K091562 · Lambda Scientific Spa · GEX · Sep 8, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091562
Device Name
DOCTOR SMILE A-810 AND B-980 LASERS
Applicant
Lambda Scientific Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 8, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Doctor Smile A-810 Laser is intended for use in the following procedure: Soft Tissue, Laser Periodontal procedures and tooth whitening. Intra- and extra oral surgery including incision, excision, vaporization, ablation, haemostasis and coagulation of oral soft tissues, Laser soft tissue curettage, whitening. Doctor Smile Diode B-980 Laser is intended for use in the following procedure: Surgical applications requiring the ablation, vaporization, excision, incision, haemostasis, or coagulation of soft tissue in medical specialties including dentistry, dermatology, general surgery, gynecology, ophthalmology, pulmonology, otolaryngology, othopedics.
Device Story
Doctor Smile A-810 and B-980 are portable diode lasers (810 nm and 980 nm wavelengths). System consists of laser console with display/software, fiber optic delivery system, handpieces, and protective eyewear. Device transmits laser energy via contact fiber optic delivery to target soft tissue. Used by clinicians in various medical specialties (dentistry, dermatology, surgery, etc.) for ablation, vaporization, excision, incision, haemostasis, and coagulation. Output is controlled laser energy; clinician directs energy to tissue to achieve surgical effect. Benefits include precise tissue management and haemostasis during procedures.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Portable diode laser system; 810 nm (A-810) and 980 nm (B-980) wavelengths. Contact fiber optic delivery system. Components: console with display/software, fiber optic, handpieces, protective eyewear. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and 21 CFR 1040.10/1040.11.
Indications for Use
Indicated for soft tissue surgical procedures including incision, excision, vaporization, ablation, haemostasis, and coagulation. Specific applications include dentistry (periodontal procedures, biopsies, frenectomy, gingivoplasty, implant recovery, pulpotomy, whitening), dermatology, general surgery, gynecology, ophthalmology, pulmonology, otolaryngology, and orthopedics.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
SEP 0 8 2009
Image /page/0/Picture/2 description: The image shows the logo for LAMBDA Scientifica SpA. The logo includes a stylized lambda symbol on the left. To the right of the symbol is the company name, "LAMBDA Scientifica SpA", with the word "LAMBDA" in a larger font size. There is also a signature above the company name.
# 510(K) SUMMARY
| 510(k) owner and submitter | LAMBDA Scientifica S.p.A.<br>Via Retrone 39<br>36077 Altavilla Vicentina (Vicenza)<br>Italy<br>+39 0444 349 165<br>+39 0444 349 954 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person | Mr Boschi Alessandro<br>+39 0444 349 165<br>boschi@lambdascientifica.com |
| The common name of the device | Diode laser |
| Trade name / Propietary names | Doctor Smile A-810, B-980 Lasers |
| Classification name | Laser surgical instrument for use in general and plastic surgery and<br>dermatology |
| Product code | Gex |
| CFR Number | 21 CFR §878.4810 |
| Predicate Devices | Doctor Smile A-810, B-980:Lasers<br>- Arc Laser GmbH Fox Q-810, Q-980 K073322<br>- Biolase Ezlase® K082938<br>- Sirona Sirolaser K053161<br>- Kavo Gentleray 980 K072262<br>- Ceralas 980 K072779<br>- Claros Nano Dental Laser System K081652 |
| Product description | Doctor Smile A-810, B-980 Lasers are the same standard diode<br>medical lasers respectively with 810, 980 nm wavelength. |
| Intended Use | Doctor Smile A-810 Laser is intended for use in the following<br>procedure: Soft Tissue, Laser Periodontal procedures and tooth<br>whitening. Intra- and extra oral surgery including incision, excision,<br>vaporization, ablation, haemostasis and coagulation of oral soft<br>tissues, Laser soft tissue curettage, whitening.<br>Doctor Smile Diode B-980 Laser is intended for use in the following<br>procedure: Surgical applications requiring the ablation, vaporization,<br>excision, incision, haemostasis, or coagulation of soft tissue in<br>medical specialties including dentistry, dermatology, general surgery,<br>gynecology, ophthalmology, pulmonology, otolaryngology, othopedics. |
{1}------------------------------------------------
| Performance standards | The Doctor Smile A-810, B-980 Lasers comply with the following<br>standards:<br>IEC 60601-1:1998+A1:1991+A2:1995<br>IEC 60601-2-22:1995<br>IEC 60825-1:1993+A1:1997+A2:2001<br>21 CFR 1040.10 and 1040.11 |
|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device technological<br>characteristics and<br>Comparison to Predicate<br>device(s) | Doctor Smile A-810, B-980 Lasers are portable instruments which<br>transmit laser energy via a contact fiber optic delivery system and<br>each unit is compromised of the following main components:<br>-a laser system console (including display panel, software and<br>control);<br>-a fiber optic delivery system;<br>-one or more handpieces<br>-protective eye ware<br>Doctor Smile Diode A-810, B-980 Lasers have the same intended use<br>and the same or substantially equivalent technical specifications and<br>mechanism of action as compared with the named predicated devices. |
| Conclusion: | The Doctor Smile A-810, B-980 Lasers have identical or similar<br>indications for use, principles of operation, overall technical and<br>functional capabilities. Therefore is substantially equivalent to the<br>predicate devices. They have similar intended uses and comply with<br>the same safety and performance standards. |
**Review:**
Boschi Alessandro
7
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, composed of three curved lines that suggest wings and a head. The bird is oriented to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
SEP 0-8 2009
Lambda Scientifica S.p.A % Mr. Boschi Alessandro Via Retrone 39 36077 Altavilla Vicentina Italy
Re: K091562
Trade/Device Name: Doctor Smile A-810 B-980 Lasers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: September 1, 2009 Received: September 1, 2009
Dear Mr. Alessandro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Boschi Alessandro
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Melkerson Director
Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for LAMBDA Scientifica SpA. The logo consists of a lambda symbol inside of a circle on the left. To the right of the symbol is the text "LAMBDA Scientifica SpA". There is a line underneath the text.
## Indications for Use
510(k) Number (if known):
Device name: Doctor Smile A-810, B-980 Lasers
#### Doctor Smile A-810 Laser is intended for use in the following procedure:
Soft Tissue, Laser Periodontal procedures and tooth whitening Intra- and extra oral surgery including incision, excision, vaporization, ablation, haemostasis and coagulation of oral soft tissues, Laser soft tissue curettage, whitening.
Following specific indications: Excisional and incisional biopsies, Exposure of unerupted teeth, Fibroma removal, Frenectomy, Frenotomy, Gingival troughing for crown impressions, Gingivoplasty, Gingival incision and excision
Haemostasis and coagulation, Implant recovery, Incision and drainage of abscess
Leukoplakia, Operculectorny, Oral papillectornies, Pulpotomy as an adjunct to root canal therary, Reduction of gingival hypertrophy, Soft tissue crown lengthening, Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa. Vestibuloplasty, Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket. Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingiyal index, gingival bleeding index, probe depth, attachment loss and tooth mobility.). Light activation of whitening/bleaching material.
### Doctor Smile B-980 Laser is intended for use in the following procedure:
Surgical applications requiring the ablation, vaporization, excision, incision, hemostasis, or coagulation of soft tissue in medical specialties including dentistry, gastroenterology, dermatology, general surgery, gynecology, opthamology, pulmonology, otolaryngology, othopedics.
#### Dental applications:
Frenectomy, Frenotomy, Biopsy, Operculectomy, Implant Recovery, Gingivoplasty, Gingivolasty, Gingival Troughing, Crown Lengthening, Haemostasis of Donor Site, Removal of Granulation Tissue, Laser-assisted Flap Surgery, Debridement of Deseased Epithelial Lining, Incisions and Draining of Abscesses, Tissue Retraction for Impressions. Papillectory, Vestibuloplasty, Excision of Lesions, Leukoplakia, Exposure of Unerupted Partially Erupted Teeth, Removal of Hyperplastic Tissues, Treatment of Aphthous Ulcers, Sulcular Debridement, Pulpotomy, Pulpotomy as an Adjunct to Root Canal Therapy, and Light Activation of Bleaching Materials
Prescription use
and/or
Over-the-counter use
(21 CFR 801 Subpart D)
Neil R. Sigh for xxx
Division of Surgical, Orthopedic, and Restorative Devices
(21 CFR 807 Subpart C)
**Review:**
Boschi Alessandro
ensori e Sistemi di Monitoraggio / Monitoring Systems and sensors 510(k) Numbe a Retrone n. 39 - 36077 Altavilla Vicentina (VI) Italia - Tel/fax +39-0444-349165 /349954 http://www.lambdascientifica.com - E-Mail: info@lambdascientifica.com Cap. Soc. Euro 500.000,00 - C.F. e P. I.V.A. 02558810244 - Registro Imprese di Vicenza n. 199230/96
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.