SOLARIS NT1 AND NT1A HANDHELD PULSE OXIMETERS WITH SENSOR ACCESSORIES
K073249 · Solaris Medical Technology, Inc. · DQA · Jul 25, 2008 · Cardiovascular
Device Facts
Record ID
K073249
Device Name
SOLARIS NT1 AND NT1A HANDHELD PULSE OXIMETERS WITH SENSOR ACCESSORIES
Applicant
Solaris Medical Technology, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Jul 25, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SOLARIS NT1 and NT1A Handheld Pulse Oximeters with sensor accessories are non-invasive, spot-check, oxygen saturation and pulse rate monitors. They operate on battery power using SOLARIS reusable Sp02 sensors for pediatric and adult patients.
Device Story
Handheld pulse oximeters (NT1, NT1A) measure SpO2 and pulse rate via reusable sensors; inputs are physiological signals from patient extremities. Device processes signals through a physiological processing board; calculates SpO2 and pulse rate; displays results on LED screen. Used in clinical settings (e.g., hospital wards) by medical/nursing staff. NT1 performs spot-checks; NT1A performs spot-checks and continuous monitoring with alarms. Output provides rapid oxygenation status; reduces need for arterial blood gas analysis; assists clinical decision-making regarding patient physiological condition.
Clinical Evidence
Bench testing only. Laboratory testing verified design specifications and compliance with IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, and ISO14971 standards. No clinical trial data presented.
Technological Characteristics
Handheld form factor; battery-powered. Composed of main control board (LED display, UART, DC/DC) and physiological signal processing board (SpO2 sensor control, IR/RED LED driver). Connectivity via UART interface. Standards: IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, ISO14971.
Indications for Use
Indicated for pediatric and adult patients requiring non-invasive, spot-check monitoring of oxygen saturation (SpO2) and pulse rate. The NT1A model is also indicated for continuous monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Exhibit #1 4073249
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 5l0(k) number is: _______________________________________________________________________________________________________________________________________________
#### Submitter's Identification: 1.
Alan P. Schwartz mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, N.Y. 11021
Date Summary Prepared: July 17, 2008
### Sponsor Company Name/Address/Contact Person 2.
Solaris Medical Technology, Inc. 400 Oyster Point Blvd., Ste. 534 South San Francisco, CA 94080 Tel: (415) 221-4350 Fax: (415) 221-4360
Rachel Cheng Contact: Director, Regulatory Affairs Position:
#### Manufacturing Facility Name and Address 3.
Newtech, Inc. R1-B1, Hi-Tech Industrial Park Nanshan District, Shenzhen Guanadong 518057 P.R. China
#### Name of the Device: 4.
SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories
{1}------------------------------------------------
#### 5. Common or Usual Name:
Oximeter. Pulse 74 DQA. 21 CFR Part 870.2700
#### 6. Predicate Device Information:
K# 001688, Datex-Ohmeda TuffSat 3000 Pulse Oximeter, Datex-Ohmeda Inc.
#### 7. Device Description:
## Purpose and Function of Device:
The subject SOLARIS NT1 and NT1A Handheld Pulse Oximeters measure pulse rate and oxygen saturation. The signals are converted into digital data and processed; the pulse oximeter examines the data and displays the data. The subject pulse oximeters also provide operating control for the user. The pulse oximeters are intended for use in spot checking; the NT1A model is intended to both spot check and perform continuous monitoring. The pulse oximeters can be used in hospital clinical areas such as general wards to provide additional information to the medical and nursing staff about the physiological condition of the patient. The subject devices are intended to be used under supervision by clinical personnel. The intended location of use is clinics.
The subject pulse oximeters provide a rapid indication of a patient's level of oxygenation which reflects the effective ventilation. The NT1A Pulse Oximeter with alarm allows continuous and instantaneous monitoring of SpO2 and both the NT1 and NT1A Pulse Oximeters reduce the need for arterial puncture and bloodgas analysis.
The subject pulse oximeters are composed of two boards, a main control board and a physiological singal processing board. The main control board is composed of a LED display and its control and driver, data memory, audio circuit, UART communication interface, and DC/DC circuits. The physiological processing board is composed of a SpO2 sensor control and IR/RED LEDs driver, data collection, signal processing algorithm execution, UART communication interface and DC/DC circuits.
The oximeter is housed in a single package. The main components of the assembly are listed as following
- Main Board .
- SpO2 Module .
{2}------------------------------------------------
- Key Pad .
- Batterv ●
#### 8. Intended Use:
The SOLARIS NT1 and NT1A Handheld Pulse Oximeters with sensor accessories are non-invasive, spot-check, oxygen saturation and pulse rate monitors. They operate on battery power using SOLARIS reusable Sp02 sensors for pediatric and adult patients.
### Comparison to Predicate Devices: 9.
The basic intended use of the subject devices and Datex-Obmeda TuffSat 3000 Pulse Oximeter are the same:
- I. Both the NT1 Pulse Oximeter and Datex-Ohmeda TuffSat 3000 Pulse Oximeter are non-invasive, spot-checking monitors. The difference between the NT1 and Datex-Ohmeda devices and the NT1A device is that in addition to being noninvasive spot-checking monitors, the NT1A device performs continuous monitoring.
- 2. The patient parameters for both the subject devices and Datex-Ohmeda TuffSat 3000 Pulse Oximeter are the same: oxygen saturation (SpO2), pulse rate (PR) and pulse strength.
- 3. The target populations range for both the subject devices and the Datex-Ohmeda Tuffsat 3000 Pulse Oximeter is the same: for pediatrics and adults.
Both the NT1 and NT1A Pulse Oximeters are substantially equivalent to Datex-Ohmeda TuffSat 3000 Pulse Oximeter. The pulse oximeters provide a means for interfacing with a patient and collecting parameter specific physiological signals. Then the signals are converted into digital data and processed, and the SpO2 and pulse rate values are calculated and displayed on LED screen.
#### 10. Testing
Laboratory testing was conducted to validate and verify that the SOLARIS NT1
{3}------------------------------------------------
and NT1A Handheld Pulse Oximeters with Sensor Accessories met all design specifications and were substantially equivalent to the predicate device, Datex-Ohmeda TuffSat 3000 Pulse Oximeter. SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories have also been tested to assure compliance to the requirements of various published standards, including IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, and ISO14971.
#### 11. Conclusions:
The conclusions drawn from clinical and laboratory testing of the SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories demonstrates that the device is as safe, as effective, and performs as well as the legally marketed predicate device, the Datex-Ohmeda TufSet 3000 Pulse Oximeter, K001688.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 25 2008
Solaries Medical Technology, Incorporated C/O Mr. Alan P. Schwartz mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K073249
Trade/Device Name: SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: July 7, 2008 Received: July 9, 2008
Dear Mr. Schwartz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Schwartz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
TN Samuels, Rnd, MD for //
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Exhibit B
Indications for Use
1 Page
510(k) Number (if known):
SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Device Name: Sensor Accessories
Indications For Use:
Solaris NT1 and NT1A Handheld Pulse Oximeters with sensor accessories are non-invasive spot-check, oxygen saturation and pulse rate monitors. They operate on battery power using SOLARIS reusable Sp02 sensors for pediatric and adult patients.
Mr. Vml
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K073249 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Prescription Use × (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.