PORT-A-CATH POWER P.A.C. IMPLANTABLE VENOUS ACCESS SYSTEMS
K093414 · Smiths Medical Asd, Inc. · LJT · Mar 1, 2010 · General Hospital
Device Facts
Record ID
K093414
Device Name
PORT-A-CATH POWER P.A.C. IMPLANTABLE VENOUS ACCESS SYSTEMS
Applicant
Smiths Medical Asd, Inc.
Product Code
LJT · General Hospital
Decision Date
Mar 1, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling. When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable Huber needle, the PORT-A-CATH® POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media.
Device Story
Implantable venous access system; consists of portal with self-sealing septa and catheter connector; placed in chest or arm. Facilitates repeated venous access for parenteral delivery of medications, fluids, nutritional solutions, and blood sampling. When paired with power-indicated Huber needle, enables power injection of contrast media. Used in clinical settings by healthcare professionals. System provides long-term vascular access, reducing need for repeated venipuncture; benefits patient through reliable, repeated delivery of therapy.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility established via material similarity to predicates. MR safety testing performed per ASTM F2503-05 at 3 Tesla, confirming MR Conditional status.
Technological Characteristics
Implantable venous access port with self-sealing septa and catheter connector. Materials biocompatible per predicate similarity. MR Conditional per ASTM F2503-05. Sterile, non-pyrogenic, single-patient use.
Indications for Use
Indicated for patients requiring repeated vascular access for injection/infusion therapy or venous blood sampling. When used with power-injectable Huber needles, indicated for power injection of contrast media at a maximum rate of 5 ml/sec.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
PORT-A-CATH® POWER P.A.C. Implantable Venous Access Systems (K070116)
PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems (K072657)
Submission Summary (Full Text)
{0}------------------------------------------------
# smiths medical
bringing technology to life
# 510(k) Number: Ko 934/ 4
Page 1 of 3
Date Prepared October 29, 2009
## Submitter Information
MAR - 1 2010
| Submitter's Name: | Smiths Medical ASD, Inc. |
|-------------------|------------------------------------------|
| Address: | 1265 Grey Fox Road<br>St. Paul, MN 55112 |
Establishment Registration: 2183502
| Contact Person: | Rachelle Parsons, RAC<br>Sr. Regulatory Affairs Associate | |
|-----------------|-----------------------------------------------------------|--|
| | Phone: (651) 628-7018<br>Fax: (651) 628-7457 | |
## Device Information
| Trade Name: | PORT-A-CATH® POWER P.A.C. Implantable Venous<br>Access Systems |
|----------------------|----------------------------------------------------------------------|
| Common Name: | Subcutaneously Implanted Intravascular Infusion Port and<br>Catheter |
| Classification Name: | Port and Catheter, Implanted, Subcutaneous, Intravascular |
| Product Code: | LJT |
| Regulation: | 21 CFR §880.5965 |
| | |
| Trade Name: | PORT-A-CATH® II POWER P.A.C. Implantable Venous<br>Access Systems |
| Common Name: | Subcutaneously Implanted Intravascular Infusion Port and<br>Catheter |
| Classification Name: | Port and Catheter, Implanted, Subcutaneous, Intravascular |
| Product Code: | LJT |
| Regulation: | 21 CFR §880.5965 |
Submission Page 19
:" ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{1}------------------------------------------------
## 510(k) Summary
#### Page 2 of 3
#### Predicate Device(s)
The predicate devices are the above currently marketed devices. The reference 510(k) numbers for these devices are as follows:
| Device | 510(k) |
|----------------------------------------------------------------|----------------|
| PORT-A-CATH® POWER P.A.C. Implantable Venous Access Systems | K070116 |
| PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems | and<br>K072657 |
#### Device Descriptions
## PORT-A-CATH®, PORT-A-CATH® II POWER P.A.C. line of Access Ports
The PORT-A-CATH® POWER P.A.C and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems are totally implantable venous access systems designed to permit repeated access to the venous system for the parenteral delivery of medications, fluids, nutritional solutions, and for the sampling of venous blood. When used with a power indicated Huber needle, the PORT-A-CATH® POWER P.A.C. and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems are indicated for power injection of contrast media.
The PORT-A-CATH® POWER P.A.C. and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems can be placed in the chest or arm, are supplied sterile, are non-pyrogenic and intended for single patient use only. The PORT-A-CATH® POWER P.A.C and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems consists of a portal with one or two self-sealing septa, and catheter connector that are assembled during implantation. The systems are offered in either a kit or a tray configuration.
#### Intended Use/Indications for Use
The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable Huber needle, the PORT-A-CATH® POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media.
{2}------------------------------------------------
## 510(k) Summary
#### Page 3 of 3
#### Summary of Non-Clinical Testing
The non-clinical testing included assessment of the physical properties of the PORT-A-CATH® POWER P.A.C. Implantable Venous Access System and its ability to achieve its intended use. Bench testing of the device confirmed the suitability of the device for its intended use.
Biocompatibility assessment of the device was performed. The purpose of the biocompatibility assessment was to ensure that biocompatibility had been established for the device. The device is biocompatible based on the similarity of the materials of construction to the predicate devices commercially marketed by Smiths Medical ASD, Inc.
MR safety testing was performed on selected ports in accordance with established American Society for Testing and Materials (ASTM) standards. This testing was performed to establish the safety of the listed medical devices in the MR environment. The devices selected represent the entire product line of the devices listed. Testing was performed in the presence of a 3 Tesla MR imaging device.
The devices listed were determined to be MR Conditional according to the terminology specified in the ASTM Designation F2503-05 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment.
#### Summary of Clinical Testing
Human clinical studies were deemed not necessary to evaluate the safety or effectiveness of the PORT-A-CATH® POWER P.A.C. Implantable Venous Access Systems.
#### Statement of Equivalence
The PORT-A-CATH® POWER P.A.C. Implantable Venous Access System is substantially equivalent to the currently marketed Power PAC Implantable Venous Access System based on a comparison of the indications for use and the technological characteristics of the device.
#### Conclusion
The PORT-A-CATH® POWER P.A.C. Implantable Venous Access System is substantially equivalent to the currently marketed Power PAC Implantable Venous Access System based on the technological characteristics of the device. Bench tests confirmed the suitability of the device for its intended use.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Rachelle Parsons Senior, Regulatory Affairs Specialist Smiths Medical ASD, Incorporated 1265 Grev Fox Road Saint Paul, Minnesota 55112
# MAR - 1 2010
Re: K093414
Trade/Device Name: PORT-A-CACH® II POWER P.A.C. Implantable Venous Access Systems, PORT-A-CATH® POWER P.A.C. Implantable Venous Access Systems, P.A.S. Port® T2 POWDER PAC Implantable Venous Access Systems Regulation Number: 21CFR 880.5965
Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT
Dated: February 17, 2010 Received: February 18, 2010
Dear Ms. Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practive, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure -
{5}------------------------------------------------
# Indications for Use
510(k) Number:
Device Name: PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems
Indications for Use:
"The PORT-A-CATH" II POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable Huber needle, the PORT-A-CATH® II POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. For power injections of contrast media, the maximum recommended infusion rate is 5 ml/sec."
Prescription Use X (Per 21 CFR 801.109)
AND/OR
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off) Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Number: K093414
{6}------------------------------------------------
# Indications for Use
510(k) Number:
Device Name: PORT-A-CATH® POWER P.A.C. Implantable Venous Access Systems
#### Indications for Use:
"The PORT-A-CATH" POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for infusion therapy and/or venous blood sampling.
When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable Huber needle, the PORT-A-CATH® POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. For power injections of contrast media, the maximum recommended infusion rate is 5 ml/sec."
Prescription Use X (Per 21 CFR 801.109)
AND/OR
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093414
{7}------------------------------------------------
# Indications for Use
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: P.A.S. Port® T2 POWER.P.A.C. Implantable Venous Access Systems
#### Indications for Use:
"The P.A.S Port® T2 POWER P.A.C. Implantable Venous Access Systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
When used with a GRIPPER PLUS® POWER P.A.C. Safety Huber Needle or other power injectable Huber needle, the P.A.S. Port® POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. For power injections of contrast media, the maximum recommended infusion rate is 5 ml/sec."
Prescription Use >< (Per 21 CFR 801.109)
AND/OR
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clas
(Division Sign-Off) Division of Anesthesiology, General Hoggifal Infection Control, Dental Devices
510(k) Number: k093414
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.