Venaflo™ II Vascular Grafts are indicated for use as subcutaneous arteriovenous conduits for blood access only.
Device Story
Venaflo II is a vascular graft used as a subcutaneous arteriovenous conduit for blood access. It features a pre-formed venous cuff designed to promote laminar flow at the venous anastomosis. The device is available in various lengths and diameters, in straight or stepped configurations, and with or without external support. Modifications from the predicate include printed trim lines on the venous cuff to optimize sizing for veins 8mm and smaller, enhancing hemodynamic performance. The graft is implanted by a surgeon to facilitate vascular access for patients requiring hemodialysis or similar blood access procedures.
Clinical Evidence
Bench testing only. Verification of the modified device was conducted in accordance with FDA guidance and industry standards, including ANSI/AAMI VP20-1994, ANSI/AAMI/ISO 7198, and ISO 10993 parts 1, 7, and 11135. All test results met predetermined acceptance criteria.
Technological Characteristics
Vascular graft prosthesis; pre-formed venous cuff; straight or stepped configurations; optional external support. Materials evaluated per ISO 10993-1 (biocompatibility) and ISO 10993-7 (EO residuals). Sterilization via ethylene oxide (ISO 11135).
Indications for Use
Indicated for use as subcutaneous arteriovenous conduits for blood access.
Regulatory Classification
Identification
A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”
{0}------------------------------------------------
## Venafio II Vascular Graft 510(k) Summary of Safety and Effectiveness 21 CFR 807.92(a).
### General Information:
| Submitter Name: | Bard Peripheral Vascular, Inc. |
|----------------------|----------------------------------------------------------------|
| Address: | 1625 W. Third Street<br>P. O. Box 1740<br>Tempe, AZ 85280-1740 |
| Telephone Number: | (480) 894-9515 ext. 2836 |
| Fax Number: | (480) 449-2546 |
| Contact Person: | Glenn Norton |
| Date of Preparation: | August 19, 2005 |
### Device Information:
| Device Name: | Venaflo™ II Vascular Graft |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Venaflo™ II |
| Common/Usual Name: | Arteriovenous Vascular Graft |
| Classification Name: | 74 DSY - Prosthesis, Vascular Graft,<br>Of 6mm and Greater Diameter<br>21 CFR 870.3450 - Class II<br>Vascular graft prosthesis |
| Classification Panel: | Cardiovascular |
### Predicate Device:
IMPRA Venaflo™ Vascular Graft, K981079, concurrence date 05/07/1998.
## Summary of Change:
The Venaflo Vascular Graft is being modified with printed trim lines on a more The Vehance Vaboutived venous cuff profile to enhance the good hemodynamic surformance of the predicate design. All other aspects of the modified device remain performance of the predicate device. The changes optimize venous cuff sizing for the the bains as are rimming the cuff for veins 8mm and smaller, and improve hemodynamic performance at the venous anastomosis.
### Device Description:
Venaflo II arteriovenous vascular grafts have a pre-formed venous cuff to promote venano II atterrovenous variate at the venous anastomosis. Venaflo II grafts are
{1}------------------------------------------------
available in various lengths and diameters, in straight and stepped configurations, with and without external support.
## Intended Use of Device:
Venaflo™ II Vascular Grafts are indicated for use as subcutaneous arteriovenous conduits for blood access only.
## Technological Comparison to Predicate Device:
The technological characteristics of Venaflo II vascular grafts are substantially equivalent to those of the predicate Venaflo vascular graft in terms of intended use, oquireation, user population, basic design, performance, labeling, packaging, and sterilization method.
## Non-Clinical Performance Data:
Design verification of the modified device was done with conformance to or evaluated based on the following FDA Guidance and industry standards:
- Guidance Document for Vascular Prosthesis 510(k) Submissions, dated . 11/01/2000
- ANSI/AAMI VP20-1994, Cardiovascular Implants Vascular Prosthesis ●
- ANSI/AAMI/ISO 7198: 1998/2001, Cardiovascular implants Tubular . vascular implants
- AAMI/ANSI/ISO 10993-1: 1997, Biological evaluation of medical devices -. Part 1: Evaluation and testing, and the FDA Modified ISO 10993 Test Profile
- AAMI/ANSI/ISO 10993-7: 1995, Biological evaluation of medical devices -. Part 1: Ethylene Oxide Sterilization Residuals
- AAMI/ANSI/ISO 11135:1994, Medical Devices -- Validation and Routine . Control of Ethylene Oxide Sterilization
All test results confirm the modified device to be substantially equivalent to the predicate device.
## Conclusions:
Venaflo II Vascular Grafts met all predetermined acceptance criteria of design verification and validation as required by the risk analysis, and as specified by applicable standards, guidance's, test protocols and/or customer inputs. Venaflo II Vascular Grafts are substantially equivalent to the legally marketed predicate device, the Venaflo Vascular Graft, K981079, concurrence date May 5, 1998.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal are three stylized lines that appear to be a symbol or logo for the department.
SEP 2 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bard Peripheral Vascular, Inc. c/o Mr. Glenn Norton 1625 W. Third Street P.O. Box 1740 Tempe, AZ 85280-1740
Re: K052282
Venaflo™ II Vascular Graft Venano - Frances 21 CFR 870.3450 Regulation Number: 21 CFR 8 70.3430
Regulation Name: Vascular Graft Prosthesis of Less than 6MM Diameter
S Regulatiory Class: Class II (two) Product Code: DSY Dated: August 19, 2005 Received: August 23, 2005
Dear Mr. Norton:
Deal 110.00 We have reviewed your Section 510(t) premarket notification of the indications
referenced above and have determined the device is substantially equivalent in interstate welered above and have determined the devices marketed in interstate
for use stated in the enclosure) to legally marketed in interstate for use stated in the enclosure) to legally market predical Device Amendents, or to
commerce prior to May 28, 1976, the enactment date of the Medical Device Amendents, or to for as a more of the mastment and of the wedness of the Federal Food. Dough, the Federal Food, Drug,
devices that have been reclassified in accordance with the provisited app devices that have been reclassified in accordance with the proval application (PMA).
devices that have been reclassified in accordance approval opproval oppiocions of the Ac de cosmetic Act (Act) that do not require approval of a previsions of the Act. The Act. The
You may, therefore, market the device, subject to the general ontrols provisions and Course, market the Act include requirements for amual registration, listing of
general controls provisions of the Act include requirements for annual misbranding and s and of the Act include requirements for millian regultents for minds regional would be
devices, good manufacturing practice, labeling, and prohibitions against misbranding adulteration.
additions is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) into either class if (special connelle)
may be subject to such additional controls. Existing major regulations affecting your device n your article of Federal Regulations, Title 21, Parts 800 to 898. In addition, PDA may
may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may
b be found in the Code of Federal Regulations, Title 21, Farls C
publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Glenn Norton
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Section 1-B
## Venaflo™ II Vascular Graft Special 510(k)
# INDICATION(S) FOR USE STATEMENT*
I state in my capacity as Regulatory Affairs Associate Project Manage of Bard I state in my capacity as regulatery - taxion [510(k)] for the Venaflo™ II Vascular Graft is indicated for the following:
" Venaflo ™ II Vascular Grafts are indicated for use as subcutaneous arteriovenous conduits for blood access only."
Signature of 510(k) Submitter: Printed Name of Submitter:
Glenn Norton
Glenn Norton
Date:
8/19/05
*Suggested language and format to meet the requirements of sections 513(i) of the *Suggested fallguage and format to meet the required and sections 807.92(a)(5) and 801.4 of the Code of Federal Regulations, Title 21.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|-------------------|---------------------------------------------------------------------------|
| | <div><img alt="Signature" src="signature.png"/></div> (Division Sign-Off) |
| 510(k) Number | k052282-Division of Cardiovascular Devices |
| | 510(k) Number k052282 |
| Division Sign-Off | |
| | Office of Device Evaluation |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.