TRACHEAL INTUBATION FIBERVIDEOSCOPE, MODELS LF-Y0004 AND LF-Y0005
K082720 · Olympus Medical Systems Corporation · EOQ · Nov 5, 2008 · Ear, Nose, Throat
Device Facts
Record ID
K082720
Device Name
TRACHEAL INTUBATION FIBERVIDEOSCOPE, MODELS LF-Y0004 AND LF-Y0005
Applicant
Olympus Medical Systems Corporation
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Nov 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These instruments have been designed to be used with a Suction Pump and EndoTherapy accessories and other ancillary equipment for airway management, which includes endoscopic treatment, diagnosis, and observation to access airway anatomy, endotracheal/endobronchial intubation and management.
Device Story
The LF-Y0004 and LF-Y0005 are all-in-one, cordless, flexible fiber-videoscopes designed for airway management. Unlike predicate devices requiring external light sources, video processors, and monitors, these units integrate these functions into a single portable device. The device captures visual input from the airway via a distal optical system and displays it on an integrated monitor. Used by clinicians for airway visualization, endotracheal/endobronchial intubation, and diagnostic/therapeutic procedures. The integrated design improves portability and simplifies the setup for airway management in clinical settings. Output allows the physician to navigate the airway and confirm tube placement, facilitating safer intubation and diagnostic observation.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and comparison of technological characteristics.
Technological Characteristics
Flexible fiber-videoscopes with integrated light source, video processor, and monitor. LF-Y0004: 3.9mm distal end, 4.1mm insertion tube. LF-Y0005: 5.1mm distal end, 5.2mm insertion tube. Both feature 600mm working length, 90-degree field of view, and forward-viewing optics. Cordless, all-in-one form factor.
Indications for Use
Indicated for airway management, including diagnosis, observation, and endoscopic treatment to access airway anatomy, and for endotracheal/endobronchial intubation in patients requiring such procedures.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
6082720
# 510(k) SUMMARY
NOV - 5 2008
# TRACHEAL INTUBATION FIBERVIDEOSCOPE LF-Y0004 and LF-Y0005
### September 8, 2008
#### 1 General Information
- 트 Applicant: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047 북 Official Correspondent: Stacy Abbatiello Kluesner, RAC
Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-5405 Fax: 484-896-7128 Email:stacy.kluesner@olympus.com Establishment Registration No: 2429304
- Aizu Olympus Co., Ltd. Manufacturer: 500 Aza-Muranishi, Ooaza-lidera, Monden-cho, Aizuwakamatsu-shi, Fukushima, Japan 965-8520 Establishment Registration No: 9610595
### Device Identification 2
월
| | Device Trade Name: | TRACHEAL INTUBATION FIBERVIDEOSCOPE<br>LF-Y0004 and LF-Y0005 |
|--|-----------------------|--------------------------------------------------------------|
| | Common Name: | TRACHEAL INTUBATION FIBERVIDEOSCOPE |
| | Regulation Number: | 21 CFR 874.4680 |
| | Regulation Name: | Bronchoscope (flexible or rigid) and accessories |
| | Regulatory Class: | II |
| | Classification Panel: | Ear Nose & Throat |
| | Product Code: | EOQ - Bronchoscope (Flexible Or Rigid) |
{1}------------------------------------------------
### 3 Predicate Device Information
The following table shows the primary components of the subject devices and the devices to which we claim substantial equivalence (predicate devices).
| Table 14-1. Primary Components & Predicate Devices | | | | | | | |
|----------------------------------------------------|--|--|--|--|--|--|--|
|----------------------------------------------------|--|--|--|--|--|--|--|
| Subject Devices<br>(Part of this submission) | Predicate Devices | PD's<br>510(k) No. |
|-------------------------------------------------|---------------------------------------------------------------------------------------------|--------------------|
| TRACHEAL INTUBATION<br>FIBERVIDEOSCOPE LF-Y0004 | Olympus LF-DP Tracheal Intubation<br>Fiberscope | K981543 |
| TRACHEAL INTUBATION<br>FIBERVIDEOSCOPE LF-Y0005 | EVIS EXERA BRONCHOFIBERVIDEO<br>SCOPE OLYPUS BF TYPE XP160F<br>Olympus Bronchoscope BF-XT40 | K033225<br>K023984 |
#### Device Description ব
The TRACHEAL INTUBATION FIBERVIDEOSCOPE LF-Y0004 and LF-Y0005 are all-in-one cordless endoscopes with great portability which include light source, video processor, and video monitor functions.
The new endoscopes are basically identical to each predicate device shown in Table 14-1 in intended use, and similar in specifications, performance and materials.
### ട് Indications for Use
These instruments have been designed to be used with a Suction Pump and EndoTherapy accessories and other ancillary equipment for airway management, which includes diagnosis, and observation endoscopic treatment. to access airway anatomy, endotracheal/endobronchial intubation and management.
### б Comparison of Technological Characteristics
The TRACHEAL INTUBATION FIBERVIDEOSCOPE LF-Y0004 and LF-Y0005 are basically identical to the predicate device in intended use, and similar in specifications except for the all-in-one unit, outer diameter, and optical system. Comparison between the subject and predicate devices is shown in Table 14-2.
{2}------------------------------------------------
Talions : TRACHEAL MTUBATION FIBERVIDEOSCOPE LE-Y0004 and LF-Y000
Trelicate Devices : TRACHEAL NEWS : Present : Prove and LE-Y000
Predicate Devices : TRACHEAL MTUBERVIOR
| Specifications | Subject Device<br>LF-Y0004 | Subject Device<br>LF-Y0005 | Predicate Device<br>LF-DP | Predicate Device<br>BF-XP160F | Predicate Device<br>BF-XT40 |
|----------------------------------------|------------------------------------------------------|------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------|-----------------------------------------------------------------------------------------|
| Field of View | 90° | 90° | 90° | 90° | 120° |
| Depth of Field | 4-50mm | 3-50mm | 3-50mm | 2-50mm | 3-50mm |
| Direction of View | 0° (Forward viewing) | 0° (Forward viewing) | 0° (Forward viewing) | 0° (Forward viewing) | 0°(Forward viewing) |
| Outer Diameter<br>of Distal End | φ3.9mm | φ5.1mm | φ3.1mm | φ2.8mm | φ6.1mm |
| Outer Diameter<br>of Insertion<br>Tube | φ4.1mm | φ5.2mm | φ3.1mm | φ2.8mm | φ6.2mm |
| Angulation<br>UP/DOWN | UP : 120°<br>DOWN : 120° | UP : 180°<br>DOWN : 130° | UP : 120°<br>DOWN : 120° | UP : 180°<br>DOWN : 130° | UP : 180°<br>DOWN : 130° |
| Working Length | 600mm | 600mm | 600mm | 600mm | 550mm |
| Source of<br>examination<br>light | These are built-in<br>functions of the<br>endoscope. | These are built-in<br>functions of the<br>endoscope. | External Light Source<br>Unit | External Light Source<br>Unit | External Light<br>Source Unit |
| Video<br>Processor | | | External Video System<br>Center Unit (when<br>connected via Video<br>Converter Unit) | External Video<br>System Center Unit | External Video<br>System Center Unit<br>(when connected<br>via Video Converter<br>Unit) |
| Monitor | | | External Video Monitor<br>Unit | External Video Monitor<br>Unit | External Video<br>Monitor Unit |
3/4
{3}------------------------------------------------
# 7
.
When compared to the predicate device, the TRACHEAL INTUBATION FIBERVIDEOSCOPE LF-Y0004 and LF-Y0005 do not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device.
:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Olympus Medical System Corporation c/o Stacy Abbatiello Kluesner Olympus America, Inc 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034
NOV - 5 2008
Re: K082720
Trade/Device Name: Tracheal Intubation Fiberscope LF-Y0004 and LF-Y0005 Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: II Product Code: EOQ Dated: September 16, 2008 Received: September 23, 2008
Dear Ms. Kluesner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
### Page 2 - Stacy Abbatiello Kluesner
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 10(t) marketed predicate device results in a classification for your device and thus, perrorits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Melvina B. Esguerra, M.D.
Malvina B. Eydelman, M.D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known):
Device Name: TRACHEAL INTUBATION FIBERVIDEOSCOPE LF-Y0004, LF-Y0005 Indications For Use:
These instruments have been designed to be used with a Suction Pump and EndoTherapy accessories and other ancillary equipment for airway management, which includes endoscopic treatment, diagnosis, and observation to access airway anatomy, endotracheal/endobronchial intubation and management.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Concur
(Division Sign-Off)
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devices
510(k) Number *K082720*
Page 1 of 1
: 032327
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.