K031051 · Intelligent Hearing Systems · GWJ · Feb 20, 2004 · Neurology
Device Facts
Record ID
K031051
Device Name
SMARTEP-ASSR, MODEL M811007
Applicant
Intelligent Hearing Systems
Product Code
GWJ · Neurology
Decision Date
Feb 20, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1900
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of the SmartEP-ASSR product is for the recording of auditory steady-state evoked potential data. The product is intended to be used as a diagnostic aid in auditory and hearing related disorders and as an adjunctive tool in the estimation of behavioral hearing thresholds. SmartEP-ASSR can be used for patients of all ages. It is intended to be used by trained personnel in a hospital, nursery, clinic, audiologist's office or other appropriate setting.
Device Story
SmartEP-ASSR is an auditory evoked potential testing system; measures auditory steady-state responses (ASSR). Inputs: physiological signals via 1-4 optically isolated channels; auditory stimuli (clicks, pure tones, multifrequency) via transducers (earphones, bone conduction, OAE probe). System processes signals using programmable artifact rejection and user-selectable digital filters; outputs ASSR data for clinical interpretation. Used in hospitals, clinics, or audiologist offices by trained personnel. Provides diagnostic aid for hearing disorders; assists in estimating behavioral hearing thresholds. Benefits patients by enabling objective hearing assessment across all ages.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and comparison of technical specifications.
Technological Characteristics
Evoked response auditory stimulator; 1-4 optically isolated channels; 16-bit A/D resolution. Stimuli: clicks, pure tones, multifrequency (500Hz-8kHz). Connectivity: USB interface. Software runs on Windows 98SE/ME/2000/XP. Features programmable artifact rejection, variable gain (30k-300k), and user-selectable digital filters (1-5k Hz).
Indications for Use
Indicated for patients of all ages with auditory or hearing-related disorders to record auditory steady-state evoked potential (ASSR) data and estimate behavioral hearing thresholds.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
{0}------------------------------------------------
K031051
FEB 2 0 2004
## 510(k) Summary
| Prepared By: | Intelligent Hearing Systems<br>7356 SW 48th Street<br>Miami, FL 33155 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone: | (305) 668-6102 |
| FAX: | (305) 668-6103 |
| Contact Person: | Edward Miskiel, Ph.D. |
| Date Summary prepared: | February 13, 2004 |
| Name of the Device: | SmartEP-ASSR |
| Common Name: | Evoked Response System |
| Classification Name: | Evoked Response Auditory Stimulator (per CFR 874.1900) |
| Predicate Device: | Bio-logic MASTER Evoked Response System (K021895) |
| Device Description: | SmartEP-ASSR is a auditory evoked potential testing device<br>that is capable of measuring auditory steady state responses<br>(ASSR). |
| Intended Use: | The intended use of the SmartEP-ASSR product is for the<br>recording of auditory steady-state evoked potential data.<br>The product is intended to be used as a diagnostic aid in<br>auditory and hearing related disorders and as an adjunctive<br>tool in the estimation of behavioral hearing thresholds.<br>SmartEP-ASSR can be used for patients of all ages. It is<br>intended to be used by trained personnel in a hospital,<br>nursery, clinic, audiologist's office or other appropriate<br>setting. |
| Technological Characteristics: | The SmartEP-ASSR device is similar to the predicate device<br>in its intended use and data processing methodologies. |
# INTELLIGENT HEARING SYSTEMS
· Miami, Florida 33155 · Telephone: 305-668-6102 · FAX: 305-668-6103 7356 S.W. 48th Street Email Address: info@ihsys.com = www.ihsys.com Page E1
{1}------------------------------------------------
### Substantial Equivalence Based on Assessment of Performance Data:
The SmartEP-ASSR is substantially equivalent to the Bio-logic MASTER Evoked Response System device marketed by Bio-logic Systems Corp. with FDA 510(k) clearance number K021895. Comparisons of technical parameters are shown in the table below.
| Parameter | Predicate Device<br>(Non-Prcamendmant Device)<br>Bio-Logic Master (K021895) | Device Under Current 510(k)<br>Review<br>SmartEP-ASSR |
|-----------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Data Acquisition | | |
| A/D Resolution | 16 bit | 16 bit |
| Artifact Rejection | Programmable | Programmable |
| Amplifiers | | |
| Channels | 1 Channel Optically Isolated | 1-4 Channels Optically Isolated |
| Gain | 10k | Variable (30k-300k) |
| Filters | | |
| Slope | 12 dB/octave | 6 dB/octave |
| LP | 1.5k, 3k, 10k, 20k Hz | 30, 100, 300, 500, 1k, 1.5k, 3k, 5k<br>Hz |
| HP | 0.1, 0.3, 1, 3, 10, 30, 100, 300Hz | 1, 10, 30, 50, 100, 150, 300, 500 Hz |
| Notch | 50/60Hz | 50/60Hz |
| Digital Filters | 1-200Hz | User Selectable (1 -5k Hz) |
| Noise Level | 0.45 μV RMS (10-3k Hz) | 0.33 μV RMS (1-3k Hz) |
| Input Impedance | 100M Ohms | 5M Ohms |
| CMR Ratio | 110 dB at 50/60 Hz | 117 dB at 60 Hz and<br>110 dB at 1k Hz |
| Impedance Test | | |
| Signal | 20 Hz Sinewave | 1k Hz Sinewave |
| Auditory Stimuli | | |
| Presentation | Monaural or Binaural | Monaural or Binaural |
| Number of Frequencies | 1-4 | 1-8 |
| Test Frequencies | 500, 750, 1k, 1.5k, 2k, 3k, 4k, 6k, 8k<br>Hz | User Selectable: Clicks, Pure Tones<br>and Multifrequency Stimuli (500Hz-<br>8k Hz) |
| Types | Sinewave | Sinewave, Tone Burst |
| Envelopes | Linear, Blackman, Gaussian,<br>Hanning Envelopes | Linear, Blackman, Gaussian,<br>Hanning, Rectangular, Triangular,<br>Trapezoidal, Exact Blackman,<br>Cosine, Cosine Squared, Cosine<br>Cubed |
| Intensity | 0-125 dB SPL<br>(132 dB Optional) | 0-125 dB SPL |
| Masking | White Noise Programmable | White Noise Programmable |
| Transducers | TDH Earphones, Insert Earphones,<br>Bone Conduction, Sound Field | TDH Earphones, Insert Earphones,<br>Bone Conduction, Sound Field, OAE<br>Probe |
| Computer Requirements | | |
| Computer Type | Personal Computer | Personal Computer |
| Operating System | Microsoft Windows | Microsoft Windows 98SE, ME,<br>2000, XP |
| Interface Connection | Serial | USB (Universal Serial Bus) |
{2}------------------------------------------------
:
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three horizontal bars that curve upwards, resembling a stylized human figure. The symbol is black, and the text is also black.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 0 2004
Edward Miskiel, Ph.D. President and CEO Intelligent Hearing Systems 7356 S.W. 48" Street Miami, Florida 33155
Rc: K031051
Trade/Device Name: SmartEP- ASSR Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked response auditory stimulator Regulatory Class: II Product Code: GWJ Dated: December 18, 2004 Received: December 23, 2004
Dear Dr. Miskiel:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bootion of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the encreased 976, the enactment date of the Medical Device Amendments, or 10 commerce print to May 2017 - 7:37
devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, de nees mat nave been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, merelore, manes of the Act include requirements for annual registration, listing of general controls provisions and prohibitions against misbranding and adulteration.
If your device is classificd (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that I Dr Has Intact a and regulations administered by other Federal agencies. You must of any I edetar statutes and regirements, including, but not limited to: registration and listing (21 Cliff Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K I art 607), latoring (21 OS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Edward Miskiel, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and A yours finding of substantial equivalence of your device to a legally premarket notication: "The Presults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific accept (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N Milken
elia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K031051
Device Name: SmartEP-ASSR
Indications For Use:
The intended use of the SmartEP-ASSR product is for the recording of auditory steadystate evoked potential data. The product is intended to be used as a diagnostic aid in auditory and hearing related disorders and as an adjunctive tool in the estimation of behavioral hearing thresholds.
SmartEP-ASSR can be used for patients of all ages. It is intended to be used by trained personnel in a hospital, nursery, clinic, audiologist's office or other appropriate setting.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div><img alt="signature" src="signature.png"/></div> |
|--|-------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of General, Restorative, |
| | and Neurological Devices |
| 510(k) Number | K031051 |
|---------------|---------|
|---------------|---------|
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.