CARDIOTECH SPIROMETRY SYSTEM, MODEL GT-105

K090646 · Chest M.I., Inc. · BZG · Jul 23, 2009 · Anesthesiology

Device Facts

Record IDK090646
Device NameCARDIOTECH SPIROMETRY SYSTEM, MODEL GT-105
ApplicantChest M.I., Inc.
Product CodeBZG · Anesthesiology
Decision DateJul 23, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1840
Device ClassClass 2
AttributesPediatric

Indications for Use

The CardioTech Spirometry System, Model GT-105, is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (4 to 17 years) and adult (18 to 99 years) patients in hospitals, physician's offices, laboratories, and occupational health environments.

Device Story

CardioTech GT-105 measures physiological parameters (SVC, FVC, MVV, MV) via flow sensor head with air resistance element; patient breathes through disposable mouthpiece/filter. Pressure transducer measures differential pressure; A/D converter digitizes signal (12-bit). Data transfers via RS-232C/USB to PC running Spirobank software. Medical practitioners (technicians) use software interface to print, save, and recall results. Device used in hospitals, physician offices, labs, and occupational health environments. Provides objective lung function data to assist clinicians in diagnosis and patient management.

Clinical Evidence

Bench testing only. Device performance tested in accordance with American Thoracic Society (ATS) standards (1994 Update). Electrical safety and electromagnetic compatibility tested per IEC 60601-1 and IEC 60601-1-2.

Technological Characteristics

Diagnostic spirometer; flow sensor with resistance element; pressure transducer; 12-bit A/D conversion; RS-232C/USB connectivity; PC-based software (Spirobank). Tested to IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).

Indications for Use

Indicated for pediatric (4-17 yrs) and adult (18-99 yrs) patients requiring measurement of maximal lung air volume and flow for diagnostic purposes in clinical, laboratory, or occupational settings.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K090646 # 510(k) Summary (per 21 CFR 807.92(c)) 1. Applicant JUL 2 3 2009 CHEST M. I., Inc. 3-6-10 Hongo Bunkyo Tokyo, Japan 113-0033 Hiroaki Hoki, General Manager, International Sales & Marketing, Export & Import Division Tel: (81) 3 3812 7251 Fax: (81) 3 3812 7220 E-mail: nw008@nifty.com Date Prepared: January 19, 2009 - 2. Device Name CardioTech GT-105 Spirometer Trade Name: Common/ Usual Name: Spirometer, Diagnostic Classification Name: Diagnostic Spirometer Regulation Number: 868.1840 Product Code: BZG Classification: II Panel: Anesthesiology ## 3. Predicate Devices The CardioTech Spirometer is substantially equivalent to the following devices: | 510(k) Number | Device | Applicant | |---------------|----------------------------------------|----------------------| | K022103 | PB700 Renaissance II Spirometry System | Puritan Bennett Inc. | | K031102 | Microlab Spirometer | Micro Direct, Inc. | ## 4. Indications for Use The CardioTech Spirometry System, Model GT-105, is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (4 to 17 years) and adult (18 to 99 years) patients in hospitals, physician's offices, laboratories, and occupational health environments. {1}------------------------------------------------ ## 5. Description of the Device The CardioTech GT-105 Spirometry System measures a wide range of physiological parameters such as SVC, FVC, MVV, and MV and consists of a spirometer and a flow sensor. The flow sensor head, which includes a flow air resistance element, is mounted on the flow sensor body. Via a disposable mouthpiece and filter, the patient breathes as instructed through the flow sensor head. The pressure transducer in the spirometer then measures the differential pressure in the flow sensor head which is digitized into 12 bit data by an A/D converter. The data can be transferred via RS-232C/USB interfaces to a PC in which Spirobank software is installed. Since the software includes a number of screens and 'buttons' that help to logically perform the function, the medical practitioner (e.g., technician) can print (reports), save and recall the data/results. #### 6. Summary of the Technical Characteristics #### Electrical Testing . The CardioTech Spirometry System was tested in accordance with the medical electrical equipment requirements defined by IEC 60601-1-2:2001 + A1:2005 and IEC 60601-1:1998 + A1:1991 + A2:1995 for electromagnetic compatibility and electrical safety, respectively. #### Performance Testing . The CardioTech Spirometry System was tested in accordance with the industry standard developed by the American Thoracic Society. 3 #### 7. Safety and Effectiveness The CardioTech GT-105 Spirometry System is a safe and effective device and is substantially equivalent to the predicate devices listed in this 510(k) submission; that is, the CardioTech GT-105 Spirometry System has the same indications for use and is similar in both design and function. Any differences in technological characteristics between the CardioTech GT-105 Spirometry System and the predicate devices `do not raise issues of safety and effectiveness. 4 American Thoracic Society Standardization of Spirometry 1994 Update. Am J Respir Crit Care Med 1995, 152: 1107-1136. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". JUL 2:3 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Chest M.I. Incorporated C/O Ms. Jean Asquith Senior Regulatory Consultant Emergo Group, Incorporated 1705 South Capital of Texas Highway, Suite 500 Austin, Texas 78746 Re: K090646 Trade/Device Name: CardioTech Spirometry System, Model GT-105 Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: July 6, 2009 Received: July 8, 2009 ## Dear Ms. Asquith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2- Ms. Asquith Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance, You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, ph far Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ## Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: CardioTech Spirometry System, Model GT-105 Indications for Use:. The CardioTech Spirometry System, Model GT-105, is a diagnostic tool to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (4 to 17 years) and adult (18 to 99 years) patients in hospitals, physician's offices, laboratories, and occupational health environments. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schulten (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K090646
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...