INTELECT SWD 100, MODEL 1600; SENIOR SOLUTIONS SHORTWAVE DIATHERMY, MODEL 1601
K083433 · Chattanooga Group · IMJ · Mar 18, 2009 · Physical Medicine
Device Facts
Record ID
K083433
Device Name
INTELECT SWD 100, MODEL 1600; SENIOR SOLUTIONS SHORTWAVE DIATHERMY, MODEL 1601
Applicant
Chattanooga Group
Product Code
IMJ · Physical Medicine
Decision Date
Mar 18, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5290
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy device for use in applying therapeutic deep heat for selected medical conditions by applying electromagnetic energy in the radio frequency band of 27.12 megahertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
Device Story
Device applies electromagnetic energy (27.12 MHz) to generate deep tissue heat for therapeutic purposes. System comprises power supply, RF oscillator, power amplifier, and output resonant tank. Clinician selects capacitive or inductive electrodes (various sizes/types) to deliver energy. Output intensity and tuning controls manage energy transfer; power output meter monitors current draw. Dosage determined by patient sensation (Dose I-IV). Used in clinical settings by healthcare providers to treat pain, muscle spasms, and joint stiffness. Benefits include increased blood flow and reduction of inflammatory conditions. Device does not treat malignancies.
Clinical Evidence
Bench testing only. Device performance verified through compliance with recognized consensus standards including UL 60601-1 and IEC 60601-1-2. No clinical data provided.
Technological Characteristics
Shortwave diathermy device operating at 27.12 MHz. Components: RF oscillator, power amplifier, resonant tank. Accessories: capacitive electrodes (round, flexible rubber) and inductive electrodes (monode, diplode). Standards: UL 60601-1, IEC 60601-1-2, IEC 60601-1-4. Energy source: electrical power supply. Form factor: clinical console with manual intensity and tuning controls.
Indications for Use
Indicated for patients requiring therapeutic deep heat for musculoskeletal disorders, muscle spasms, joint stiffness, contractures, and chronic inflammatory or infective conditions including tenosynovitis, bursitis, synovitis, and chronic inflammatory pelvic diseases. Contraindicated for the treatment of malignancies.
Regulatory Classification
Identification
A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.
Special Controls
*Classification: Class II (special controls).* The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible.
(ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined:
(A) Peak output power;
(B) Pulse width;
(C) Pulse frequency;
(D) Duty cycle;
(E) Characteristics of other types of modulation that may be used;
(F) Average measured output powered into the RF antenna/applicator;
(G) Specific absorption rates in saline gel test load or other appropriate model;
(H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and
(I) Characterization of the deposited energy density in saline gel test load or other appropriate model.
(iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use.
(v) Labeling must include the following:
(A) Output characteristics of the device;
(B) Recommended treatment regimes, including duration of use; and
(C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications.
(vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
{0}------------------------------------------------
#### K083433 510(k) #
Page 1 of 3
#### Chattanooga Group SUBMITTER: 4717 Adams Road Hixson, TN 37343
#### ESTABLISHMENT 1022819 REGISTRATION:
- CONTACT: Michael Treas. Director of Regulatory Affairs
DATE PREPARED: November 17, 2008
- PROPRIETARY NAME: Intelect® SWD 100/ Senior Solutions®
CLASSIFICATION: Class II
#### PRODUCT CODES: IMJ
#### 21 CFR 890.5290 REGULATION NUMBER Shortwave Diathermy AND COMMON NAMES:
PANEL: Physical Medicine
## Indications for Use:
Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffness, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases.
Generally accepted indications for use:
- . Pain Relief
- . Reduce Muscle Spasm
- Decrease Joint Stiffness ●
- Contractures
- Increase Blood Flow .
- Chronic Inflammatory Conditions .
- Bursitis .
- Tenosynovitis .
- Synovitis .
- Chronic Inflammatory Pelvic Disease ●
{1}------------------------------------------------
## 510(k) # K083433
# Intended Use:
Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy device for use in applying therapeutic deep heat for selected medical conditions by applying electromagnetic energy in the radio frequency band of 27.12 megahertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
# Description :
The Intelect® SWD 100/ Senior Shortwave Diathermy devices consists of a power supply that provides power to a radio frequency oscillator. This radio frequency oscillator provides stable, drift-free oscillations at the required frequency. The power amplifier generates the power required to drive the different types of electrodes. The output resonant tank tunes in the patient as part of the circuit and allows maximum power to be transferred to the patient.
The output intensity controls the percentage of maximum power transferred to the patient circuit. The tuning control adjusts the output circuit for maximum energy transfer from the radio frequency oscillator. The power output meter monitors only the current that is drawn from the power supply and not the energy being delivered to the patient.
The power output produces sufficient energy to raise the tissue temperature into a therapeutic range. The specific absorption rate (SAR) represents the rate of energy absorbed per unit area of tissue mass.
Patient sensation provides the basis for recommendations of continuous shortwave diathermy dosage and thus varies considerably with different patients. The following are recommended dosage guidelines: Dose I (lowest): No sensation of heat
Dose II (low): Mild heating sensation
Dose III (medium): Moderate (pleasant) heating sensation
Dose IV (heavy): Vigorous heating that is tolerable below the pain threshold
Therapy may be delivered to the patient via either capacitance or induction techniques. The Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy devices use the following accessories:
- Capacitive Electrodes
- Pair of 80mm round
- Pair of 120mm round
- Pair of 165mm round
- 180mm x 120mm flexible rubber electrodes
- 250mm x 145mm flexible rubber electrodes
- Inductive Electrodes
- Monode 14cm
- Diplode 18cm x 39cm
Each of these techniques can affect different biological tissues, therefore selection of the appropriate electrodes is essential for effective treatment.
{2}------------------------------------------------
# 510(k) Summary per 21 CFR 807.92(c)
## 510(k) # K083433
Page 3 of 3
| 510(k) # | Proprietary<br>Name | Predicate Device<br>Manufacture's | Device<br>Classification<br>Names | Regulation<br>Numbers | Classification(s)<br>and Product<br>Codes |
|----------|---------------------|--------------------------------------|-----------------------------------|-----------------------|-------------------------------------------|
| K042554 | Auto Therm 390 | Mettler Electroinics,<br>Corporation | Shortwave<br>Diathermy | 890.5290 | Class II, IMJ |
| K030382 | SeliTherm | Selicor, Inc. | Shortwave<br>Diathermy | 890.5290 | Class II, IMJ |
| K022458 | Auto Therm 395 | Mettler Electronics,<br>Corporation | Shortwave<br>Diathermy | 890.5290 | Class II, IMJ |
| K973732 | Megapulse II | PTI, Inc. | Shortwave<br>Diathermy | 890.5290 | Class II, IMJ |
# Substantially Equivalent Predicate Devices
# Declarations of Conformity
The Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy devices comply with the following FDA recognized Consensus Standards:
UL 60601-1: 2003, Standards for Medical Equipment Part 1: General Requirements for Safety, 1* Edition
IEC 60601-1-2: 2001, Medical Electrical Equipment Part 1 - 2: General requirements for Safety -Collateral Standard, Electromagnetic Compatibility - Requirements and Tests, 2nd Edition
IEC 60601-2-3: 1991, Medical Electrical Equipment – Part 2: Particular Requirements for the Safety of Short-Wave Therapy Equipment; Amendment 1 – 1998, 2nd Edition (The United States FDA does not recognize IEC 60601-2-3)
IEC 60601-1-4: 2000, Medical Electrical Equipment -- Part 1 -- 4: General Requircments for Safety -Collateral Standard: Programmable Medical Electrical Systems Consolidated with Amendment 1, 1999
# Truthful and Accurate Statement
A statement attesting to the truthfulness and accuracy of the information was included in the premarket submission.
# Further Information
Contact:
Chattanooga Group 4717 Adams Road Hixson, TN 37343 U.S.A.
Michael Treas, Director of Regulatory Affairs Telephone: (423) 870-2281 ext. 7207 Fax: (423) 870-7404 Email: michael.treas@djoglobal.com
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is facing left and has a sharp beak and talons.
### Public Health Service
MAR 1 8 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Chattanooga Group % Mr. Michael Treas Director of Regulatory Affairs 4717 Adams Road Hixson, Tennessee 37343
Re: K083433
Trade/Device Name: Intelect® SWD 100/Senior Solutions® Shortwave Diathermy Regulation Number: 21 CFR 890.5290 Regulation Name: Shortwave diathermy Regulatory Class: II Product Code: IMJ Dated: February 17, 2009 Received: February 18, 2009
Dear Mr. Treas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices markcted in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
# Page 2 - Mr. Michael Treas
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device. to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number:
Device Name: Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy
Indications for Use:
Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffiess, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases.
Generally accepted indications for use:
- Pain Relief .
- Reduce Muscle Spasm .
- . Decrease Joint Stiffness
- . Contractures
- Increase Blood Flow �
- Chronic Inflammatory Conditions .
- . Bursitis
- Tenosynovitis �
- Synovitis .
- Chronic Inflammatory Pelvic Disease
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | |
|-------------------------------------------------|--|
| AND/OR | |
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------------------------------|----------|
| Division of General, Restorative, and Neurological Devices | |
| 510(k) Number | 14083433 |
Submission Page 12 of 905 Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.