K083526 · Respironics Inc., Sleep & Home Respiratory Group · CBK · Mar 13, 2009 · Anesthesiology
Device Facts
Record ID
K083526
Device Name
TRILOGY 100 VENTILATOR
Applicant
Respironics Inc., Sleep & Home Respiratory Group
Product Code
CBK · Anesthesiology
Decision Date
Mar 13, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Respironics Trilogy100 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.
Device Story
Microprocessor-controlled, blower-based ventilator; provides pressure support, pressure control, or volume-controlled ventilation. Inputs include user-defined clinical parameters via interface; device monitors patient-ventilator interaction. Outputs include delivered air/oxygen mixture via invasive (ET tube) or non-invasive (mask) interfaces; real-time pressure display; clinical data storage. Used in home, hospital, and portable settings (wheelchairs/gurneys) by clinicians or patients/caregivers. Alarms alert users to treatment changes. Supports multiple modes: CPAP, S, S/T, T, PC, PC-SIMV, AC, CV, SIMV. Benefits patients by providing necessary respiratory support for those unable to breathe adequately on their own.
Clinical Evidence
Bench testing only. Performance testing conducted per ASTM F1100-90, F1246-91, ISO 10651-2, and ISO 10651-6. EMC testing per IEC 60601-1-2. Electrical, mechanical, and environmental testing performed per FDA Draft Reviewers Guidance (1993). Software validation performed per FDA Guidance (2005). Results confirm design and system requirements met.
Technological Characteristics
Microprocessor-controlled blower-based system. Supports CPAP, S, S/T, T, PC, PC-SIMV, AC, CV, SIMV modes. Features Bi-Flex and AVAPS therapy. Connectivity includes serial interface and remote data access. Powered by electrical source. Designed for invasive/non-invasive use. Complies with ASTM F1100-90, F1246-91, ISO 10651-2/6, and IEC 60601-1-2 standards.
Indications for Use
Indicated for pediatric through adult patients weighing at least 5 kg (11 lbs.) requiring continuous or intermittent mechanical ventilatory support. Suitable for invasive or non-invasive use in home, institutional, or portable settings. Not for use as a transport ventilator.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
30 November 2008
Official Contact
Date of Submission
Zita A. Yurko Director, Regulatory Affairs Respironics, Inc. 1740 Golden Mile Highway Monroeville, PA 15146 Zita.yurko@respironics.com
724-387-4120 t 724-882-4120 c 724-387-7490 f
21 CFR 868.5895
Classification Reference
Product Code
NOU - Continuous Ventilator, Home Use
CBK - Continuous Ventilator, Facility Use
Common/Usual Name Ventilator, continuous, non-life supporting
New Device
Proprietary Name Respironics Trilogy 100 Ventilator Ventilatory Support System
Respironics BiPAP Synchrony with AVAPS (K070328)
Respironics PLV Continuum P2000 (K022750)
Pulmonetic LTV-1000 (K984056/K051767)
Newport Medical HT-50 (K9912133)
Reason for submission
Predicate Device(s)
000017
{1}------------------------------------------------
# Substantia! Equivalence
The Trilogy 100 Ventilator has the following similarities to the selected cleared predicate devices:
- D Same intended use.
- Same operating principle.
- Same technology. 0
- Same manufacturing process. ם
The modes of ventilation on Trilogy 100 are similar to those on other currently marketed continuous ventilator. Trilogy 100 has similar performance characteristics to the predicate devices, as listed below, as such; the Trilogy 100 ventilator does not raise any new questions of safety or effectiveness,
The following features are either unchanged or similar to the Respironics BiPAP Synchrony with AVAPS (K070328):
- Environment of Use o
- o Physical Characteristics of the Design
- Physical Characteristics of the Manufacturing Process 0
- Physical Characteristics of the Energy Delivered 0
- Physical Characteristics of the Materials o
- Physical Characteristics of the Anatomical Sites O
- Physical Characteristics of the Energy Source ಂ
- Rise Time O
- Ramp 0
- Triggering o
- Pressure Regulation method o
- Alarm/Power Control Indicators o
- Patient Alarms (for Pressure Support modalities) 0
- System Error Alarms (for Pressure Support Modalities) o
{2}------------------------------------------------
- Tab 5 510(k) summary
- Required Alarms: FDA Reviewers Guidance for Ventilators and ASTM F1100-90 (for MNS class ಿ ventilators)
- Device Settings (Tidal Volume) 0
- Inspiratory Time Setting Method 0
- Rise Time Setting Method ం
- Degree of protection against electric shock 0
- Degree of protection patient applied part o
- Pressure Support Modes of Operation (CPAP, S, S/T, T, PC) o
- Therapy features (Bi-Flex and AVAPS) o
The following features are either unchanged or similar to the Respironics BiPAP PLV Continuum (K034032):
- Patient Population 0
- Environment of Use o
- Product Code 0
- Physical Characteristics of the Design o
- Physical Characteristics of the Manufacturing Process ర
- Physical Characteristics of the Energy Delivered O
- Physical Characteristics of the Materials 0
- o Physical Characteristics of the Anatomical Sites
- Physical Characteristics of the Modem o
- Physical Characteristics of the Energy Source O
- Rise Time 0
- Ramp O
- Triggering 0
- O Pressure Regulation method
{3}------------------------------------------------
- Alarm/Power Control Indicators ರ
- Patient Alarms for CBK devices ಂ
- System Error Alarms for CBK devices o
- Required Alarms: FDA Reviewers Guidance for Ventilators and ASTM F1100-90 (for CBK class 0 ventilators)
- L
- Data Storage o
- Rise Time Setting Method o
- Serial Interface O
- Degree of protection against electric shock 0
- Degree of protection patient applied part o
- Modes of Operation (Positive Pressure Ventilation, Assist Control, Volume Controlled, Pressure o Controlled and Pressure Support)
The following features are either unchanged or similar to the Pulmonetic LTV-1000 (K984056/K051767):
- Patient Population o
- Environment of Use 0
- Product Code o
- Physical Characteristics of the Design o
- Physical Characteristics of the Manufacturing Process 0
- Physical Characteristics of the Energy Delivered o
- Physical Characteristics of the Materials o
- Physical Characteristics of the Anatomical Sites o
- Physical Characteristics of the Modem 0
- Rise Time o
- Ramp 0
{4}------------------------------------------------
- Tab 5 510(k) summary
- Triggering O
- Pressure Regulation method 0
- Alarm/Power Control Indicators ਼
- Patient Alarms o
- System Error Alarms ు
- Required Alarms: FDA Reviewers Guidance for Ventilators and ASTM F1100-90 (for CBK class 0 ventilators)
- Humidifier o
- Humidifier Interface O
- Remote Data Access ಂ
- Data Storage 0
- User Interface 0
- Real-Time Pressure Display 0
- Device Settings display (Pressure) O
- Device Settings display (Rate) o
- Device Settings (Tidal Volume) 0
- Inspiratory Time Setting Method 0
- Rise Time Setting Method O
- Serial Interface 0
- Degree of protection against electric shock o
- Degree of protection patient applied part o
- Modes of Operation (Positive Pressure Ventilation, Assist/Control, SIMV, CPAP) o
The following features are either unchanged or similar to the Newport Medical HT-50 (K992133):
{5}------------------------------------------------
### Abbreviated 510(k)
- Patient Population O
- Environment of Use o
- Product Code 0
- Physical Characteristics of the Design O
- Physical Characteristics of the Manufacturing Process o
- Physical Characteristics of the Energy Delivered o
- Physical Characteristics of the Materials o
- Physical Characteristics of the Anatomical Sites 0
- Physical Characteristics of the Modem o
- 0 Physical Characteristics of the Energy Source
- CPAP Pressure Range O
- 0 IPAP Pressure Range
- EPAP Pressure Range o
- Inspiratory Time Range 0
- Rise Time ం
- Ramp ಂ
- Triggering 0
- Pressure Regulation method O
- Alarm/Power Control Indicators o
- Patient Alarms o
- System Error Alarms o
- Required Alarms: FDA Reviewers Guidance for Ventilators and ASTM F1100-90 (for CBK class o ventilators)
- Remote Data Access 0
- Data Storage 0
{6}------------------------------------------------
- Real-Time Pressure Display o
- Device Settings display (Pressure) o
- Device Settings display (Rate) 0
- Device Settings (Tidal Volume) O
- Inspiratory Time Setting Method O
- Rise Time Setting Method 0
- Serial Interface 0
- Oxygen Safety Valve ಂ
- Degree of protection against electric shock o
- o Degree of protection patient applied part
- Modes of Operation (Positive Pressure ventilation and assist/control, SIMV and CPAP) o
Bench testing has confirmed that the Trilogy 100 Ventilator performs equivalently to the cited device predicates. Performance testing was conducted per the applicable sections of ASTM F1100-90, F1246-91, ISO 10651-2 and ISO 10651-6. EMC testing was performed per IEC 60601-1-2. Electrical, mechanical and environmental testing was performed in accordance with the FDA Draft Reviewers Guidance for Pre-market Notification Submission (1993). Software validation testing was performed per FDA's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices (2005). The results of all testing demonstrate that all design and system requirements for the Trilogy 100 have been met.
Respironics has followed the FDA's Guidance for Industry and FDA Staff document "pre-market assessment of pediatric medical devices" and applied the principle of FDA's Least Burdensome Approach to demonstrate the Substantial Equivalence of the Trilogy 100 Ventilator to its predicate devices. We conclude that the existing and cleared predicate device indications for use can be safety and effectively applied to the Trilogy 100 ventilator.
## Intended Use
The Respironics Trilogy100 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.).
{7}------------------------------------------------
The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.
# Device Description
The Respironics Trilogy 100 Ventilatory Support System is a microprocessor controlled blower based pressure support, pressure control or volume controlled ventilator intended for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in the institutional, home and transport settings and is applicable for adults and pediatric patients weighing at least 5 kg (11 lbs) who require the following types of ventilatory support.
CPAP – Continuous Positive Airway Pressure
- S Spontaneous Ventilation
- S/T Spontaneous and Timed Ventilation
- T Timed Ventilation
- PC Pressure Control Ventilation
- PC-SIMV Pressure Controlled Synchronized Intermittent Mandatory Ventilation
- AC Assist Control Ventilation
- CV Control Ventilation
- SIMV Synchronized Intermittent Mandatory Ventilation
A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.
The Trilogy 100 Ventilator is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.
The Trilogy 100 Ventilator can deliver invasive (via ET tube) or non-invasively (via a mask)
(End of Tab.)
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 3 2009
Ms. Zita A. Yurko Respironics, Incorporated Sleep and Home Respiratory Group 1740 Golden Mile Highway Monroeville, Pennsylvania 15146
Re: K083526
> Trade/Device Name: Trilogy 100 Ventilator Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: February 27, 2009 Received: March 2, 2009
Dear Ms. Yurko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
## Page 2 - Ms. Yurko
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours
\$\xi\$
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{10}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Trilogy 100 Ventilator Device Name: __
The Respironics Trilogy100 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.).
The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.
Prescription Use Over-The-Counter Use X 1 Teschiption Ose ____________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sine Page
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page) 0 0 0 1 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.