K060792 · Photoactif, LLC · ILY · Dec 1, 2006 · Physical Medicine
Device Facts
Record ID
K060792
Device Name
ILLUMIMED
Applicant
Photoactif, LLC
Product Code
ILY · Physical Medicine
Decision Date
Dec 1, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The illumiMed device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum is intended to reduce mild to moderate acne vulgaris. Use of the combination of blue, red and near infrared regions of the spectrum is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
Device Story
illumiMed is a mobile, wheel-mounted LED phototherapy system. It utilizes super-luminous LEDs at 430 nm (blue), 660 nm (red), and 940 nm (near-infrared) to deliver light energy or topical heat. The device features a control panel with push-pads for selecting treatment protocols and an adjustable arm holding up to five LED panels; accessory wands and pads are also available. Operated by clinicians, the device provides targeted dermatological treatment or therapeutic heating. Software controls the treatment protocols and allows for manufacturer-provided updates. The output is used by healthcare providers to manage acne or musculoskeletal conditions, potentially benefiting patients through non-invasive, pain-free symptom relief and skin condition management.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by performance testing and validation studies comparing the device's technological characteristics and output specifications to those of the predicate devices.
Technological Characteristics
Mobile unit on wheels; 120V AC power. LED wavelengths: 430 nm (blue), 660 nm (red), 940 nm (near-infrared). Modular design with panel arm, wands, and pads. Software-controlled treatment protocols with push-button interface. Software is field-upgradable by the manufacturer.
Indications for Use
Indicated for patients with mild to moderate acne vulgaris (blue light) and for patients requiring topical heating for temporary relief of minor muscle/joint pain, arthritis, muscle spasm, stiffness, and to increase local blood circulation (red/near-infrared light).
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
Predicate Devices
Omnilux Revive and Omnilux Plus Combination (K050216)
{0}------------------------------------------------
# DEC - 1 2006
# PREMARKET NOTIFICATION 510(k) SUMMARY As required by §807.92
# Device Name - as required by 807.92(a)(2):
| Trade Name: | illumiMed™ | | | | | | | |
|-----------------------------------|---------------------------------------------------------------------------|--|--|--|--|--|--|--|
| Common/Classification Name: | LED Phototherapy/Laser Instrument, Surgical, Powered<br>and Infrared lamp | | | | | | | |
| Classification Regulation(s): | 878.4810 and 890.5500 | | | | | | | |
| Device Class: | Class II | | | | | | | |
| Product Codes (Procode): | GEX and ILY | | | | | | | |
| Premarket Notification submitter: | | | | | | | | |
| Company Name: | PhotoActif, LLC | | | | | | | |
| Company Address: | 1959 South Power Road, Suite 103-313 | | | | | | | |
| | Mesa, Arizona 85206 | | | | | | | |
| Company Phone: | (480) 827-1212 | | | | | | | |
| Company FAX: | (480) 827-1213 | | | | | | | |
| Contact: | Jeremy Shellman, President and CEO | | | | | | | |
| Preparation Date: | March 8, 2006 (amended October 19, 2006) | | | | | | | |
# LEGALLY MARKETED PREDICATE DEVICE – as required by A. 807.92(a)(3). For details, see Exhibit 13.
The identified legally marketed GEX predicate devices identified by the submitter are:
- Omnilux Revive and Omnilux Plus Combination, K050216, from . Photo Therapeutics Limited, Cheshire, United Kingdom
- Omnilux Blue, K030883, from Photo Therapeutics Limited, Cheshire, . United Kingdom
- RevLight®, Skincare System, K042630, from Skincare Technology Inc., . Chicago, IL
- AcneLift, K041103, from Inner Act LLC, Reno, Nevada. .
One legally marketed ILY predicate device is:
- Omnilux Plus, K043317, from Photo Therapeutics Limited, Cheshire, . United Kingdom.
{1}------------------------------------------------
### B. DEVICE DESCRIPTION - as required by 807.92(a)(4)
The submitted device, illumiMed, is a treatment method using specific wavclengths of polychromatic energy produced by super-luminous LEDs (light emitting diodes) to treat a variety of skin and body conditions. It is an affordable, pain-free means to address a variety of dermatological concerns. The IR spectrum is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promotion the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum is intended to reduce mild to moderate acne vulgaris. Use of the combination of blue, red and near infrared regions of the spectrum is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
The control panel displays to the operator what treatment is being performed and provides the user with a choice of using the LED panels on the arm, LED attachments or both for any chosen treatment protocol.
The software component of the device controls an appropriate number of treatment protocols and each protocol can be enabled by a single action push-pad or button on the control panel. The software can be upgraded by the manufacturer.
### C. DEVICE CLAIMS - as required by 807.92(a)(4)
The illumiMed device is intended to emit energy to treat dermatological conditions including mild to moderate acne vulgaris. The illumiMed is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
# PRODUCT AND TECHNICAL SPECIFICATIONS - as required by D. 807.92(a)(4)
The illumiMed device is powered by grounded 120 volt electrical power with an operator controlled on/off switch. The device is housed in a unit that is mounted on four wheels allowing it to be moved to any location desired by the operator.
{2}------------------------------------------------
The control panel has push-pads or buttons that control its operation and software functionality. The software functionally can be upgraded by the manufacturer. The device has a panel arm extension that can hold between one (1) and five (5) LED panels and can be cantilevered in any direction for LED treatment of a patient in a prone, seated or standing position. The LED panels can be combinations of LEDs at 7660 nm ("red"), y 940-950 nm (near infrared), and (430 nm ("blue"). The panels can be configured as an individual panel or in combinations up to five panels of LEDs. Accessory LED pads and wands are available for additional treatment modalities with 660 nm, 940 nm and 430 nm LED diodes.
### E. INTENDED USE - as required by 807.92(a)(5)
The illumiMed device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum is intended to reduce mild to moderate acne vulgaris. Use of the combination of blue, red and near infrared regions of the spectrum is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
### F. INDICATIONS FOR USE - See Exhibit 3.
Use of 430 nm "blue" LEDs is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
Use of the 660 nm "red" LEDs and of the 940 nm "near infrared" LEDs are indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Use of the accessory 38 blue LED Wand is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
{3}------------------------------------------------
Use of the accessory 38 red/near infrared LED Wand is indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Use of the accessory 240 red/near infrared LED Pad is indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain. arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Use of the combination of 430 nm "blue" LEDs, 660 nm "red" LEDs and of the 940 nm "near infrared" LEDs is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
### G. LEVEL OF CONCERN - as requested by recent FDA guidance
The FDA guidance document "Guidance For The Content of Premarket Submissions For Software Contained In Medical Devices," May 11, 2005, provides tables that the submitter used to identify the Level of Concern for the submitted device. See Exhibit 4 - Level of Concern.
The submitter, following approved procedures, assessed the device's software component for Level of Concern and completed the submitter's form F 0014F, Level of Concern Assessment and Certification.
Since the submitted device's software is neither a Major Level of Concern or a Moderate Level of Concern, PhotoActif's completed form F 0014F documents and the submitter claims the determination that the illumiMed device's software component can only be a MINOR Level of Concern.
## H. TECHNOLOGICAL CHARACTERISTICS SUMMARY - as required by 807.92(a)(6)
The technological characteristics available in illumiMed device include:
The submitted device provides a treatment method, using specific wavelengths of polychromatic energy produced by super-luminous LEDs (light emitting diodes), to treat a variety of skin and body conditions.
The device operates on grounded 110 volt power and is intended to be moved on its wheels to any appropriate location for use. The device has storage space for optional attachments and topicals.
{4}------------------------------------------------
A control panel provides for manual and software controlled treatment protocols, specifically with the:
- 1. Use of the four hundred ninety two (492) LED's panel-mounted on the panel arm with:
Four hundred nincty two (492) 430 nm "blue" LEDs is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
- 2. Use of the five hundred seventy six (576) LED's pancl-mounted on the panel arm with:
Two hundred eighty eight (288) 660 nm "red" LEDs and two hundred eighty eight (288) 940 nm "near infrared" LEDs is indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
- 3. The use of the accessory thirty eight (38) LED Wand with:
Twenty (20) 660 nm "red" LEDs and Eighteen (18) 940 nm "near infrared" LEDs is indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
- 4. The use of the accessory thirty eight (38) LED Wand with:
Thirty eight (38) 430 nm "blue" LEDs is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
- 5. The use of the accessory two hundred forty (240) LED Pad with:
One hundred twenty (120) 660 nm "red" LEDs and one hundred twenty (120) 940 nm "near infrared" LEDs is indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
{5}------------------------------------------------
- 6. Use of the combination of 430 nm "blue" LEDs, 660 nm "red" LEDs and of the 940 nm "near infrared" LEDs:
Is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris
Currently, the device has software protocols that are activated by pushbutton controls on the Control Panel for:
- Acne
The device's Control Panel also provides for manual operation and has three (3) buttons, "A," "B," and "C," for future treatments that would be provided by the submitter as software updates to the current software version.
{6}------------------------------------------------
| W A V A 1 2 1 2 4 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2<br>1<br>F + 0 / 1<br>0 | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| | |
| | THE The Reduline seconding mans pass on personnes on<br>Mark Pu |
.60792
# ABLE OF SIMILARITIES AND DIFFERENCES BETWEEN
| | GEX/ILY | GEX | GEX | ILY | | |
|----------------------------------------------------------------------------------------------------------|--------------------------------------|---------------------------------------------------|---------------------|---------------------|-------------------------|----------------------------|
| Characteristic | illumiMed | Omnilux Revive &<br>Omnilux Plus<br>Combo K050216 | RevLight<br>K042630 | AcneLift<br>K041103 | Omnilux Blue<br>K030883 | Omnilux<br>Plus<br>K043317 |
| Electrical powered grounded 110 - 120 volts | Yes | Yes | Yes | Yes | Yes | Yes |
| LED diodes/infrared light | 430, 660, 940 nm | Revive: 633 nm<br>Plus: 830 nm | "420 - 940 nm" | 405 nm | 417 nm | 830 nm |
| Interchangeable panels/modules | Yes | Yes | Yes | No | No | No |
| Wand Attachment | Yes | No | No | No | No | No |
| Pad Attachment | Yes | No | No | No | No | No |
| Energy Output in Joules | 430nm = 46<br>660nm &<br>940nm = 61 | *Revive = 126<br>*Plus = 66 | *48 | #50 | *48 | *66 |
| Energy Output in mw | 430nm = 3.5<br>660nm &<br>940nm = 31 | *Revive = 105<br>*Plus = 55 | *4200 | #45 | *40 | *55 |
| Software component | Yes | Yes | No | No | Yes | Yes |
| Software upgradeable | Yes | Unknown | n/a | No | Unknown | Unknown |
| Action to provide temporary relief of<br>minor muscle aches and pain | Yes | Yes | Yes | No | No | Yes |
| Action to improve circulation | Yes | Yes | Yes | No | No | Yes |
| Action to provide temporary relief<br>from arthritis and muscle spasm | Yes | Yes | Yes | No | No | No |
| Action to provide temporary<br>relaxation of muscle tissue | Yes | Yes | Yes | No | No | Yes |
| Action to provide for photo-aged<br>skin rejuvenation | Yes | Yes | Yes | No | No | Yes |
| Action to treat dermatological<br>conditions and specifically for<br>moderate inflammatory acne vulgaris | Yes | No | Yes | Yes | Yes | No |
8 of 9
{7}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE SUMMARY I.
The submitted device, illumiMed, has the same indications for use as the predicate devices.
illumiMed has the same or very similar technological characteristics as the identified predicate devices. However, while the submitter believes the characteristics are sufficiently precise to assure equivalence, the submitter has carried out validation and performance testing to further document substantial equivalence. The results of this testing substantiates that illumiMed performs as well as the predicate devices.
### .J. CONCLUSIONS
The performance testing and validation studies document that illumiMed is substantially equivalent to the GEX predicates Omnilux Revive and Omnilux Plus Combination [K050216], Omnilux Blue [K030883], RevLight, Skincare System [K042630], and AcneLift [K041103]; and the ILY predicate Olmuilro Revive [K030426].
The submitter's claim the submitted device is safe and effective for its intended use as both a prescription and non-prescription device.
{8}------------------------------------------------
Public Health Service
Image /page/8/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing and tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
PhotoActif, LLC % Mr. Jeremy Shellman President and CEO 1959 South Power Road, Suite 103-313 Mesa, Arizona 85206
Re: K060792 Trade/Device Name: illumiMed™ Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY, GEX Dated: October 18, 2006 Received: October 20, 2006
Dear Mr. Shellman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
DEC = I 2006
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{9}------------------------------------------------
Page 2 - Mr. Jeremy Shellman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerel Myours.
Kol P
Mark N. Malkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): K060792
Device Name: illumiMedTM
# Indications For Use:
Use of 430 ±10 nm "blue" LEDs is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
Use of the 660 ±10 nm "red" LEDs and of the 940 ±10 nm "near infrared" LEDs are indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase blood circulation.
Use of the accessory "38 blue LED Wand", 430 ±10 nm "blue", is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate acne vulgaris.
Use of the accessory "38 red/near infrared LED Wand", 660 ± 10nm "red" LEDs and of the 940 ±10 nm "near infrared" LEDs, is indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature: for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase blood circulation.
Use of the accessory "240 red/near infrared LED Pad", 660 ± 10nm "red" LEDs and of the 940 ±10 mm "near infrared" LEDs, is indicated for delivery of heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase blood circulation.
{11}------------------------------------------------
510 (k) Number: K060792
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use_ (Per 21 CFR 801,109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
OFF
(Division Sign-Off) (Division Sign of General, Restorative, and Neurological Devices
**510(k) Number** L060792
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.