The Ergosonix and Modulo Ultrasound Scanners are intended for use in obstetrics/gynecology, general radiology, cardiac examinations by a qualified physician, to aid in the diagnosis and evaluation of soft tissues, by generating 2 dimensional images, time motion images and biometric studies. The specific intended uses of this system include: abdominal, small parts, peripheral vascular, musculo-skeletal (conventional), musculo-skeletal (superficial), cephalic, small organ (breast, thyroid, testicle), trans-vaginal, trans-rectal, pediatric and fetal imaging, cardiac (adult) and cardiac (pediatric), transcranial, transesophageal.
Device Story
Mobile, software-controlled diagnostic ultrasound system; generates 2D, M-mode, Pulsed/CW Doppler, Color Flow, and Amplitude Doppler images. Inputs: ultrasound signals from linear, convex, microconvex, or phased array transducers; ECG signals via 3-lead assembly. Processing: piezoelectric transducer converts reflected sound waves to electrical signals; programmable beam-forming architecture uses reconfigurable building blocks for imaging modes. Outputs: real-time images, biometric measurements, audio Doppler, cine review, and DICOM storage. Used in clinics/hospitals by physicians to aid diagnosis/evaluation of soft tissues; supports biopsy, labeling, and image storage. Benefits: provides non-invasive anatomical and physiological assessment, aiding clinical decision-making through real-time visualization and quantitative analysis.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-2-37, and AIUM acoustic output standards. Acoustic output limits (ISPTA 720mW/cm2, MI 1.9) are consistent with Track 3 predicate devices. No clinical data required for this clearance.
Technological Characteristics
Piezoelectric transducer technology; 2-15MHz frequency range. Modes: 2D, M-mode, PW/CW Doppler, Color Flow, Power Doppler. Programmable beam-forming architecture using reconfigurable hardware blocks. Connectivity: DICOM module. Safety: IEC 60601-1, IEC 60601-2-37, AIUM standards. Biocompatible patient contact materials.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in obstetrics, gynecology, radiology, and cardiology. Patient population includes adults and pediatrics. Applications: abdominal, small parts, peripheral vascular, musculoskeletal, cephalic, breast, thyroid, testicle, transvaginal, transrectal, fetal, cardiac, transcranial, and transesophageal.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
# SEP 1 6 2004 510 (k) Summary for the Ultrasonix Ergosonix 500 Ultrasound
This summary of safety and effectiveness is provided as part of this Premarket Notification in rnio Gammery the Safe Medical Devices Act of 1990 revisions to 21 CFR, Part 807.92, Content and format of a 510(k) summary.
### 1.0 Submitter Information
### 1.1 Submitter
Ultrasonix Medical Corporation 310-3480 Gilmore Way Burnaby, British Columbia Canada V5G 4Y1 (t) 604-437-9500 (f) 604-437-9502
### 1.2 Contact
lulia Nuca, Quality Assurance (t) 604-437-9500 x 112 (f) 604-437-9502 (e) iulia@ultrasonix.com
### 1.3 Date Prepared
August, 2004
### 2.0 Device Name
### 2.1 Common Name
Ultrasound Imaging System
### 2.2 Proprietary Name
Ergosonix Ultrasound Scanner Modulo Ultrasound Scanner
### 2.3 Classification Name
| | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
{1}------------------------------------------------
### 2.4 Classification
Class Ila
#### 2.5 Predicate Device:
Ultrasonix Ergosonix 500 Ultrasound Scanner (K020630) ATL HDI 5000 System (K002003) Acuson Sequoia (K973767)
### 2.6 Reason for submission:
# Clearance request for new transducers, new applications and a new mode
| Transducers: | Applications: |
|--------------|----------------|
| PA4-2 | cardiac |
| PA3-2 | interventional |
| PA7-4 | intraoperative |
| L9-4 | transcranial |
| L12-5 | transesophagea |
| L12-5W | |
| C5-1 60 | |
| C5-1 40 | |
| ER7 | |
| EC9-5 | |
| MC7 | |
| C7-4 | |
| L15-8 | |
| T7-4 | |
| 3DEC9-5 | |
| 3DC5-1 | |
| CC5-1 | |
| IOT7-4 | |
| IOJ7-4 | |
lmaging mode:
CW Doppler
#### Name change request
"Ultrasonix Ergosonix 500 Ultrasound Scanner", changed to "Ergosonix Ultrasound Scanner"
### New product clearance
"MODULO Ultrasound Scanner"
{2}------------------------------------------------
### 2.7 Device description
The Ergosonix Ultrasound Scanner is a highly mobile, software-controlled, diagnostic ultrasound system capable of the following operating modes: 2D B-mode, M, Pulsed and CW Doppler, Color Flow (including amplitude Doppler). The system can generate real-time compound images and harmonic images.
The Modulo Ultrasound Scanner is identical to the Ergosonix Ultrasound Scanner, except that it has a different, smaller, enclosure. The Modulo Ultrasound Scanner has the same software and hardware as the Ergosonix Ultrasound Scanner, except for logo displays.
These systems have an electrocardiography (ECG) display feature and support a 3-lead ECG cable assembly. The systems provide measurement capabilities for anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The system has a PA and CW audio output feature and cine review, image zoom, labeling, biopsy, measurements and calculations, image storage and review, printing, and recording capabilities. The systems include a Digital Imaging and Communications (DICOM) module which enables storage
The system is designed for use in linear, convex and phased array scanning modes, and supports linear, convex, microconvex and phased array probes.
| Frequency Range | 2-15MHz |
|------------------|-------------------|
| Transducer types | Linear array |
| | Curved array |
| | Intracavity array |
| | Phased array |
{3}------------------------------------------------
The Egosonix and Modulo Ultrasound Scanners are designed to comply with the following standards:
| EN 60601-1 | European Norm, Medical Electrical Equipment |
|----------------|--------------------------------------------------------------------------------------|
| UL 2601-1 | Underwriters Laboratories Standards, Medical<br>Electrical Equipment |
| C22-2 No 601-1 | Canadian Standards Association, Medical<br>Electrical Equipment |
| EM 60601-1-1-2 | European Norm, Collateral Standard,<br>Electromagnetic Compatibility |
| IEC 60601-2-37 | Particular requirements for the safety of<br>ultrasonic medical diagnostic equipment |
| AIUM | Acoustic Output Labeling Standard for Diagnostic<br>Ultrasound Equipment - Jan 1998 |
| AIUM | Standard for Real-Time Display of Thermal and<br>Mechanical Acoustic Output Indices |
### 3.0 Summary of Intended Uses
The Ergosonix and Modulo Ultrasound Scanners are intended for use in obstetrics/gynecology, general radiology, cardiac examinations by a qualified physician, to aid in the diagnosis and evaluation of soft tissues, by generating 2 dimensional images, time motion images and biometric studies. The specific intended uses of this system include: abdominal, small parts, peripheral vascular, musculo-skeletal (conventional), musculo-skeletal (superficial), cephalic, small organ (breast, thyroid, testicle), trans-vaginal, trans-rectal, pediatric and fetal imaging, cardiac (adult) and cardiac (pediatric), transcranial, transesophageal.
### 4.0 Comparison to Predicate Device
The Ergosonix and Modulo Ultrasound Scanners are substantially equivalent to the predicate devices with respect to intended use/indications for use, principles of operation and technological characteristics.
### 5.0 Technological characteristics
The technological characteristics are substantially similar to that of the predicates. The device operates identically to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sounds waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as 2D or M-mode images. Doppler shift caused by blood flow is displayed as Color Flow, or as spectrum analysis. The modes of this device (2D, PW Doppler, Color Flow Mapping Doppler, Power Doppler) are the same as the predicate devices identified in item 2.5. Transducer patient contact materials are biocompatible.
The beam forming architecture is very similar to that of the predicate devices. The receiving and processing hardware is similar but innovative in that it is a programmable system made of 2 building blocks, which can be reconfigured to operate the system in any imaging mode.
{4}------------------------------------------------
The parameters used to adjust image quality are the same as that seen in the predicates. This includes the use of TGC gain sliders, depth control, base control and angling, among others.
#### 6.0 Safety considerations
As track 3 ultrasound devices, the Ergosonix and Modulo Ultrasound Scanners are designed to comply with the "Standard For Real Time Display Of Thermal And Mechanical Acoustic Output Indices On Diagnostic Ultrasound Equipment (1992)" published by the National Electrical Manufacturers Association as UD-3.
With respect to limits on acoustic outputs, the Ergosonix and Modulo Ultrasound Scanners comply with the guideline limits set in the September 30, 1997 revision of 510(k) Diagnostic Ultrasound Guidance.
With regard to general safety, the Ergosonix and Modulo Ultrasound Scanners are designed to comply with IEC 601-1 (1988) Medical Electrical Equipment, Part 1: General Requirements for Safety, and IEC 60601-2-37: Particular Requirements For The Safety Of Ultrasonic Medical Diagnostic And Monitoring Equipment.
The devices' acoustic output limits are:
| ISPTA (d) | 720mW/cm2 |
|-----------------------|----------------------|
| TIS/TIB/TIC | 0.1 - 4.0 (Range) |
| Mechanical Index (MI) | 1.9 (Maximum) |
| ISPPA (d) | 0 - 700W/cm2 (Range) |
The limits are the same as predicate Track 3 devices.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms extending upwards, representing health, services, and people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### SEP 1 6 2004
Ms. Iulia Nuca Quality Assurance Ultrasonix Medical Corp. 310-3480 Gilmore Way Burnaby, BC, V5G 4YI . . CANADA
Re: K042326
Trade Name: Ergosonix / Modulo Ultrasound Scanners Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic trasducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: August 25, 2004 Received: August 27, 2004
Dear Ms. Nuca:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bour determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to I edelul I ood, Drog, and sions of the Act. The general controls provisions of the Act include the general connents provision, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Ergosonix / Modulo Ultrasound Scanners, as described in your premarket notification:
Transducer Model Number
C5-1 40 convex 1/5MHz 40mm radius transducer C5-1 60 convex 1/5MHz 60mm radius transducer
{6}------------------------------------------------
L12-5 linear 5/12MHz 38mm transducer L12-5W linear 5/12MHz 50mm transducer L9-4 linear 4/9MHz 38mm transducer PA3-2 phased array 2/3MHz transducer PA4-2 phased array 2/4MHz transducer PA3-2 phased array 2/4MHz transducer . PA7-4 phased array 4/7MHz transducer ER7 biplane endocavity 5/9MHz transducer EC9-5microconvex endocavity 5/9MHz 10mm radius transducer MC7 microconvex 5/9MHz 10mm radius transducer L15-8 linear 8/15MHz 29mm transducer C7-4 convex 4/7MHz 40mm radius transducer CC5-1 microconvex 1/5MHz 15mm radius transducer T7-4 TEE phased 4/7MHz transducer 3DC5-1 motorized convex 1/5MHz 40mm radius transducer 3DEC9-5 motorized microconvex 5/9MHz 10mm radius transducer IOT7-4 convex 4/7MHz 40mm radius intraoperational transducer IOJ7-4 linear 4/7MHz 38mm radius intraoperational transducer
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may or below of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny 1 vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Or N Fat 80 77 tucemes (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
{7}------------------------------------------------
Page 3 - Ms. Nuca
Enclosure(s)
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed noutication. The I 27 Imaling of succion for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please rryou destre specific at (301) 594-4591. Additionally, for questions on the contact the Office of Compilanour device, please contact the Office of Compliance at (301) 594promonon and advertising or Joulation entitled, "Misbranding by reference to premarket 4059. Also, please now the regaral information on your responsibilities under the houthcation (21 Orient and overy) ion of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at t (301) 594-1212.
Sincerely yours,
David R. Segerson
Nancy C. Brogdon
Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{8}------------------------------------------------
Prescription Use (Per 221 CFR 801.109)
# Ergosonix Ultrasound Scanner Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | | Mode of Operation | | | | | | | | |
|-----------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Abdominal | | P | P | P | N | P | P | P | P (*1) | P (*2) |
| Intraoperative (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative Neurological | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Pediatric | | P | P | P | N | P | P | P | P (*1) | P (*2) |
| Small Organ (specify) | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Neonatal Cephalic | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Adult Cephalic | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transrectal | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Transvaginal | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Leparoscopic | | | | | | | | | P (*1) | P (*2) |
| MSK Conventional | | P | P | P | | P | P | P | P (*1) | P (*2) |
| MSK Superficial | | P | P | P | | P | P | P | P (*1) | P (*2) |
| Other (specify) (*3) | | N | N | N | | N | N | N | N (*1) | N (*2) |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
### Additional Comments:
- 'r B/M, B/PWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFMCWD, B/CFMCWD, B/DPD/CWD
B/WD, B/C/WD, B/CFM/PWD, B/AD/PWD, B/CFMCND, B/CF/C/P/D/
- 1 BIM, BIPWD, B/OVD, B/OHD, B/O/ M/X WB, Directional Power Doppler (DPD)
- *3 Transcranial Doppler
FLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF WEEDEDI
Concurrence of CDRH, Office of Device Evaluation (ODE)
{9}------------------------------------------------
# MODULO Ultrasound Scanner
# Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | P | P (*1) | P (*2) | |
| Abdominal | | P | P | P | | N | P | P | P | P (*1) | P (*2) |
| Intraoperative (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Intraoperative Neurological | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Pediatric | | P | P | P | | N | P | P | P | P (*1) | P (*2) |
| Small Organ (specify) | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| Neonatal Cephalic | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| Adult Cephalic | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| Cardiac | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Transesophageal | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Transrectal | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| Transvaginal | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| Leparoscopic | | | | | | | | | | | |
| MSK Conventional | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| MSK Superficial | | P | P | P | | | P | P | P | P (*1) | P (*2) |
| Other (specify) (*3) | | N | N | N | | | N | N | N | N (*1) | N (*2) |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
- 1 B/M, B/PWD, B/CWD, B/CFM/PWD, B/D/PWD, B/DPD/PWD, B/CFM/CWD, B/D/CWD, B/DPD/CWD
- *2 Freehand 3D imaging, live 3D imaging, Directional Power Doppler (DPD)
- *3 Transcranial Doppler
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (Division of General
510(k) Nu
{10}------------------------------------------------
# C5-1 40 convex 1/5MHz 40mm radius transducer Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
*1 BIM, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
*2 Freehand 3D imaging, Directional Power Doppler (DPD)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ØDE)
ion (ODE)
David A. Seymour
vision Sian-Off Division of Reproduc and Radiological De
{11}------------------------------------------------
| C5-1 60 convex 1/5MHz 60mm radius transducer |
|----------------------------------------------|
|----------------------------------------------|
Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | | Mode of Operation | | | | | | | | | |
|-----------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Leparoscopic | | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) | |
| Other (specify) | | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
*1 BM, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
*2 Freehand 3D imaging, Directional Power Doppler (DPD)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evalue
Evaluation (ODE)
Daniel A. Symm
and 1 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 /
{12}------------------------------------------------
# L12-5 linear 5/12MHz 38mm transducer
# Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
"1 BM, B/PWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
*2 Freehand 3D imaging, Directional Power Doppler (DPD)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluati
uation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K042326
{13}------------------------------------------------
# L12-5W linear 5/12MHz 50mm transducer
# Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
*1 B/M, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
*2 Freehand 3D imaging, Directional Power Doppler (DPD)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evalua
David h. Seymm
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K042326
510(k) Number
{14}------------------------------------------------
## L9-4 linear 4/9MHz 38mm transducer
### Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
B/M, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD ﮩ
- *2 Freehand 3D imaging, Directional Power Doppler (DPD)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
ion(ODE)
David R. Syverson
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K042326
510(k) Number
{15}------------------------------------------------
# PA3-2 phased array 2/3MHz transducer Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Other (specify) (*3) | | N | N | N | N | N | N | N | N (*1) | N (*2) |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
^1 B/M, B/PWD, B/CWD, B/CFM/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
- *2 Freehand 3D imaging, Directional Power Doppler (DPD)
- *3 Transcranial Doppler
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
uation (ODE)
David A. Sypm
Division Sign Off
(Division Sign-O Division of Reproducti and Radiological Dev E 1 011-1 11,
{16}------------------------------------------------
# PA4-2 phased array 2/4MHz transducer
# Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Other (specify) (*3) | | N | N | N | N | N | N | N | N (*1) | N (*2) |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
- B/M, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/DPD/CWD, B/DPD/CWD 、
- *2 Freehand 3D imaging, Directional Power Doppler (DPD)
- *3 Transcranial Doppler
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
valuation (ODE)
David R. Segerson
(Division Sign-Off)
Division of Reproductive, and Radiological Devic
{17}------------------------------------------------
# PA3-2 phased array 2/4MHz transducer
## Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Other (specify) (*3) | | N | N | N | | N | N | N | N (*1) | N (*2) |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
- B/M, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD で
- *2 Freehand 3D imaging, Directional Power Doppler (DPD)
- *3 Transcranial Doppler
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (
David R. Lysem
(Division Sign-Off) Division of Reproductive, and Radiological Devices
{18}------------------------------------------------
Prescription Use (Per 221 CFR 801.109)
# PA7-4 phased array 4/7MHz transducer Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Other (specify) (*3) | | N | N | N | | N | N | N | N (*1) | N (*2) |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
- 1 B/M, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
- *2 Freehand 3D imaging, Directional Power Doppler (DPD)
- *3 Transcranial Doppler
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation {
tion (QDE)
(Division Sign-Off)
(Division Sign-Off, Division of Reproductive, and Radiological Devices
{19}------------------------------------------------
# ER7 biplane endocavity 5/9MHz transducer Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transvaginal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Leparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Other (specify) (*3) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
#### Additional Comments:
*1 B/M, B/PWD, B/C/ND, B/CFM/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
*2 Freehand 3D imaging, Directional Power Doppler (DPD)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Lygass
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices SAHAR
inal
{20}------------------------------------------------
# EC9-5 microconvex endocavity 5/9MHz 10mm radius transducer Diagnostic Ultrasound Indications for Use Form
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transvaginal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transurethral | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.