DIGITAL ULTRASOUND IMAGING SYSTEM, CONVEX ARRAY TRANSDUCER, LINEAR ARRAY TRANSDUCER, CONVEX ARRAY TRANSDUCER, PHASED ARR
K092907 · Shantou Institute of Ultrasonic Instuments · IYN · Mar 5, 2010 · Radiology
Device Facts
Record ID
K092907
Device Name
DIGITAL ULTRASOUND IMAGING SYSTEM, CONVEX ARRAY TRANSDUCER, LINEAR ARRAY TRANSDUCER, CONVEX ARRAY TRANSDUCER, PHASED ARR
Applicant
Shantou Institute of Ultrasonic Instuments
Product Code
IYN · Radiology
Decision Date
Mar 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended for ultrasonic pulsed echo imaging and measurement for abdominal, pediatric, small organs, musculo-skeletal, cardiac and peripheral vascular applications.
Device Story
Digital ultrasound imaging system; supports linear, convex, phased array, and 3D transducers. Inputs: ultrasonic echoes from patient anatomy. Processing: system performs signal processing to generate 2D (B-mode), M-mode, PWD, Color Doppler, and 3D images. Features include cine review, image zoom, measurements, calculations, storage, and printing. Used in clinical settings by healthcare professionals for diagnostic imaging. Output: visual ultrasound images and quantitative measurements displayed on-screen for clinician review. Clinical utility: assists in diagnosis and assessment of abdominal, pediatric, small organ, musculo-skeletal, cardiac, and peripheral vascular conditions.
Clinical Evidence
Bench testing only. The device was tested per FDA guidance 'Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers' (2008). Acoustic output measured per NEMA UD 2:2004. Conformance to safety standards IEC 60601-1, ISO 10993-5, and ISO 10993-10 verified.
Technological Characteristics
Digital ultrasound imaging system; supports linear, convex, phased array, and 3D transducers. Operating modes: B, M, PWD, Color Doppler, 3D. Safety standards: IEC 60601-1, ISO 10993-5, ISO 10993-10. Acoustic output per NEMA UD 2:2004.
Indications for Use
Indicated for diagnostic ultrasonic imaging in abdominal, pediatric, small organ, musculo-skeletal, cardiac, and peripheral vascular applications using B, M, PWD, Color Doppler, and 3D imaging modes. Contraindications not specified.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
# 510(k) Summary
MAR - 5 2010
This summary of 510(k) safety and effectiveness information is provided in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).
The assigned 510(k) number is: K092907
1. 510(k) Owner:
Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI) 77 Jinsha Road, Shantou, Guangdong 515041, China Tel: 86-754-882501 50 Fax: 86-754-88251499
#### Contact Person:
Flower Cai
Shantou Institute of Ultrasonic Instruments Co., Ltd. 77 Jinsha Road, Shantou, Guangdong 515041, China
Date Prepared: February 4, 2010
#### 2. Device/Trade Name:
CTS-8800 Digital Ultrasound Imaging System
#### Classification Name:
Regulatory Class: II
| Ultrasonic Pulsed Doppler Imaging System | 90-IYN (per 21 CFR 892.1550) |
|------------------------------------------|------------------------------|
| Ultrasonic Pulsed Echo Imaging System | 90-IYO (per 21 CFR 892.1560) |
| Diagnostic Ultrasound Transducer | 90-ITX (per 21 CFR 892.1570) |
{1}------------------------------------------------
#### 3. Predicate Device:
The subject device is substantially equivalent to the device currently having FDA 510(k) clearance Ultrasonix Ergosonix 500 Ultrasound Scanner (K042326) with respect to intended use, principles of operation and technological characteristics.
#### 4. Device Description:
The SIUI CTS-8800 is a Digital Ultrasound Imaging System capable of the following operating modes: 2D (B mode), M, Doppler (PWD mode), Color (CFM mode) and 3D. The system is designed for use in linear, convex, phased array and 3D scanning modes and supports linear, convex, phased array and 3D transducers. The system has cine review, image zoom, measurements and calculations, image storage and review, printing and recording capabilities.
#### 5. Intended Use:
The device is intended for ultrasonic pulsed echo imaging and measurement for abdominal, pediatric, small organs, musculo-skeletal, cardiac and peripheral vascular applications.
#### 6. Safety Considerations:
The CTS-8800 Digital Ultrasound Imaging System has been tested per the FDA Guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" issued in September 2008. The acoustic output is measured and calculated per NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment: 2004. The device conforms to applicable medical device safety standards, such as IEC 60601-1, ISO10993-5 and ISO 10993-10.
#### 7. Conclusion:
The conclusions drawn from testing of the CTS-8800 Digital Ultrasound Imaging System demonstrates that the device is as safe and effective as the legally marketed predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
# APR - 1 2010
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI) % Mr. Bob Leiker ORS Representative Quality and Regulatory Services, Inc. 7263 Cronin Circle DUBLIN CA 94568
Re: K092907
Trade/Device Name: CTS-8800 Digital Ultrasound Imaging System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: February 4, 2010 Received: February 16, 2010
Dear Mr. Leiker:
This letter corrects our substantially equivalent letter of March 5, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the CTS-8800 Digital Ultrasound Imaging System, as described in your premarket notification:
Transducer Model Number
Convex Array C3L60G Linear Array L7L38G Convex Array 4DL40G Phased Array P3F14G
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours
Donald JAB
Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
(
SIUI CTS-8800 Digital Ultrasound Imaging System
### Indications for Use Statement
Device Name:
CTS-8800 Digital Ultrasound Imaging System with
Convex Array Transducer C3L60G
Linear Array Transducer L7L38G
Convex Array Transducer 4DL40G
Phased Array Transducer P3F14G
Indications for Use:
Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculo-skeletal, cardi ;
peripheral vascular applications in B, M, PWD, Color Doppler and 3D imaging modes.
Prescription Use (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IP NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K092907
: 让
{5}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## 3.1 System Indications for Usc Form System: CTS-8800
| Clinical Application | Mode of Operation | | | | | | | | |
|-----------------------------|------------------------------------|---|---|-----|-----|------------------|-----------------------|--|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | | O: er*<br>(S ecify) |
| Ophthalmic | Ophthalmic | | | | | | | | N |
| Fetal<br>Imaging<br>& Other | Fetal | N | N | | | | | | |
| | Abdominal | N | N | N | | N | | | N |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | N | N | N | | N | | | |
| | Small Organ (Specify) | N | N | N | | N | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeleral<br>(Conventional) | N | N | N | | N | | | |
| | Musculo-skeletal<br>(Superficial) | N | N | N | | N | | | |
| | Intravascular | | | | | | | | |
| | Other (Specify) | N | N | N | | N | | | N |
| Cardiac | Cardiac Adult | N | N | N | | N | | | |
| | Cardiac Pediatric | N | N | N | | N | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | | | |
| | Other (Specify) | | | | | | | | |
N = new indication: P = previously cleared by FDA: E = added under this appendix
* Other modes of oneration include: 3-D Imaging:
· Other modes of Oncranon a lines include: Prosuste, Kidney, Uterus, Ovary Small organs include: Thyroid, Testes, Breast
Prescription Usc (Per 2) CFR 801.109)
9)
Division of Radiolo Offica of In
510K K692907
{6}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
# 3.2 Transducer Indications for Use Form Transducer: Convex Array C3L60G
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Ot er*<br>(S cify) |
| Ophthalmic | Ophthalmic | | | | | | | - |
| Fetal<br>Imaging<br>& Other | Fetal | N | N | | | | | - |
| | Abdominal | N | N | N | . | N | | - |
| | Intra-operative (Specify) | | | | | | | - |
| | Intra-operative (Neuro) | | | | | | | - |
| | Laparoscopic | | | | | | | - |
| | Pediatric | | | | | | | - |
| | Small Organ (Specify) | | | | | | | - |
| | Neonatal Cephalic | | | | | | | - |
| | Adult Cephalic | | | | | | | - |
| | Trans-rectal | | | | | | | - |
| | Trans-vaginal | | | | | | | - |
| | Trans-urethral | | | | | | | - |
| | Trans-esoph. (non-Card.) | | | | | | | - |
| | Musculo-skeletal<br>(Conventional) | | | | | | | - |
| | Musculo-skeletal<br>(Superficial) | | | | | | | - |
| | Intravascular | | | | | | | - |
| | Other (Specify) | N | N | N | | N | | - |
| Cardiac | Cardiac Adult | | | | | | | - |
| | Cardiac Pediatric | | | | | | | - |
| | Intravascular (Cardiac) | | | | | | | - |
| | Trans-esoph. (Cardiac) | | | | | | | - |
| | Intra-cardiac | | | | | | | - |
| | Other (Specify) | | | | | | | - |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | - |
| | Other (Specify) | | | | | | | - |
Other (Specity)
N = new indication; P = previously cleared by FDA: E = addch under this annendix Additional Comments: Other uses include: Prostate, Kidney, Utcrus, Qvary
Prescription Use (Per 21 CFR 801.109)
DAPO
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
S10K. K092907
Indications For Use
{7}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
# 3.3 Transducer Indications for Use Form Transduccr: Lincar Array L7L38G
| Clinical Application | | Mode of Operation | | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | N | N | N | | N | | | |
| | Small Organ (Specify) | N | N | N | | N | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-csoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | N | N | | N | | | |
| | Musculo-skeletal<br>(Superficial) | N | N | N | | N | | | |
| | Intravascular | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | | | |
| | Other (Specify) | | | | | | | | |
N = new indication: P = previously cleared by FDA: E = added under this appeadix N = new indication: P = previously cleared by 12711 = 1211 = 1211 = 161
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safet
S10K 6042461
Indications l'or Use
{8}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
# 3.4 Transducer Indications for Use Form Transducer: Conyex Array 4DL40G
| Clinical Application | | Mode of Operation | | | | | | | Ot er*<br>(Specify) |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | | |
| Ophthalmic | Ophthalmic | | | | | | | | N |
| Fetal<br>Imaging<br>& Other | Fetal | | N | | | | | | |
| | Abdominal | | N | | | | | | N |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Ncuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (Specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skelctal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | N |
| | Other (Specify) | N | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (Specify) | | | | | | | | |
N = new indication: P = previously cleared by FDA: E = added under this apnendix
· Other modes include: 3-D Imaging;
· Other mouts include: 5-2 millions Prostate. Kidney, Uterus, Ovary
Prescription Use (Per 21 CFR 801.109)
Division of Rediological Office of in Vitro Diagn
610K K692907
Indications For Use
{9}------------------------------------------------
#### ടുപ്പു പോട്-8800 Digital Ultrasound Imaging System
## Diagnostic Ultrasound Indications for Use Form
# 3.5 Transducer Indications for Use Form Transducer: Phased Array P3F14G
| Clinical Application | | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|-------|-----------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Otl * | (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | | |
| | Small Organ (Specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-csoph. (non-Card.) | | | | | | | | |
| | Musculo-skolctal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| | Cardiac Adult | N | N | N | | N | | | |
| Cardiac | Cardiac Pediatric | N | N | N | | N | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | | |
| Vessel | Other (Specify) | | | | | | | | |
Outer (Specify)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Sefaty
610K. K092907
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.