafts. Device does not use active sensors or software; flow induction is purely mechanical based on graft geometry.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (ISO 7198 physical testing, water permeability, tensile strength, suture retention, dynamic compliance), computational fluid dynamics (CFD), flow rig analysis, and animal studies (mini-pigs, sheep, dogs) evaluating safety, blood flow models, and performance of the spiral flow design.
Technological Characteristics
Straight tubular ePTFE vascular graft with external spiral flow inducer and indicator ring. Inducer/indicator material: ChronoFlex C-80A (biodurable medical grade polyurethane). Diameters: 5, 6, 7mm. Mechanical flow induction principle. Sterilization method not specified. No software or active energy source.
Indications for Use
Indicated for patients requiring subcutaneous arteriovenous conduit for hemodialysis vascular access. Implantation must be performed by trained physicians/surgeons in a hospital operating theatre.
Regulatory Classification
Identification
A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”
{0}------------------------------------------------
K०५५०५५
MAR 2 3 2010
. i
## Tayside Flow Technologies Ltd. Traditional 510(k)
For the Spiral Laminar Flow™ Vascular Anteriovenous Graft 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
Submitter's Name:
Tayside Flow Technologies Limited
Submitter's Address:
Tayside Flow Technologies Ltd: Unit 22, Prospect Business Centre Technology Park, Dundee Scotland, DD2 1TY
+44 (0) 1382 598 532 Telephone Fax +44 (0) 1382 598 528
Establishment Registration Number:
3007676031
Contact Person:
Edwin Lindsay
Telephone +44 (0) 7917134922
Date Prepared:
18" March 2010
{1}------------------------------------------------
For the Spiral Laminar Flow™ Vascular Arteriovenous Graft 510(k) Summary
### Device Classification Information:
| Regulation<br>Number | Device Name | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|---------------------------------------------------------------|-----------------|-----------------|-------------------------|
| 870.3450 | Prosthesis, Vascular Graft,<br>Of 6mm And Greater<br>Diameter | Class 2 | DSY | Cardiovascular |
| 870.3450 | Prosthesis, Vascular Graft,<br>Of Less Then 6mm<br>Diameter | Class 2 | DYF | Cardiovascular |
### Device Trade Name:
Spiral Laminar Flow™ Vascular Arteriovenous Graft
### Device Common Name:
ePTFE Vascular Graft with SLF™
### Intended Use:
The Spiral Laminar Flow™ Vascular Arteriovenous Graft is a vascular prosthesis, which is intended for use as a subcutaneous arteriovenous conduit for vascular access during hemodialysis.
ONLY trained and qualified physicians and/or surgeons, under the controlled conditions of a hospital operating theatre environment are indicated for use of this device for implantation.
### Summary of Substantial Equivalence:
The Tayside Flow Technologies (TFT) Spiral Laminar Flow™ Vascular Arteriovenous Graft is ട്രവിവിച്ചിരിക്കുന്നു. അവലംബം Limited SwitlGraft™ Graft (K051312/K060741), Bard IMPRA Venaflo® Vascular Access Graft (K052282), Bard IMPRA CenterFlex® Vascular Access Graft (K924360) and Vascutek PTFE Supported ePTFE Vascular Prostheses (K043552).
### Device Description:
The TFT Spiral Laminar Flow™ Vascular Arteriovenous Graft is to be used as an arteriovenous conduit for hemodialysis access. The graft has a specially designed section which is intended to induce spiral laminar flow.
This section is designed to propagate spiral flow though the graft and into the distal circulation. TFT Spiral Laminar Flow™ Vascular Arteriovenous Graft is manufactured from a straight tubular expanded polytetrafluoroethylene (ePTFE) vascular graft. The straight graft is combined with TFT's unique SLF™ external spiral flow inducer and inducer indicator, both made from ChronoFlex® C-80A; a Biodurable Medical Grade polyurethane.
The inducer indicator is a palpable ring over the proximal end of the spiral flow inducer. Its purpose is to indicate to the user where the spiral inducer segment begins since it is intended that cannulation in this segment should be avoided.
{2}------------------------------------------------
For the Spiral Laminar Flow™ Vascular Arteriovenous Graft
Technological Characteristics:
A comparative review of the TFT Spiral Laminar Flow™ Vascular Arteriovenous Graft with the predicate devices found that the technological characteristics, performance and principle of operation were substantially equivalent.
A comparison is presented in the table below:
| Property | New Device:<br>TFT SLF™<br>Vascular<br>Arteriovenous<br>Graft | TFT SLF™<br>Vascular<br>Peripheral<br>Graft | Veryan Medical<br>SwirlGraft™<br>Vascular Access Graft | Bard Venaflo<br>Vascular<br>Access Graft | Bard IMPRA<br>CenterFlex<br>Vascular Access<br>Graft | Vascutek<br>Arteriovenous<br>Graft |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------|------------------------------------------------------------------|
| Device<br>Manufacturer | Tayside Flow<br>Technologies<br>Ltd | Tayside Flow<br>Technologies<br>Ltd | Veryan Medical<br>Ltd. | Bard IMPRA | Bard IMPRA | Vascutek Ltd. |
| Device Trade<br>Name | Spiral Laminar<br>Flow™<br>Vascular<br>Arteriovenous<br>Graft | Spiral Laminar<br>Flow™<br>Vascular<br>Peripheral Graft | SwirlGraft™<br>Vascular Access<br>Graft | Venaflo™ II<br>Vascular Graft | IMPRA CenterFlex<br>Grafts | Maxiflo With<br>PTFE Support<br>Seal PTFE with<br>PTFE Support |
| 510(K) Number | N/A | K083169 | K051312/<br>K060741 | K052282 | K924360 | K043552 |
| Device<br>Common<br>Name | ePTFE<br>Vascular Graft<br>with SLF™ | ePTFE<br>Vascular Graft<br>with SLF™ | ePTFE Vascular<br>Graft | ePTFE<br>Vascular Graft<br>Prosthesis | ePTFE Vascular<br>Graft Prosthesis | ePTFE Vascular<br>Graft Prosthesis |
| Device<br>Classification<br>name | Prosthesis,<br>Vascular Graft,<br>Of 6mm and<br>Greater<br>Diameter<br>and<br>Prosthesis,<br>Vascular Graft,<br>Of Less Then<br>6mm Diameter | Prosthesis,<br>Vascular Graft,<br>Of 6mm and<br>Greater<br>Diameter<br>and<br>Prosthesis,<br>Vascular Graft,<br>Of Less Then<br>6mm Diameter | Prosthesis,<br>Vascular Graft,<br>Of 6mm and<br>Greater<br>Diameter | Prosthesis,<br>Vascular<br>Graft, Of 6mm<br>and Greater<br>Diameter | Prosthesis, Vascular<br>Graft, Of 6mm and<br>Greater Diameter | Prosthesis,<br>Vascular Graft, Of<br>6mm and Greater<br>Diameter |
| Device<br>Classification | Class II | Class II | Class II | Class II | Class II | Class II |
| Materials | ePTFE and PU | ePTFE, PTFE<br>and PU | ePTFE | ePTFE and<br>PTFE | ePTFE and PTFE | ePTFE and PTFE |
| Tube | External spiral<br>inducer 6cm<br>long at the<br>distal end of<br>tube | External spiral<br>inducer 10cm<br>long at the<br>distal end of<br>tube | Small amplitude<br>helical geometry<br>along the tube<br>length | Straight tube<br>distal end<br>diameter<br>radially<br>expanded | Straight tube | Straight tube |
| Beading | No | Yes | No | Partial<br>beading<br>optional | Partial beading<br>optional | Yes |
| Distal End<br>Modification | Yes | Yes | No | Yes | No | No |
| Diameter | 5, 6 and 7mm | 5, 6, 7 and 8mm | 6mm | 6, 7 and 8mm | 6, 7 and 8mm | 6, 7 8 and 10 mm |
| Intended Use | Subcutaneous<br>arteriovenous | Bypass or<br>reconstruction | Subcutaneous | Subcutaneous | Subcutaneous | Creation of<br>subcutaneous |
{3}------------------------------------------------
For the Spiral Laminar Flow™ Vascular Anteriovenous Graft 510(k) Summary
| Property | New Device:<br>TFT SLFTM<br>Vascular<br>Arteriovenous<br>Graft | TFT SLFTM<br>Vascular<br>Peripheral<br>Graft | Veryan Medical<br>SwirlGraftTM<br>Vascular<br>Access Graft | Bard Venaflo<br>Vascular<br>Access Graft | Bard IMPRA<br>CenterFlex<br>Vascular Access<br>Graft | Vascutek<br>Arteriovenous<br>Graft |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| | conduits for<br>hemodialysis<br>access | of occluded or<br>diseased<br>peripheral<br>arterial blood<br>vessels. | conduit for<br>vascular access. | conduits for<br>blood access<br>only | conduits for blood<br>access only | arteriovenous<br>conduits for blood<br>access, bypass or<br>reconstruction of<br>occluded or<br>diseased arterial<br>blood vessels |
| Intended<br>Location of<br>Use | Forearm,<br>upper arm,<br>thigh | Either above or<br>below the knee | Forearm, upper<br>arm, thigh | Forearm,<br>upper arm.<br>thigh | Forearm, upper arm,<br>thigh | Forearm, upper<br>arm, thigh, below<br>knee |
| Device<br>Description | An ePTFE<br>graft with a<br>pre-trimmed<br>cuff and unique<br>SLFTM external<br>spiral flow<br>inducer and<br>inducer<br>indicator made<br>from Medical<br>Grade<br>polyurethane<br>(PU) at the<br>distal end . | An ePTFE<br>supported graft<br>with a pre-<br>trimmed cuff<br>and and unique<br>SLFTM external<br>spiral made<br>from Medical<br>Grade<br>polyurethane<br>(PU) at the<br>distal end. | A 6mm<br>expanded<br>ePTFE vascular<br>graft that is<br>manufactured<br>with a small<br>amplitude helical<br>geometry along<br>its entire length.<br>The Swirl Graft<br>Vascular Access<br>Graft is a<br>standard wall<br>ePTFE<br>construction with<br>no external<br>support. | An expanded<br>ePTFE<br>vascular graft<br>with a radially<br>expanded pre-<br>formed venous<br>cuff at the<br>distal end.<br>The grafts are<br>available in<br>various<br>lengths and<br>diameters, in<br>straight,<br>stepped and<br>tapered<br>configurations,<br>with and<br>without<br>external<br>support and<br>with or without<br>a carbon lining | An expanded<br>ePTFE vascular<br>graft<br>The grafts are<br>available in various<br>lengths and<br>diameters, in<br>straight, tapered and<br>stepped<br>configurations, with<br>and without external<br>support and with or<br>without a carbon<br>lining | A supported or<br>unsupported<br>ePTFE vascular<br>graft |
### Performance Data:
Bench testing and animal data demonstrated that the safety and effectiveness of the TFT Spiral Laminar Flow™ Vascular Arteriovenous Graft is equivalent to the predicate devices.
- . Biocompatibility Testing
Tayside Flow Technologies (TFT) Spiral Laminar Flow™ Vascular Arteriovenous Graft is a straight tubular vascular graft made from expanded polytetrafluoroethylene (ePTFE). The unique SLF™ spiral flow inducer and inducer indicator are injection molded onto the outer surface of the straight graft. The inducer and indicator are made from ChronoFlex® C-80A; a Biodurable Medical Grade polyurethane, further information is available in as section 11.
To determine the biocompatibility testing required for the TFT graft materials the following was taken into account:
{4}------------------------------------------------
### For the Spiral Laminar Flow™ Vascular Arteriovenous Graft 510(k) Summary
- The requirements of ISO 10993 Part1 and especially Annex A of this part which . provides guidance on the selection of tests
- The fact that the materials used in the TFT graft are already well-characterised and . are approved for use as long-term vascular implants. Section 11 includes a list of vascular devices which make use of these materials and which are 510(k) cleared.
Based on this analysis it was determined that the tests specified in the following parts of ISO 10993 were appropriate: parts 4, 5, 6, 10, 11 and 13.
These tests performed in compliance with GLP, confirmed that the biocompatibility of the Spiral Laminar Flow™ Vascular Arteriovenous Graft is sufficient for their intended use.
These test results are further supported by the fact that these materials have been in clinical use in implant applications for many years with good results.
Further proof of the biocompatibility of the materials used in the TFT graft was provided by the fact that both the materials used have been cleared by FDA for vascular graft use as already mentioned. The details of the 510(k)s are given in section 11.
#### Performance Testing .
The determination of the optimum configuration for the TFT graft was based on 1) a literature review, 2) computational fluid dynamics (CFD) and 3) flow rig work. A number of design areas were evaluated:
- 1. Clinical Literature Review (Appendix B)
- 2. Helical angle (computational fluid dynamics and flow rig)
- 3. Number of ridges (computational fluid dynamics and flow rig)
- Height of ridge (computational fluid dynamics and flow rig) ধ
- Ridge profile (computational fluid dynamics and flow rig) u
- Graft profile in polyester (flow rig) 6.
Good correlation of CFD data, flow rig data and in vivo data confirmed the suitability of the design.
### Haemodynamic Testing: Effect of Diameter
As 6mm and 8mm grafts are commercially available for infrainguinal bypass, there is no 'industry standard' as far as diameter is concerned. However, 6mm through 8mm grafts are commonly used for arteriovenous access. To verify that 6mm and 8mm grafts have comparable blood flow characteristics, in-house flow rig and computational fluid dynamic work has been carried out.
### Physical Testing
Characterisation and Physical testing has been carried out to ISO7198 Cardiovascular implants - tubular vascular prosthesis. This includes :-
- . Water permeability
- Circumferential tensile strength
{5}------------------------------------------------
For the Spiral Laminar Flow™ Vascular Arteriovenous Graft 510(k) Summary
- Longitudinal tensile strength .
- Probe burst strength .
- Usable length of formed material .
- Relaxed internal diameter .
- . Wall thickness
- . Pressurised internal diameter
- . Suture retention strength
- . Kink diameter / radius
- . Dynamic compliance
The test results demonstrate that the TFT Spiral Laminar Flow™ Vascular Arteriovenous Graft has sufficient strength and physical properties to perform as intended under the expected in vivo loading conditions.
### Animal Testing
### Development of SLF grafts
TFT has completed the following animal studies during the development of the TFT Spiral Laminar Flow™ ePTFE Vascular Grafts.
| STUDY | Length of<br>Recovery<br>Study | Species Used | Description | Number In<br>Each Group | Purpose |
|----------------------------------------|--------------------------------|--------------|----------------------------------------------------------------------------|-------------------------|-----------------------------------------------------------|
| HSAW (100) | 6 Months | Mini-Pig | Bard Control vs.<br>Bard+TFT Spiral<br>POLYESTER 8 mm | 8 | |
| HSAW (103) | 1 Week | Mini-Pig | Bard Control vs.<br>Bard+TFT Spiral<br>POLYESTER 8 mm | 1 | |
| HSAW (104) | 1 Months | Mini-Pig | Bard Control vs.<br>Bard+TFT Spiral vs.<br>TFT (McMurry)<br>POLYESTER 8 mm | 2 | Blood flow<br>Model<br>Development |
| NPIMR (pilot) | 1 Month | Sheep | TFT (McMurry)<br>POLYESTER 8 mm | 1 | |
| NPIMR (SH-<br>02) | 3 Month | Sheep | Bard Control vs.<br>Bard+TFT Spiral<br>POLYESTER 8 mm<br>Model Development | 5 | |
| CHUM (TFT-<br>CHUM pilot)<br>TFT-8-004 | 2 Weeks | Dog | Bard Control vs. TFT<br>(McMurry)<br>POLYESTER 8 mm | 2 | Blood flow<br>Model<br>Development<br>AND Safety |
| CHUM (TFT-<br>8-005) | 6 Weeks | Dog | Bard Control vs. TFT<br>(McMurry)<br>POLYESTER 8 mm | 2 | Study<br>Development |
| CHUM and<br>MHI<br>(TFT-8-006) | 20 Weeks | Dog | Bard Control vs. TFT<br>(McMurry)<br>POLYESTER<br>8 mm | 10 | Safety Study/<br>Proof of<br>principle and<br>Performance |
| TFT-8-007 | 14 Weeks | Sheep | BARD ePTFE<br>8mm | 2 | Proof of<br>principle<br>study |
{6}------------------------------------------------
For the Spiral Laminar Flow™ Vascular Arteriovenous Graft 510(k) Summary
| STUDY | Length of<br>Recovery Study | Species Used | Description | Number In<br>Each Group | Purpose |
|-----------------|-----------------------------|--------------|---------------------------------------------------------------------------|-------------------------|----------------------------------------------------------------------------------|
| NPIMR<br>(SH03) | 2 Weeks | Sheep | 4 mm Polyester Spiral<br>Grafts | 4 | Proof of<br>principle study<br>(Performance<br>Study 4 mm<br>Vascular<br>Grafts) |
| G0003-09 | 2 Weeks | Pig | 6mm ePTFE Spiral<br>Access Grafts v 6mm<br>ePTFE Control<br>Access Grafts | 2 | Proof of<br>principle study<br>(Performance<br>Study 6 mm<br>Vascular<br>Grafts) |
### Safety and Effectiveness:
The TFT SLF™ Vascular Arteriovenous Graft utilises similar technology currently found in legally marketed predicate devices. Based on testing and comparison with the predicate devices, the TFT SLFTM Vascular Arteriovenous Graft indicated no adverse indications or results. It is our determination that the TFT Spiral Laminar Flow™ Vascular Arteriovenous Graft is safe, effective and performs within its design specifications and is substantially equivalent to the predicate device.
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Tayside Flow Technologies LTD c/o Mr. Rudy Mounia 1141 E. Hawken Way Chandler, AZ 85286
## MAR 2 3 2010
Re: K094044
Trade Name: Spiral Laminar Flow™M Vascular Arteriovenous Graft Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular graft prosthesis Regulatory Class: II (two) Product Code: DSY Dated: December 28, 2009 Received: December 31, 2009
Dear Mr. Mounia:
..
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Rudy Mounia
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Dma R. Vi. Amer
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Tayside Flow Technologies Ltd.
Traditional 510(k)
For the Spiral Laminar Flow™ Vascular Arteriovenous Graft
### Indications for Use
510(k) Number (if known): 《0 9 40 4-4
Device Name: Spiral Laminar Flow™ Vascular Arteriovenous Graft
Indications for Use:
The Spiral Laminar Flow™ Vascular Arteriovenous Graft is a vascular prosthesis, which is intended for use as a subcutaneous arteriovenous conduit for vascular access during hemodialysis.
ONLY trained and qualified physicians and/or surgeons, under the controlled conditions of a hospital operating theatre environment are indicated for use of this device for implantation,
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
~~~~~~~~~ R ~~~~~~~~
(Division Sign-Off)
Division of Cardiovascular Devices
Page 1 of 1
510(k) Number_ko94044
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.