POWER PORT-A-CATH AND POWER PORT-A-CATH II; IMPLANTABLE VENOUS ACCESS SYSTEMS AND POWER GRIPPER PLUS NEEDLE
K070116 · Smiths Medical MD, Inc. · LJT · May 23, 2007 · General Hospital
Device Facts
Record ID
K070116
Device Name
POWER PORT-A-CATH AND POWER PORT-A-CATH II; IMPLANTABLE VENOUS ACCESS SYSTEMS AND POWER GRIPPER PLUS NEEDLE
Applicant
Smiths Medical MD, Inc.
Product Code
LJT · General Hospital
Decision Date
May 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are implantable vascular access devices designed to provide long term repeated access to the vascular system. The GRIPPER PLUS® POWER P.A.C. needle is designed for the administration into or withdrawal of fluids from implanted ports.
Device Story
Implantable venous access system consisting of portal with self-sealing septa and catheter; used for repeated vascular access, medication/fluid delivery, and blood sampling. System allows power injection of contrast media when used with specific GRIPPER PLUS® POWER P.A.C. needle. Needle features passive safety mechanism to reduce needlestick injury risk. Used in clinical settings by healthcare professionals. Output is vascular access for infusion or contrast injection. Benefits include reliable long-term venous access and capability for high-pressure contrast imaging.
Clinical Evidence
No clinical data. Clinical studies were deemed unnecessary due to the similarity in materials, design, and function to existing commercially available systems. Evidence is based on bench testing and design verification.
Technological Characteristics
Totally implantable venous access system; portal with one or two self-sealing septa; single or dual lumen catheter; non-coring needle. Materials and sterilization methods are unchanged from predicate devices. Passive needle stick protection feature included in needle.
Indications for Use
Indicated for patients requiring repeated vascular access for injection/infusion therapy or venous blood sampling. When used with the GRIPPER PLUS® POWER P.A.C. needle, indicated for power injection of contrast media at max rates of 5 ml/sec (19/20G) or 3 ml/sec (22G). The needle is also indicated for fluid administration/withdrawal and protection against accidental needlestick injuries.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
{0}------------------------------------------------
070116
PORT-A-CATH® POWER P.A.C. Implantable Venous Access System, PORT-A-CATH® II POWER P.A.C. Implantable Venous Access System, and GRIPPER PLUS® POWER P.A.C. Needle
### 510(k) Summary
MAY 2 3 2007
## I. Applicant (Sponsor) Name and Address
Smiths Medical MD, Inc. 1265 Grey Fox Road St. Paul, MN 55112 Establishment Reg. No.: 2183502
### II. Contact Name and Phone
Phil Neururer
Senior Regulatory Affairs Specialist
Company Phone: (651) 628-7592
Company Fax: (651) 628-7457
### III. Device Trade/Proprietary Name
- Catheter/Port
PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems
- Needle
GRIPPER PLUS® POWER P.A.C. needle
## IV. Device Classification/Common Name/Panel
Catheter/Port
21 CFR Reference: §880.5965 21 CFR Common Name: Implanted Infusion Port Classification: Class II Product Code: LJT Panel: General Hospital
Needle
21 CFR Reference: §880.5440 21 CFR Common Name: Huber Needle Intravascular Administration Set Classification: Class II Product Code: FPA Review Panel: General Hospital and Personal Use
## V. Identification of Predicate Device
Smiths Medical MD, Inc. believes the PORT-A-CATH® POWER P.A.C. implantable venous access system, PORT-A-CATH® II POWER P.A.C. implantable venous access system, and GRIPPER PLUS® POWER P.A.C. needle are substantially equivalent to the following devices and thus may market the devices under these equivalencies.
Confidential
{1}------------------------------------------------
K070116
PORT-A-CATH® POWER P.A.C. Implantable Venous Access System, PORT-A-CATH® II POWER P.A.C. Implantable Venous Access System, and GRIPPER PLUS® POWER P.A.C. Needle
### 510(k) Summary
Catheter/Port
Device Name/510(k) number:
PORT-A-CATH® implantable venous access system (K864446, K875276, K932095, K942024. and K060036) PORT-A-CATH® II implantable venous access system (K932840, K942024, K962695, and K060036) PowerPort™ Implanted Titanium Port with 8Fr. ChronoFlex® (K060812)
Needle
Device Name/510(k) number:
GRIPPER PLUS® needle (K021999)
## VI. Device Description
PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems
The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are similar to the current PORT-A-CATH® and PORT-A-CATH® II implantable venous access systems. They both are totally implantable venous access systems designed to permit repeated access to the venous system for the parenteral delivery of medications, fluids, nutritional solutions, and for the sampling of venous blood. However, when used with a GRIPPER PLUS® POWER P.A.C. needle, the PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems can be placed in the chest or arm.
The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are supplied sterile and non-pyrogenic. All PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are designed and intended for single patient use only. A system consists of a portal with one or two self-sealing septa and a single or dual lumen catheter and is accessible by percutaneous puncture with a non-coring needle.
## GRIPPER PLUS® POWER P.A.C. Needle
The GRIPPER PLUS® POWER P.A.C. needle is similar to the current GRIPPER PLUS® needle, with the addition of an added indication. Both are supplied sterile and non-pyrogenic and are intended for the administration into or withdrawal of fluids from implanted ports. In addition, they both have a passive needle stick protection feature that is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick injuries. It does not protect against other routes of bloodborne pathogen transmission.
The difference between the two however, is when used with a PORT-A-CATH® and/or PORT-A-CATH® II POWER P.A.C. implantable venous access system; the GRIPPER PLUS® POWER P.A.C. needle is indicated for power injection of contrast media.
{2}------------------------------------------------
K070116
PORT-A-CATH® POWER P.A.C. Implantable Venous Access System, PORT-A-CATH® II POWER P.A.C. Implantable Venous Access System, and GRIPPER PLUS® POWER P.A.C. Needle
## 510(k) Summary
### VII. Intended Use of the Device
The intended use of each device has not changed from that of the predicate.
PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems
The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are implantable vascular access devices designed to provide long term repeated access to the vascular system.
GRIPPER PLUS® POWER P.A.C. Needle
The GRIPPER PLUS® POWER P.A.C. needle is designed for the administration into or withdrawal of fluids from implanted ports.
### VIII. Indications for Use
The Indications for Use of each device has changed from that of the predicate with the addition of power injection.
PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems
The PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
When used with a GRIPPER PLUS® POWER P.A.C. needle, the PORT-A-CATH® and PORT-A-CATH® II POWER P.A.C. implantable venous access systems are indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec.
GRIPPER PLUS® POWER P.A.C. Needle
The GRIPPER PLUS® POWER P.A.C. needle is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick injuries.
When used with the PORT-A-CATH® and/or PORT-A-CATH® II POWER P.A.C. implantable venous access system, the GRIPPER PLUS® POWER P.A.C. needle is indicated for power injection of contrast media into the central venous system. The maximum recommended infusion rate is 5 ml/sec. for 19 and 20 gauge needles and 3ml/sec. for 22 gauge needles.
{3}------------------------------------------------
K670116
PORT-A-CATH® POWER P.A.C. Implantable Venous Access System, PORT-A-CATH® II POWER P.A.C. Implantable Venous Access System, and GRIPPER PLUS® POWER P.A.C. Needle
## 510(k) Summary
## IX. Summary of Studies
#### A. Functional Testing
The PORT-A-CATH® POWER P.A.C. implantable venous access system, PORT-A-CATH® II POWER P.A.C. implantable venous access system, and GRIPPER PLUS® POWER P.A.C. needle met all established acceptance criteria for performance testing and design verification testing.
Packaging and Sterilization systems are unchanged from predicate devices.
#### B. Clinical Studies
Clinical studies for the PORT-A-CATH® POWER P.A.C. implantable venous access system, PORT-A-CATH® II POWER P.A.C. implantable venous access system, and GRIPPER PLUS® POWER P.A.C. needle were deemed not necessary due to their similarity in materials, design and function to current Smiths Medical MD, INC. devices and other commercially available systems.
#### C. Conclusions Drawn from the Studies
Based upon the information provided; the PORT-A-CATH® POWER P.A.C. implantable venous access system, PORT-A-CATH® II POWER P.A.C. implantable venous access system, and GRIPPER PLUS® POWER P.A.C. needle met all acceptance criteria for performance testing and design verification testing. Therefore, these products are considered acceptable for human use.
{4}------------------------------------------------
Image /page/4/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 2 3 2007
Mr. Phil Neururer Senior Regulatory Affairs Specialist Smiths Medical MD, Incorporation 1265 Grey Fox Road St. Paul, Minnesota 55112
Re: K070116
Trade/Device Name: PORT-A-CATH and PORT-A-CATH II Power P.A.C. Implantable Venous Access Systems and Gripper Plus Power P.A.C. Needle Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT, FPA Dated: April 18, 2007 Received: April 19, 2007
Dear Mr. Neururer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Neururer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chris Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# PORT-A-CATH® POWER P.A.C. Implantable Venous Access System, PORT-A-CATH® II POWER P.A.C. Implantable Venous Access System, and GRIPPER PLUS® POWER P.A.C. Needle
# Indications for Use
510(k) Number (if known): K070116
Device Name: PORT-A-CATH® POWER P.A.C. Implantable Venous Access Systems
Indications for Use:
"The PORT-A-CATH® POWER P.A.C. implantable venous access systems are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
When used with a GRIPPER PLUS® POWER P.A.C. needle, the PORT-A-CATH® POWER P.A.C. implantable venous access system is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec."
OR X Prescription Use 21 CFR 801.109)
Over-The Counter Use ------Per
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cinta D. Munoz
ംാ വ്യൂർ-Off) Ision of Anesthesiology, General Hospital, Jon Control, Dental Devices
3(k) Number: k 474116
Confidential
{7}------------------------------------------------
# PORT-A-CATH® POWER P.A.C. Implantable Venous Access System, PORT-A-CATH® II POWER P.A.C. Implantable Venous Access System, and GRIPPER PLUS® POWER P.A.C. Needle
# Indications for Use
## 510(k) Number (if known): K070116
## Device Name: PORT-A-CATH® II POWER P.A.C. Implantable Venous Access Systems
Indications for Use:
"The PORT-A-CATH® II POWER P.A.C. implantable venous access system is indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
When used with a GRIPPER PLUS® POWER P.A.C. needle, the PORT-A-CATH® II POWER P.A.C. implantable venous access system is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec."
Prescription Use ※ 21 CFR 801.109)
OR
Over-The Counter Use Per
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cutan Van
300 PLOBOR FOR tion of Anes ----------------------------------------------------------------------------------------------------------------------------------------------------------------John Control, Dencer Devices
Number. Ky7911
TAB 2 - Page 2 of 3
Confidential
{8}------------------------------------------------
# PORT-A-CATH® POWER P.A.C. Implantable Venous Access System, PORT-A-CATH® II POWER P.A.C. Implantable Venous Access System, and GRIPPER PLUS® POWER P.A.C. Needle
## Indications for Use
## 510(k) Number (if known): K070116
## Device Name: GRIPPER PLUSTM POWER P.A.C. Needle
Indications for Use:
"The GRIPPER PLUS® POWER P.A.C. needle is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick injuries.
When used with the PORT-A-CATH® POWER P.A.C. and POWER PORT-A-CATH® II POWER P.A.C. implantable venous access systems, the GRIPPER PLUS® POWER P.A.C. needle is also indicated for power injection of contrast media into the central venous system. The maximum recommended infusion rate is 5 ml/sec. for 19 and 20 gauge needles and 3ml/sec. for 22 gauge needles."
Prescription Use 21 CFR 801.109)
OR
Over-The Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cunten
ia Cign-OW) tion of Anesthasiology, General Hospital. Jon Control, Dental Devices
10(k) Number _________________________________________________________________________________________________________________________________________________________________
Confidential
TAB 2 - Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.