K072262 · Kavo America · GEX · Oct 26, 2007 · General, Plastic Surgery
Device Facts
Record ID
K072262
Device Name
GENTLERAY 980 DIODE LASER SYSTEM
Applicant
Kavo America
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 26, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue. The GENTLEray 980 Diode Dental Laser System is intended for use in the following procedures: Frenectomy, Frenotomy, Biopsy, Operculectomy, Implant Recovery, Gingivectomy, Gingivoplasty, Gingival Troughing, Crown Lengthening, Hemostasis of Donor Site, Removal of Granulation Tissue, Laser-assisted Flap Surgery, Debridement of Deseased Epithelial Lining, Incisions and Draining of Abscesses, Tissue Retraction for Impressions, Papillectomy, Vestibuloplasty, Excision of Lesions, Leukoplakia, Exposure of Unerupted/Partially Erupted Teeth, Removal of Hyperplastic Tissues, Treatment of Aphthous Ulcers, Sulcular Debridement, Pulpotomy, Pulpotomy as an Adjunct to Root Canal Therapy, and Light Activation of Bleaching Materials.
Device Story
GENTLEray 980 is a portable diode laser system for intraoral soft tissue surgery. Device uses 980 nm wavelength laser energy delivered via contact fiber optic system to ablate, incise, excise, and coagulate tissue. System includes peristaltic pump for irrigation. Operated by dental professionals in clinical settings. Clinician controls laser parameters (power, pulse duration, frequency) via device interface; output is delivered to target tissue to achieve surgical goals like hemostasis or tissue removal. Benefits include precise tissue management and reduced bleeding during dental procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Validation included electrical safety, electromagnetic compatibility, and hardware/software functional testing to demonstrate performance equivalent to predicate devices.
Technological Characteristics
Diode laser system; 980 nm wavelength; 0.4-12W output power; pulsed or continuous operating modes; 635 nm red aiming beam; air-cooled; peristaltic pump for irrigation; steam autoclave sterilization for accessories; portable form factor (10.5" x 7" x 6"); 100-240V power requirement.
Indications for Use
Indicated for intra- and extra-oral soft tissue surgery including incision, excision, hemostasis, coagulation, and vaporization in adult and pediatric patients. Procedures include frenectomy, biopsy, gingivectomy, implant recovery, abscess drainage, and tooth whitening.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
KO 72262
Image /page/0/Picture/1 description: The image contains two logos, one on the left and one on the right. The logo on the left is a stylized design, while the logo on the right contains the words "GENDEX" on the top line and "DEXIS" on the bottom line. The text in the logos is in a simple, sans-serif font.
| Applicant: | KaVo Dental Corporation | OCT 26 2007 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Address: | 340 East Main Street<br>Lake Zurich, IL 60047<br>USA | |
| Phone Number:<br>Fax Number: | 847-550-6800<br>847-550-6825 | |
| Contact Person: | Mari Lambert | |
| Summary Prepared: | October 4, 2007 | |
| Name of Device: | KaVO GENTLEray 980 | |
| Trade Name: | GENTLEray 980 | |
| Common Name: | GENTLEray | |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology | |
| CFR Number: | 21 CFR §878.4810 | |
| Product Code: | GEX | |
| Product Description: | The GENTLEray 980 Diode Dental Laser System is a portable<br>instrument intended for ablating, incising, excising, and<br>coagulating intraoral soft tissue (including the marginal and<br>interdental gingiva) using a contact fiber optic delivery system. | |
| Intended Use: | Intra- and extra-oral surgery including incision, excision,<br>hemostasis, coagulation and vaporization of soft tissue.<br>The GENTLEray 980 Diode Dental Laser System is intended for<br>use in the following procedures: Frenectomy, Frenotomy,<br>Biopsy, Operculectomy, Implant Recovery, Gingivectomy,<br>Gingivoplasty, Gingival Troughing, Crown Lengthening,<br>Hemostasis of Donor Site, Removal of Granulation Tissue,<br>Laser-assisted Flap Surgery, Debridement of Deseased Epithelial<br>Lining, Incisions and Draining of Abscesses, Tissue Retraction<br>for Impressions, Papillectomy, Vestibuloplasty, Excision of<br>Lesions, Leukoplakia, Exposure of Unerupted/Partially Erupted<br>Teeth, Removal of Hyperplastic Tissues, Treatment of Aphthous<br>Ulcers, Sulcular Debridement, Pulpotomy, Pulpotomy as an<br>Adjunct to Root Canal Therapy, and Light Activation of | |
| Performance<br>Standards: | The GENTLEray 980 Diode Dental Laser System complies with<br>the appropriate sections of 21 CFR §1010 and 21 CFR §1040. | |
| Substantial<br>Equivalence: | The GENTLEray 980 Diode Dental Laser System has the same<br>intended use and the same or substantially equivalent technical<br>specifications and mechanism of action as compared with the<br>named predicated devices. The KaVo GENTLEray 980 Diode<br>Laser is substantially equivalent to the Ceramoptec Ceralas D15<br>(K983058, K991891) sold latterly under the BioLitec brand<br>name SmilePro™ 980, the ADT Diolase 980 D Laser System<br>(K023547) and the Elexxion Claros (K063152). Performance<br>testing to validate the safety and effectiveness of the<br>GENTLEray 980 Diode Dental Laser System includes electrical<br>safety, electromagnetic compatibility, and validation testing of<br>both hardware and software functions. The comparison of<br>specifications are as follows: | |
{1}------------------------------------------------
KO72262
| Specifications | Proposed Device | Predicate Device | Predicate Device | Predicate Device |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Wavelength | KaVo GENTLEray 980 | Biolitec Smilepro 980 | ADT Diolase 980 D | Elexxion Claros |
| | 980 nm | 980 nm | 980 nm | 810 nm |
| Output Power | 12W | 15W | 15 W | 30 W |
| Power Range | .4 - 12W | 1-15W | .2 - 15 W | 0.01W - 30 W |
| Increments | .2 - 1W | 1W | .2 - 1W | 0.01W - 1 W |
| Operating Modes | Pulsed or Continuous | Pulsed or Continuous | Pulsed or Continuous | Pulsed or Continuous |
| Pulse Duration ON | 25μs to 99.9 Sec. | 0.01 to 99.9 Sec. | 0.01 to 99.9 Sec | 25μs to 99.9 Sec. |
| Pulse Duration OFF | 25μs to 99.9 Sec. | 0.01 to 99.9 Sec. | 0.01 to 99.9 Sec | 25μs to 99.9 Sec. |
| Frequency | 20.000 Hz | 100 Hz | 100 Hz | 20.000 Hz |
| Aiming beam | 635 nm, <1mW; Red | 635 nm, 4mW; Red | 635 nm, 4mW; Red | 635 nm, 4mW; Red |
| Cooling | Air Cooled | Air Cooled | Air Cooled | Air Cooled |
| Weight | 9 lbs. (4.5kg) | 15 lbs. (9kg) | 11 lbs. (5kg) | 48 lbs. (22kg) |
| Dimensions | 10.5" x 7" x 6" | 14" x 9" x 7" | 14" x 9" x 3" | 33" x 18" x 20" |
| Power<br>Requirements | 100 - 240 V | 110/220 V | 110/220 V | 110/220 V |
| Sterilization<br>Methods | Steam Autoclave | Steam Autoclave | Steam Autoclave | Steam Autoclave |
| Irrigant Supply | Peristaltic Pump | Peristaltic Pump | Peristaltic Pump | None |
| Indications for Use | Intra- and extra-oral<br>surgery including incision,<br>excision, hemostasis,<br>coagulation and<br>vaporization of soft tissue.<br>The device is intended for<br>use in the following<br>procedures: Frenectomy,<br>Frenotomy, Biopsy,<br>Operculectomy, Implant<br>Recovery, Gingivectomy,<br>Gingivoplasty, Gingival<br>Troughing, Crown<br>Lengthening, Hemostasis<br>of Donor Site, Removal of | Intra- and extra-oral<br>surgery including incision,<br>excision, hemostasis,<br>coagulation and<br>vaporization of soft tissue.<br>The device is intended for<br>use in the following<br>procedures: Frenectomy,<br>Frenotomy, Biopsy,<br>Operculectomy, Implant<br>Recovery, Gingivectomy,<br>Gingivoplasty, Gingival<br>Troughing, Crown<br>Lengthening, Hemostasis<br>of Donor Site, Removal of | Intra- and extra-oral<br>surgery including incision,<br>excision, hemostasis,<br>coagulation and<br>vaporization of soft tissue.<br>The device is intended for<br>use in the following<br>procedures: Frenectomy,<br>Frenotomy, Biopsy,<br>Operculectomy, Implant<br>Recovery, Gingivectomy,<br>Gingivoplasty, Gingival<br>Troughing, Crown<br>Lengthening, Hemostasis<br>of Donor Site, Removal of | Dental Soft Tissue<br>Indications Including<br>Pulpal Tissues*<br>Incision, excision,<br>vaporization, ablation and<br>coagulation of oral soft<br>tissues, including:<br>Excisional and incisional<br>biopsies, Exposure of<br>unerupted teeth, Fibroma<br>removal, Frenectomy,<br>Frenotomy, Gingival<br>troughing for crown<br>impressions,<br>Gingivectomy |
| | Laser-assisted Flap<br>Surgery, Debridement of<br>Deseased Epithelial<br>Lining, Incisions and<br>Draining of Abscesses,<br>Tissue Retraction for<br>Impressions,<br>Papillectomy,<br>Vestibuloplasty, Excision<br>of Lesions, Leukoplakia,<br>Exposure of<br>Unerupted/Partially<br>Erupted Teeth, Removal<br>of Hyperplastic Tissues,<br>Treatment of Aphthous<br>Ulcers, Sulcular<br>Debridement, Pulpotomy,<br>Pulpotomy as an Adjunct<br>to Root Canal Therapy,<br>and Light Activation of<br>Bleaching Materials. | Laser-assisted Flap<br>Surgery, Debridement of<br>Deseased Epithelial<br>Lining, Incisions and<br>Draining of Abscesses,<br>Tissue Retraction for<br>Impressions,<br>Papillectomy,<br>Vestibuloplasty, Excision<br>of Lesions, Leukoplakia,<br>Exposure of<br>Unerupted/Partially<br>Erupted Teeth, Removal<br>of Hyperplastic Tissues,<br>Treatment of Aphthous<br>Ulcers, Sulcular<br>Debridement, Pulpotomy,<br>Pulpotomy as an Adjunct<br>to Root Canal Therapy,<br>and Light Activation of<br>Bleaching Materials. | Laser-assisted Flap<br>Surgery, Debridement of<br>Deseased Epithelial<br>Lining, Incisions and<br>Draining of Abscesses,<br>Tissue Retraction for<br>Impressions,<br>Papillectomy,<br>Vestibuloplasty, Excision<br>of Lesions, Leukoplakia,<br>Exposure of<br>Unerupted/Partially<br>Erupted Teeth, Removal<br>of Hyperplastic Tissues,<br>Treatment of Aphthous<br>Ulcers, Sulcular<br>Debridement, Pulpotomy,<br>Pulpotomy as an Adjunct<br>to Root Canal Therapy,<br>and Light Activation of<br>Bleaching Materials. | incision and excision,<br>Hemostasis and<br>coagulation, Implant<br>recovery, Incision and<br>drainage of abscesses,<br>Leukoplakia,<br>Operculectomy, Oral<br>papillectomies,<br>Pulpotomy, Pulpotomy as<br>an adjunct to root canal<br>therapy, Reduction of<br>gingival hypertrophy, Soft<br>tissue crown lengthening,<br>Treatment of canker<br>sores, herpetic and<br>aphthous ulcers of the<br>oral mucosa, and<br>Vestibuloplasty.<br>*For use on adult and<br>pediatric patients<br>Laser Periodontal<br>Procedures<br>Laser soft tissue<br>curettage, Laser removal<br>of diseased, infected,<br>inflamed and necrosed<br>soft tissue within the<br>periodontal pocket,<br>Removal of highly<br>inflamed edematous<br>tissue affected by bacteria<br>penetration of the pocket<br>lining and junctional<br>epithelium, Sulcular<br>debridement (removal of<br>diseased or inflamed soft<br>tissue in the periodontal<br>pocket to improve clinical<br>indices including gingival<br>index, gingival bleeding<br>index, probe depth,<br>attachment loss and tooth<br>mobility)<br>Tooth Whitening<br>Laser assisted<br>whitening/bleaching of<br>teeth, Light activation for<br>bleaching materials for<br>teeth. |
| | | | | |
:
:
{2}------------------------------------------------
02262
Conclusion:
After analyzing both bench and user testing data, it is the conclusion of Kavo that the GENTLEray 980 Diode Laser System is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kavo America % Intertek Testing Services Mr. Daniel W. Lehtonen 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
OCT 2 6 2007
Re: K072262 Trade/Device Name: GENTLEray 980 Diode Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and and in dermatology Regulatory Class: II Product Code: GEX Dated: October 10, 2007 Received: October 11, 2007
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Daniel W. Lehtonen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/suport/index.html
Sincerely yours,
Mark A. Milkeron
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): k022262
Device Name: GENTLEray 980 Diode Laser System
Indications for Use:
Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue.
The GENTLEray 980 Diode Dental Laser System is intended for use in the following procedures: Frenectomy, Frenotomy, Biopsy, Operculectomy, Implant Recovery, Gingivectomy, Gingivoplasty, Gingival Troughing, Crown Lengthening, Hemostasis of Donor Site, Removal of Granulation Tissue, Laser-assisted Flan Singgry, Debridement of Deseased Epithelial Lining, Incisions and Draining of Abscesses, Tissue Retraction for Impressions, Papillectomy, Vestibuloplasty, Excision of Lesions, Leukoplakia, Exposure of Unerupted/Partially Erupted Teeth, Removal of Hyperplastic Tissues, Treatment of Aphthous Ulcers, Sulcular Debridement, Pulpotomy, Pulpotomy as an Adjunct to Root Canal Therapy, and Light Activation of Bleaching Materials.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arlene A. Millerson
(Division of General, Restorative, and Neurological Dev
KO 72262
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.