SOLARIS MODELS NT1 AND NT1A HANDHELD PULSE OXIMETERS WITH SENSOR ACCESSORIES
K092950 · Solaris Medical Technology, Inc. · DQA · Mar 12, 2010 · Cardiovascular
Device Facts
Record ID
K092950
Device Name
SOLARIS MODELS NT1 AND NT1A HANDHELD PULSE OXIMETERS WITH SENSOR ACCESSORIES
Applicant
Solaris Medical Technology, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Mar 12, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Solaris NT1 Handheld Pulse Oximeter with Sensor Accessories is a noninvasive spot-check, functional arterial oxygen saturation and pulse rate monitor. The device operates on battery power using SOLARIS reusable Sp02 sensors. The device is intended for pediatric and adult patients and can be used in hospital clinical areas by medical and nursing staff. The Solaris NT1A Handheld Pulse Oximeter with Sensor Accessories is a noninvasive spot-check or continuous monitoring, functional arterial oxygen saturation and pulse rate monitor. The device has data storage and data transfer functionality via a USB hub. The device operates on battery power using SOLARIS reusable Sp02 sensors. The device is intended for pediatric and adult patients and can be used in hospital clinical areas by medical and nursing staff.
Device Story
Handheld pulse oximeters (NT1/NT1A) measure SpO2 and pulse rate via reusable sensors; signals converted to digital data; processed; displayed on device. NT1A includes USB hub for data storage/transfer to external databases/third-party analysis. Used in hospital clinical areas; operated by medical/nursing staff. Provides rapid oxygenation status; reduces need for arterial blood gas analysis. NT1A supports continuous monitoring with alarms. Benefits include non-invasive physiological assessment; improved data management via USB connectivity.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed compliance with IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, and ISO14971. Performance testing for NT1A confirmed that accuracy and measurement range remained unaffected by hardware/software design changes.
Technological Characteristics
Handheld pulse oximeter; battery-powered; uses reusable SpO2 sensors. Features USB hub for data storage and transfer. Standards: IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, ISO14971. Software includes data processing and storage algorithms.
Indications for Use
Indicated for pediatric and adult patients requiring non-invasive monitoring of functional arterial oxygen saturation (SpO2) and pulse rate. NT1 is for spot-check use; NT1A is for spot-check or continuous monitoring. Used in hospital clinical areas by medical and nursing staff.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories (K073249)
Submission Summary (Full Text)
{0}------------------------------------------------
# MAR 1 2 2010
# Exhibit #1
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:_097956
#### 1. Submitter's Identification:
Susan Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd. Great Neck, N.Y. 11021
Date Summary Prepared: September 22, 2009
#### Sponsor Company Name/Address/Contact Person 2.
Solaris Medical Technology, Inc. 400 Oyster Point Blvd., Ste. 534 South San Francisco, CA 94080 Tel: (650) 588-3980 Fax: (650) 588-3988
Rachel Cheng Contact: Position: Director, Requlatory Affairs
#### 3. Manufacturing Facility Name and Address
Newtech, Inc. R1-B1, Hi-Tech Industrial Park Nanshan District, Shenzhen Guangdong 518057 P.R. China
#### 4. Name of the Device:
SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories
{1}------------------------------------------------
# 5. Common or Usual Name:
Oximeter. Pulse 74 DQA. 21 CFR Part 870.2700
### 6. Predicate Device Information:
K073249. SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories, Solaris Medical Technology, Inc.
### 7. Device Description:
The subject SOLARIS NT1 and NT1A Handheld Pulse Oximeters measure pulse rate and oxygen saturation. The signals are converted into digital data and processed; the pulse oximeter examines the data and displays the data. The subject pulse oximeters also provide operating control for the user. The pulse oximeters are intended for use in spot checking; the NT1A model is intended to both spot check and perform continuous monitoring. The pulse oximeters can be used in hospital clinical areas such as general wards to provide additional information to the medical and nursing staff about the physiological condition of the patient. The subject devices are intended to be used under supervision by clinical personnel. The intended location of use is clinics.
The subject pulse oximeters provide a rapid indication of a patient's level of oxygenation which reflects the effective ventilation. The NT1A Pulse Oximeter with alarm allows continuous and instantaneous monitoring of SpO2 and both the NT1 and NT1A Pulse Oximeters reduce the need for arterial puncture and bloodgas analysis.
#### 8. Intended Use:
The Solaris NT1 Handheld Pulse Oximeter with Sensor Accessories is a noninvasive spot-check, functional arterial oxygen saturation and pulse rate monitor. The device operates on battery power using SOLARIS reusable Sp02 sensors. The device is intended for pediatric and adult patients and can be used in hospital clinical areas by medical and nursing staff.
The Solaris NT1A Handheld Pulse Oximeter with Sensor Accessories is a noninvasive spot-check or continuous monitoring, functional arterial oxygen saturation and pulse rate monitor. The device has data storage and data transfer functionality via a USB hub. The device operates on battery power using SOLARIS reusable Sp02 sensors. The device is intended for pediatric and adult patients and can be used in hospital clinical areas by medical and nursing staff.
{2}------------------------------------------------
# ഗ് Comparison to Predicate Devices:
The modified SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories have the same "Indications for Use Statement" and are similar in design to our original 510(k) cleared device predicates.
Based on new market requirements for NT1 and NT1A Handheld Pulse Oximeters, we are adding USB hub, test data storage and output and external storage functions, making operation and data processing more convenient and allowing implementation of external database processing and third-party analysis.
Modifications made to the original device design (NT1 and NT1A) do not affect any change in structure and appearance. Design improvements on the hardware circuit and software has been made by adding USB hub, data storage and data output functions.
### 10. Testing
Laboratory testing was conducted to validate and verify that the SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories met all design specifications and were substantially equivalent to the predicate device. SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories have also been tested to assure compliance to the requirements of various published standards, including IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, and ISO14971. Performance testing for the NT1A device was conducted to show that the accuracy and measurement range of the device is not affected by the device re-design. Performance testing documentation was included in the submission under "NT1A Sp02 Oximeter Testing Reports".
# 11. Conclusions:
The conclusions drawn from testing of the SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories demonstrates that the device is as safe, as effective, and performs as well as the legally marketed predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wave-like lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Solaris Medical Technology, Incorporated C/O Ms. Susan D. Goldstein-Falk Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
MAR 1 2 2010
Re: K092950
Trade/Device Name: SOLARIS NT1 Handheld Pulse Oximeter with Sensor Accessories, SOLARIS NT1A Handheld Pulse Oximeter with Sensor Accessories Regulation Number: 21CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA Dated: March 4, 2010 Received: March 5, 2010
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Ms. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Exhibit C
# Indications for Use
Page _ 1 of of ______________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known):
Device Name: SOLARIS NT1 Handheld Pulse Oximeter with Sensor Accessories
Indications For Use:
The Solaris NT1 Handheld Pulse Oximeter with Sensor Accessories is a non-invasive spot-check, functional arterial oxygen saturation and pulse rate monitor. The device operates on battery power using SOLARIS reusable Sp02 sensors. The device is intended for pediatric and adult patients and can be used in hospital clinical areas by medical and nursing staff.
Prescription Use × (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
L Schult
(Division Sign-Off) (Division Sign-Off) ————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————— ofection Control, Dental Devices
510(k) Number: K092750
{6}------------------------------------------------
Exhibit C
Indications for Use
Page __ 2 of of 2
510(k) Number (if known):
SOLARIS NT1A Handheld Pulse Oximeter with Device Name: Sensor Accessories
Indications For Use:
The Solaris NT1A Handheld Pulse Oximeter with Sensor Accessories is a non-invasive spot-check or continuous monitoring, functional arterial oxygen saturation and pulse rate monitor. The device has data storage and data transfer functionality via a USB hub. The device operates on battery power using SOLARIS reusable Sp02 sensors. The device is intended for pediatric and adult patients and can be used in hospital clinical areas by medical and nursing staff.
Prescription Use X (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
L- Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital .. Ifection Control, Dental Devices
510(k) Number: K092950
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.