VALIDATE CM2 CALIBRATION VERIFICATION / LINEARITY TEST SET, MODEL 402
Applicant
Maine Standards Co.
Product Code
JJY · Clinical Chemistry
Decision Date
Apr 16, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
VALIDATE CM2 Calibration Verification/Linearity Test Set solutions are intended for in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi-automated, and manual chemistry systems. Each VALIDATE CM2 Calibration Verification/Linearity Test Set consists of two sets of bottles. Set 1 contains BNP, hs-CRP, Troponin-I, and MPO. Set 2 contains NT-proBNP, hs-CRP, Troponin-T and MPO.
Device Story
VALIDATE® CM2 Calibration Verification / Linearity Test Set consists of human serum-based solutions containing multiple levels of analytes (TnI, TnT, BNP, NT-proBNP, hsCRP, MPO). Used in clinical laboratories by trained professionals to verify the performance of automated, semi-automated, and manual chemistry analyzers. Device establishes relationship between theoretical operation and actual performance of analytes; confirms linearity and reportable ranges. Not for routine quality control or calibration. Healthcare providers use output to validate analyzer accuracy and ensure reliable patient test results.
Clinical Evidence
No clinical data. Bench testing only. Stability testing protocols and acceptance criteria were reviewed and found acceptable. Value assignment process involves triplicate assays of high/low levels and final bottled product to ensure linear relationship.
Technological Characteristics
Human serum-based matrix; liquid, ready-to-use form factor. Multi-level test set (2 sets of 6 levels). Storage at -10 to -20°C. No electronic components, software, or energy sources.
Indications for Use
Indicated for in vitro diagnostic use in clinical chemistry systems to verify linearity, calibration, and reportable ranges for Troponin-I, Troponin-T, BNP, NT-proBNP, hsCRP, and MPO. Not for use as routine quality control or calibration materials. Not for use on systems employing reflectance spectroscopy.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
VALIDATE® CM1 Calibration Verification / Linearity Test Set (K090475)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k090475
B. Purpose for Submission:
New device
C. Measurand:
Calibrator/Control material for Troponin-I (TnI), Troponin-T (TnT), BNP, the N-amino terminus of brain naturiuretic peptide (NT-proBNP), high sensitive C-reactive protein (hsCRP), and Myeloperoxidase (MPO).
D. Type of Test:
Control Material
E. Applicant:
Maine Standards Company
F. Proprietary and Established Names:
VALIDATE CM2 Calibration Verification/Linearity Test Set
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1660, Quality Control Material (assayed and unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJY, Multi-analyte controls, all kinds (assayed)
4. Panel:
Clinical Chemistry (75)
{1}
Page 2 of 5
## H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indications(s) for use:
VALIDATE CM2 Calibration Verification/Linearity Test Set solutions are intended for *in vitro* diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi-automated, and manual chemistry systems. Each VALIDATE CM2 Calibration Verification/Linearity Test Set consists of two sets of bottles. Set 1 contains BNP, hs-CRP, Troponin-I, and MPO. Set 2 contains NT-proBNP, hs-CRP, Troponin-T and MPO.
3. Special conditions for use statement(s):
VALIDATE CM2 Calibration Verification/Linearity Test solutions are not intended for use as routine quality control materials or as calibration materials.
These solutions are not intended for use on systems employing reflectance spectroscopy.
4. Special instrument requirements:
Automated, semi-automated, and manual chemistry systems.
## I. Device Description:
VALIDATE CM2 Calibration Verification/Linearity Test Set is in a human serum matrix that is compatible with chemistry systems for measuring Troponin-I (TnI), Troponin-T (TnT), BNP, the N-amino terminus of brain naturiuretic peptide (NT-proBNP), high sensitive C-reactive protein (hsCRP), and Myeloperoxidase (MPO). Each test set contains two sets of bottles. Set 1 contains TnI, BNP, hsCRP, and MPO; Set 2 contains TnT, NT-proBNP, hsCRP, and MPO. There is a bottle of base matrix included with each set. Each bottle contains 2.0 mL of solution. There exists a linear relationship among Levels 1 through 5.
Material of human origin used in the manufacture of this test set has been tested using FDA approved methods or methods cleared in compliance with the European Directive 98/79/EC, Annex I, III, or IV and found to be non-reactive for HBsAg and antibodies to HCV and HIV-1/2.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
VALIDATE CM1 Calibration Verification/Linearity Test Set.
2. Predicate 510(k) number(s):
{2}
k053247
Page 3 of 5
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device: VALIDATE CM2 | Predicate: VALIDATE CM1 |
| Intended Use | VALIDATE CM2 Calibration Verification/Linearity Test Set solutions are intended for in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi-automated, and manual chemistry systems. | The Validate Cardiac Marker Calibration Verification Test Sets are used for quantitatively verifying calibration, validating reportable ranges, and determining linearity in automated, semi-automated, and manual chemistry analyzers. |
| Matrix | Human Serum | Human Serum |
| Preparation | Liquid, ready to use | Liquid, ready to use |
| Stability | Until expiration | Until expiration |
| Storage | -10 to -20°C | -10 to -20°C |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analytes | Set 1: BNP, hs-CRP, TnI, MPO
Set2: NT-proBNP, hsCRP, TnT, MPO | CK-MB, MYO |
| Number of Levels | 2 sets of 6 including base matrix | 5 including a 0 |
| Packaging | 12 x 2.0 mL | 5 x 3.0 mL |
K. Standard/ Guidance Document Referenced (if applicable):
CLSI document EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline.
L. Test Principle:
Not Applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable.
3
{3}
Page 4 of 5
b. Linearity/assay reportable range:
Not Applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: No traceability claims were made.
Stability: Stability testing protocols and acceptance criteria were reviewed and found to be acceptable. Test sets are stable until the expiration date printed on the bottle when stored at -10° to -20°C and handled according to instructions. A maximum of four (4) freeze-thaw cycles are recommended.
Value Assignment: The raw materials are prepared at high concentration, analyzed, and are spiked into the human serum matrix. A linear relationship is achieved by mixing the low and high levels to produce five levels with equally spaced concentrations. For value assignment the high and low levels (1 and 5) are assayed in triplicate. Once all the levels are diluted all 5 levels are run again in triplicate, and then again in triplicate once bottled. Appropriate controls for a given system are run with each determination and must fall within the cited acceptance range. It is recommended that each laboratory establish its own values and acceptable non-linearity.
c. Detection limit:
Not Applicable.
d. Analytical specificity:
Not Applicable.
e. Assay cut-off:
Not Applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable.
b. Matrix comparison:
Not Applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not Applicable.
{4}
Page 5 of 5
5. Expected values/Reference range:
Not Applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.