K093416 · Respironics, Inc. · CBK · Jan 29, 2010 · Anesthesiology
Device Facts
Record ID
K093416
Device Name
TRILOGY 200 VENTILATOR
Applicant
Respironics, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jan 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Respironics Trilogy200 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy200 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation.
Device Story
Trilogy 200 is a microprocessor-controlled, blower-based ventilator providing pressure support, pressure control, or volume-controlled ventilation. It supports CPAP, S, S/T, T, PC, PC-SIMV, AC, CV, and SIMV modes. The device delivers therapy invasively (via ET tube) or non-invasively (via masks) to patients weighing ≥ 5 kg. Used in hospitals, homes, and transport settings (wheelchairs/gurneys). Operators include clinicians or caregivers. The device features a user interface for setting clinical parameters and displaying data, plus fixed and adjustable alarms to alert users to treatment changes. It functions as a modified version of the Trilogy 100, incorporating a universal porting block. The system assists patients requiring mechanical ventilation, potentially improving respiratory outcomes through precise pressure/volume delivery.
Clinical Evidence
Bench testing only. Performance and functionality were verified through extensive waveform performance, triggering, and alarm functionality testing. All tests met established acceptance criteria, demonstrating that modifications did not negatively impact the device's performance compared to the predicate.
Technological Characteristics
Microprocessor-controlled, blower-based ventilator. Supports pressure support, pressure control, and volume-controlled ventilation. Features a universal porting block. Operates via AC/DC power. Designed for invasive and non-invasive delivery. Includes user interface and alarm system.
Indications for Use
Indicated for pediatric through adult patients weighing at least 5 kg (11 lbs) requiring continuous or intermittent mechanical ventilatory support. Suitable for invasive or non-invasive use in home, hospital, or portable settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
## TAB 3
.........
# الم
# JAN 2 9 2010
## 510(K) SUMMARY OF SAFETY & EFFECTIVENESS
| Original Date of Submission | 30 October 2009 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name | Trilogy 200 |
| Common/Usual Name | Ventilator, continuous, life supporting |
| Establishment Registration # | 2518422 |
| Address of Mfr. Facility | Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668<br>(724) 387-4120 FAX (724)-387-7490 CELL (724) 882-4120 |
| Classification | Class II device |
| Classification Panel | Anesthesiology Devices |
| Classification Reference | 21 CFR 868.5895 |
| Product Code | CBK - Continuous Ventilator, Facility Use |
| Predicate Device(s) | Respironics Trilogy 100 (K083526) |
| Labeling | Draft Labeling can be found in Tab 5. |
| Indications for Use | The Respironics Trilogy200 system provides continuous or intermittent<br>ventilatory support for the care of individuals who require mechanical<br>ventilation. Trilogy200 is intended for pediatric through adult patients<br>weighing at least 5 kg (11 lbs.). |
| | The device is intended to be used in home, institution/hospital, and<br>portable applications such as wheelchairs and gurneys, and may be<br>used for both invasive and non-invasive ventilation. |
| Reason for Submission | Modified device |
{1}------------------------------------------------
### Substantial Equivalence
The Trilogy 200 system has the following similarities to the previously cleared predicate device:
- O Same intended use.
- D Same operating principle.
- O Same technology.
- O Same manufacturing process.
There is no change to the intended use, operating principle, technology or manufacturing process for the Trilogy 200. There are minor mechanical changes to the Trilogy 200 to accommodate the universal porting block modification discussed in this submission. All other modifications did not change the electromechanical platform of the Trilogy 100, cleared in K083526.
To demonstrate performance and functionality of the Trilogy was unaffected as a result of these changes, extensive waveform performance, triggering and alarm functionality testing was performed. All tests were verified to meet the required acceptance criteria. Results of this testing concluded that the verification testing raises no new issues of safety or effectiveness.
Respironics has followed the FDA's Guidance for Industry and FDA Staff document "pre-market assessment of pediatric medical devices" and applied the principle of FDA's Least Burdensome Approach to demonstrate the Substantial Equivalence of the Trilogy 200 system. As a result we conclude that the existing indications for use can be safely and effectively applied to this device.
## Device Description
The Respironics Trilogy 200 Ventilatory Support System is a microprocessor controlled blower based pressure support, pressure control or volume controlled ventilator intended for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in the institutional/hospital, home and transport settings and is applicable for adults and pediatric patients weighing at least 5 kg (11 lbs) who require the following types of ventilatory support.
- CPAP Continuous Positive Airway Pressure
- S Spontaneous Ventilation
- S/T Spontaneous and Timed Ventilation
- T Timed Ventilation
- PC Pressure Control Ventilation
- PC-SIMV Pressure Controlled Synchronized Intermittent Mandatory Ventilation
{2}------------------------------------------------
AC – Assist Control Ventilation
CV - Control Ventilation
SIMV - Synchronized Intermittent Mandatory Ventilation
A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.
The Trilogy 200 Ventilator is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.
The Trilogy 200 Ventilator can deliver therapy to patients > 5kg (11 Ibs), invasively (via ET tube) or noninvasively (via masks that are cleared for use with this patient population including small child, pediatric and adult sizes).
(End of Tab.)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The bird is composed of three curved lines that create a sense of movement and flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Respironics, Incorporated Ms. Zita A: Yurko Director, Regulatory Affairs Sleep & Home Respiratory Group 1740 Golden Mile Highway Monroeville, Pennsylvania 15146
JAN 29 2010
Re: K093416
Trade/Device Name: Trilogy 200 Regulation Number: 21CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: October 30, 2009 Received: November 2, 2009
Dear Ms. Yurko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Ms. Yurko
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/Medica]Devices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
Page _ 1_ of__1
#### Indications for Use
510(k) Number (if known): K093416
Device Name:
The Respironics Trilogy 200 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy200 is intended for pediatric through adult patients weighing at least 5 kg (11 Ibs.).
The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation.
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
hubtt
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k093416
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.