K051417 · C.R. Bard, Inc. · LJS · Jun 30, 2005 · General Hospital
Device Facts
Record ID
K051417
Device Name
6 FR FL POWERHOHN AND POWERLINE CATHETER
Applicant
C.R. Bard, Inc.
Product Code
LJS · General Hospital
Decision Date
Jun 30, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerHohn" and PowerLine™ Catheters are intended for short or long term access to the central venous system for intravenous therapy and blood sampling.
Device Story
Open-ended, radiopaque polyurethane central venous catheter (CVC); 6 Fr DL size; 40-50 cm length; reverse taper design; depth indicators on shaft. Used for IV therapy, blood sampling, and power injection of contrast media. Operated by clinicians in clinical settings. Purple colorant added for differentiation; extension leg, junction, and clamp ID tag include identification markings. Device provides central venous access; enables high-pressure contrast media delivery (up to 300 psi) for diagnostic imaging. Benefits include reliable, long-term vascular access for various therapeutic and diagnostic procedures.
Clinical Evidence
Bench testing only. Performance verification and validation testing conducted per FDA guidance for intravascular catheters and ISO 10555 standards. Biocompatibility testing performed per ISO 10993. FMEA analysis used to evaluate risks. No clinical data presented.
Technological Characteristics
Radiopaque polyurethane construction; 6 Fr DL diameter; 40-50 cm length; reverse taper design. Sensing/actuation: passive fluid delivery/withdrawal. Energy source: external power injector (max 300 psi). Sterilization: Ethylene Oxide (AAMI/ANSI/ISO 11135). Connectivity: None. Standards: ISO 10555-1, ISO 10555-3, ISO 10993.
Indications for Use
Indicated for short or long term central venous system access in patients requiring IV fluid, blood product, drug, or parenteral nutrition administration, blood withdrawal, or power injection of contrast media. Max infusion rate 5cc/sec; max power injector pressure 300 psi.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
JUN 3 0 2005
6 Fr DL PowerH Special 510(k)
page 1 of 3
00012
-31 એ
Section 6
### 510(k) Summary
## 6 Fr DL PowerHohn" and PowerLine" Catheter
### 510(k) Summary of Safety and Effectiveness Information 21CFR 807.92
#### 6.1 Submitter Information
| Submitter Name: | Bard Access Systems, Inc. (BAS)<br>[Subsidiary of C.R. Bard, Inc.] |
|----------------------|--------------------------------------------------------------------|
| Address: | 5425 W. Amelia Earhart Drive<br>Salt Lake City, UT 84116 |
| Telephone Number: | (801) 595-0700, Ext. 4903 |
| Fax Number: | (801) 595-5425 |
| Contact Person: | Peggy Keiffer |
| Date of Preparation: | May 26, 2005 |
#### 6.2 Device Name
| Device Name: | 6 Fr DL PowerHohn™ and PowerLine™ Catheter |
|-----------------------|--------------------------------------------|
| Trade Name: | 6 Fr DL PowerHohn™ and PowerLine™ Catheter |
| Common/Usual Name: | Central Venous Catheter (CVC) |
| Classification Panel: | General Hospital |
| Classification Name: | 80LJS - Long Term Intravascular Catheter |
| | 21 CFR 880.5970, Class II |
| | Central Venous Catheter (CVC) |
#### 6.3 Predicate Device Name(s)
| Device Name: | PowerHohnTM and PowerLineTM Catheter |
|-------------------------|-----------------------------------------------|
| Trade Name: | PowerHohnTM and PowerLineTM Catheter |
| Common/Usual Name: | Central Catheter (CVC) |
| Classification Name: | Long Term Intravascular Catheter (80 LJS) |
| Premarket Notification: | K050185, concurrence date -May 26, 2005. |
| Device Name: | Poly Per-Q-Cath PICC Catheter |
| Trade Name: | Poly Per-Q-Cath PICC Catheter |
| Common/Usual Name: | Peripherally Inserted Central Catheter (PICC) |
| Classification Name: | Long Term Intravascular Catheter (80 LJS) |
| Premarket Notification: | K034019, concurrence date - January 21, 2005 |
#### 6.4 Device Description
- The PowerHohn" and PowerLine™ Catheters are open-ended radiopaque polyurethane � catheters.
- Catheter size is 6 Fr DL with 50 cm (PowerLine) or 40 cm (PowerHohn) usable length. .
- . The catheter has a reverse taper design.
- . Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point.
- . Catheters are provided sterile in basic radiology configurations.
{1}------------------------------------------------
6 Fr DL PowerHo
3
- . Purple colorants were added to the catheter materials to provide the cath appearance that allows the end user to differentiate the PowerHohn" and PowerLine other CVC catheters.
- . The catheter extension leg. junction and clamp ID tag were printed with markings to identify the catheter as PowerHohn" and PowerLine" and to include information to facilitate proper use of the device.
#### 6.5 Intended Use
The PowerHohn" and PowerLine™ Catheters are intended for short or long term access to the central venous system for intravenous therapy and blood sampling.
The intended use has not changed.
#### 6.6 Indications for Use
The indications for use have not changed from the predicate PowerHohn" and PowerLine"" catheter (K050185).
PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and power injection of contrast media. The maximum recommended infusion rate is 5cc/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi.
#### Summary of Technological Characteristics in Relation to the Predicate Device 6.7
Does the new device have the same technological characteristics, e.g. design, material, etc.?
Not in all regards. The 6 Fr DL PowerHohn™ and PowerLine™ catheter has some minor differences from the predicate 5 Fr SL PowerHohn" and PowerLine" catheter. However, the basic fundamental scientific technology of the catheter has not changed.
### Could the new characteristics affect safety or effectiveness?
Yes. The new characteristics could affect safety or effectivencss of the device.
### Do the new characteristics raise new types of safety and effectiveness questions?
No. There are no new types of issues of safety and effectiveness.
## Do accepted scientific methods exist for assessing effects of the new characteristics?
Yes. The following FDA guidance documents and international standards were used to evaluate the device's performance:
- Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term ● Intravascular Catheters, March 16, 1995
- . ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements
- . ISO 10555-1:1997, Sterile, Single-Use Intravascular Catheters, General requirements, Amendment I
- ISO 10555-3:1997, Sterile, Single-Use Intravascular Catheters, Central venous catheters .
- AAMI/ANSI/ISO 11135:1994, Medical Devices Validation and Routine Control of Ethylene . Oxide Sterilization
{2}------------------------------------------------
No new design validation was required. The design validation for the predicate device issues pertaining to the subject device
6 Fr DL PowerHohn
Biocompatibility requirements of ISO 10993 Biological Evaluation of Medical Devices Part-1: Evaluation and Testing and the FDA Modified ISO 10993 Test Profile for externally communicating, blood contacting, long-term devices were met. All materials used in the manufacture of the subject device were previously cleared for similar applications by Bard Access Systems.
### Are performance data available to assess effects of new characteristics?
Yes. Verification and validation testing was performed according to protocols based on the above referenced guidance document recommendations and standards, as well as in accordance with inhouse protocols.
Do performance data demonstrate equivalence?
Yes. Performance data gathered in design verification testing demonstrated that the 6 Fr DL PowerHohn" and PowerLine" catheter is substantially equivalent to the predicate 5 Fr SL PowerHohn"" and PowerLine"" catheter, and the risks associated with use of the new device were found acceptable when evaluated by FMEA.
#### 6.8 Conclusion
The 6 Fr DL PowerHohn" and PowerLine " catheter meets all the predetermined performance acceptance criteria of the testing performed and, based on FDA 's decision tree, is substantially equivalent to the predicate device 5 Fr SL PowerHohn" and PowerLine" catheter, K050185, cleared May 26, 2005. and the Poly Per-Q-Cath PICC Catheters. cleared under K034019.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three wavy lines, which are meant to represent the human form.
Public Health Service
JUN 3 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Peggy Keiffer Senior Regulatory Affairs Manager C.R. Bard, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116
Re: K051417
Trade/Device Name: 6 Fr DL PowerHohn™ and PowerLine™ Catheter Regulation Number: 21 CFR 880.5970 Regulation Name: Long Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: May 31, 2005 Received: June 1, 2005
Dear Ms. Keiffer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Keiffer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sulette y. Michael Oms.
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
6 Fr DL Po and PowerLine
Section 1.2
Special 510(k)
- Page 1 of 1
# Indications for Use Statement
510(k) Number (if known):
Device Name: 6 Fr DL PowerHohn" and PowerLine™ Catheter
Indications For Use:
PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and power injection of contrast media. The maximum recommended infusion rate is 5cc/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Auu Suliom
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, D 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.