The SmartScreener-Plus2 device is indicated for use in the recording and automated analysis of human physiological data (auditory brainstem responses and/or otoacoustic emissions) necessary for the diagnosis of auditory and hearing-related disorders. The device is especially indicated for use in the screening of infants to determine hearing loss. Sites appropriate for use include the well-baby nursery, neonatal intensive care unit (NICU), mother's bedside, outpatient clinic, audiologist's office, or doctor's office. The device is simple to operate. It does not require special technical skills or interpretation of results by the device operator. Basic training with the device is sufficient to learn how to operate it. The SmartScreener-Plus2 device can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults. The possible anatomical sites of contact are the patient's ear, ear canal, and/or external skin surrounding the ear (with the contact object being a sound delivery eartip, headphone cushion, or an acoustic ear coupler such as a circumaural ear cup, intra-aural ear bud, or supra-aural ear patch), the patient's head (with the contact object being a bone vibrator), and the patient's head, nape of the neck and shoulder (with the contact object being electrodes that are capable of measuring biopotentials). The SmartScreener-Plus2 device is for prescription use.
Device Story
SmartScreener-Plus2 is a noninvasive diagnostic device for hearing and neurological abnormalities. Inputs: electrical signals (AEPs/ABR) via skin electrodes and acoustic signals (TEOAEs/DPOAEs) via probe microphone. Operation: device delivers auditory stimuli; records physiological responses; uses signal processing to isolate signals from background noise; applies statistical algorithms to determine presence of response. Output: automated analysis of hearing status. Used in NICUs, nurseries, clinics, and offices by trained operators (nurses/technicians). Simple operation; no complex interpretation required. Benefits: enables objective hearing screening in populations unable to perform behavioral tests.
Clinical Evidence
Bench testing only. Device specifications were verified and validated against predicate devices. No clinical study data provided.
Technological Characteristics
System includes main hardware unit, transducers, acoustic couplers, electrodes, cables, and computer. Sensing: electrodes for biopotentials, microphone for acoustic emissions. Energy: auditory stimulation. Connectivity: computer-based. Electrical safety: meets IEC60601-1, UL2601-1, CSA-C22.2 No. 601.1, AAMI.ES1, EN60601-1-2, EN60601-2-40, EN55011, EN61000-3-2, EN61000-3-3.
Indications for Use
Indicated for patients of all ages, including infants, children, adults, and geriatric patients, for recording and automated analysis of auditory brainstem responses (ABR) and otoacoustic emissions (OAE) to diagnose auditory/hearing disorders. Especially indicated for screening infants and individuals where behavioral audiometry is unreliable (e.g., cognitively impaired or uncooperative adults).
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Predicate Devices
Bio-logic ABaer Cub with integrated OAE and ABR functions (K021801)
Intelligent Hearing Systems SmartEP-ASSR (K031051)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Intelligent Hearing Systems. The logo has a graphic to the left of the words "Intelligent Hearing" stacked on top of "Systems". The graphic is a square with rounded corners and a series of concentric arcs inside. The letters in "Systems" are each in their own square.
# 0(k) Summarv
JUL - 7 2009
| Submitter: | Intelligent Hearing Systems<br>6860 SW 81st Street<br>Miami, FL 33143 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone:<br>FAX:<br>Contact Person: | (305) 668-6102<br>(305) 668-6103<br>Edward Miskiel, Ph.D. |
| Date Summary Prepared: | June 18, 2009 |
| Trade Name: | SmartScreener-Plus2 |
| Common Name(s): | Auditory Evoked Potential Device, Otoacoustic Emissions Device,<br>Automated Hearing Screener. |
| Classification Name: | Auditory Evoked Response Stimulator (per CFR.882.1900)<br>& Audiometer (per CFR 874.1050) |
| Predicate Device(s): | Bio-logic ABaer Cub with integrated OAE and ABR functions<br>510(k) #K021801<br>Intelligent Hearing Systems SmartEP-ASSR 510(k) #K031051 |
### Description of the Device:
The SmartScreener-Plus2 is a noninvasive device used to test for hearing and related neurological abnormalities. The device is especially indicated for use in the screening of infants to determine hearing loss.
The SmartScreener-Plus2 device records and detects auditory evoked potentials (AEPs) and otoacoustic emissions (OAEs) which are generated in response to a series of auditory stimuli delivered to the patient. AEPs are electrical signals produced by the patient's nourological activity and are picked up by electrodes attached to the patient's skin. The portion of the AEP that originates from the anatomical structures of the inner ear to the auditory brainstem is commonly called the Auditory Brainstem Response (ABR). Otoacoustic emissions are acoustic signals produced by the patient's inner ear and are picked up by a sensitive microphone located in an OAE probe. The SmartScreener-Plus2 device is capable of recording Transient Evoked OAEs (TEOAEs or TrOAEs), and Distortion Product OAEs (DPOAEs)
The SmartScreener-Plus2 uses advanced signal processing technology to separate AEPs and OAEs from background noise and other physiological activity. The device uses proven statistical algorithms to automatically determine if there is an AEP or OAE response present to the auditory stimuli delivered to the patient. A user, with device administrator approved password security clearance, is able to set the output level of the stimulus. AEP stimulus levels
{1}------------------------------------------------
between 30 to 40 dB nHL, and TEOAE stimulus level of 85 dB SPL, and DPOAE stimulus levels of 65 and 55 dB SPL are commonly used for newborn hearing screening applications.
The SmartScreen-Plus2 device is simple to operate. It does not require any special technical skills or interpretation of results. Basic training is sufficient to learn how to use the device.
## Indications for Use:
. The SmartScreener-Plus2 device is indicated for use in the recording and automated analysis of human physiological data necessary (auditory brainstem responses and/or otoacoustic emissions) for the diagnosis of auditory and hearing-related disorders. The device is especially indicated for use in the screening of infants to determine hearing loss.
Sites appropriate for use include the well-baby nursery, neonatal intensive care unit (NICU), mother's bedside, outpatient clinic, audiologist's office, or doctor's office. The device is simple to operate. It does not require special technical skills or interpretation of results by the device operator. Basic training with the device is sufficient to learn how to operate it:
The SmartScreener-Plus2 device can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults.
The possible anatomical sites of contact are the patient's ear, ear canal, and/or external skin surrounding the ear (with the contact object being a sound delivery eartip, headphone cushion, or an acoustic ear coupler such as a circumaural ear cup, intra-aural ear bud, or supra-aural ear patch), the patient's head (with the contact object being a bone vibrator), and the patient's head, nape of the neck and shoulder (with the contact object being electrodes that are capable of measuring bio-potentials).
The SmartScreener-Plus2 device is for prescription use.
#### Comparison with Predicate Device:
Comparisons of technological characteristics of the SmartScreener-Plus2 with the predicate devices (Bio-logic ABaer Cub with integrated OAE and ABR functions and the Intelligent Hearing Systems SmartEP-ASSR) are given in the table below:
| Parameter for Comparison | Predicate Device<br>Bio-logic ABaer Cub with integrated<br>OAE and ABR functions (K021801) | Predicate Device<br>IHS SmartEP-ASSR<br>(K031051) | Similarity or<br>Difference to<br>Predicate Devices |
|--------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------|
| Target Population | Patients of all ages. | Same. | Same. |
| Where Used | Clinical setting. | Same. | Same. |
| Anatomical Sites | Scalp, ear and other skin sites. | Same. | Same. |
| Sterility | None required. | Same. | Same. |
| Biocompatibility | Completely biocompatible. | Same. | Same. |
{2}------------------------------------------------
| Design | Main hardware unit (housing<br>circuitry, supplies, and equipment)<br>connected to (1) patient cables and<br>transducers and (2) a portable<br>computer (with installed testing<br>software). | Same. | Same. |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Components/<br>Materials | Main hardware unit, transducers,<br>acoustic couplers, electrodes, cables,<br>computer, printer and cart. | Same. | Same. |
| Human Factors | Simple, easy-to-follow instructions<br>and software user interface. | Same. | Same. |
| Standards. | None known. | Same. | Same. |
| Energy Used and/or<br>Delivered | Stimulation of auditory system. | Same. | Same. |
| Chemical Safety | No chemicals are involved in the use<br>of this device. | Same. | Same. |
| Mechanical Safety | No mechanical parts in contact with<br>patient. | Same. | Same. |
| Electrical Safety | Meets IEC60601-1, UL2601-1,<br>CSA-C22.2 No. 601.1, AAMI.ES1,<br>and CE Mark. | Meets EN60601-1,<br>EN60601-1-2,<br>EN60601-2-40,<br>EN55011,<br>EN61000-3-2,<br>EN61000-3-3, and<br>CE Mark. | Same as SmartEP-<br>ASSR (K031051). |
| Thermal Safety | Device is not thermal in nature. | Same. | Same. |
### Substantial Equivalence:
Based on the information presented here, the SmartScreener-Plus2 device is substantially equivalent to the Bio-logic ABaer Cub with Integrated OAE and ABR and the Intelligent Hearing Systems SmartEP-ASSR devices.
#### Performance Testing:
This submission includes a comparison table of specifications for both the SmartScreener-Plus2 device and the predicate devices. The SmartScreener-Plus2 device specifications were verified and validated.
### Conclusions:
Based on the information presented in this submission, the SmartScreener-Plus2 device is as safe and effective as the identified predicate devices, the Bio-logic ABaer Cub with integrated OAE and ABR and the Intelligent Hearing Systems SmartEP-ASSR.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with outstretched wings, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple and iconic, representing the HHS's mission to protect the health of all Americans.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Intelligent Hearing Systems % Edward Miskiel, Ph.D. President & CEO 6860 SW 81 51 Street Miami, Florida 33143
JUL - 7 2009
Re: K090579
Trade/Device Name: SmartScreener-Plus2 Regulation Number: 21 CFR 882.1900 Regulation Name: Auditory Evoked Response Stimulator Regulatory Class: Class II Product Code: GWJ, EWO Dated: June 18, 2009 Received: June 24, 2009
Dear Dr. Miskiel:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
MAJ. R. E. H. M.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number: K090579
Device Name: SmartScreener-Plus2
Indications For Use:
The SmartScreener-Plus2 device is indicated for use in the recording and automated analysis of human physiological data (auditory brainstem responses and/or otoacoustic emissions) necessary for the diagnosis of auditory and hearingrelated disorders. The device is especially indicated for use in the screening of infants to determine hearing loss.
Sites appropriate for use include the well-baby nursery, neonatal intensive care unit (NICU), mother's bedside, outpatient clinic, audiologist's office, or doctor's office. The device is simple to operate. It does not require special technical skills or interpretation of results by the device operator. Basic training with the device is sufficient to learn how to operate it.
The SmartScreener-Plus2 device can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults.
The possible anatomical sites of contact are the patient's ear, ear canal, and/or external skin surrounding the ear (with the contact object being a sound delivery eartip, headphone cushion, or an acoustic ear coupler such as a circumaural ear cup, intra-aural ear bud, or supra-aural ear patch), the patient's head (with the contact object being a bone vibrator), and the patient's head, nape of the neck and shoulder (with the contact object being electrodes that are capable of measuring biopotentials).
The SmartScreener-Plus2 device is for prescription use.
510(k) Number
:
Prescription Use _______X (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------|--|
| Division of Ophthalmic, Neurological and Ear, | |
| Nose and Throat Devices | |
Page 1 of 1K090579
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.