K090221 · Viora , Ltd. · NUV · Jul 1, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090221
Device Name
REACTION SYSTEM
Applicant
Viora , Ltd.
Product Code
NUV · General, Plastic Surgery
Decision Date
Jul 1, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reaction™ system is intended for the treatment of the following medical conditions, using the B-contour and F-contour applicators for delivering non thermal RF combined with massage: - relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation and - temporary reduction in the appearance of cellulite. Using the ST applicator for delivering RF, the Reaction™ system is intended for the treatment of relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation.
Device Story
Reaction™ System combines CORE™ (Channeling Optimized RF Energy) technology with mechanical vacuum massage. System comprises console (power supply, vacuum pump, RF generator, CPU, display) and three applicators (B-Contour, F-Contour, ST). Operator adjusts RF power and vacuum intensity via handpiece or applicator panels. Used in clinical settings by trained operators. RF energy provides heating/therapeutic effects; vacuum massage aids circulation and tissue manipulation. Output is thermal/mechanical energy delivered to skin/muscle to alleviate pain, spasms, and improve circulation; cellulite reduction achieved via combined RF/massage. Benefits include non-invasive symptom relief and aesthetic improvement.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
System utilizes CORE™ (Channeling Optimized RF Energy) and mechanical vacuum massage. Includes console, handpiece, and three applicators (B-Contour, F-Contour, ST). Energy source: RF generator and vacuum pump. Connectivity: manual adjustment via operator panels on handpiece/applicators. Software: embedded control for RF power and vacuum intensity.
Indications for Use
Indicated for patients requiring relief of minor muscle aches/pain, relief of muscle spasms, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K090221
Image /page/0/Picture/1 description: The image shows the word "VIORA" in a bold, sans-serif font. Above the word is a stylized graphic that resembles a cursive letter or a decorative flourish. The graphic is composed of curved lines that form a loop-like shape, adding a touch of elegance to the overall design.
Kugo221
ર-I
page (1) of (3)
JUL - 1 2009
# 510(k) Summary:
Reaction™ System
Company Name: Viora Ltd.
Contact Person: Omri Kessler
COO
Telephone: +972-3-5622248 +972-3-5622247 Fax: E-mail: omri@vioramed.com
## Authorized US Agent:
Josef Luzon - CEO Viora Inc. 30 Montgomery Street, Suite 660 Jersey City, NJ 07302
Telephone: (201) 332-4100 Fax: (201) 332-4200 E-mail: josef(@vioramed.com
Date prepared: January 23, 2009
Trade Name: Reaction™ System
Classification name: Laser surgical instrument for use in-general and plastic surgery and in dermatology
Common/usual name: Massager, vacuum, light induced heating
Product Code: NUV, ISA
Regulation No.: 878.4810
Class: II
Panel identification: General and Plastic Surgery Panel.
-2-
{1}------------------------------------------------
090221
Page ② of ⑬
Image /page/1/Picture/2 description: The image shows the word "VIORA" in a bold, sans-serif font. Above the word is a stylized graphic that resembles an infinity symbol with a flourish extending upwards from the left side. The text and graphic are both black against a white background.
## Predicate Device:
VelaShape from Syneron Medical Ltd, Yokneam Elite, Israel, cleared under 510(k) # K071872;
5-2
VelaSmooth, Shaper from Syneron Medical Ltd, Yokneam Elite, Israel, cleared under 510(k) # K050397;
LPG Therapeutic Massager from LPG USA Inc, Fort Lauderdale, Fl, USA, cleared under 510(k) # K990445 and
### Description of the device:
The Reaction™ system combines CORE™ (Channeling Optimized RF Energy) technology with mechanical vacuum massage of the skin. The Reaction™ treatment is indicated for the treatment of selected medical conditions such as relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation, temporary reduction in the appearance of cellulite
The Reaction™ system provides the treatments using 3 specially-designed applicators that treat various areas of the body, including large and small areas.
The system includes five core modules: the Console, the hand piece, and 3 treatment applicators: B-Contour, F-Contour and ST.
The Reaction™ console houses the following components:
- Power supply
- Hand piece
- Vacuum pump
- RF Generator
- Main CPU
- Display unit
The hand piece allows the operator to adjust the treatment parameters of the RF power and intensity of mechanical vacuum massage.
The applicators incorporate operator panels and are used to adjust the treatment parameters of the RF power and mechanical vacuum manipulation.
-3-
{2}------------------------------------------------
K09J221/
Page 3 of 3
Image /page/2/Picture/2 description: The image shows the word "VIORA" in a serif font. Above the word is a stylized graphic that resembles an infinity symbol with a line extending upwards from the left side of the symbol. The text and graphic are in black and white.
### Indications for Use:
The Reaction™ system is intended for the treatment of the fellowing medical conditions. using the B-contour and F-contour applicators for delivering non thermal RF combined with massage:
5-3
- relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation and
- temporary reduction in the appearance of cellulite.
Using the ST applicator for delivering RF, the Reaction™ system is intended for the treatment of relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation.
#### Substantial Equivalence:
The Reaction 130 System using the B-contour and the F-contour applicators, has the same intended use and the same performance characteristics as the fotlowing predicate devices: VelaShape from Syneron Medical Ltd, Yokneam Elite, Israel, cleared under 510(k) # K071872; VelaSmooth, Shaper from Syneron Medical Ltd, Yokneam Elite, Israel, cleared under 510(k) # K050397 and LPG Therapeutic Massager from LPG USA Inc, Fort Lauderdale, Fl, USA, cleared under 510(k) # K990445.
When using the ST applicator, the Reaction 130 System has the same intended use and the same performance characteristics as the RF application of Cutera CMMCD from Cutera Inc in Brisbane, CA, USA, cleared under 510(k) # K080300 and the following predicate devices: Intelect® SWD 100/ Senior Solutions from Chattanooga Group in Hixson, TN. USA, cleared under 510(k) # K083433 and Auto*Therm® 390, Model ME 390 from Mettler Electronics Corp in Anaheim, CA, USA, cleared under 510(k) # K042554.
The Reaction TM System is therefore substantially equivalent to those devices for each of the intended applications.
#### Conclusion -
The evaluation of the Reaction " System does not raise any additional concerns regarding safety and effectiveness and may therefore be considered substantially equivalent to the predicate device.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and strength. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
JUL - 1 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Viora Ltd. % Viora Inc. Mr. Josef Luzon Chief Executive Officer 30 Montgomery Street, Suite 660 Jersey City, New Jersey 07302
Re: K090221
Trade/Device Name: Reaction™ System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: NUV, ISA, ILY Dated: June 18, 2009 Received: June 19, 2009
Dear Mr. Luzon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Josef Luzon
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
fren Nillson
Mark N.Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known):
K090221
Device Name:
Reaction™ System
Indications for Use:
The Reaction™ system is intended for the treatment of the following medical conditions, using the B-contour and F-contour applicators for delivering non thermal RF combined with massage: - relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation and - temporary reduction in the appearance of cellulite.
Using the ST applicator for delivering RF, the Reaction™ system is intended for the treatment of relief of minor musele aches and pain, relief of muscle spasm, temporary improvement of local blood circulation.
Prescription Use X ાર (Part 21 CFR 801 Subpart D)
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil Re Qdr for nkm
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K090221
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.