K092734 · Espansione Marketing S.P.A · GEX · Sep 20, 2010 · General, Plastic Surgery
Device Facts
Record ID
K092734
Device Name
E-LIGHT LINE
Applicant
Espansione Marketing S.P.A
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 20, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The E-Light Line devices are intended for dermatological use by physicians and healthcare professionals as the following: IPL Technology is intended for: · Treatment of Acne (from 390 to 1200nm filter) · Treatment of vascular and benign pigmented lesions, cutaneous lesions including warts, scars, striae , and facial and leg veins (from 550 to 1200nm filter). · Removal of unwanted hair in all skin types (from 570- to 1200nm filter) LED Technology is intended: · Yellow LED 594 nm for treatment of periorbital wrinkles and rhytides · Blue LED 428nm to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris · Red LED 632nm for treatment of superficial, benign vascular and pigmented lesions VACUUM Technology is intended to: · Improve Lymphatic circulation in the treated area
Device Story
E-Light Line is a multi-technology dermatological system comprising a main control unit and interchangeable handpieces for IPL, LED, and vacuum therapy. The main unit manages power, capacitor charging, and electronic control. IPL handpieces emit light (390-1200nm) for hair removal, acne, and vascular/pigmented lesions. LED handpieces provide specific wavelengths (Yellow 594nm, Blue 428nm, Red 632nm) for wrinkles, acne, and skin lesions. Vacuum handpieces provide aspiration for lymphatic circulation. Operated by physicians and healthcare professionals in clinical settings. Only one output source functions at a time. System aids clinical decision-making by providing targeted energy-based therapies for various dermatological conditions, potentially improving patient skin appearance and lymphatic health.
Clinical Evidence
No clinical study was submitted. Evidence is based on bench testing, including electrical safety (EN 60601-1:2004) and electromagnetic compatibility (IEC 60601-1-2:2004) testing, and comparison of technical specifications to predicate devices.
Technological Characteristics
System includes main control unit with microcontroller and capacitor charge system. Handpieces: IPL (390-1200nm), LED (Yellow 594nm, Blue 428nm, Red 632nm), and Vacuum. Standards: EN 60601-1 (safety), IEC 60601-1-2 (EMC), EN 980 (symbols), EN ISO 14971 (risk management).
Indications for Use
Indicated for dermatological use by physicians and healthcare professionals for acne treatment, vascular/pigmented lesion removal, wart/scar/striae treatment, hair removal, periorbital wrinkle reduction, and lymphatic circulation improvement.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a black and white logo. The logo appears to be a stylized design that incorporates geometric shapes and letterforms. On the left side, there is a circle with a white triangle inside. The right side of the logo features a stylized letter 'e'.
K 092734
**ESPANSIONE MARKETING S.P.A.**
Blocco 27 Via Orefici 152
Centergross 40050 Funo BOLOGNA - ITALY
Tel. +39 051 8901611 Fax +39 051 863400
SEP 2 0 2010
# 510(k) Summary for the E-Light Line
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
# 1. General Information
Submitter:
Espansione Marketing Spa is located at: Blocco 27 - Via Orefici 152 Centergross 40050 Funo - Bologna ITALY
Contact Person:
Guido Bonapace ISEMED srl Via Borgo Santa Cristina 12 40026 Imola (BO) Italy Mob.phone: +39-335-5378686 Telephone: +39-051-6527315 Fax: +39-051-6284344 Email: gbonapace@isemed.eu
Summary Preparation Date:
August 28, 2009
- 2. Names
Device Family Name: Device models: Common Name: Classification Name:
Product Code: Regulation number:
E-light Line Trinity Plus, EPI-C Plus, Evoluzione Plus IPL, LED and Vacuum equipment II, Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology GEX /ISA 878.4810
### 3. Predicate Devices
The E-light line is substantially equivalent to the following legally marketed device:
1
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo that appears to be a stylized representation of the letters 'ae' or a symbol resembling an infinity sign with a pointed element on the left side. The logo is black and white, with the shapes defined by solid black areas against a white background. The design is simple and modern, with clean lines and a bold appearance.
**ESPANSIONE MARKETING S.P.A.**
Blocco 27 Via Orefici 152
Centergross 40050 Funo BOLOGNA - ITALY
Tel. +39 051 8901611 Fax +39 051 863400
info@espansione.it
www.espansione.it
| Applicant | Device name | 510(k) Number |
|-----------------------------|------------------------------------------|---------------|
| Espansione Marketing S.p.A. | EPI-C Plus | K070494 |
| Photonova of Sweden AB | Photonova Family of pulsed light systems | K073477 |
| Light BioScience LLC | GentleWaves Led Photomodulation | K031425 |
| Skincare Technology Inc. | RevLigth | K042630 |
| Photo Therapeutics Limited | Ominilux Revive | K030426 |
| Photo Therapeutics Limited | Ominilux Blue | K030883 |
| Quantel Derma Gmbh | LEDA | K090762 |
| PhotoActif LLC | IllumiMed | K060792 |
| LPG One Inc. | CELLU M6 Keymodule | K053225 |
# 4. Device Description
The E-light Line is a family of devices with three types of output emitted from different handpieces. The outputs of the E-light Line are Intense Pulsed Light (IPL), LED and Vacuum. These outputs may be issued from different handpieces. The output handpieces are:
- IPL handpiece
- Led handpiece
- 이 Led Body Band
- 0 Led Masks
r
- Vacuum handpiece
Only one output source works at a time. All the E-Light line devices include a Main Unit that controls and manages the IPL, LED or Vacuum handpiece. The Main Unit contains all circuitry to control the device as the capacitor charge system, the microcontroller and all the electronic parts. The E-light line devices are:
- . Trinity Plus
- l Evoluzione Plus
- l EPI-C Plus
The E-Light Trinity Plus supports all the technologies (IPL, LED and Vacuum) and therefore all the handpieces. The E-Light Evoluzione Plus and EPI-C plus support the IPL and LED Technologies and therefore all the handpiece with the exception of the Vacuum handpieces.
#### 5. Indications for Use
The E-Light Line devices are intended for dermatological use by physicians and healthcare professionals as the following:
IPL Technology is intended for:
- · Treatment of Acne (from 390 to 1200nm filter)
- · Treatment of vascular and benign pigmented lesions, cutaneous lesions including warts, scars, striae , and facial and leg veins (from 550 to 1200nm filter).
2
{2}------------------------------------------------
Blocco 27 Via Orefici 152
Centergross 40050 Funo BOLOGNA - ITALY
Tel. +39 051 8901611 Fax +39 051 863400
info@espansione. it
· Removal of unwanted hair in all skin types (from 570- to 1200nm filter) LED Technology is intended:
- · Yellow LED 594 nm for treatment of periorbital wrinkles and rhytides
- · Blue LED 428nm to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris
- · Red LED 632nm for treatment of superficial, benign vascular and pigmented lesions
VACUUM Technology is intended to:
i
- · Improve Lymphatic circulation in the treated area
## 6. Substantial equivalence description
The E-light line is a family of device with difference energy emissions: intense pulsed light (IPL), light emitting diode (LED) and aspiration massager (Vacuum).
The IPL technology is the same of the Espansione Marketing Epi-c Plus (K070494) both for intended use and for technological characteristics, in fact the E-light line family was developed with the same hardware of the Espansione Marketing Epi-c Plus (K070494). The IPL technology is also similar to the Photonova - Photonova Family of pulsed light systems (K073477), in fact the intended use and the emission parameter are very similar to E-light line.
The LED technology is equivalent to the predicate devices K031425 - Light Bioscience LLC -Gentlewaves Led Photomodulation, K042630 - Skincare Technology Inc. - RevLight, K030426 -Photo Therapeutics Limited - Ominilux Revive, K030883 - Photo Therapeutics Limited -Ominilux Blue, K090762 - Quantel Derma Gmbh - LEDA, K060792 - PhotoActif LLC -IllumiMed. The outputs of the Led technology of the E-light line are yellow, red and blue.
The Espansione Marketing E-light line is equivalent to the K031425 - Light Bioscience LLC -Gentlewaves Led Photomodulation and to the K090762 - Quantel Derma Gmbh - LEDA because the E-light line has yellow led with the same wavelength, similar output parameters (energy density, treatment time, dose range), have the same type of application.
The Espansione Marketing E-light line is equivalent to the K030426 - Photo Therapeutics Limited - Ominilux Revive, K042630 - Skincare Technology Inc. - RevLight and to the K090762 -Quantel Derma Gmbh -- LEDA because the E-light line has red led with the same wavelength, similar output parameters (energy density, treatment time, dose range), have the same type of application.
The Espansione Marketing E-light line is equivalent to the K030883 - Photo Therapeutics Limited - Ominilux Blue, the K042630 - Skincare Technology Inc. - RevLight, and to the K060792 -PhotoActif LLC - IllumiMed because the E-light line has red led with the same wavelength, similar output parameters (energy density, treatment time, dose range), have the same body application.
The Vacuum technology is equivalent to LPG One device (K053225). In fact the intended use of the E-light line massager is included in the LPG One device (K053225) intended use. The output parameters of E-light line are included in the LPG One device (K053225) parameters and the method of application is very similar for both devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo that appears to be a stylized representation of the letters 'de'. The 'd' is formed by a circle with a triangle cut out of it, and the 'e' is a simple, sans-serif letter. The logo is black and white, and the overall design is modern and minimalist. The logo is likely used to represent a company or organization with a name that starts with the letters 'de'.
**ESPANSIONE MARKETING S.P.A.**
Blocco 27 Via Drefici 152
Centergross 40050 Funo BOLOGNA - ITALY
Tel. +39 051 8901611 Fax +39 051 863400
info@espansione.it
www.espansione.it.
Then we believe that the Espansione Marketing E-light line is substantially equivalent to the predicate devices, because it have the same intended use, the similar technology characteristics and the similar application as the predicate devices and no significant differences exists between the predicate devices.
#### 7. Performance Data
The E-Light Line devices have been successfully passed the electric safety tests in accordance with the EN 60601-1:2004 Medical Electrical Equipment - Part 1: General Requirements for Safety and the electromagnetic compatibility tests in accordance with the IEC 60601-1-2:2004 Medical Electrical Equipment - Part 1: General Requirements for Safety; Electromagnetic Compatibility --Requirements and Tests.
The E-Light Line devices have been developed according to the following international standards too:
o EN 980:2004 Graphical symbols for use in the labeling of medical devices
o EN ISO 14971:2007 Medical devices - Application of risk management to medical devices
No clinical study was been submitted.
After an analysis of the indications and intended uses, the technology characteristics, outputs, performance and safety (both electric and electromagnetic), Espansione Marketing believes that no significant differences exist between the predicate devices listed in Section 3, above.
The E-light line 510(k) Summary includes only information that is also covered in the body of the 510(k), it does not contain any puffery or unsubstantiated labeling claims, it does not contain any raw data and contains only summary data, it does not contain any trade secret or confidential commercial information, it does not contain any patient identification information.
4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Espansione Marketing Spa % Isemed S.R.L. Mr. Guido Bonapace Vin Borgo Santa Cristina 12 40026 Imola, Bologna, Italy
SEP 2 0 2010
Re: K092734
Trade/Device Name: E-light Line Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general surgery and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ISA, OPA Dated: August 30, 2010 Received: September 01, 2010
Dear Mr. Bonapace:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Guido Bonapace
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Malkerson
Mark N. Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Espansione Marketing S.p.A. 510(k) Notification
E-light Line
SEP 2 0 2010
#### Indications for Use
510(k) Number (if known): K092734
Device Name:
ﻟﺮﻱ ﺍ
E-light Line
Indications for Use:
The E-Light Line devices are intended for dermatological use by physicians and healthcare professionals as the following:
IPL Technology is intended for:
- · Treatment of Acne (from 390 to 1200nm filter)
- · Treatment of vascular and benign pigmented lesions, cutaneous lesions including warts, scars, striae , and facial and leg veins (from 550 to 1200nm filter).
- · Removal of unwanted hair in all skin types (from 570- to 1200nm filter)
- LED Technology is intended:
- · Yellow LED 594 nm for treatment of periorbital wrinkles and rhytides
- · Blue LED 428nm to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris
- · Red LED 632nm for treatment of superficial, benign vascular and pigmented lessons VACUUM Technology is intended to:
· Improve Lymphatic circulation in the treated area
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil Rf. Ryden formten
Division Sign-Of (Division Sign-On)
Division of Surgical, Orthopedic, Division of Storative Devices
510(k) Number K092734
000007
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.