MICROLAB SPIROMETER

K031102 · Micro Direct, Inc. · BZG · Dec 1, 2003 · Anesthesiology

Device Facts

Record IDK031102
Device NameMICROLAB SPIROMETER
ApplicantMicro Direct, Inc.
Product CodeBZG · Anesthesiology
Decision DateDec 1, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1840
Device ClassClass 2
AttributesPediatric

Indications for Use

The Microlab spirometer is intended, for prescription use only, to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (4 to 17 years) and adult (18 to 99 years) patients in hospitals, physician offices, laboratories and occupational health testing environments.

Device Story

Microlab Spirometer measures lung function via digital volume transducer; turbine sensor captures airflow; electronic integration converts flow to volume; system calculates clinically relevant parameters; compares results to published predicted normal values. Used in hospitals, physician offices, laboratories, and occupational health settings by clinicians. Output displayed for provider assessment of lung disease detection and monitoring.

Clinical Evidence

Bench testing only. Performance testing conducted to demonstrate compliance with American Thoracic Society accuracy and precision standards. Electrical safety testing performed per IEC60601-1-2.

Technological Characteristics

Turbine sensor for airflow measurement; electronic integration for volume calculation. Includes digital volume transducer, housing, AC power adapter, and disposable cardboard mouthpieces. Connectivity via serial cable for external printer. Electrical safety per IEC60601-1-2.

Indications for Use

Indicated for pediatric (4-17 years) and adult (18-99 years) patients for measuring maximal volume and flow of air in/out of lungs to detect and monitor lung diseases. Prescription use only.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEC - 1 2003 K03112 Non-Confidential Summary of Safety and Effectiveness April 4, 2003 | Micro Direct, Inc.<br>803 Webster Street<br>Lewiston, ME 04240 | Tel - (207) 786-7808 | Fax - (207) 786-7280 | |----------------------------------------------------------------|----------------------|----------------------| |----------------------------------------------------------------|----------------------|----------------------| | Official contact: | David R. Staszak, President | |----------------------------|--------------------------------------------| | Proprietary or Trade Name: | Microlab Spirometer | | Common/Usual Name: | Spirometer | | Classification Name: | 73 BZG – Diagnostic Spirometer | | Intended device: | Diagnostic Spirometer | | Predicate devices: | Micro Medical MicroDL Diarycard Spirometer | Device description: The Microlab spirometry systems consists of the Microlab spirometer, digital volume transducer, transducer housing, AC power adapter, power supply cord, disposable cardboard mouthpieces, carrying case, quick start guide and operators manual. Optional accessories include calibration syringe, nose clips, serial cable and a serial to parallel converter for use of an external compatible printer. Intended use: The Microlab spirometry system is intended, for prescription use only, to measure the maximal volume and flow of air that can be moved in and out of a patient's lungs. The testing can be used in the detection and monitoring of certain lung diseases. The system is intended for use with pediatric and adult patients (4 to 99 years old) in hospitals, physician offices, laboratories and occupational health testing environments. Comparison to predicate device: The Microlab spirometry system has the same technological characteristics of the referenced predicate device. In both devices, air flow is measured directly via a turbine sensor and then electronically integrated to obtain volume. Both devices then perform calculations to express the volume in clinically relevant terms, including comparisons to predicted normal values from published literature. The intended use for both devices is the same. Summary of Testing: Performance testing was conducted to demonstrate compliance with the accuracy and precision standards set by the American Thoracic Society. The device was tested to demonstrate conformance with IEC60601-1-2 requirements for electrical safety. - Conclusion: The Microlab spirometry systems does not raise new questions of safety or effectiveness when compared to the legally marketed predicate device and is substantially equivalent to the predicate device. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle, with three stylized wing shapes. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 1 2003 Mr. David R. Staszak President Micro Direct, Incorporated 803 Webster Street Lewiston, Maine 04240 Rc: K031102 Trade/Device Name: Microlab Spirometer Regulation Number: 21 CFR 868.1840 Regulation Name: Spirometer, Diagnostic Regulatory Class: II Product Code: BZG Dated: September 24, 2003 Received: September 25, 2003 Dear Mr. Staszak: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. David R. Staszak Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Susan Runion Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Statement Ko31102 (To be assigned) 510(k) Number Device Name: Microlab Spirometer Indications for Use: The Microlab spirometer is intended, for prescription use rno Mioreleb ope the maximal volume and flow of air that can be moved in and out of a patient's lungs. The system is intended for use with pediatric (4 to 17 years) and adult (18 to 99 years) patients in hospitals, physician offices, laboratories and occupational health testing environments. Concurrence of CDRH, Office of Device Evaluation (ODE) Q. Nartorhe for JXH 11/25/03 ision Sign-Off ion of Anesthesiology, General Hospital, Infection Control, Dental Device 510(k) Number: _K Prescription Use √ Over-The-Counter Use OR (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...